The ADVIA 2120/2120i auto-analyzer Body Fluid Application is an in vitro diagnostic test for the enumeration of the total nucleated cell (TNC) count and RBC count for pleural, peritoneal, and peritoneal dialysis (PD) specimens collected in K2 or K3 EDTA.
Device Story
ADVIA 2120/2120i hematology analyzer modified for body fluid analysis; uses existing CBC/Diff TIMEpac reagents. System aspirates, dilutes, and analyzes fluid samples via Basophil/Lobularity and RBC/Platelet channels. Basophil channel uses surfactant/phthalic acid to lyse RBCs and strip leukocyte cytoplasm; laser diode light scatter (2°-3° low-angle, 5°-15° high-angle) counts TNCs. RBC/Platelet channel uses sodium dodecyl sulfate/glutaraldehyde to sphere/fix RBCs for counting via same light scatter principle. Operated by laboratory personnel in clinical settings. Output includes TNC and RBC counts, displayed on cytograms. TNC count derived from Basophil channel; flagged if inconsistent with Peroxidase channel WBC count. Results assist clinicians in evaluating body fluid cellular composition. Benefits include automated, standardized quantification of cells in serous fluids.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of technological characteristics to the predicate device.
Indicated for the enumeration of total nucleated cell (TNC) and RBC counts in pleural, peritoneal, and peritoneal dialysis (PD) fluid specimens collected in K2 or K3 EDTA.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
COULTER® LH 750 Body Fluids Application (k030606)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k090346
B. Purpose for Submission:
Modification for body fluid application for ADVIA 2120 and the ADVIA 2120i.
C. Manufacturer and Instrument Name:
Siemens Healthcare Diagnostics, Inc.
Body Fluid Application for ADVIA 2120 and the ADVIA 2120i.
D. Type of Test or Tests Performed:
Quantitative, Automated Hematology Analyzer
WBC, NEUT%/#, LYMPH%/#, MONO%/#, EOS%/#, BASO%/#, LUC, RBC, HCT, MCV, RDW, CHCM, HDW, HGB, MCH, MCHC, PLT, MPV, RETIC%/#, MCVg, MCVr, CHCMg, CHCMr. CHg, CHr
Body Fluid Application:
Total nucleated cell (TNC) count and RBC count
E. System Descriptions:
1. Device Description:
The Body Fluid Application can be run on the ADVIA 2120 and the ADVIA 2120i hematology analyzers. The ADVIA 2120/ADVIA 2120i Body Fluid Application uses the ADVIA 120 CBC TIMEpac and ADVIA 120 Diff TIMEpac reagents to determine the TNC and RBC counts. No additional Body Fluid Application specific reagents are needed.
The ADVIA 2120 Hematology system consists of the following: an analytical module that aspirates, dilutes, and analyzes whole blood samples; an autosampler that automatically mixes, identifies, and presents the samples for processing; a computer workstation that controls the instrument, provides primary user interface with the instrument and manages the data produced by the instrument; a printer that optionally generates reports based on the instrument results and an autoslide module.
2. Principles of Operation:
The ADVIA 2120/2120i Body Fluid Application uses the Basophil/Lobularity and RBC/Platelet channels to enumerate the TNC and RBC counts. The TNC count is derived from the Basophil/Lobularity channel. The ADVIA 2120/2120i BASO reagent contains surfactant and phthalic acid which, in the presence of low heat in the Baso channel reaction chamber, lyses RBCs and strips the cytoplasmic membrane from all leukocytes except basophils. This cell suspension is subsequently passed through the flowcell. The cell suspension is intercepted by light from the laser diode where the low-angle light scatter (2° to 3°) and high-angle light scatter (5° to 15°) signals of each cell are counted.
The RBC count is derived from the RBC/Platelet channel. The ADVIA 2120/2120i
{1}
RBC/Platelet reagent uses sodium dodecyl sulfate and glutaraldehyde to sphere and fix the RBCs. This cell suspension uses the same flowcell and low-angle and high-angle light scatter signals as the Baso channel to count the RBCs. The system automatically reports the TNC and RBC counts in conventional or SI units appropriate for body fluid samples.
The Basophil/Lobularity cytochemical reactions consist of two steps:
Step 1 - Red blood cells and platelets are lysed using the ADVIA 120 BASO reagent.
Step 2 - All white blood cells except basophils are stripped of their cytoplasm using the ADVIA 120 BASO reagent and the increased temperature in the reaction chamber.
