← Product Code [LLZ](/productcode/LLZ) · K090431

# DEXIS SOFTWARE (K090431)

_Dexis, LLC · LLZ · Jun 22, 2009 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K090431

## Device Facts

- **Applicant:** Dexis, LLC
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jun 22, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The DEXIS software is a software program for general dental and maxillofacial diagnostic imaging. It controls capture, display, enhancement and saving of X-ray digital images from digital imaging systems. It can also handle other types of images acquired by digitizing film with a flat bed scanner, or color images from intra-oral and extra-oral dental cameras.

## Device Story

DEXIS Software is a dental imaging management program; controls capture, display, enhancement, and storage of X-ray images from digital systems, flatbed scanners, and intra/extra-oral cameras. Used in dental clinics by practitioners; facilitates patient database management and image manipulation to improve diagnostic visibility. Software processes images via dental-specific tools to enhance diagnostic value; allows export/import in standard formats. Benefits include streamlined diagnostic workflow and improved image analysis for dental/maxillofacial assessment.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Software-based dental imaging system; 32-bit architecture; compatible with Microsoft Windows XP, 2000, 98, NT, and ME. Functions include image capture, display, enhancement, and database management. Connectivity via video capture cards and standard digital imaging interfaces.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- DEXIS Model 601 ([K962631](/device/K962631.md))
- Gendex Dental Systems, VixWin Pro ([K060178](/device/K060178.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K090431
Page 1 of 2

Image /page/0/Picture/1 description: The image shows a logo for Sydron Dental Specialties. The logo features the letters 'sds' in a stylized font, with a curved line above the letters. Below the letters, the words 'SYDRON DENTAL SPECIALTIES' are written in a smaller font. The logo appears to be for a dental company.

# Section III - 510(k) Summary of Safety and Effectiveness

Submitter:

JUN 2222009

Sybron Dental Specialties, Inc. 1717 West Collins Drive Orange, CA 92656 Claudia Ortiz - Contact Person

Date Summary Prepared: February 4, 2009

#### Device Name:

- Trade Name DEXIS Software .
- Common Name Dental Imaging Device .
- Classification Name Picture Archiving and Communication Systems, per 21 CFR § . 892.2050.

### Devices for Which Substantial Equivalence is Claimed:

- Fair Image, Inc., (now owned by DEXIS, LLC.), DEXIS Model 601 (K962631) .
- Gendex Dental Systems, VixWin Pro (K060178) .

#### Device Description:

The DEXIS Software is a software program for general dental and maxillo-facial diagnostic imaging. It controls capture, display, enhancement and saving of x-ray digital images from digital imaging systems. It can also handle other types of images acquired by digitizing film with a flat bed scanner, or color images from intraoral or extraoral dental cameras. DEXIS Software allows the following functionality:

- Control scanning and intake of x-ray images from imaging plates with the * DenOptix scanner.
- Control the direct capture of x-ray images from the intraoral sensor Visualix / GX-. S. KaVo Dig eXam, Orthoralix DPI & DDE series systems.
- View and capture color images from certain cameras, such as dental camera, via a . suitable video capture card.
- Export and import digital images (such as those obtained by scanning a film) in . several standard file formats.
- Process images with dental specific tools, to enhance their diagnostic value. •
- Analyze and manipulate images in order to gather additional diagnostic . information which may not be immediately apparent on initial visual inspection.
- Create a database of patients and easily store images in patient folders. .

1717 West Cullins Avenue, Orange, CA 92867 800-537-7824 714-516-7400

{1}------------------------------------------------

K890431
Page 2 of 2

#### Intended Use of the Device:

The DEXIS software is a software program for general dental and maxillofacial diagnostic imaging. It controls capture, display, enhancement and saving of X-ray digital images from digital imaging systems. It can also handle other types of images acquired by digitizing film with a flat bed scanner, or color images from intra-oral and extra-oral dental cameras.

# Substantial Equivalence:

The DEXIS Software is an existing device which was granted market clearance by the FDA following the submission of a 510(k) pre-market notification (K962631). The modified DEXIS Software has been rewritten in 32-bit code and will now run in 'native' format with Microsoft® Windows XP® (home or professional) and Windows® 2000. The modified device also retains 'backward compatibility' with earlier versions including Windows® 98, NT and non-upgrade versions of ME. There will be no change to the intended use or technical charactcristics.

The DEXIS Software is substantially equivalent in intended use and technical characteristics as the VixWin Pro software marketed by Gendex Dental Systems (K060178).

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

JUN 222009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### DEXIS LLC

%Ms. Claudia Ortiz Compliance Director, Regulatory Affairs & Quality Assurance Sybron Dental Specialties 1717 West Collins Avenue ORANGE CA 92867

Re: K090431

Trade/Device Name: DEXIS Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 27, 2009 Received: June 2, 2009

Dear Ms. Ortiz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrb/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

Device Name: DEXIS Software

# Indications for Use:

The DEXIS software is a software program for general dental and maxillofacial diagnostic imaging. It controls capture, display, enhancement and saving of X-ray digital images from digital imaging systems. It can also handle other types of images acquired by digitizing film with a flat bed scanner, or color images from intra-oral and extra-oral dental cameras.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

fori mthey

f Reproductive, Abdominal,
ological Devices
Number: K090431

Page 1 of 1

રજુના

---

**Source:** [https://fda-staging.innolitics.com/device/K090431](https://fda-staging.innolitics.com/device/K090431)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).