← Product Code [GEI](/productcode/GEI) · K090393

# ARTHROCARE SYSTEM 15000 CONTOLLER, MODEL 72202149, ARTHROCARE ARTHROWANDS, MODEL 72202139,72202141, 72202143, 72202144 (K090393)

_Arthrocare Corp. · GEI · Jun 23, 2009 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K090393

## Device Facts

- **Applicant:** Arthrocare Corp.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Jun 23, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The ArthroCare System 15000 is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures: [List of procedures including ACL/PCL, acromioplasty, articular cartilage, bursectomy, chondroplasty, fascia, ligament, notchplasty, scar tissue, soft tissue, subacromial decompression, synovectomy, tendon, acetabular labrum, articular labrum, capsule, capsular release, cartilage flaps, cysts, discoid meniscus, frozen shoulder release, glenoidale labrum, lateral release, loose bodies, meniscal cystectomy, meniscectomy, plica removal, synovial membrane, triangular fibrocartilage (TFCC), villusectomy, carpal ligaments, glenohumeral capsule, medial retinaculum, rotator cuff, and wrist tendons across various joints including ankle, elbow, hip, knee, shoulder, and wrist].

## Device Story

ArthroCare System 15000 is a bipolar, high-frequency electrosurgical system; comprises a controller (generator), disposable bipolar wands, and foot control. Used in arthroscopic and orthopedic procedures; operated by surgeons. Device delivers high-frequency energy to wands to perform soft tissue ablation, resection, coagulation, and hemostasis. Output allows precise tissue removal and bleeding control; enhances surgical efficiency and patient outcomes by minimizing thermal damage to surrounding tissues. Used in clinical/OR settings.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Bipolar electrosurgical system; high-frequency generator (Controller); disposable bipolar single-use Wands; foot-controlled operation. System utilizes high-frequency energy for tissue ablation and coagulation.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- ArthroCare System 12000 ([K071709](/device/K071709.md), [K082666](/device/K082666.md))
- ArthroCare ArthroWands ([K070958](/device/K070958.md), [K082980](/device/K082980.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K090393

JUN 23 2009

## 510(k) Summary

# ArthroCare Corporation ArthroCare® System 15000

### General Information

Submitter Name/Address:

Establishment Registration Number:

Contact Person:

Date Prepared:

Device Description

Trade Name:

Generic/Common Name:

Classification Name:

ArthroCare Corporation 680 Vaqueros Avenue Sunnyvale, CA 94085-3523

2951580

Valerie Defiesta-Ng Director, Regulatory Affairs

February 13, 2009

ArthroCare® System 15000

Electrosurgical Device and Accessories

Electrosurgical Cutting and Coagulation Device and Accessories (Class II, 21 CFR 878.4400, Product Code GEI)

#### Predicate Devices

ArthroCare System 12000 ArthroCare ArthroWands

K071709 and K082666 K070958 and K082980

#### Product Description

The ArthroCare System 15000 consists of a bipolar, high frequency, electrosurgical generator called the Controller, a family of disposable, bipolar, single use Wands and Foot Control.

{1}------------------------------------------------

## Intended Uses

.

The ArthroCare System 15000 is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

| Arthroscopic and Orthopedic Procedures | Joint Specific or All Joints (ankle, elbow, hip, knee, shoulder, and wrist) |
|----------------------------------------|-----------------------------------------------------------------------------|
| Ablation and Debridement               |                                                                             |
| • ACL/PCL                              | Knee                                                                        |
| • Acromioplasty                        | Shoulder                                                                    |
| • Articular Cartilage                  | All Joints                                                                  |
| • Bursectomy                           | All Joints                                                                  |
| • Chondroplasty                        | All Joints                                                                  |
| • Fascia                               | All Joints                                                                  |
| • Ligament                             | All Joints                                                                  |
| • Notchplasty                          | Knee                                                                        |
| • Scar Tissue                          | All Joints                                                                  |
| • Soft Tissue                          | All Joints                                                                  |
| • Subacromial Decompression            | Shoulder                                                                    |
| • Synovectomy                          | All Joints                                                                  |
| • Tendon                               | All Joints                                                                  |
| Excision and Resection                 |                                                                             |
| • Acetabular Labrum                    | Hip                                                                         |
| • Articular Labrum                     | All Joints                                                                  |
| • Capsule                              | All Joints                                                                  |
| • Capsular Release                     | Knee                                                                        |
| • Cartilage Flaps                      | Knee                                                                        |
| • Cysts                                | All Joints                                                                  |
| • Discoid Meniscus                     | Knee                                                                        |
| • Frozen Shoulder Release              | Shoulder                                                                    |
| • Glenoidale Labrum                    | Shoulder                                                                    |
| • Lateral Release                      | Knee                                                                        |
| • Ligament                             | All Joints                                                                  |
| • Loose Bodies                         | All Joints                                                                  |
| • Meniscal Cystectomy                  | Knee                                                                        |

.

 ·

·

.

:

:

.