The Body Fluid Application default Run Screen, displays 3 cytograms:
Peroxidase, Basophil/Lobularity and RBC Scatter cytograms. The Basophil/Lobularity and Peroxidase channel cytograms have been modified specifically for the light scatter characteristics of the cells in body fluid samples. The RBC cytogram is the same as the cytogram for whole blood analysis.
The reported TNC count is derived from the Basophil/Lobularity channel. The TNC count is the sum of the events above the Baso Y-channel Noise threshold plus the events left of Baso X-channel 49. The events in the Baso channel Noise region for the Body Fluid Application are excluded from the TNC count.
The WBC count from the Peroxidase channel (WBCP) is compared to the TNC count from the Basophil/Lobularity channel. When the TNC does not agree with the WBCP count within limits specific to the Body Fluid Application, the TNC count is flagged with an asterisk. Determination of the RBC count is the same as for whole blood samples. The Header information includes the body fluid specimen type.
3. Modes of Operation:
ADVIA 2120/2120i system is a random access analyzer which has three sampling modes: [1] Manual open tube sampler, [2] manual closed tube sampler and [3] automated closed tube sampler (Autosampler).
4. Specimen Identification:
Sample identification is done by following modes: [1] Manually entered sample ID [2] barcode [3] rack/position numbers (for autosampler only) and [4] instrument automatic incremental numbering.
5. Specimen Sampling and Handling:
Body fluids samples must be manually mixed by gentle inversion for the manual open and closed tube modes. The automated closed tube sampler mixer rotates first to +45° then to +135°, staying at each orientation for a minimum of 0.6 seconds and a maximum of 15 seconds. A total of 25 mixer-cycles are required before aspiration. Fully mixed samples which have a dwell of over 15 seconds require an additional 20 mixer cycles before aspiration. If a dwell exceeds 5 minutes, the samples must be mixed again for 45 cycles before aspiration.
6. Calibration:
ADVIA 2120/2120i systems calibration for all reportable CBC, Diff. and retic. (except % retic.) parameters require the ADVIA SETpoint Calibrator and one
2
{2}
whole blood sample.
7. Quality Control:
It is recommended that the system be controlled using ADVIA TESTpoint or 3·in·1 Hematology Controls. These controls are intended to be integrated into a clinical laboratory's own quality control program and procedures.
8. Software:
FDA has reviewed applicant's Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X ☐ or No ☐
F. Regulatory Information:
1. Regulation section:
21 CFR 864.5220 - Automated differential cell counter
2. Classification:
Class II
3. Product code:
GKZ - Counter, Differential Cell
4. Panel:
81 Hematology
G. Intended Use:
1. Indication(s) for Use:
The ADVIA 2120/2120i auto-analyzer Body Fluid Application is an in vitro diagnostic test for the enumeration of the total nucleated cell (TNC) count and RBC count for pleural, peritoneal, and peritoneal dialysis (PD) specimens collected in K2 or K3 EDTA.
2. Special Conditions for Use Statement(s):
None
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
COULTER® LH 750 Body Fluids Application - k030606
2. Comparison with Predicate Device:
| Similarities | | |
| --- | --- | --- |
| Item | Predicate Device | Device |
| | COULTER® LH 750 Body Fluids Application | Siemens ADVIA 2120/2120i Body Fluids Application |
| Instrument | Automated Hematology Analyzer | Same |
| Intended Use | Enumeration of the total nucleated cell (TNC) count and RBC count | Same |
| Dilution | Automated dilution | Same |
| Counts | Automated calculation of counts | Same |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Predicate Device | Device |
| | COULTER® LH 750 Body Fluids Application | Siemens ADVIA 2120/2120i Body Fluids Application |
| Sample Type | Pleural, peritoneal, and Peritoneal Dialysis fluids | Cerebrospinal fluid, serous fluid, and synovial fluid |
| Detection | Impedance technology | Light scatter technology |
I. Special Control/Guidance Document Referenced:
1. CLSI Document. EP5-A2. Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline—Second Edition". 2004. CLSI, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898.
2. CLSI Document H56-A. "Body Fluid Analysis for Cellular Composition; Approved Guideline". 2006. CLSI, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898.
3. CLSI Document EP17-A. "Protocol for determination of Limits of Detection and Limits of Quantitation". 2004. CLSI, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898.
4. CLSI Document GP10. "Assessment of the Clinical Accuracy of Laboratory Tests Using Receiver Operating Characteristic (ROC) Plots". CLSI, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898.
