TROPAZONE LOTION

K090337 · Midlothian Laboratories · FRO · Aug 13, 2009 · SU

Device Facts

Record IDK090337
Device NameTROPAZONE LOTION
ApplicantMidlothian Laboratories
Product CodeFRO · SU
Decision DateAug 13, 2009
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

Tropazone Lotion is used to manage and relieve the burning and itching experienced with various types of dermatoses, including radiation dermatitis, atopic dermatitis, and allergic contact dermatitis. It helps relieve dry waxy skin by maintaining a moist wound and skin environment, which is beneficial to the healing process.

Device Story

Tropazone Lotion is a non-sterile, semi-viscous, oil-in-water emulsion hydrogel wound dressing. It is applied topically to the skin to manage and relieve symptoms of dermatoses. The device functions by maintaining a moist wound and skin environment, which supports the healing process. It is intended for prescription use, requiring a physician's diagnosis of the underlying disease state. The lotion is applied to the affected area three times per day or as needed. It provides symptomatic relief from burning and itching and addresses dry, waxy skin conditions.

Clinical Evidence

No clinical data provided. Evidence consists of non-clinical performance testing, including a L929 Agar overlay cytotoxicity study showing a mild/moderate reaction and skin irritation/sensitization testing under occlusion indicating the device is a non-primary irritant and non-primary sensitizer.

Technological Characteristics

Non-sterile, semi-viscous oil-in-water emulsion. Ingredients include purified water, liquid paraffin (mineral oil), petrolatum, glyceryl stearate, PEG-100 stearate, paraffin, lecithin, polysorbate 60, DEA-Cetyl Phosphate, dimethicone, carbomer, imidazolidinyl urea, methylparaben, propylparaben, triethanolamine, and fragrance. Emulsion thickened with polyacrylic acid polymer (carbomer).

