K090334 · Siemens Medi Cal Solutions, Inc. · IYN · Feb 20, 2009 · Radiology
Device Facts
Record ID
K090334
Device Name
ACUSON S2000 DIAGNOSTIC ULTRASOUND SYSTEM
Applicant
Siemens Medi Cal Solutions, Inc.
Product Code
IYN · Radiology
Decision Date
Feb 20, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The S2000™ ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system also provides the ability to measure anatomical structures (fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-esophageal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The Arterial Health Package (AHP) software provides the physician with the capability to measure intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imagino".
Device Story
Acuson S2000 is a diagnostic ultrasound system for imaging and fluid flow analysis. It processes acoustic signals from various transducers (curved, linear, phased array, TEE) to produce real-time 2D, M-mode, Doppler (PWD, CWD, Color, Amplitude), and 3D images. Used in clinical settings (e.g., OR, clinic) by physicians and sonographers. Includes calculation packages for anatomical measurements and the Arterial Health Package (AHP) for Intima Media Thickness (IMT) measurement. AHP allows comparison against normative tables to assist in cardiovascular risk communication. Output is displayed for clinician review to support diagnosis. Benefits include non-invasive visualization and quantitative assessment of anatomical structures and vascular health.
Clinical Evidence
Bench testing only. System safety and performance verified against standards including UL 60601-1, IEC 60601-2-37, and AIUM/NEMA UD-2/UD-3. AHP software validated against peer-reviewed normative tables.
Technological Characteristics
Diagnostic ultrasound system using pulsed echo and Doppler imaging. Transducers: curved, linear, phased array, TEE. Connectivity: standard ultrasound interfaces. Software: includes measurement and calculation packages (AHP). Biocompatibility: ISO 10993-1. Safety: IEC 60601-1, IEC 60601-2-37.
Indications for Use
Indicated for fetal, abdominal, intraoperative, pediatric, small parts, transcranial, OB/GYN, cardiac, pelvic, neonatal/adult cephalic, vascular, musculoskeletal, superficial musculoskeletal, and peripheral vascular ultrasound imaging and fluid flow analysis in patients of all ages.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Acuson Antares Ultrasound System
Submission Summary (Full Text)
{0}------------------------------------------------
IC0{0334
S2000 Ultrasound System 510(k) Submission
## 510(k) Summarv Prepared February 3, 2009
FEB 2 0 2009 Sponsor: Siemens Medical Solutions, Inc., Ultrasound Division 1230 Shorebird Way Mountain View, California 94043
Contact Person: Shelly Pearce Telephone: (650) 694-5988 Fax: (650) 694-5580
Submission Date: February 3, 2009
Device Name: Acuson S2000™ Diagnostic Ultrasound System
Common Name: Diagnostic Ultrasound System
## Classification:
Regulatory Class: 11 Review Category: Tier II Classification Panel: Radiology
Ultrasonic Pulsed Doppler Imaging System FR # 892.1550 Product Code 90-IYN Ultrasonic Pulsed Echo Imaging System FR # 892,1560 Product Code 90-1YO Diagnostic Ultrasound Transducer FR # 892.1570 Product Code 90-ITX
## A. Legally Marketed Predicate Devices
The Acuson S2000™Ultrasound System is substantially equivalent to the Acuson Antares Ultrasound System.
## B. Device Description:
The Acuson S2000™ has been designed to meet the following product safety standards:
- UL 60601-1, Safety Requirements for Medical Equipment ■
- IEC 60601-2-37 Diagnostic Ultrasound Safety Standards 1
- CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment 해
- AIUM/NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical 를 Acoustic Output Indices on Diagnostic Ultrasound Equipment
- AIUMNEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound 내
- 트 93/42/EEC Medical Devices Directive
- Safety and EMC Requirements for Medical Equipment 파
- EN/IEC 60601-1 내
- EN/IEC 60601-1-1
- EN/IEC 60601-1-2
- IEC 1157 Declaration of Acoustic Power 체
- ISO 10993-1 Biocompatibility 비
{1}------------------------------------------------
## C. Intended Use
The S2000™ ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures {fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalic, adult cephalic, cardiac, trans-esophageal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imagino".
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the USA. The seal features a stylized eagle with outstretched wings, symbolizing protection and service. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
.
Public Health Service
FEB 2 0 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Siemens Medical Solution USA, Inc., Ultrasound Group % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K090334 . .
