K090422 · Jolife AB · DRM · Jun 17, 2009 · Cardiovascular
Device Facts
Record ID
K090422
Device Name
LUCAS 2
Applicant
Jolife AB
Product Code
DRM · Cardiovascular
Decision Date
Jun 17, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LUCAS 2 Chest Compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness. LUCAS 2 must only be used in cases where chest compressions are likely to help the patient.
Device Story
LUCAS 2 is an electrically powered mechanical chest compression system; provides automated, consistent, and uninterrupted compressions for adult patients in acute circulatory arrest. Device consists of an upper unit with an electrically driven piston rod and pressure pad/suction cup, mounted on a back plate. Used in clinical and transport settings by emergency medical personnel; allows for hands-free operation and patient safety during transport. System delivers compressions according to AHA guidelines; replaces manual compressions to ensure consistent circulation. Benefits include improved consistency of compressions, safety for personnel during transport, and reduced interruption times during application.
Clinical Evidence
Bench testing only. Functional testing was conducted under various operating conditions to demonstrate that the device meets product specifications and safety standards. No clinical data provided.
Technological Characteristics
Electrically powered mechanical chest compression system. Components include an upper unit with an electrically driven piston rod, pressure pad with suction cup, and back plate. Battery-powered. Materials are identical to those used in the predicate LUCAS 1 (K062401).
Indications for Use
Indicated for adult patients experiencing acute circulatory arrest, characterized by absence of spontaneous breathing, pulse, and loss of consciousness, where chest compressions are clinically indicated.
Regulatory Classification
Identification
An external cardiac compressor is an externally applied prescription device that is electrically, pneumatically, or manually powered and is used to compress the chest periodically in the region of the heart to provide blood flow during cardiac arrest. External cardiac compressor devices are used as an adjunct to manual cardiopulmonary resuscitation (CPR) when effective manual CPR is not possible (e.g., during patient transport or extended CPR when fatigue may prohibit the delivery of effective/consistent compressions to the victim, or when insufficient EMS personnel are available to provide effective CPR).
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Nonclinical performance testing under simulated physiological conditions must demonstrate the reliability of the delivery of specific compression depth and rate over the intended duration of use.
(2) Labeling must include the following:
(i) The clinical training necessary for the safe use of this device;
(ii) Adjunctive use only indication prominently displayed on labels physically placed on the device and in any device manuals or other labeling;
(iii) Information on the patient population for which the device has been demonstrated to be effective (including patient size and/or age limitations,
*e.g.,* adult, pediatric and/or infant); and(iv) Information on the time necessary to deploy the device as demonstrated in the performance testing.
(3) For devices that incorporate electrical components, appropriate analysis and testing must demonstrate that the device is electrically safe and electromagnetically compatible in its intended use environment.
(4) Human factors testing and analysis must validate that the device design and labeling are sufficient for effective use by the intended user, including an evaluation for the time necessary to deploy the device.
(5) For devices containing software, software verification, validation, and hazard analysis must be performed.
(6) Components of the device that come into human contact must be demonstrated to be biocompatible.
{0}------------------------------------------------
2009-02-16
jolife
1090422
JUN 1 7 2009
# 510(k) SUMMARY
### Jolife's LUCAS 2
This document provides a brief summary of the LUCAS 2 device and its supporting information.
## 1) GENERAL DATA
510(k) submitter
Jolife AB Ingmar Malm Ideon Scheelevägen 17 SE-223 70 Lund Sweden
Phone: + 46 46 286 50 00 Cell phone: + 46 70 721 22 24 Fax: + 46 46 286 50 10
e-mail: ingmar.malm@jolife.com
Contact person
Howard Holstein Hogan & Hartson Columbia Square 555 Thirteenth street NW Washington DC 20004-1109
Phone: (202) 637-5813
Date when prepared
February 13, 2009
Trade name
LUCAS 2
DRM
Common name
Chest Compression System
Classification name
Product code
Predicate devices
● Lucas 1 cleared under K062401
- Thumper model 1008 cleared under K073079
External cardiac compressor (21 CFR 870.5200)
Autopulse model 100 cleared under K072527 ●
JOLIFE AB, Ideon Science Park, SE-223 70 Lund, Swedon, Visiting Address: Scheelevagen 17 Tel: +46 (0) 46 286 50 00, Fax: +46 (0) 46 286 50 10, E-mail: info@jotife.com, jolife.com 1 (3)
0017
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image shows the word "jolife" in a stylized font, with a circular graphic to the left of the word. The graphic consists of a circle with a line passing through it. The word "jolife" is written in a font that appears to be textured or dotted. The image is in black and white.
#### 2) DESCRIPTION OF THE DEVICE
LUCAS 2 is an electrically powered mechanical chest compression system providing controlled automated chest compressions on adult patients who have acute circulatory arrest,
LUCAS 2 consists of an upper part containing the electrically driven piston rod, which acts on the patients chest via a pressure pad. The pressure pad is surrounded by a suction cup.