Indications for Use

Indicated for patients experiencing burning and itching associated with dermatoses, including radiation dermatitis, atopic dermatitis, and allergic contact dermatitis. Used to relieve dry waxy skin by maintaining a moist environment.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary ### Submitter of the Application Midlothian Laboratories 760 Industrial Park Blvd., Unit C Montgomery, AL 36117 Phone: (334) 288-8661 Fax: (334) 288-8651 Toll Free: (800) 344-8661 #### Trade Name Tropazone Lotion #### Common Name Hydrogel wound dressing ### Device Classification Hydrogel wound dressing with preservatives Unclassified, FRO #### Substantial Equivalence/ Predicate Device Tropazone Lotion is substantially equivalent to the currently marketed devices, MimyX cream, cleared under K041342, Zenieva cleared under K073246 and Biafine cleared under K964240. #### Device Description Tropazone Lotion is a non-sterile, semi-viscous emulsion intended for topical application. It is presented for prescription (requires a physician diagnosis of disease state) use. The product is formulated as an oil-in-water emulsion containing a polyacrylic acid polymer as the thickening agent. The oil composition of Tropazone lotion is composed of mineral oil, lecithin, fatty acids and a silicon-based organic polymer. ### Intended Use of the Device Tropazone Lotion is used to manage and relieve the burning and itching experienced with various types of dermatoses, including radiation dermatitis, atopic dermatitis, and allergic contact dermatitis. It helps relieve dry waxy skin by maintaining a moist wound and skin environment, which is beneficial to the healing process. Intended use is identical to that of MimyX and Zenieva. AUG 1 3 2009 {1}------------------------------------------------ K090337 pg 2 of 3 : ## Table 1 - Technological Comparison . . . : | Product Name | Tropazone | MimyX | Zenieva | Biafine | |---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) | K090000 | K041342 | K073246 | K964240 | | Ingredients | Purified water,<br>liquid paraffin<br>(mineral oil),<br>petrolatum,<br>alcohol, glyceryl<br>stearate, PEG-100<br>stearate, paraffin,<br>lecithin,<br>polysorbate 60,<br>DEA-Cetyl<br>Phosphate,<br>dimethicone,<br>carbomer,<br>imidazolidinyl urea,<br>methylparaben,<br>propylparaben,<br>triethanolamine,<br>fragrance | Purified water,<br>olive oil, glycerin,<br>pentylene glycol,<br>palm glycerides,<br>vegetable oil,<br>hydrogenated<br>lecithin, squalene,<br>betaine,<br>palmitamide MEA,<br>sarcosine,<br>acetamide MEA,<br>hydroxyethyl<br>cellulose, sodium<br>carbomer, xanthan<br>gum | Purified water,<br>olive oil, glycerin,<br>pentylene glycol,<br>palm glycerides,<br>vegetable oil,<br>hydrogenated<br>lecithin, squalene,<br>betaine,<br>palmitamide MEA,<br>sarcosine,<br>acetamide MEA,<br>hydroxyethyl<br>cellulose, sodium<br>carbomer, xanthan<br>gum | Purified water, liquid<br>paraffin, ethylene<br>glycol monostearate,<br>stearic acid, propylene<br>glycol, paraffin wax,<br>squalene, avocado oil,<br>trolamine/sodium<br>alginate,<br>triethanolamine, cetyl<br>palmitate,<br>methylparaben<br>(sodium salt), sorbic<br>acid (potassium salt),<br>polyparaben (sodium<br>salt) and fragrance | | # applications<br>per day | 3 times per day<br>or as needed | 3 times per day<br>or as needed | 3 times per day<br>or as needed | 3 times per day | | Claim | Tropazone Lotion is<br>used to manage<br>and relieve the<br>burning and itching<br>experienced with<br>various types of<br>dermatoses,<br>including radiation<br>dermatitis, atopic<br>dermatitis, and<br>allergic contact<br>dermatitis. It helps<br>relieve dry waxy<br>skin by maintaining<br>a moist wound and<br>skin environment,<br>which is beneficial<br>to the healing<br>process. | MimyX is<br>used to manage<br>and relieve the<br>burning and itching<br>experienced with<br>various types of<br>dermatoses,<br>including radiation<br>dermatitis, atopic<br>dermatitis, and<br>allergic contact<br>dermatitis. It helps<br>relieve dry waxy<br>skin by maintaining<br>a moist wound and<br>skin environment,<br>which is beneficial<br>to the healing<br>process. | Zenieva is<br>used to manage<br>and relieve the<br>burning and itching<br>experienced with<br>various types of<br>dermatoses,<br>including radiation<br>dermatitis, atopic<br>dermatitis, and<br>allergic contact<br>dermatitis. It helps<br>relieve dry waxy<br>skin by maintaining<br>a moist wound and<br>skin environment,<br>which is beneficial<br>to the healing<br>process. | Biafine is for the<br>dressing and<br>management of<br>superficial wounds,<br>minor abrasions,<br>dermal ulcers, donor<br>sites, 1st and 2nd degree<br>burns, including<br>sunburns, and<br>radiation dermatitis.<br>When applied properly<br>to a wound, Biafine<br>provides an optimum<br>moist environment for<br>the healing process and<br>isolates the wound<br>from harmful germs<br>and other external<br>contamination. | | Product<br>Description | Water-based<br>emulsion | Water-based<br>emulsion | Water-based<br>emulsion | Water-based emulsion | | Physical<br>Properties | Non-sterile white to<br>off-white lotion | Non-sterile white to<br>off-white thick<br>cream | Non-sterile white to<br>off-white thick<br>cream | Non-sterile white to<br>off-white thick cream | : : : {2}------------------------------------------------ K090337 pg 3 of 3 ## Clinical Performance Data Tropazone when tested under occlusion may be considered a non-primary irritant and a non-primary sensitizer to skin. . ### Non-Clinical Performance Data In a L929 Agar overlay cytotoxicity study using Tropazone, the cells exhibited a mild/moderate reaction. Carlos Concession ### Conclusion Tropazone is substantially equivalent to the predicate devices listed in the table above. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. In the center of the circle is an abstract image of an eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002 ## AUG 1 8 2009 Midlothian Laboratories % Mr. Bryce Harvey President 760 Industrial Park Boulevard, Unit C Montgomery, Alabama 36117 Re: K090337 Trade/Device Name: Tropazone Lotion Regulatory Class: Unclassified Product Code: FRO Dated: June 23, 2009 Received: June 24, 2009 Dear Mr. Harvey: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ ## Page 2 - Mr. Bryce Harvey If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark M. Millbern Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K090337 # Indications for Use 510(k) Number (if known): Ko 90337 Device Name: Tropazone Lotion_ Indications for Use: Tropazone Lotion is used to manage and relieve the burning and itching experienced with various types of dermatoses, including radiation dermatitis, atopic dermatitis, and allergic contact dermatitis. It helps relieve dry waxy skin by maintaining a moist wound and skin environment, which is beneficial to the healing process. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) ## Concurrence of CDRH, Office of Device Evaluation (ODE) Nil RPOden for nxn (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices Page | of h 510(k) Number K090337
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...