The SUN-RAIN Ultrasonic Therapy generates deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.
Device Story
Ultrasonic therapy device generates high-frequency sound waves via a round-headed probe. Sound waves penetrate skin to create gentle deep heat in soft tissues; heat induces vasodilatation, increasing blood flow to target areas. Used for pain relief, muscle spasms, and joint contractures. Operated by healthcare professionals in clinical settings. Device consists of console and probe; SU-400/600/800 models include 3MHz frequency capability. Output is thermal energy; clinical benefit is symptomatic relief of musculoskeletal conditions.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (IEC/EN 60601-1), electromagnetic compatibility (IEC/EN 60601-1-2), and ultrasonic physiotherapy equipment safety (IEC 60601-2-5, IEC 61689).
Technological Characteristics
Ultrasonic physiotherapy equipment; generates high-frequency sound waves. Models SU-300 through SU-900. Includes 3MHz frequency options for specific models. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-5, and IEC 61689. Uses internal PCB control circuitry.
Indications for Use
Indicated for patients requiring deep heat therapy for relief of pain, muscle spasms, and joint contractures. Contraindicated for the treatment of malignancies.
Regulatory Classification
Identification
An ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a landscape with trees and a cloudy sky. The text "SUN-RAIN" is written in bold letters at the bottom of the image. The image appears to be a logo or a graphic design element, possibly for a company or product with a name related to sun and rain.
KO90542
OCT 2 8 2009
## 5. 510(K) SUMMARY
# SUMMARY OF SAFETY AND EFFECTIVENESS for SUN-RAIN Ultrasonic Therapy
| DATE OF SUBMISSION: | February 23, 2009 |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| SUBMITTER: | SUN-RAIN System Corp.<br>8F, No.125, Lane 235, Paochiao Road,<br>Sindian City, Taiwan, ROC, 23145<br>TEL: 886-2-89191180<br>FAX: 886-2-89191190 |
ESTABLISHMENT REGISTRATION NO: 9616346
OFFICIAL Dr. JEN, KE-MIN CONTACT: No. 58, Fu-Chiun Street. Hsin-Chu City, TAIWAN, ROC, 30067 TEL: 886-3-5208829 FAX:886-3-5209783
TRADE NAME: SUN-RAIN Ultrasonic Therapy, SU-300, SU-400. SU-500, SU-600, SU-700, SU-800, SU-900
COMMON/USUAL Ultrasonic Therapy
NAME:
IMI, Class II
CLASSIFICATION CODE:
REGULATION 890.5300 NUMBER:
PREDICATED DEVICE:
SUN-RAIN Ultrasonic Therapy, SU-100 K024013
- INTENDED USE: The SUN-RAIN Ultrasonic Therapy generates deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "SUN-RAIN" in a bold, sans-serif font. Above the text, there is a blurry, pixelated area that appears to be a faded or obscured image. The text is the most prominent feature, suggesting it is the main subject of the image.
Description of Ultrasonic equipment generates high frequency Device: sound waves that are transferred to a specific body area via a round-headed probe. The sound waves travel deep into tissue and create gentle heat. As the probe glides over the skin's surface, sound waves penetrate the skin's surface causing soft tissues to creating deep heat. In turn, the heat induces vasodilatation: drawing blood into the target tissues. The generated deep heat is found to help relieve pain and reduce muscle spasms.
#### Non-Clinical Tests Submitted:
The Ultrasonic Therapy equipment has been tested in accordance with applicable standards for medical device electrical safety, electromagnetic compatibility, and the particular requirements for safety of ultrasonic physiotherapy equipment.
The relevant standards including:
- 1. IEC/EN 60601-1 : Medical electrical equipment Part 1. General requirements for safety, 1996.
- 2. IEC/EN 60601-1-2 : Medical electrical equipment, Part 2. Electromagnetic compatibility -Requirements and tests, 2004.
- 3. IEC 60601-2-5 : Safety particular requirements for ultrasonic physiotherapy equipment, 2005.
#### None
- 1. IEC 61689 : Ultrasonics Physiotherapy systems -Field specifications and methods of measurement in the frequency range.
- 2. Reliability Test report.
The subject new devices and the predicate device have the same indications for use and same power input. The major difference is new device of SU-400/SU-600/SU-800 is with 3MHz of frequency. Besides, the size and the dimension for the subject new devices are different. This is due to the different design and choice for the clients. Although the appearance and probe size are different but we use the same PCB and the specification is almost the This is not related to the safety same. or effectiveness.
Thus the new device is substantially equivalent to the predicate devices in this aspect.
- Clinical Tests Submitted:
Performance Tests Submitted:
Conclusion:
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three overlapping, curved lines that resemble a bird in flight or a stylized human figure. The overall design is simple and conveys a sense of movement and progress.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
OCT 2 3 2009
SUN-RAIN System Corporation % ROC Chinese-European Industrial Research Society Dr. Ke-Min Jen No. 58, Fu Chiun Street Hsin Chu City, Taiwan, ROC 30067
Re: K090542
Trade/Device Name: SUN-RAIN Ultrasonic Therapy, SU-300, SU-400, SU-500, SU-600, SU-700, SU-800, SU-900 Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic diathermy Regulatory Class: II Product Code: IMI Dated: September 30, 2009 Received: October 6, 2009
Dear Dr. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Dr. Ke-Min Jen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CcntersOffices/CDRHOffices/ucm1117801cc11ccm11178001.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, Jease note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Preu 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark M Millham
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
K_____________________________________________________________________________________________________________________________________________________________________________ 510(k) Number:
Device Name:
...
.... ..
SUN-RAIN Ultrasonic Therapy, SU-300, SU-400, SU-500, SU-600, SU-700, SU-800, SU-900
Indications for Use :
The SUN-RAIN Ultrasonic Therapy generates deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
KATHLEEN M. HOGAN FOR M.MEKERSON
(Division Sign of Surgical, Orthopedic, Divisionative Devices
Page I of 1
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.