VRAD PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM SOFTWARE
Applicant
Virtual Radiologic Corporation
Product Code
LLZ · Radiology
Decision Date
Mar 27, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
The vRad™ PACS software is used with general purpose computing hardware, which meets or exceeds minimum specifications. vRad™ PACS software is intended to receive, transmit, store and display images for clinical purposes. The vRad™ PACS Viewer component is intended for installation on an off-the-shelf PC, meeting or exceeding minimum specifications and networked with vRad™ PACS Storage component. The vRad™ PACS Viewer is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The vRad™ PACS Viewer can process medical images from DICOM modalities such as X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine, and images from other DICOM compliant modalities. The vRad™ PACS system should not be used for Mammography primary image diagnosis.
Device Story
vRad™ PACS is a software-only system for electronic transmission, storage, and display of radiological images. It accepts DICOM-compliant inputs from various modalities (X-ray, CT, MRI, ultrasound, nuclear medicine). The software provides tools for image manipulation, enhancement, compression, quantification, and visualization, including 3D Maximum Intensity Projection (MIP) and Multi-Planar Reformatting (MPR). It is intended for use on off-the-shelf PCs by trained radiologists in clinical settings to facilitate image interpretation. The system acts as a primary user interface for viewing and processing images, aiding clinicians in diagnostic decision-making by providing optimized image presentation.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software-only PACS; DICOM-compliant; supports 3D MIP and MPR visualization; runs on off-the-shelf general-purpose computing hardware; network-based storage and viewer architecture.
Indications for Use
Indicated for clinical receipt, transmission, storage, and display of medical images from DICOM-compliant modalities (X-ray, CT, MRI, ultrasound, nuclear medicine) for interpretation by qualified radiologists. Contraindicated for primary diagnosis of mammography images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
FuiiFilm Medical Systems, Synapse Workstation Software (K051553)
Fuiifilm Medical Systems, Synapse Image Visualization Software (MIP/MPR) Obliquus (K061672)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
MAR 2 7 2009
February 25, 2009 Date of Summary Preparation:
- Virtual Radiologic Corporation Submitter: 11995 Singletree Lane Suite 500 Eden Prairie, MN 55344-5349
- Company Contact: Kimberly Tokach Requiatory Compliance Manager Virtual Radiologic Corporation Manufacturer:
- 11995 Singletree Lane Suite 500 Eden Prairie, MN 55344-5349
- vRad™ Picture Archiving and Communications Device Name: System, (PACS)
- Medical Image Processing Software Common/Usual Name:
- Picture Archiving and Communications System Classification Name: LLZ Product Code: 21 CFR 892.2050 Regulation Number: Device Classification: Class II FuiiFilm Medical Systems, Synapse Workstation Predicate Devices: Software, K051553 Fuiifilm Medical Systems, Synapse Image
Visualization Software (MIP/MPR) Obliquus, K061672
### Device Description:
vRad™ PACS is a device which consists solely of software and allows electronic transmission of radiological patient images from one location to another. The device has the capability to accept, transfer, display, store, and digitally process medical images to trained and qualified radiologists for the purposes of providing digital diaging The software provides functions for performing operations interpretation services. related to image manipulation, enhancement, compression, and quantification of medical images (except mammography images). The software enables the user to display 3D
{1}------------------------------------------------
maximum intensity projection (MIP) and Multi-Planar Reformatting (MPR) visualization of study images.
### Intended Use:
The vRad™ PACS software is used with general purpose computing hardware, which meets or exceeds minimum specifications. vRad™ PACS software is intended to receive, transmit, store and display images for clinical purposes. The vRad™ PACS Viewer component is intended for installation on an off-the-shelf PC, meeting or exceeding minimum specifications and networked with vRad™ PACS Storage component. The vRad™ PACS Viewer is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The vRad™ PACS Viewer can process medical images from DICOM modalities such as X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine, and images from other DICOM compliant modalities.
The vRad™ PACS system should not be used for Mammography primary image diagnosis.
## Comparison of Technological Characteristics:
The vRad™ PACS shares the same technological characteristics as the predicate devices. These characteristics include similar design, technical requirements and intended use.
# Substantial Equivalence:
The vRad™ PACS is substantially equivalent to the predicate devices in design, technical requirements and intended use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Virtual Radiologic Corporation
# MAR 2 7 2009
- % Mr. Mark Job
- Responsible Third Party Official
- Regulatory Technology Services LLC
- 1394 25th Street NW
- BUFFALO MN 55313
Re: K090649
Trade/Device Name: vRad™ Picture Archiving and Communications System (PACS) Software Regulation Number: 21 CFR 892.2050
Regulation Name: Picture archiving and communications system
Regulatory Class: II
Product Code: LLZ
· Dated: March 10, 2009
Received: March 11, 2009
#### Dear Mr. Job:
We.have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and . adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse everts (Medical Device Reporting (MDR)), please contact the Division of Surveillanee-Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.support/index.btml.
Sincerely yours,
Janine M. Morris
anine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
# Current 510(k) Number:
#### Device Name:
vRad M1 Picture Archiving and Communications System (PACS) Software
909014
#### Indications for Use:
The vRad™ PACS software is used with general purpose computing hardware, which meets or exceeds minimum specifications. vRad™ PACS software is intended to receive, transmit, store and display images for clinical purposes. The vRad™ PACS Viewer component is intended for installation on an off-the-shelf PC, meeting or exceeding minimum specifications and networked with vRad™ PACS Storage component. The vRad™ PACS Viewer is intended to serve as the primary user interface for the processing of medical images for presentation on displays appropriate to the medical task being performed. The vRad™ PACS Viewer can process medical images from DICOM modalities such as X-ray radiography, X-ray computed tomography, magnetic resonance imaging, ultrasound, nuclear medicine, and images from other DICOM compliant modalities.
The vRad™ PACS system should not be used for Mammography primary image diagnosis.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|--------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal and Radiological Devices | |
| 510(k) Number | K090649 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.