Trade/Device Name: ACUSON S2000m Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: February 9, 2009 Received: February 10, 2009
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, buting of devices, good manifacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the ACUSON S2000™ Diagnostic Ultrasound System, as described in your premarket notification:
#### Transducer Model Number
CW2 Probe CW5 Probe EC9-4 Curved Array 9L4 Linear Array 14L5 Multi-D Array
4P1 Phased Array 6C2 Curved Array 4C1 Curved Array 4V1 Phased Array 10V4 Phased Array
{3}------------------------------------------------
Enclosure(s)
14L5 SP Linear Array 7CF2 Curved Array 9EVF4 Curved Array V 5Ms Multiplane TEE 17L5HDS Linear Array 18L6 HD Linear Array 8V3 Phased Array 4V1c Phased Array 613 EV8C4
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this totterspiesses contact that an (240) 276-3566: ·
Sincerely yours,
Heelt Lewen
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
#### 1.3 Indications for Use
510(k) Number (if known):
K090334
Device Name: S2000™Diagnostic Ultrasound System
Indications for Use: K090334
The S2000™ ultrasound imaging systems are intended for the following applications: Fetal, Abdominal, Intraoperative, Pediatric, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications.
The system also provides the ability to measure anatomical structures (fetal, abdominal, intraoperative, intraoperative neurological, pediatric, small organ, neonatal cephalio, adult cephalic, cardiac, trans-esophageal, transvaginal, peripheral vessel, musculoskeletal (conventional), musculo-skeletal (superficial) and neonatal cardiac} and calculation packages that provide information that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
The Arterial Health Package (AHP) software provides the physician with the capability to measure Intima Media Thickness and the option to reference normative tables that have been validated and published in peer-reviewed studies. The information is intended to provide the physician with an easily understood tool for communicating with patients regarding state of their cardiovascular system. This feature should be utilized according to the "ASE Consensus Statement; Use of Carotid Ultrasound to Identify Subclinical Vascular Disease and Evaluate Cardiovascular Disease Risk: A Consensus Statement from the American Association of Echocardiography; Carotid Intima-Media Thickness Task Force, Endorsed by the Society for Vascular Imaging".
Prescription Use (Part 21 CFR 801 Subpart.D) AND/OR
Over-The-Counter Use (21 CER 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K090334
S2000 510(k) Submission
Page 12 of 46
{5}------------------------------------------------
## S2000 Ultrasound System 510(k) Submission
## 1.3 Indications for Use Forms
## Diagnostic Ultrasound Indications for Use Form
.
510 (k) Number (if known):
Device Name: Intended Use:
# ACUSON S2000 Ultrasound System
Ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 13 |
| Abdominal | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 13 |
| Intraoperative<br>(Note 9) | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14 |
| Intraoperative<br>Neurological | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14 |
| Pediatric | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 |
| Small Organ<br>(Note 1) | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14 |
| Neonatal Cephalic | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 |
| Adult Cephalic | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 |
| Cardiac | P | P | P | P | P | P | P | | BMDC | Note<br>2,3,4,5,6,7,8,10,15 |
| Trans-esophageal | P | P | P | P | P | P | P | | BMDC | |
| Transrectal | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14 |
| Transvaginal | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | P | P | P | P | P | P | P | | BMDC | Note2,3,4,5,6,7,8,10,<br>11,14,15 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | P | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14 |
| Musculo-skeletal<br>Superficial | P | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14 |
| Other (specify)<br>Neonatal Cardiac | P | P | P | P | P | P | P | | BMDC | Note 3,4,6 |
N = new indication; P = previously cleared by FDA, K072786
Note 1 For example: breast, testes, thyroid, penis, prostate, etc.
Note 2 Ensemble tissue harmonic imaging
- Note 3 SieClear multi-view spatial compounding
- Tissue Equalization Technology Note 4
- Note 5 3-Scape real-time 3D imaging
- Note 6 Cadence contrast agent imaging
- Note 7 B&W SieScape panoramic imaging .
- Power SieScape panoramic imaging Note 8
- Note 9 For example: vascular, abdominal Note 10
Clarify VE vascular enhancement technology Note 11 Advanced Sieclear spatial compounding
- Note 13 STIC
- eSie™ Touch elasticity imaging Note 14
etc. Note 15 AHP
(Division Sign-Off)
Division
Division of Reproductive, Abdominal and
Reproductive, Abdominal and
Radiological Devices
510(k) No.
