This device is an electric powered wheelchair for indoor and/or outdoor use, to be used by partially disabled individuals capable of operating a few essential hand controls.
Device Story
Extreme 4x4-X8 is a battery-powered, four-wheel drive electric wheelchair; features welded steel frame, four 24V direct-drive motors, joystick controller, seat, armrests, and footrests. Designed for indoor/outdoor use on various surfaces including soft/rough terrain and inclines. Operated by patient via joystick; provides mobility for individuals with limited physical capability. Controller (Dynamic DX2-PMA70L) is programmable to meet individual user needs; includes battery charge gauge. Safety features include electromagnetic brakes engaged when stationary and drive inhibition during certain seat positioning functions. Benefits include improved obstacle climbing and handling via articulated frame and increased speed/range compared to predecessor.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Welded steel frame; four 24V 4-pole direct-drive gear in-line motors; Dynamic DX2-PMA70L electronic controller; two 73 Amp/hour batteries. Dimensions: 45.25" length, 28" width. Features electromagnetic braking system. Upholstery meets California 117 fire retardancy specifications.
Indications for Use
Indicated for partially disabled individuals, limited to a sitting position, who possess the capability to operate a powered wheelchair for indoor and/or outdoor mobility.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
K090350
Image /page/0/Picture/1 description: The image shows a logo with a spiral design on the left and the words "Innovation In Motion" on the right. The spiral design has small lines radiating outward, resembling a sun. The text is in a simple, sans-serif font, with "Innovation" on the top line and "In Motion" on the bottom line.
PO ROX 507 gola, IN 46703 USA h. 260.665.2769 260.665.3047
# 510(k) Summary Extreme 4x4-X8 Power Wheelchair Innovation In Motion 510(k) Premarket Notification
APR 1 0 2009
C-1
Applicant's Name, Address, Telephone, Fax, and E-mail information:
Innovation In Motion 201 Growth Parkway P0 Box 507 Angola, IN 46703 Phone: 260.665.2769 Fax: 260.665.3047 E-mail: rick@vestil.com
Preparer's Name, Address, Telephone, Fax, Contact Name, E-mail information, and Date Prepared:
Innovation In Motion 201 Growth Parkway P0 Box 507 Angola, IN 46703 Phone: 260.665.2769 Fax: 260.665.3047 Contact Name: Rick Michael Vice President E-mail: rick@vestil.com Date: February 2009
Manufacturer's Name, Address, Telephone, and Fax Information:
Magic International Pty. Ltd. 2/16 Viewtech Place Rowville, Victoria 3178 Australia Phone: 011.613.9755.8100 Fax: 011.613.9755.8111
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the words "Innovation In Motion". To the left of the text is a spiral design with three lines extending from the left side. The text is in a simple, sans-serif font. The overall design is clean and modern.
PO BOX 507
Angola, IN 46703 USA
Ph. 260.665.2769
F. 260.665.3047
mobility-usa.com
Device Trade, Common, and Classification names:
Device Name: Extreme 4x4- X8
Common Name: Power Wheelchair
Classification Name: Restorative Devices
Predicate Device:
Extreme 4x4- X4 (K000796)
Device Description:
The Extreme 4x4- X8 is a battery powered, electric motor driven device with the intended function of providing mobility to persons limited to a sitting position, that have the capability of operating a powered wheelchair. It is a four-wheel drive format wheelchair as its basic configuration with direct drive motors on all four wheels. The wheelchair frame is a non-folding type.
The frame of the unit is of welded steel construction. The wheelchair consists of the following basic components: frame w/four motors, joystick, seat, armrests, and footrests.
Intended Use:
This device is an electric powered wheelchair for indoor and/or outdoor use, to be used by partially disabled individuals capable of operating a few essential hand controls.
