RET CAM II, RETCAM 3 AND RETCAM SHUTTLE AND PORTABLE OPHTHALMIC IMAGING SYSTEMS
K090326 · Clarity Medical Systems · HKI · Nov 25, 2009 · Ophthalmic
Device Facts
Record ID
K090326
Device Name
RET CAM II, RETCAM 3 AND RETCAM SHUTTLE AND PORTABLE OPHTHALMIC IMAGING SYSTEMS
Applicant
Clarity Medical Systems
Product Code
HKI · Ophthalmic
Decision Date
Nov 25, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Attributes
Pediatric
Indications for Use
General ophthalmic imaging including retinal, corneal, and external imaging. . Photodocumentation of pediatric ocular diseases including retinopathy of prematurity (ROP). . Screening for Type 2 pre-threshold retinopathy of prematurity (ROP) (zone 1, stage 1 or 2, . without plus disease, or zone 2, stage 3, without plus disease) or treatment-requiring ROP, defined as Type 1 ROP (zone 1, any stage, with plus disease; zone 1, stage 3 without plus disease; or zone 2, stage 2 or 3, with plus disease) or threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2, with plus disease)* in 35-37 week postmenstrual infants.
Device Story
RetCam Ophthalmic Imaging Systems (RetCam II, 3, Shuttle, Portable) capture digital color images of the eye (retinal, corneal, external). System uses a digital camera handpiece with multiple field-of-view lenses and a standard Halogen light source (optional Fluorescein source for RetCam II/3). Operation controlled via footswitch or keyboard; images displayed on console or laptop monitors. Proprietary software manages image capture, storage, retrieval, and export. Used in clinical settings by trained personnel (e.g., neonatal nurses) to screen for ROP in infants. Clinicians review images to diagnose and document ocular disease, facilitating clinical decision-making regarding treatment for ROP. Benefits include standardized, high-quality wide-field digital imaging for remote or expert review, aiding in the timely detection of sight-threatening conditions.
Clinical Evidence
Prospective trial (Chiang et al., 2007) evaluated RetCam wide-field digital imaging for ROP screening in 67 infants (31-37 weeks PMA). Images captured by neonatal nurses were interpreted by three retinal specialists and compared to binocular indirect ophthalmoscopy (BIO). Results for Type 2 threshold ROP or worse: Sensitivity was 1.0 (95% CI: 0.868, 1.0) for all graders; Specificity ranged from 0.851 to 0.883 (95% CI: 0.765, 0.933). Image quality was rated 'adequate' or 'possibly adequate' in 93.3% to 100% of cases.
Technological Characteristics
Digital ophthalmic camera system with handpiece and multiple field-of-view lenses. Illumination via standard Halogen light source (optional Fluorescein source). Image processing and management via proprietary software on integrated computer (RetCam II/3) or laptop (Shuttle/Portable). Connectivity includes image export capabilities. Footswitch/keyboard control for focus, light intensity, and capture. Class II device.
