# EQUIVABONE (K090310)

_Etex Corp. · MQV · Mar 24, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K090310

## Device Facts

- **Applicant:** Etex Corp.
- **Product Code:** [MQV](/productcode/MQV.md)
- **Decision Date:** Mar 24, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3045
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

EquivaBone is intended for use in filling bone voids or defects of the skeletal system (such as the extremities, spine, and the pelvis) that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. It is a bone graft substitute that resorbs and is replaced with new bone during the healing process.

## Device Story

EquivaBone is a biocompatible bone graft substitute; composed of synthetic calcium phosphate, carboxymethyl cellulose (CMC), and human demineralized bone matrix (DBM). Supplied as a single-use kit containing sterile powders and hydration solution. Mixed in the operating room by clinical staff to form a flowable putty; implanted manually or via syringe into bone voids. Hardens at body temperature; resorbs and remodels into new bone during healing. DBM component assayed for osteoinductive potential in athymic nude mouse model. Used by surgeons to fill non-structural skeletal defects.

## Clinical Evidence

Bench testing only. Regression testing conducted per FDA guidance for Resorbable Calcium Salt Bone Void Filler Devices to demonstrate that modifications do not adversely affect safety or performance. DBM osteoinductive potential verified via athymic nude mouse model.

## Technological Characteristics

Materials: Synthetic calcium phosphate, carboxymethyl cellulose (CMC), and human demineralized bone matrix (DBM). Form: Flowable putty formed by mixing sterile powders and hydration solution. Mechanism: Resorbable bone void filler. Hardens at body temperature.

## Regulatory Identification

A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.

## Predicate Devices

- CaP Plus ([K063050](/device/K063050.md))
- CaP Plus ([K080329](/device/K080329.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAR 2 4 2009 .

K0903K Special 510(k) Submission - EquivaBone Osteoinductive Bone Graft Substitute

## 5. 510(k) Summary

### Submitter:

ETEX Corporation 38 Sidney Street Cambridge, MA 02139 Registration No .: 1225112 Owner/Operator No .: 9014709

Contact Person:

Christopher Klaczyk Regulatory Affairs Manager Office: (617) 577-7270 x160 Mobile: (617) 710-8091 Fax: (617) 577-7170 E-Mail: cklaczyk@etexcorp.com

Date Prepared: February 5, 2009

Product Code(s): MBP (21 CFR 888.3045)

Device Class: II (21 CFR 888.3045)

Classification Panel: Orthopacdics

FDA Panel Number: 87

Classification Name: Filler, Bone Void, Osteoinductive (21 CFR 888.3045)

Proprietary Name: EquivaBone Osteoinductive Bone Graft Substitute

Predicate Device(s): CaP Plus (K063050) CaP Plus (K080329)

- Device Description: EquivaBone is a biocompatible bone graft substitute material consisting of synthetic calcium phosphate, carboxymethyl cellulose (CMC) and human demineralized bone matrix (DBM). It is supplied in a single use kit as sterile powders and hydration solution that are mixed together at the time of use in the operating room to form flowable putty which is implanted manually or can be extruded through a syringe. After implantation the product hardens at body temperature and resorbs and remodels during the healing process. Each lot of DBM contained within EquivaBone is assayed for osteoinductive potential in an athymic nude mouse model. This may or may not be predictive of EquivaBone osteoinductivity in humans.

{1}------------------------------------------------

Special 510(k) Submission - EquivaBone Osteoinductive Bone Graft Substitute

Intended Use: EquivaBone is intended for use in filling bone voids or defects of the skeletal system (such as the extremities, spine, and the pelvis) that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. It is a bone graft substitute that resorbs and is replaced with new bone during the healing process.

Materials: Synthetic calcium phosphate and demineralized bone matrix (DBM)

Performance Data: Regression testing consistent with Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA Staff (dated June 2, 2003) has been submitted to show that the proposed changes to the predicate devices do not affect the risk profile of the devices.

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized eagle symbol. The eagle is facing to the right and has three stripes across its body.

MAR 2 4 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

ETEX Corporation % Mr. Christopher Klaczyk Regulatory Affairs Manager 38 Sidney Street Cambridge, Massachusetts 02139

Re: K090310

Trade/Device Name: EquivaBone Osteoinductive Bone Graft Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: II Product Code: MQV, MBP Dated: February 5, 2009 Received: February 6, 2009

#### Dear Mr. Klaczyk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

# Page 2 - Mr. Christopher Klaczyk

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

fr Ai, m.D. fo.

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Special 510(k) Submission -- EquivaBone Osteoinductive Bone Graft Substitute

# 4. Indications For Use

510(k) Number (if known):

Device Name:

EquivaBone Osteoinductive Bone Graft Substitute (originally cleared as CaP Plus)

Indications for Use:

EquivaBone is intended for use in filling bone voids or defects of the skeletal system (such as the extremities, spine, and the pelvis) that are not intrinsic to the stability of the bony structure. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. It is a bone graft substitute that resorbs and is replaced with new bone during the healing process.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW TIIIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Deyige, Evaluation (ODE)

(Division Sign-Off) Division of General. Restorative. and Neurological Devices

1090310 510(k) Number_

ETEX Corporation

Page 4-1

---

**Source:** [https://fda-staging.innolitics.com/device/K090310](https://fda-staging.innolitics.com/device/K090310)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).