{2}------------------------------------------------

| Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) |
|----------------------------------------|--------------------------------------------------------------------------------------|
| • Plica Removal                        | All Joints                                                                           |
| • Scar Tissue                          | All Joints                                                                           |
| • Soft Tissue                          | All Joints                                                                           |
| • Synovial Membrane                    | All Joints                                                                           |
| • Tendon                               | All Joints                                                                           |
| • Triangular Fibrocartilage (TFCC)     | Wrist                                                                                |
| • Villusectomy                         | Knee                                                                                 |
| Coagulation                            |                                                                                      |
| • ACL/PCL                              | Knee                                                                                 |
| • Articular Cartilage                  | All Joints                                                                           |
| • Carpal Ligaments                     | Wrist                                                                                |
| • Glenohumeral Capsule                 | Shoulder                                                                             |
| • Ligament                             | All Joints                                                                           |
| • Medial Retinaculum                   | Knee                                                                                 |
| • Rotator Cuff                         | Shoulder                                                                             |
| • Tendon                               | All Joints                                                                           |
| • Wrist Tendons                        | Wrist                                                                                |

#### Substantial Equivalence

In establishing substantial equivalence to the predicate devices, ArthroCare compared the indications for use, dimensional specifications, and performance specifications of the subject device and the predicate device. Additionally, performance testing has been completed to demonstrate the substantial equivalence of the ArthroCare System 15000 to the predicate device. The performance testing and device comparison demonstrated that the subject devices are substantially equivalent to the predicate devices, and is safe and effective for its intended use.

### Summary of Safety and Effectiveness

The ArthroCare System 15000, as described in this premarket notification 510(k), is substantially equivalent to the predicate devices. The differences in performance specifications, and labeling are not substantial changes or modifications, and do not significantly affect the safety or efficacy of the System.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with flowing lines, positioned to the right of the department's name. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the left side of the emblem.

#### JUN 23 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

ArthroCare Corporation % Ms. Valerie Defiesta-Ng Director, Regulatory Affairs 680 Vaqueros Avenue Sunnyvale, California 94085-3523

Re: K090393

Trade/Device Name: ArthroCare® System15000 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: II Product Code: GEI Dated: May 19, 2009 Received: May 20, 2009

Dear Ms. Defiesta-Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class 111 (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing

{4}------------------------------------------------

Page 2-Ms. Valerie Defiesta-Ng

practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/cdrh/comp/ for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark H. Milliken

Mark N. Melkerson Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use Statement ·

Device Name

ArthroCare® System 15000

510(k) Number: K

Indications for Use:

The ArthroCare System 15000 is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:

| Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) |
|----------------------------------------|--------------------------------------------------------------------------------------|
| <b>Ablation and Debridement</b>        |                                                                                      |
| • ACL/PCL                              | Knee                                                                                 |
| • Acromioplasty                        | Shoulder                                                                             |
| • Articular Cartilage                  | All Joints                                                                           |
| • Bursectomy                           | All Joints                                                                           |
| • Chondroplasty                        | All Joints                                                                           |
| • Fascia                               | All Joints                                                                           |
| • Ligament                             | All Joints                                                                           |
| • Notchplasty                          | Knee                                                                                 |
| • Scar Tissue                          | All Joints                                                                           |
| • Soft Tissue                          | All Joints                                                                           |
| • Subacromial Decompression            | Shoulder                                                                             |
| • Synovectomy                          | All Joints                                                                           |
| • Tendon                               | All Joints                                                                           |
| <b>Excision and Resection</b>          |                                                                                      |
| • Acetabular Labrum                    | Hip                                                                                  |
| • Articular Labrum                     | All Joints                                                                           |
| • Capsule                              | All Joints                                                                           |
| • Capsular Release                     | Knee                                                                                 |
| • Cartilage Flaps                      | Knee                                                                                 |
| • Cysts                                | All Joints                                                                           |
| • Discoid Meniscus                     | Knee                                                                                 |
| • Frozen Shoulder Release              | Shoulder                                                                             |
| • Glenoidale Labrum                    | Shoulder                                                                             |
| • Lateral Release                      | Knee                                                                                 |
| • Ligament                             | All Joints                                                                           |
| • Loose Bodies                         | All Joints                                                                           |
| • Meniscal Cystectomy                  | Knee                                                                                 |
| • Meniscectomy                         | Knee                                                                                 |

*(Division Sign-Off)*

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number_

{6}------------------------------------------------

| Continued                              |                                                                             |
|----------------------------------------|-----------------------------------------------------------------------------|
| Arthroscopic and Orthopedic Procedures | Joint Specific or All Joints (ankle, elbow, hip, knee, shoulder, and wrist) |
| • Plica Removal                        | All Joints                                                                  |
| • Scar Tissue                          | All Joints                                                                  |
| • Soft Tissue                          | All Joints                                                                  |
| • Synovial Membrane                    | All Joints                                                                  |
| • Tendon                               | All Joints                                                                  |
| • Triangular Fibrocartilage (TFCC)     | Wrist                                                                       |
| • Villusectomy                         | Knee                                                                        |
| Coagulation                            |                                                                             |
| • ACL/PCL                              | Knee                                                                        |
| • Articular Cartilage                  | All Joints                                                                  |
| • Carpal Ligaments                     | Wrist                                                                       |
| • Glenohumeral Capsule                 | Shoulder                                                                    |
| • Ligament                             | All Joints                                                                  |
| • Medial Retinaculum                   | Knee                                                                        |
| • Rotator Cuff                         | Shoulder                                                                    |
| • Tendon                               | All Joints                                                                  |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

X

Over-the-Counter Use

Division Sign Off

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K090393

. .

---

**Source:** [https://fda-staging.innolitics.com/device/K090393](https://fda-staging.innolitics.com/device/K090393)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).