← Product Code [HKI](/productcode/HKI) · K090326

# RET CAM II, RETCAM 3 AND RETCAM SHUTTLE AND PORTABLE OPHTHALMIC IMAGING SYSTEMS (K090326)

_Clarity Medical Systems · HKI · Nov 25, 2009 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K090326

## Device Facts

- **Applicant:** Clarity Medical Systems
- **Product Code:** [HKI](/productcode/HKI.md)
- **Decision Date:** Nov 25, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1120
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Pediatric

## Indications for Use

General ophthalmic imaging including retinal, corneal, and external imaging. . Photodocumentation of pediatric ocular diseases including retinopathy of prematurity (ROP). . Screening for Type 2 pre-threshold retinopathy of prematurity (ROP) (zone 1, stage 1 or 2, . without plus disease, or zone 2, stage 3, without plus disease) or treatment-requiring ROP, defined as Type 1 ROP (zone 1, any stage, with plus disease; zone 1, stage 3 without plus disease; or zone 2, stage 2 or 3, with plus disease) or threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2, with plus disease)* in 35-37 week postmenstrual infants.

## Device Story

RetCam Ophthalmic Imaging Systems (RetCam II, 3, Shuttle, Portable) capture digital color images of the eye (retinal, corneal, external). System uses a digital camera handpiece with multiple field-of-view lenses and a standard Halogen light source (optional Fluorescein source for RetCam II/3). Operation controlled via footswitch or keyboard; images displayed on console or laptop monitors. Proprietary software manages image capture, storage, retrieval, and export. Used in clinical settings by trained personnel (e.g., neonatal nurses) to screen for ROP in infants. Clinicians review images to diagnose and document ocular disease, facilitating clinical decision-making regarding treatment for ROP. Benefits include standardized, high-quality wide-field digital imaging for remote or expert review, aiding in the timely detection of sight-threatening conditions.

## Clinical Evidence

Prospective trial (Chiang et al., 2007) evaluated RetCam wide-field digital imaging for ROP screening in 67 infants (31-37 weeks PMA). Images captured by neonatal nurses were interpreted by three retinal specialists and compared to binocular indirect ophthalmoscopy (BIO). Results for Type 2 threshold ROP or worse: Sensitivity was 1.0 (95% CI: 0.868, 1.0) for all graders; Specificity ranged from 0.851 to 0.883 (95% CI: 0.765, 0.933). Image quality was rated 'adequate' or 'possibly adequate' in 93.3% to 100% of cases.

## Technological Characteristics

Digital ophthalmic camera system with handpiece and multiple field-of-view lenses. Illumination via standard Halogen light source (optional Fluorescein source). Image processing and management via proprietary software on integrated computer (RetCam II/3) or laptop (Shuttle/Portable). Connectivity includes image export capabilities. Footswitch/keyboard control for focus, light intensity, and capture. Class II device.

## Regulatory Identification

An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- RetCam II Ophthalmic Imaging System ([K081858](/device/K081858.md))
- RetCam 3 Ophthalmic Imaging System ([K081858](/device/K081858.md))
- RetCam Shuttle Ophthalmic Imaging System ([K081858](/device/K081858.md))
- RetCam Portable Ophthalmic Imaging System ([K083771](/device/K083771.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K090326

## NOV 2 5 2009

| 510(K) Summary                |                                                                                                                                                                                                        | CLARITY MEDICAL SYSTEMS, INC. |  |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--|
| Regulatory Authority:         | Safe Medical Devices Act of 1990, 21 CFR 807.92                                                                                                                                                        |                               |  |
| Device Name:                  | RetCam II Ophthalmic Imaging System<br>RetCam 3 Ophthalmic Imaging System<br>RetCam Shuttle Ophthalmic Imaging System<br>RetCam Portable Ophthalmic Imaging System                                     |                               |  |
| Common Name(s):               | Ophthalmic Imaging System                                                                                                                                                                              |                               |  |
| Classification Name(s):       | Ophthalmic Camera                                                                                                                                                                                      |                               |  |
| Manufacturer:                 | Clarity Medical Systems, Inc.                                                                                                                                                                          |                               |  |
| Reg. Number:                  | 2952489                                                                                                                                                                                                |                               |  |
| Address:                      | 5775 W. Las Positas Blvd.<br>Pleasanton, CA 94588-4084                                                                                                                                                 |                               |  |
| Telephone:                    | (925) 463-7984                                                                                                                                                                                         |                               |  |
| Classification(s):            |                                                                                                                                                                                                        |                               |  |
| Device Class:                 | Class II                                                                                                                                                                                               |                               |  |
| Classification Panel:         | Ophthalmology                                                                                                                                                                                          |                               |  |
| Product Code(s):              | HKI                                                                                                                                                                                                    |                               |  |
| Equivalent Predicate Devices: | RetCam II Ophthalmic Imaging System, K081858<br>RetCam 3 Ophthalmic Imaging System, K081858<br>RetCam Shuttle Ophthalmic Imaging System, K081858<br>RetCam Portable Ophthalmic Imaging System, K083771 |                               |  |

