The Huntleigh Healthcare Ltd Sonicaid FM820 and FM830 Encore fetal monitors are indicated for use by trained healthcare professionals in non-invasive monitoring of physiological parameters in pregnant adult human females, and fetuses, during the intrapartum and antepartum periods of pregnancy. The devices are intended for use in clinical and hospital-type facilities. They are not intended for use in intensive care units, operating rooms or in transport monitoring applications. Sonicaid FM820E is suitable for use when there is a need to monitor the following physiological parameters: - Single or twin fetal heart rates by means of ultrasound - Fetal or maternal heart rate via ECG - Uterine activity externally or internally sensed - Fetal movement - maternally sensed. Sonicaid FM830E is suitable for use when there is a need to monitor the following physiological parameters: - Single or twin fetal heart rates by means of ultrasound - Fetal or maternal heart rate via ECG - Uterine activity externally or internally sensed - Fetal movement - maternally sensed - Maternal heart rate and oxygen saturation via pulse oximetry - Maternal non-invasive blood pressure
Device Story
Bedside fetal/maternal monitors (FM820E/FM830E) for clinical/hospital use. Inputs: ultrasound transducers (1MHz) for fetal heart rate (FHR); Toco/IUP sensors for uterine activity; ECG electrodes for maternal/fetal heart rate; cabled switch for fetal movement; SpO2 finger clip (FM830E); NIBP cuff (FM830E). Operation: Ultrasound uses Doppler shift detection; Toco/IUP uses strain gauge bridges; ECG uses active reference drive. Signals are digitized and processed to extract vital signs. Output: Color LCD display of parameters/waveforms; thermal printer for hard-copy CTG traces; audio feedback via loudspeaker. Cross-checks maternal HR (from ECG/SpO2) against FHR to prevent false monitoring. Clinicians use output to assess fetal/maternal well-being and guide clinical decisions. FM830E includes NIBP fail-safe microprocessor. Connectivity: RS232/Ethernet to Central Monitoring System.
Clinical Evidence
Bench testing only. Comparative bench testing confirmed performance is at least as good as the predicate. No clinical trials were conducted for substantial equivalence; however, the submission references clinical trials performed by OEM providers for SpO2 sensors. User evaluation trials were proposed to verify effectiveness in clinical settings.
Indicated for pregnant adult females and fetuses during antepartum and intrapartum periods. Used by healthcare professionals in clinical/hospital settings. Contraindicated for use in ICU, OR, transport, patients with cardiac pacemakers, during defibrillation, surgery, or MRI.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
Predicate Devices
Sonicaid FM830 (k002150)
RDT Limited Tempus IC™ Patient Monitor (k082718)
Reference Devices
Intran Plus IUP 400 (k955443)
Safelinc Fetal Spiral Electrode (k904745)
GE Quik-Connect Spiral Electrode (k911657)
Suntech APC series (k051904)
BCI Adult Comfort Clip 3444 (k974697)
Nellcor Durasensor DS-100A (k863784)
Submission Summary (Full Text)
{0}------------------------------------------------
# sonicaid FM800
## Fetal Monitor
# FDA 510(k) Premarket Notification
## Section 5
# 510(k) Summary
Huntleigh Healthcare Ltd.
پ
Page 5.1 FM800E FDA 510(k) Premarket Notification
{1}------------------------------------------------
# 510(k) Summar
Diagnostic Products Division
### 5.1 APPLICANT INFORMATION
| Submitter: | Huntleigh Healthcare Limited<br>Diagnostic Products Division |
|------------|--------------------------------------------------------------------|
| Address: | Unit 35, Portmanmoor Road<br>Cardiff<br>CF24 5HN<br>United Kingdom |
| Telephone: | (44) 2920 485885 |
| Fax: | (44) 2920 492520 |
| Contact: | Graham Booth |
| Prepared: | 26 January 2009 |
#### 5.2 DEVICE INFORMATION
| Proprietary Name: | Sonicaid FM820E and FM830E (E=Encore) |
|-----------------------------|-------------------------------------------|
| Common/Classification Name: | Perinatal Monitoring Device & Accessories |
#### 5.3 IDENTIFICATION OF LEGALLY MARKETED DEVICES TO WHICH EQUIVALENCE IS CLAIMED
| Predicate Device 1: | Sonicaid FM830 (k002150) |
|---------------------|--------------------------------------------------|
| Predicate Device 2: | RDT Limited Tempus IC™ Patient Monitor (k082718) |
#### 5.4 NEW DEVICE DESCRIPTION
#### 5.4.1 Outline
The FM820E and FM830E are fetal/maternal monitors designed for use in clinical and hospital environments during the ante-partum and intra-partum phases of pregnancy.