J. Performance Characteristics:
1. Analytical Performance:
a. Accuracy:
The Body Fluid Application TNC count was evaluated by comparing the hemocytometer count to the ADVIA 2120/2120i count for 434 body fluid samples. The Body Fluid Application RBC count was evaluated by comparing the manual chamber count to the ADVIA 2120/2120i count for 78 body fluid samples. Samples were pleural, peritoneal, and peritoneal dialysis specimens collected in EDTA tubes with no visible clotting. The results are considered acceptable if they meet the design input requirements for the method comparison study. The design input requirements are >0.85 for r, 1 +/- 0.15 for slope and 0 +/- 100 for intercept.
| Parameter | n | r | Slope | Intercept | 95% CI Slope | 95% CI Intercept |
| --- | --- | --- | --- | --- | --- | --- |
| TNC | 434 | 0.942 | 0.90 | 65 | 0.87 to 0.93 | -22 to 153 |
| RBC | 78 | 0.964 | 0.99 | -0.5 | 0.93 to 1.05 | -14.6 to 13.5 |
Conclusion: The results of the method comparison studies indicated comparable performance relative to the reference method.
b. Precision/Reproducibility:
Precision Studies
{4}
A precision study was performed based on the guidelines contained in CLSI EP5-A2. The TNC and RBC pools were assayed over a 20 day period with two runs per day; two replicates per run, for total of 80 replicates for each pool.
Stabilized pools were prepared with target TNC concentrations of 50, 100, 500, 1000, and 10000 cells/uL in plasma diluted 1:2 with phosphate buffer.
Stabilized pools were prepared with target RBC concentrations of 50, 100, 500, and 1000 x 10³ cells/uL in plasma diluted 1:2 with phosphate buffer.
The results were analyzed using nested analysis of variance consistent with the guidance contained in CLSI EP5-A2.
TNC Precision
| Within Run | | | | Between Run | | Between Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Level | Mean | SD | CV | SD | CV | SD | CV | SD | CV |
| 50 | 58.6 | 6.887 | 11.7% | 3.001 | 5.1% | 1.103 | 1.9% | 7.6 | 12.9% |
| 100 | 118.33 | 10.908 | 9.2% | 0 | 0.0% | 3.47 | 2.9% | 11.4 | 9.7% |
| 500 | 593.48 | 26.32 | 4.4% | 0 | 0.0% | 8.304 | 1.4% | 27.6 | 4.7% |
| 1000 | 1179.5 | 35.152 | 3.0% | 17.329 | 1.5% | 19.276 | 1.6% | 43.7 | 3.7% |
| 10000 | 116.27 | 196.994 | 1.7% | 61.866 | 0.5% | 136.538 | 1.2% | 247.542 | 2.1% |
RBC Precision
| Within Run | | | | Between Run | | Between Day | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Level | Mean | SD | CV | SD | Mean | SD | CV | SD | CV |
| 50 | 52.664 | 2.202 | 4.2% | 0 | 0.0% | 1.001 | 1.9% | 2.4 | 4.6% |
| 100 | 104.99 | 2.994 | 2.8% | 0 | 0.0% | 1.4 | 1.3% | 3.3 | 3.1% |
| 500 | 518.48 | 6.405 | 1.2% | 2.04 | 0.4% | 3.497 | 0.7% | 7.6 | 1.5% |
| 1000 | 1023.7 | 13.055 | 1.3% | 6.404 | 0.6% | 0 | 0.0% | 14.5 | 1.4% |
The results are considered acceptable if they meet the design input requirements for within-run precision. The design input requirements are ≤ 15% CV at 500 cells/μL for the TNC count and ≤ 10% CV at 50 x 10³ cells/μL for the RBC count. Conclusions: The results of the precision studies using stabilized specimens indicated acceptable precision over the time period tested.