#### Device Description:

RetCam Ophthalmic Imaging Systems utilize a digital camera in a handpiece with multiple field of view lenses to capture color ophthalmic digital images including retinal, corneal, and external images. An on board computer (RetCam II and RetCam 3) or laptop computer (RetCam Shuttle and RetCam Portable) is used to store, view, retrieve, and export the digital ophthalmic images. A standard Halogen light source is used in all RetCam configurations to provide illumination to the eye through the handpiece. An optional Fluorescein light source is also available with the RetCam II and RetCam 3 configurations. Light intensity, camera focus, and image capture are controlled by the use of a footswitch on all configurations and can also be controlled by a keyboard on the RetCam II and RetCam 3 consoles. Console monitors are provided with the RetCam II and RetCam 3 configurations for viewing images. The laptop monitor is used for viewing images with the RetCam Shuttle and RetCam Portable. Proprietary software is installed on the computers to capture, store, view, retrieve, and export ophthalmic images.

{1}------------------------------------------------

## Indication for Use:

- General ophthalmic imaging including retinal, corneal, and external imaging. .
- Photodocumentation of pediatric ocular diseases including retinopathy of prematurity (ROP). .
- Screening for Type 2 pre-threshold retinopathy of prematurity (ROP) (zone 1, stage 1 or 2, . without plus disease, or zone 2, stage 3, without plus disease) or treatment-requiring ROP, defined as Type 1 ROP (zone 1, any stage, with plus disease; zone 1, stage 3 without plus disease; or zone 2, stage 2 or 3, with plus disease) or threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2, with plus disease)* in 35-37 week postmenstrual infants.

*References:

- 1. Cryotherapy for Retinopathy of Prematurity Cooperative Group. Multicenter trial of cryotherapy for retinopathy of prematurity: preliminary results. Archives of Ophthalmology 1988: 106(4):471-479.
- 2. Early Treatment for Retinopathy of Prematurity Cooperative Group. Revised indications for the treatment of retinopathy of prematurity: results of the Early Treatment for Retinopathy of Prematurity Randomized Trial. Archives of Ophthalmology 2003; 121(12):1684-1694.

## Clinical Performance Data:

Chiang, et. al. (Telemedical Retinopathy of Prematurity Diagnosis Accuracy and, Reliability, and Image Quality: Archives of Ophthalmology; 2007;125(11):1531-1538) reported the results of a prospective trial to measure the accuracy, reliability, and image quality of RetCam wide-field digital images to screen for ROP. Eyes from 67 consecutive infants underwent RetCam wide-field digital retinal imaging by trained neonatal nurses using a standard protocol of 3 images per eye and the 130 degree lens. The infants were 31 to 33 weeks and/or 35 to 37 weeks postmenstrual age (PMA).

