BEGO SEMADOS S-LINE
Device Facts
| Record ID | K090716 |
|---|---|
| Device Name | BEGO SEMADOS S-LINE |
| Applicant | Bego Implant Systems GmbH & Co. KG |
| Product Code | DZE · Dental |
| Decision Date | Oct 6, 2009 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3640 |
| Device Class | Class 2 |
Indications for Use
The Bego Semados® S-line threaded endosseous dental implants are indicated for restorations in the upper and lower jaw (single tooth replacement, abutments for bridgework, partial or complete edentolism)
Device Story
BEGO Semados® S-Line is a threaded endosseous dental implant system featuring TiPure™155 surface treatment. Designed for subgingival placement in upper or lower jaws via one- or two-stage surgical procedures. Used by dental professionals for tooth replacement and prosthetic support. Provides structural foundation for single crowns, bridges, or partial/complete dentures. Benefits patients by restoring oral function and aesthetics in edentulous or partially edentulous sites.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Threaded endosseous dental implant; TiPure™155 surface treatment; titanium-based construction; designed for one- or two-stage surgical placement.
Indications for Use
Indicated for patients requiring dental restorations in the upper or lower jaw, including single tooth replacement, bridgework abutments, or treatment for partial or complete edentulism.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
- XIVE Dental Implant System® (K032158)