{2}------------------------------------------------
### ాevice Comparison Table:
# COMPARISON TABLE WITH MARKETED DEVICE
··
| K000796 | | K000796 | |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | X4 | | X8 |
| Type | Wheelchair, Power | | Wheelchair, Power |
| Intended Use | To provide mobility to persons limited to a<br>sitting position, who have the capability of<br>operating a power wheelchair | | To provide mobility to persons limited to a<br>sitting position, who have the capability of<br>operating a power wheelchair |
| Where Used | Designed to operate indoors and over a<br>variety of outdoor surfaces including<br>soft/rough terrain with inclines up to 1 in 14 | | Designed to operate indoors and over a<br>variety of outdoor surfaces including<br>soft/rough terrain with inclines up to 1 in 14 |
| Design | Compact, rigid | | Compact, with front frame articulation |
| Construction | Steel frame | | Steel frame |
| Dimensions | Overall length: 43.3"<br>Overall width: 28" | | Overall length: 45.25"<br>Overall width: 28" |
| Wheel Size | 4 x Ø14" | | 4 x Ø14" |
| Rolling Base Weight<br>(No Seat, No<br>Batteries) | 159 lbs. | | 145 lbs. |
| Range | 19 miles | | 25 miles |
| Maximum Speed | 5 mph | | 6.5 mph |
| Maximum Weight<br>Capacity | 400 lbs | | 400lbs |
| Controller | Dynamic DX-PMB2 electronic control and<br>joystick<br>*Programmable to meet individual needs<br>*Battery charge level gauge<br>*Attaches to either side | | Dynamic DX2-PMA70L electronic control<br>and joystick<br>*Programmable to meet individual needs<br>*Battery charge level gauge<br>*Attaches to either side |
| Motors | Four 24V, 4 pole direct drive motors/gearbox<br>*Freewheeling lever/motor lock releases and<br>engages front and rear motor brakes | | Four 24V, 4 pole direct drive gear in line<br>motors/gearbox<br>*Freewheeling lever/motor lock releases and<br>engages rear motor brakes |
| Suspension | None | | None |
| Batteries | Two 73 Amp/hour | | Two 73 Amp/hour |
| Seat Width | 16"-24" | | 16"-24" |
| Seating | Upholstery meets California 117<br>specifications for fire retardancy | | Upholstery meets California 117<br>specifications for fire retardancy |
| Footrest | 1 piece rigid footplate, flip up footplates and<br>others | | 1 or 2 piece, fixed or flip up, angle and<br>height adjustable rigid footplates, and others |
| Armrest | Height adjustable, removable | | Height adjustable, removable |
| Safety Features | Electromagnetic brake engaged when chair<br>is stationary. When electromagnetic brakes<br>are disengaged, chair cannot be driven | | Electromagnetic brake engaged when chair<br>is stationary. When electromagnetic brakes<br>are disengaged, chair cannot be driven<br>Drive is inhibited with some seat positioning<br>functions |
・・・..
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the words "Innovation In Motion". To the left of the text is a circular design that resembles a sun or a spiral. The text is stacked, with "Innovation" on top and "In Motion" below. The font appears to be handwritten or stylized, giving the logo a unique and informal look.
PO BOX 507 Angola, IN 46703 USA Ph. 260.665.2769 F. 260.665.3047 mobility-usa.com
Substantial Equivalence:
The X8 has the same intended function and use as the X4. The X8 is the next generation of the X4 with minor modifications:
- . Articulated frame for improved obstacle climbing and handling
- High Speed, high torque high efficiency gear in-line motors .
- Latest edition of Dynamic Control electronic software/hardware .
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three overlapping curved lines forming its body and wings. The symbol is enclosed within a circular border, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged around the upper half of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Vestil Manufacturing Coporation % Innovation in Motion Mr. Rick Michael Vice President 201 Growth Parkway Angola, Indiana 46703
Re: K090350
Trade Name: Extreme 4x4 - X8 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: II Product Code: ITI Dated: March 16, 2009 Received: March 20, 2009
Dear Mr. Michael:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
APR 1 0 2009
:
{5}------------------------------------------------
#### Page 2 - Mr. Rick Michael
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please . contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A. Wilkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: EXTREME 4x4-X8
Indications For Use:
inclo:lite provide The intended use of the XB power wheelchair is ಕಿ position , whoshave the capability persons limited to ಗಿ ડાં તેતાવુ র্ন power bis heal chair opevativ ત્ત્વ
Prescription Use (Part 21 CFR 301 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark N. Milliken
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
- Page : xi
510(k) Number K
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.