## Repeatability Testing
Specimens were utilized from the stability study, described below. One specimen of each body fluid type assayed 10 times in one run on a single day. Additional assays were not possible because of sample volume and sample stability limitations. The mean, standard deviation (SD), and coefficient of variation are calculated for each sample included in the study. The results are considered acceptable if they meet the design input requirements for within-run precision. The design input requirements are ≤ 15% CV at 500 cells/μL for the TNC count and ≤ 10% CV at 50 x 10³ cells/μL for the RBC count.
| | TNC | | | RBC | | |
| --- | --- | --- | --- | --- | --- | --- |
| Fluid | Mean | SD | CV | Mean | SD | CV |
| Pleural | 93492 | 8749.5 | 9.4 | 80 | 6.9 | 8.6 |
| Peritoneal | 274 | 22.5 | 8.2 | 41 | 1.6 | 3.9 |
| PD | 64 | 6.5 | 10.2 | 28 | 1.7 | 6.1 |
{5}
Conclusion: The results from the repeatability study demonstrate that within-run precision of the body fluid counts meets the design input requirements of ≤ 15% CV at 500 cells/μL for the TNC count and ≤ 10% CV at 50 × 10³ cells/μL for the RBC count.
## Reproducibility/Specimen Stability
The goal of the testing the Body Fluid Software is to demonstrate stability of fluids under different conditions using the system software. This was accomplished by using remnant body fluid samples collected in K2 or K3 EDTA at two external sites. These sites collected pleural, peritoneal and peritoneal dialysate body fluids. The samples were run on ADVIA 2120/2120i using the body fluid application. These fluid samples were tested within 2-hours to represent the baseline or zero (0) time result, as well at different time points, and different conditions (room temperature and at 2-8°C). Specimens included:
Five (5) samples with RBC counts that fall in the range from ≥ 10,000 cells/uL to 6.76 × 10⁶ cells/μL for each type of fluid.
Five (5) samples with TNC counts that fall in the range of 0.030 to 400 × 10³ cells/μL (30 to 400,000) for each type of fluid.
Conclusion: The observed regression slopes all meet the acceptance criteria of falling within the range of 0.9 to 1.1. It can be concluded that body fluid samples are stable at room temperature for 8 hours and when refrigerated for 24 hours.
## c. Linearity:
The objective of the linearity protocol was to verify low level linearity of the body fluid TNC and RBC counts. High level linearity of the counting methods has been established previously (k042251) in whole blood sampling modes. Low level pools with a target WBC concentration of 500 cells/uL and an RBC target concentration of 500 thousand cells/uL were created. Serial dilutions of these pools were made with cell free materials (plasma diluted in PBS) over the low level range of the assays. The WBC pools were aspirated 13 times each in body fluid mode, and the RBC pools were aspirated 12 times each.
TNC Linearity
| % of High Pool | Observed TNC (cells/uL) | Expected TNC (cells/uL) | Obs vs Exp % | Obs vs Exp Abs |
| --- | --- | --- | --- | --- |
| 0% | 1 | 0 | | 1 |
| 4% | 19 | 20 | 95.0 | -1 |
| 8% | 41 | 40 | 102.5 | 1 |
| 12% | 67 | 60 | 111.7 | 7 |
| 16% | 80 | 80 | 100.0 | 0 |
| 20% | 105 | 99 | 106.1 | 6 |
| 100% | 497 | 497 | 100.0 | 0 |
{6}
RBC Linearity
| % of High Pool | Observed RBC (10E3cells/uL) | Expected RBC (10E3cells/uL) | Obs vs Exp % | Obs vs Exp Abs |
| --- | --- | --- | --- | --- |
| 0% | 0 | 0 | | 0 |
| 2% | 11 | 10 | 110.0 | 1 |
| 6% | 30 | 31 | 96.8 | -1 |
| 10% | 52 | 52 | 100.0 | 0 |
| 15% | 77 | 77 | 100.0 | 0 |
| 20% | 106 | 103 | 102.9 | 3 |
| 100% | 515 | 515 | 100.0 | 0 |
Conclusion: The results meet the design requirements that the deviations are < 10% or within 10 cells.
d. Carryover:
The carryover protocol was run in all three sampling modes: manual open tube, manual closed tube, and automated sampling. Carryover was tested by four different methods in each sampling mode:
A normal whole blood assayed in CBC/Diff mode followed by three aspirations of a saline sample in body fluid mode.
A manipulated whole blood sample with elevated WBC counts (approximately 30,000 cells/uL) and RBC counts (approximately 6 million cells/uL) assayed in CBC/Diff mode followed by three aspirations of a saline sample in body fluid mode.
A normal whole blood assayed in body fluid mode followed by three aspirations of a saline sample in body fluid mode.
A manipulated whole blood sample with elevated WBC counts (approximately 30,000 cells/uL) and RBC counts (approximately 6 million cells/uL) assayed in body fluid mode followed by three aspirations of a saline sample in body fluid mode.