Images were interpreted by three expert retinal specialist graders who provided a diagnosis and evaluation of image quality. Findings were compared with a reference standard of binocular indirect ophthalmoscopy (BIO) by experienced pediatric ophthalmologists. The target condition (referral warranted disease-the presence of Type 2 pre-threshold or worse ROP) in this study that supports the use of RetCam as an ROP screening tool is Type 2 Pre-threshold ROP (Zone 1, Stage 1 or 2, without plus disease, or Zone 2, Stage 3, without plus disease) or treatment requiring ROP (Zone 1, any stage, with Plus disease; Zone 1, stage 3 without plus disease; or Zone 2, stage 2 or 3 with Plus disease) or threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2. with plus disease) at 35-37 weeks PMA.

| Performance For Type 2 Threshold ROP Or Worse |             |                            |               |                            |
|-----------------------------------------------|-------------|----------------------------|---------------|----------------------------|
| Grader                                        | Sensitivity | ~95% CI for<br>Sensitivity | Specificity   | ~95% CI for<br>Specificity |
| A                                             | 1.0 (26/26) | (0.868, 1.0)               | 0.883 (83/94) | (0.802, 0.933)             |
| B                                             | 1.0 (26/26) | (0.868, 1.0)               | 0.851 (80/94) | (0.765, 0.909)             |
| C                                             | 1.0 (26/26) | (0.868, 1.0)               | 0.851 (80/94) | (0.765, 0.909)             |

#### Sensitivity and Specificity Statistics by Grader

{2}------------------------------------------------

#### Image Quality

Before providing a diagnosis for each image set, the 3 retinal specialist graders assessed the technical quality of the images for "adequate", "possibly adequate" or "inadequate". Each of them found that images taken at 35-37 weeks PMA by trained NICU nurses were technically "adequate" or "possibly adequate" at a rate of 93.3% to 100%.

#### Conclusion

The Clinical Performance data note above was gathered using the RetCam II Ophthalmic Imaging System and supports the revised Indication for Use for the RetCam II Ophthalmic Imaging System, the RetCam 3 Ophthalmic Imaging System, the RetCam Shuttle Ophthalmic Imaging System, and the RetCam Portable Ophthalmic Imaging System.

#### Company Contact:

· Gary A. Seeger Vice President, Quality Assurance and Regulatory Affairs Clarity Medical Systems, Inc.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird with three overlapping wing-like shapes.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

#### NOV 2 5 2009

Clarity Medical Systems, Inc c/o Mr. Gary A. Seeger Vice President, Quality Assurance and Regulatory Affairs 5775 W. Las Positas Blvd. Pleasanton CA 94588

Re: K090326

Trade/Device Name: RetCam II, RetCam 3, RetCam Shuttle, RetCam Portable Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic camera Regulatory Class: Class II Product Code: HKI Dated: September 10, 2009 Received: September 14, 2009

Dear Mr. Seeger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Mr. Gary A. Seeger

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Le R. Ramm, in for

Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure .

{5}------------------------------------------------

## Indications for Use

510(k) Number:

RetCam II Ophthalmic Imaging System Device Name(s): RetCam 3 Ophthalmic Imaging System RetCam Shuttle Ophthalmic Imaging System RetCam Portable Ophthalmic Imaging System

### Indications for Use:

Each and all above listed RetCam Systems are indicated for:

- General ophthalmic imaging including retinal, corneal, and external imaging; .
- Photodocumentation of pediatric ocular diseases including retinopathy of prematurity (ROP). .
- Screening for Type 2 pre-threshold retinopathy of prematurity (ROP) (zone 1, stage 1 or 2, . without plus disease, or zone 2, stage 3, without plus disease) or treatment-requiring ROP, defined as Type 1 ROP (zone 1, any stage, with plus disease; zone 1, stage 3 without plus disease; or zone 2, stage 2 or 3, with plus disease) or threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2, with plus disease)* in 35-37 week postmenstrual infants.

*References:

- 1. Cryotherapy for Retinopathy of Prematurity Cooperative Group. Multicenter trial of cryotherapy for retinopathy of prematurity: preliminary results. Archives of Ophthalmology 1988: 106(4):471-479.
- 2. Early Treatment for Retinopathy of Prematurity Cooperative Group. Revised indications for the treatment of retinopathy of prematurity: results of the Early Treatment for Retinopathy of Prematurity Randomized Trial. Archives of Ophthalmology 2003; 121(12):1684-1694.

X Over the Counter Use Or Prescription Use PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NECESSARY

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Per 21 CFR 801.109)

Sean.

(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

510(k) Number K090326

---

**Source:** [https://fda-staging.innolitics.com/device/K090326](https://fda-staging.innolitics.com/device/K090326)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