510(k) Number
K010334
S2000 510(k) Submission
Page 18 of 1326
. W
{6}------------------------------------------------
## S2000 Ultrasound System 510(k) Submission
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.10
Diagnostic Ultrasound Indications for Use Form
510 (k) Number (if known):
Device Name: Intended Use:
#### CW2 Probe for use with ACUSON S2000
Ultrasound imaging or fluid flow analysis of the human body as follows:
14.73
| | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | P | | | | | | |
| Abdominal | | | | | P | | | | | | |
| Intraoperative<br>(Note 9) | | | | | P | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | P | | | | | | |
| Small Organ<br>(Note 1) | | | | | P | | | | | | |
| Neonatal Cephalic | | | | | P | | | | | | |
| Adult Cephalic | | | | | P | | | | | | |
| Cardiac | | | | | P | | | | | | |
| Trans-esophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | | | | P | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | P | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | P | | | | | | |
| Other (specify) | | | | | | | | | | | |
Nia such indication; P = 5 rowlously gleared by FDA ill 088803, KO72785
Additional Comments:
Note 1 For example: breast, testes, thyroid, penis, prostate, etc.
Note 9 For example: vascular, abdominal
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Hula Levin
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K090334
S2000 510(k) Submission
Page 19 of 1326
{7}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
#### 510 (k) Number (if known):
Device Name: Intended Use:
## CW5 Probe for use with ACUSON S2000
Ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | P | | | | | |
| Abdominal | | | | | P | | | | | |
| Intraoperative<br>(Note 9) | | | | | P | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | P | | | | | |
| Small Organ<br>(Note 1) | | | | | P | | | | | |
| Neonatal Cephalic | | | | | P | | | | | |
| Adult Cephalic | | | | | P | | | | | |
| Cardiac | | | | | P | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | P | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | P | | | | | |
| Musculo-skeletal<br>Superficial | | | | | P | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K# 063803, K072786
Additionshomments:
· Forexample: breast, testes, thyroid, panis, prostate, For example: vascular, abdominal Note 9
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
> > Prescription Use (Per 21 CFR 801.109)
---
(Division Sign-Off)
Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Page 20 of 1326
{8}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
510 (k) Number (if known):
Device Name: Intended Use:
#### EC9-4 Curved Array Transducer for use with ACUSON S2000 Ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 | |
| Abdominal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,6,7,8,10, 11, | |
| Intraoperative<br>Abdominal | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11,14 | |
| Neonatal Cephalic | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10 | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | | |
| Transrectal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5, 6, 7,8,10,<br>11,14 | |
| Transvaginal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K# 063803, K072788
Accitional Comments:
: Note 'i' For example: breast, testes, thyreigl, gents, prostate, atc
Ensemble tissue harmonic imaging Note 2
Note 3 SieClear multi-view spatial compounding
Note 4 Tissue Equalization Technology
Note 5 3-Scape real-time 3D imaging
Cadence contrast agent imaging Note 6
Note 7 B&W SieScape panoramic imaging
Note 8 Power SieScape panoramic imaging
Note 10 Clarify VE vascular enhancement technology
Advanced Sieclear spatial compounding Note 11
eSie™ Touch elasticity imaging Note 14
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | |
|---------------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal and, Radiological Devices | |
| 510(k) Number | K090334 |
| S2000 510(k) Submission | Page 16 of 46 |
|-------------------------|---------------|
| Page 21 of 1326 | |
{9}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
్లో ప్రా
510 (k) Number (if known): ·
Device Name: Intended Use:
9L4 Linear Array Transducer for use with ACUSON S2000 Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|--------------------------------------|
| | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 |
| Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,6,7,8,10, 11,14 |
| Neonatal. Cephalic | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10, 11 |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | N | N | | N | N | | BMDC | Note 15 |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,6, 7,8,10, 11,<br>14,15 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,6,7,8,10, 11, 14 |
| Musculo-skeletal<br>Superficial | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,6,7,8,10, 11, 14 |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K# 063085, K072786
Additional Commento:
For example: breast, testes, thyrold, panis, prostate, Note - i . -
- Ensemble tissue harmonic imaging Note 2
- Note 3 SieClear multi-view spatial compounding
- Tissue Equalization Technology Note 4
- Note 5 3-Scape real-time 3D imaging
- Note 6 Cadence contrast agent imaging
- Note 7 B&W SieScape panoramic imaging
- Power SieScape panoramic imaging Note B
- Clarify VE vascular enhancement technology Note 10
- Note 11 Advanced Sieclear spatial compounding
- Not
Not
| ote 14 | eSie™ Touch elasticity imaging |
|--------|--------------------------------|
| ote 15 | AHP |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal and Radiological Devices
K 90324
510(k) Number
S2000 510(k) Submission
Page 22 of 1326
Page 17 of 46
{10}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
510 (k) Number (if known):
Device Name: Intended Use: 14L5 Multi-D Array Transducer for use with ACUSON S2000 Ultrasound imaging or fluid flow analysis of the human body as follows:
| | | | | | | | Mode of Operation | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14 | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,6,<br>7,8,10, 11, 14 | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14 | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K# 063085, K072786
Actificital Coginants:
Note 1 For example: breast, testes, thyroid, penis, prostate, etc.