Indications for Use
Indicated for general ophthalmic imaging (retinal, corneal, external) and photodocumentation of pediatric ocular diseases, specifically retinopathy of prematurity (ROP) in 35-37 week postmenstrual infants.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
RetCam Shuttle Ophthalmic Imaging System (K081858)
RetCam Portable Ophthalmic Imaging System (K083771)
Submission Summary (Full Text)
{0}------------------------------------------------
# K090326
## NOV 2 5 2009
| 510(K) Summary | | CLARITY MEDICAL SYSTEMS, INC. | |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|--|
| Regulatory Authority: | Safe Medical Devices Act of 1990, 21 CFR 807.92 | | |
| Device Name: | RetCam II Ophthalmic Imaging System<br>RetCam 3 Ophthalmic Imaging System<br>RetCam Shuttle Ophthalmic Imaging System<br>RetCam Portable Ophthalmic Imaging System | | |
| Common Name(s): | Ophthalmic Imaging System | | |
| Classification Name(s): | Ophthalmic Camera | | |
| Manufacturer: | Clarity Medical Systems, Inc. | | |
| Reg. Number: | 2952489 | | |
| Address: | 5775 W. Las Positas Blvd.<br>Pleasanton, CA 94588-4084 | | |
| Telephone: | (925) 463-7984 | | |
| Classification(s): | | | |
| Device Class: | Class II | | |
| Classification Panel: | Ophthalmology | | |
| Product Code(s): | HKI | | |
| Equivalent Predicate Devices: | RetCam II Ophthalmic Imaging System, K081858<br>RetCam 3 Ophthalmic Imaging System, K081858<br>RetCam Shuttle Ophthalmic Imaging System, K081858<br>RetCam Portable Ophthalmic Imaging System, K083771 | | |
#### Device Description:
RetCam Ophthalmic Imaging Systems utilize a digital camera in a handpiece with multiple field of view lenses to capture color ophthalmic digital images including retinal, corneal, and external images. An on board computer (RetCam II and RetCam 3) or laptop computer (RetCam Shuttle and RetCam Portable) is used to store, view, retrieve, and export the digital ophthalmic images. A standard Halogen light source is used in all RetCam configurations to provide illumination to the eye through the handpiece. An optional Fluorescein light source is also available with the RetCam II and RetCam 3 configurations. Light intensity, camera focus, and image capture are controlled by the use of a footswitch on all configurations and can also be controlled by a keyboard on the RetCam II and RetCam 3 consoles. Console monitors are provided with the RetCam II and RetCam 3 configurations for viewing images. The laptop monitor is used for viewing images with the RetCam Shuttle and RetCam Portable. Proprietary software is installed on the computers to capture, store, view, retrieve, and export ophthalmic images.
{1}------------------------------------------------
## Indication for Use:
- General ophthalmic imaging including retinal, corneal, and external imaging. .
- Photodocumentation of pediatric ocular diseases including retinopathy of prematurity (ROP). .
- Screening for Type 2 pre-threshold retinopathy of prematurity (ROP) (zone 1, stage 1 or 2, . without plus disease, or zone 2, stage 3, without plus disease) or treatment-requiring ROP, defined as Type 1 ROP (zone 1, any stage, with plus disease; zone 1, stage 3 without plus disease; or zone 2, stage 2 or 3, with plus disease) or threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2, with plus disease)* in 35-37 week postmenstrual infants.
*References:
- 1. Cryotherapy for Retinopathy of Prematurity Cooperative Group. Multicenter trial of cryotherapy for retinopathy of prematurity: preliminary results. Archives of Ophthalmology 1988: 106(4):471-479.
- 2. Early Treatment for Retinopathy of Prematurity Cooperative Group. Revised indications for the treatment of retinopathy of prematurity: results of the Early Treatment for Retinopathy of Prematurity Randomized Trial. Archives of Ophthalmology 2003; 121(12):1684-1694.
## Clinical Performance Data:
Chiang, et. al. (Telemedical Retinopathy of Prematurity Diagnosis Accuracy and, Reliability, and Image Quality: Archives of Ophthalmology; 2007;125(11):1531-1538) reported the results of a prospective trial to measure the accuracy, reliability, and image quality of RetCam wide-field digital images to screen for ROP. Eyes from 67 consecutive infants underwent RetCam wide-field digital retinal imaging by trained neonatal nurses using a standard protocol of 3 images per eye and the 130 degree lens. The infants were 31 to 33 weeks and/or 35 to 37 weeks postmenstrual age (PMA).