Five sets of carryover sequences were run for each type of carryover test within each sampling mode. A saline primer was assayed before each body fluid sample when using the Autosampler as instructed in the body fluids instructions for use.
Carryover results are listed in the Tables below. Each Table entry is the average of the 5 carryover sequences run in each configuration.
Open Tube Sampling Carryover
{7}
| Sample | TNC % Carryover | RBC % Carryover |
| --- | --- | --- |
| Normal WBC/RBC CBC/Diff | 0.17% | 0.11% |
| Elevated WBC/RBC CBC/Diff | 0.16% | 0.16% |
| Normal WBC/RBC BF | 0.27% | 0.26% |
| Elevated WBC/RBC BF | 0.38% | 0.31% |
Manual Closed Tube Sampling Carryover
| Sample | TNC % Carryover | RBC % Carryover |
| --- | --- | --- |
| Normal WBC/RBC CBC/Diff | 0.11% | 0.11% |
| Elevated WBC/RBC CBC/Diff | 0.14% | 0.10% |
| Normal WBC/RBC BF | 0.36% | 0.25% |
| Elevated WBC/RBC BF | 0.29% | 0.25% |
Automated Sampling Carryover
| Sample | TNC % Carryover | RBC % Carryover |
| --- | --- | --- |
| Normal WBC/RBC CBC/Diff | 0.00% | 0.01% |
| Elevated WBC/RBC CBC/Diff | 0.00% | 0.01% |
| Normal WBC/RBC BF | 0.00% | 0.00% |
| Elevated WBC/RBC BF | 0.00% | 0.00% |
Conclusion: The carryover results meet the specification of $< 0.4\%$ for the TNC, and $< 0.3\%$ for the RBC count for Manual Closed Tube Sampler (MCTS) and the Manual Open Tube Sampler (MOTS), and $< 0.1\%$ for both the TNC and RBC counts for the Autosampler.
# e. Interfering Substances:
If TNC and RBC counts from patient samples where the manual differential shows a proportional neutrophil count $\geq 80\%$ , crystals, the presence of lipids and/or chylomicrons, cellular deterioration and disintegration, or the patient is under treatment with a sclerosing agent, the results should be confirmed by an alternate method.
# 2. Other Supportive Instrument Performance Data Not Covered Above:
The range of reportable values study included a limit of blank (LoB) study, a limit of detection study (LoD), and a limit of quantitation (LoQ) study as defined in CLSI EP17 CLSI EP17-A to determine the lowest concentration of TNC and RBC counts that can be detected with at least $95\%$ probability.
{8}
9
## Limit of Blank
Cell free samples were prepared by mixing equal amounts of cell free plasma and PBS. These samples were assayed in body fluid mode over a 4-day period. A total of 80 assays were performed on the blank samples.
The results are analyzed non-parametrically as recommended in section 4.1.1 of CLSI EP17-A. The TNC LoB is four (4) cells, and the RBC LoB is 1,000 cells. The limit of blank was determined to be 4.24 cells / uL for TNC and $1.43 \times 10^{3}$ cells/uL for RBC.
## Limit of Detection
Limit of detection (LoD) was estimated by preparing pools with a TNC concentration of approximately 20 cells/uL and an RBC concentration of approximately $10 \times 10^{3}$ cells/uL. Based on the LoB results, five samples prepared with a TNC concentration of approximately 20 cells/uL were assayed 13 times each day for a total of four days. Likewise, five samples prepared with an RBC concentration of approximately $10 - 11 \times 10^{3}$ cells/uL were assayed 13 times each day for a total of four days. LoD is calculated as $\mathrm{LoD} = \mathrm{LoB} + (\mathrm{c - beta})\mathrm{x(SD)}$ as recommended in section 4.3.2 of CLSI EP17-A. The limit of detection was determined to be 12.8 cells/uL for TNC and $2.8 \times 10^{3}$ cells/uL for RBC.
## Limit of Quantitation
Pools were prepared with target TNC concentrations of 0, 20, 40, 60, 80, 100 and 500 cells/uL. Each pool was assayed 13 times each. Pools were prepared with target RBC concentrations of 0, 10, 30, 50, 75, 100, and $500 \times 10^{3}$ cells/uL. Each pool was assayed 12 times each. LoQ is determined as the analyte level where a $20\%$ total CV is obtained. The LoQ for the TNC count is 27 cells/uL and the LoQ for the RBC count is $4.94 \times 10^{3}$ cells/uL.
## K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
## L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.