The support legs of the upper part are fastened to the back plate prior to starting compressions.
LUCAS 2 is designed to:
- · Provide effective, consistent and uninterrupted compressions according to the guidelines given by American Heart Association (AHA);
- · Provide good circulation during the patient transport process;
- · Provide safety during transportation for both emergency medical personnel and patient, allowing emergency medical personnel to wear safety belts during transportation while LUCAS 2 delivers continuous, consistent and uninterrupted compressions;
- · Provide hands-free compressions in any situation.
- · Be applied to the patient with interruption of less than 20 seconds.
#### 3) INTENDED USE / INDICATIONS FOR USE
LUCAS 2 Chest Compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness.
LUCAS 2 must only be used in cases where chest compressions are likely to help the patient.
#### 4) COMPARISON TO PREDICATE DEVICES
The LUCAS 2 and its predicates have basically the same performance characteristics.
The LUCAS 2 has the same performance characteristics as the previuosly cleared LUCAS 1 device.
The LUCAS 2 and Thumper 1008 act on the chest via a pressure pad, while Autopulse acts on the chest via a circumferential belt.
The LUCAS 2 and Autopulse are battery powered, while the Thumper 1008 is powered by compressed oxygen while the previously cleared LUCAS 1 is powered by compressed air.
JOLIFE AB, Ideon Science Park, SE-223 70 Lund, Sweden, Visiting Address: Scheelevägen
Tel: +46 (0) 46 286 50 00, Fax: +46 (0) 46 286 50 10, E-mail: info@jolife.com. jolife.co
2 (3)
**0018**
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the word "jolife" in a stylized font, with a logo to the left of the word. The logo consists of a circle with lines running through it. The letters in the word "jolife" are outlined in black, with a white fill.
# 5) SUMMARY OF SUBSTANTIAL EQUIVALENCE
LUCAS 2 is substantially equivalent to the predicate devices. The LUCAS 2 has the same intended use and substantially similar indications for use, basic overall function, and performance as its predicate devices.
#### 6) MATERIALS
All materials used in the manufacture of LUCAS 2 are suitable for its purpose and are well known and are the same as those materials and intended use as the materials included in the 510(k)-cleared LUCAS I (K062401). .
#### 7) TESTING
. (
Appropriate product testing was conducted and included a number of function tests during different operating conditions. These tests demonstrated that the functionality, safety, efficacy and capability of LUCAS 2 comply with the product specifications and safety standards and support substantial equivalence to predicate devices.
In all instances, the LUCAS 2 functioned as intended and all results observed were as expected,
JOLIFE AB, Ideon Science Park, SE-223 70 Lund, Sweden, Visiting Address: Scheelevagen 17 Tel: +46 (0) 46 286 50 00, Fax: +48 (0) 46 286 50 10, E-mail: Info@jolife.com, Jolife.com 3 (3)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle or bird symbol, with three curved lines forming the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAR 3 1 2011
Jolife AB c/o Mr. Howard M. Holstein Hogan Lovells US LLP Columbia Square 555 13" Street, NW Washington, DC 20004
Re: K090422
LUCAS 2 Regulation Number: 21 CFR 870.5200 Regulation Name: External Cardiac Compressor Regulatory Class: Class III Product Code: DRM Dated: May 18, 2009 Received: May 18, 2009
Dear Mr. Holstein:
This letter corrects our substantially equivalent letter of June 17, 2009.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{4}------------------------------------------------
Page 2 - Mr. Howard M. Holstein
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
M. S. Kelleher
An Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) LUCAS 2
Image /page/5/Picture/2 description: The image shows the word "jolife" next to a circular graphic. The graphic is a circle with a line going through it. The text is in a sans-serif font and appears to be slightly pixelated. The image is black and white.
# Indications for Use ·
510(k) Number (if known): _ Ko9 0 422
Device Name: LUCAS 2
Indications for Use:
LUCAS Chest Compression System is to be used for performing external cardiac compressions on adult patients who have acute circulatory arrest defined as absence of spontaneous breathing and pulse, and loss of consciousness.
LUCAS must only be used in cases where chest compressions are likely to help the patient.
| Prescription Use | X | AND/OR | Over-The-Counter Use |
|-----------------------------|---|--------|------------------------|
| (Part 21 CFR 801 Subpart D) | | | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF . NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
| 510(k) Number | K090922 |
|---------------|---------|
|---------------|---------|
JOLIFE AB, Ideon Science Park, SE-223 70 Lund, Sweden, Visiting Address: Scheelevägen 17
Tel: +46 (0) 46 286 50 00, Fax: +46 (0) 46 286 50 10, E-mail: info@lolife.com, www.jolife.com
1 (1)
3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.