Images were interpreted by three expert retinal specialist graders who provided a diagnosis and evaluation of image quality. Findings were compared with a reference standard of binocular indirect ophthalmoscopy (BIO) by experienced pediatric ophthalmologists. The target condition (referral warranted disease-the presence of Type 2 pre-threshold or worse ROP) in this study that supports the use of RetCam as an ROP screening tool is Type 2 Pre-threshold ROP (Zone 1, Stage 1 or 2, without plus disease, or Zone 2, Stage 3, without plus disease) or treatment requiring ROP (Zone 1, any stage, with Plus disease; Zone 1, stage 3 without plus disease; or Zone 2, stage 2 or 3 with Plus disease) or threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2. with plus disease) at 35-37 weeks PMA.
| Performance For Type 2 Threshold ROP Or Worse | | | | |
|-----------------------------------------------|-------------|----------------------------|---------------|----------------------------|
| Grader | Sensitivity | ~95% CI for<br>Sensitivity | Specificity | ~95% CI for<br>Specificity |
| A | 1.0 (26/26) | (0.868, 1.0) | 0.883 (83/94) | (0.802, 0.933) |
| B | 1.0 (26/26) | (0.868, 1.0) | 0.851 (80/94) | (0.765, 0.909) |
| C | 1.0 (26/26) | (0.868, 1.0) | 0.851 (80/94) | (0.765, 0.909) |
#### Sensitivity and Specificity Statistics by Grader
{2}------------------------------------------------
#### Image Quality
Before providing a diagnosis for each image set, the 3 retinal specialist graders assessed the technical quality of the images for "adequate", "possibly adequate" or "inadequate". Each of them found that images taken at 35-37 weeks PMA by trained NICU nurses were technically "adequate" or "possibly adequate" at a rate of 93.3% to 100%.
#### Conclusion
The Clinical Performance data note above was gathered using the RetCam II Ophthalmic Imaging System and supports the revised Indication for Use for the RetCam II Ophthalmic Imaging System, the RetCam 3 Ophthalmic Imaging System, the RetCam Shuttle Ophthalmic Imaging System, and the RetCam Portable Ophthalmic Imaging System.
#### Company Contact:
· Gary A. Seeger Vice President, Quality Assurance and Regulatory Affairs Clarity Medical Systems, Inc.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird with three overlapping wing-like shapes.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
#### NOV 2 5 2009
Clarity Medical Systems, Inc c/o Mr. Gary A. Seeger Vice President, Quality Assurance and Regulatory Affairs 5775 W. Las Positas Blvd. Pleasanton CA 94588
Re: K090326
Trade/Device Name: RetCam II, RetCam 3, RetCam Shuttle, RetCam Portable Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic camera Regulatory Class: Class II Product Code: HKI Dated: September 10, 2009 Received: September 14, 2009
Dear Mr. Seeger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Mr. Gary A. Seeger
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Le R. Ramm, in for
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure .
{5}------------------------------------------------
## Indications for Use
510(k) Number:
RetCam II Ophthalmic Imaging System Device Name(s): RetCam 3 Ophthalmic Imaging System RetCam Shuttle Ophthalmic Imaging System RetCam Portable Ophthalmic Imaging System
### Indications for Use:
Each and all above listed RetCam Systems are indicated for:
- General ophthalmic imaging including retinal, corneal, and external imaging; .
- Photodocumentation of pediatric ocular diseases including retinopathy of prematurity (ROP). .
- Screening for Type 2 pre-threshold retinopathy of prematurity (ROP) (zone 1, stage 1 or 2, . without plus disease, or zone 2, stage 3, without plus disease) or treatment-requiring ROP, defined as Type 1 ROP (zone 1, any stage, with plus disease; zone 1, stage 3 without plus disease; or zone 2, stage 2 or 3, with plus disease) or threshold ROP (at least 5 contiguous or 8 non-contiguous clock hours of stage 3 in zone 1 or 2, with plus disease)* in 35-37 week postmenstrual infants.
*References:
- 1. Cryotherapy for Retinopathy of Prematurity Cooperative Group. Multicenter trial of cryotherapy for retinopathy of prematurity: preliminary results. Archives of Ophthalmology 1988: 106(4):471-479.
- 2. Early Treatment for Retinopathy of Prematurity Cooperative Group. Revised indications for the treatment of retinopathy of prematurity: results of the Early Treatment for Retinopathy of Prematurity Randomized Trial. Archives of Ophthalmology 2003; 121(12):1684-1694.
X Over the Counter Use Or Prescription Use PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NECESSARY
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Per 21 CFR 801.109)
Sean.
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K090326
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.