- Ensemble tissue harmonic imaging Note 2
- SieClear multi-view spatial compounding Note 3
- Tissue Equalization Technology Note 4
- Note 5 3-Scape real-time 3D imaging
- Note 6 Cadence contrast agent imaging
- B&W SieScape panoramic imaging Note 7 Note 8
- Power SieScape panoramic imaging Clarify VE vascular enhancement technology Note 10
- Advanced Sieclear spatial compounding Note 11
- eSie™ Touch elasticity imaging Note 14
-
Halstein
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
S2000 510(k) Submission
Page 23 of 1326
{11}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
510 (k) Number (if known):
Device Name: Intended Use:
4P1 Phased Array Transducer for use with ACUSON S2000 Ultrasound imaging or fluid flow analysis of the human body as follows:
| | | | | | | | Mode of Operation | | | | |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|-----------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 | |
| Abdominal | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 | |
| Intraoperative<br>Abdominal | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 | |
| Cardiac | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,6,7,8,10 | |
| Trans-esophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K# 063803, K072786
Additional Comments:
Ensemble tiesus hamonia imaging NOVER 2
Note. 3 SieClear mult - tew spatiat compounding
Tissua Equalization Technology Note. J. ..
Note 5 3-Scape real-time 3D imaging
Note 6 Cadence contrast agent imaging
B&W SieScape panoramic imaging Note 7
Note 8 Power SieScape panoramic imaging
Clarify VE vascular enhancement technology Note 10
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)Prescription Use (Per 21 CFR 801
(Division' Sign-Off)
Division of Reproductive, Abdominal and Radiological Devices
510(k) Number < 0
S2000 510(k) Submission
Page 24 of 1326
{12}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
510 (k) Number (if known):
Device Name:
Intended Use:
6C2 Curved Array Transducer for use with ACUSON S2000 Ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 |
| Abdominal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 14 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously claared by FDA K# 063085, K072733
Additional Comments:
- Ensemble tissue harmonic imaging Note 2
- Note 3 SieClear multi-view spatial compounding
- Note 4 Tissue Equalization Technology
- 3-Scape real-time 3D imaging Note 5
- Note 7 B&W SieScape panoramic imaging
- Note 8 Power SieScape panoramic imaging
- Note 10 Clarify VE vascular enhancement technology
- Advanced Sieclear spatial compounding Note 11
eSie™ Touch elasticity imaging Note 14
Hestem
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
S2000 510(k) Submission
Page 25 of 1326
{13}------------------------------------------------
( . . . . .
#### Diagnostic Ultrasound Indications for Use Form
510 (k) Number (if known):
Device Name: Intended Use:
#### 4C1 Curved Array Transducer for use with ACUSON S2000 Ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation Color Color Amplitude Combined Other Clinical Application B PWD A M CWD Velocity Doppler Doppler (Specify) (Specify) Imaging Ophthalmic P P P P P P Note 2,3,4,5,7,8,10, Fetal BMDC 11 P P P b P P Note2,3,4,5,6,7,8, Abdominal BMDC 10, 11, 14 intraoperative . -Abdominal Intraoperative Neurological Pediatric Small Organ P P P P P P BMDC Neonatal Cephalic Adult Cephalic Cardiac P P P P P P BMDC Trans-esophageal Transrectal Transvaginal Transurethral Intravascular Peripheral vessel P P P P P P BMDC Laparoscopic Musculo-skeletal Conventional Musculo-skeletal Superficial Other (specify)
N = new indication; P = previously cleared by FDA K# 063085, K072786, K032114
Additional Comments:
Ensemble tissue harmonic imaging Note 2
Nota 3 SieClear multi-view spatial compounding
Note 4 Tissue Equalization Technology
Note 5 3-Scape real-time 3D imaging
Note 6 Cadence contrast agent imaging
B&W SieScape panoramic imaging Note 7
Note 8 Power SieScape panoramic imaging
Clarify VE vascular enhancement technology Note 10
Note 11 Advanced Sieclear spatial compounding
Note 14 eSie™ Touch elasticity imaging
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use (Per 21 CFR 801.109) | (Division Sign-Off) |
|--------------------------------------------------------------|---------------------|
| Division of Reproductive, Abdominal and Radiological Devices | |
| 510(k) Number | K090334 |
S2000 510(k) Submission Page 26 of 1326
Page 21 of 46
{14}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
510 (k) Number (if known):
Device Name: Intended Use:
4V1 Phased Array Transducer for use with ACUSON S2000 Ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|----------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10 |
| Abdominal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>14 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K# 063085, K072786
Additional Comments:
t
Note 2 Ensemble tissue harmonic imagina
Note 3 SieClear multi view spatial compounding
Tissue Equalization Technology Note 4
3-Scape real-time 3D imaging Note 5
B&W SieScape panoramic imaging Note 7
Power SieScape panoramic imaging Note 8
Note 10 Clarify VE vascular enhancement technology
Advanced Sieclear spatial compounding Note 11
Note 14 eSie™ Touch elasticity imaging
> (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)
> > Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
S2000 510(k) Submission
Page 27 of 1326
{15}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
510 (k) Number (if known):
Device Name: Intended Use:
10V4 Phased Array Transducer for use with ACUSON S2000 Ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|---------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 |
| Abdominal | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | BMDC | Note 3,4 |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | BMDC | Note 2,3,4,5,7,8,10 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K# 063085, K072786
Additional Comments:
Note 2 Ensemble tissue harmonic imaging
Nord 3 SieClear multi view spetial compounding
Noté 4 Tissue Equalization Technology
Note 5 3-Scape real-time 3D imaging
B&W SieScape panoramic imaging Nota 7
Note 8 Power SieScape panoramic imaging
Clarify VE vascular enhancement technology Note 10
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devic 510(k) Number
S2000 510(k) Submission
Page 28 of 1326
{16}------------------------------------------------
## Diagnostic Ultrasound Indications for Use Form
510 (k) Number (if known):
Device Name: Indications For Use: 1 4L5 SP Linear Array Transducer for use with ACUSON S2000 Diagnostic imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|------------------------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Note 9) | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10 |
| Intraoperative<br>Neurological | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11 |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Note 1) | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14 |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | N | N | N | | N | N | | BMDC | Note 15 |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC | Note2,3,4,5,6<br>,7,8,10, 11,14,15 |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,14 |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication; P = previously cleared by FDA K# 063085, K072
Additional Commondes: s
- Note T Forvexample: breast, hestes, thyroid, penis, prostate,
- Note 2 Ensemble tissue harmonic imaging
- Note 3 SieClear multi-view spatial compounding
- Note 4 Tissue Equalization Technology
- Note 5 3-Scape real-time 3D imaging
- Note 6 Cadence contrast agent imaging
- Note 7 B&W SieScape panoramic imaging
- Power SieScape panoramic imaging Note 8
- For example: vascular, abdominal Note 9
- Note 10 Clarify VE vascular enhancement technology
- Note 11 Advanced Sieclear spatial compounding
- eSie™ Touch elasticity imaging Note 14
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON AND THE RACK IS NEEDED) Concurrence of CDRH, Office of Device Evaluation CODE
(Division Sign-Off)
Prescription Use (Per 21 CFR 80) Dillistion of Reproductive, Ab-
Radiological Devices
510(k) Number
S2000 510(k) Submission
Page 29 of 1326
Page 24 of 46
Note 15 AHP
Note 15 AHF
{17}------------------------------------------------
Diagnostic Ultrasound Indications for Use Form
## Siemens Medical Solutions, Inc. Ultrasound Division
510 (k) Number (if known):
Device Name: Intended Use: 7CF2 Curved array mechanical 3D transducer for use with ACUSON S2000 Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | Other<br>(Specify) | |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|---------------------------|--------------------|--------------------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color Velocity<br>Imaging | | Combined<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11,13 |
| Abdominal | | P | P | P | | P | P | | BMDC | Note 2,3,4,5,7,8,10,<br>11, 13 |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.