Both units are designed for use at the bedside and the range includes a wall mounting bracket and a trolley for fixed or transportable use. The units may also be used free-standing on a work surface.
The FM820E and FM830E are powered from the local mains electrical supply.
The FM820E includes the following facilities:
- Monitoring of one or two fetal heart rates via two independent ultrasound transducers. .
- Monitoring of maternal uterine activity either via external (Toco) or internal (IUP) transducers. .
- Monitoring of maternal or fetal heart rate via ECG. �
- . Capture of maternally sensed fetal movements via cabled switch.
- Display of vital signs parameters via colour LCD screen. .
- t Control interface via combination of dedicated and "soft" function buttons in conjunction with on-screen prompts.
- . Chart printout via inbuilt thermal printer.
- Connection to Central Monitoring System possible via RS232 or Ethernet. .
{2}------------------------------------------------
- Audio and visual alerts (user set limits).
The following facilities are provided on the FM830E model in addition to the above:
- . Monitoring of maternal oxygen saturation and heart rate via pulse oximetry sensor.
- Monitoring of maternal Non-Invasive Blood Pressure (NIBP) via inflatable cuff. .
The connections for the additional facilities are incorporated in a "side pod", which extends the area of the front panel. Field upgrade of the FM820E to FM830E specification is not allowed.
#### 5.4.2 Technology & Operating Principles
#### 5.4.2.1 Ultrasound
The ultrasound channels use identical transducers, each incorporating 8 ceramic piezo elements electrically connected in parallel. The operating principle is as follows:
The monitor drives the transducer with an excitation signal, comprising a "burst" of ac electrical current at the nominal operating frequency of the transducer. The piezo elements have a characteristic response of a tuned circuit, leading to highest efficiency at a particular nominal frequency – in this case 1MHz.
The choice of frequency, the geometry and size of the individual elements, and their arrangement within the transducer affects the achievable degree of penetration, beam width and discrimination.
After each burst, the monitor "listens" for signals picked up by the transducer, which now acts as a pickup device, with peak response at the characteristic frequency. These signals will be acoustic reflections of the original transmitted burst, delayed in time according to the distance travelled within the body of the mother, and the fetus. The Doppler Effect causes reflections from moving body parts to reflect signals with a shift in frequency; it is these frequency-shifted reflections that are detected and processed within the unit to determine fetal heart activity. Audible feedback is provided to enable the operator to direct the "beam" of ultrasound so as to provide optimal pickup of the fetal heart activity, recognisable by a characteristic sound. Once the optimum location has been found, the transducer is held in place by an elasticated belt.
The second ultrasound transducer is similarly used in the case of a twin pregnancy.
The signals obtained from the ultrasound channels are digitally processed to extract the fetal heart rate(s).
The ultrasound channels rely on the transducer insulation from the circuitry of the FM800E, which is referenced to mains earth.
#### 5.4.2.2 Toco/IUP
The tocograph transducer is a pressure sensitive device that is strapped to the mother's stomach with an elasticated belt. The central part of the contact area is in the form of an elastic membrane that is free to move with respect to the main body of the device. Contractions of the uterus cause changes in pressure on the membrane and these are translated to electrical signals by means of strain gauges mounted on an internal steel spring.
The IUP sensor is an alternative method of picking up contractions. A pressure-sensitive probe is inserted into the birth canal and directly monitors changes in uterine pressure. The IUP probe is a presterilised single use device.
The tocograph and IUP sensors use similar principles (strain gauge bridges). Being two alternatives for measuring the same parameter, they are mutually exclusive. The two devices use the same input connector and signal processing circuitry. The type of transducer in use (Toco or IUP) is identified by electrical coding of the connections.
{3}------------------------------------------------
The signals derived from the tocograph and IUP sensor are digitised and processes display.
The Toco/IUP channel is electrically isolated from the main circuitry of the FM800E by transformer and optical isolators.
#### 5.4.2.3 ECG
The FM800E has one ECG channel, which can be used for monitoring either maternal or fetal heart rates.
Maternal ECG signals are derived from two contact pads placed on the mother's torso. A "reference" connection is also made - either via a third pad, or an electrode strapped to the leg.
Fetal ECG signals are derived from a scalp clip, which provides the two signal connections. The reference connection is made via an electrode strapped to the mother's leg.
The FM800E uses an "active reference drive", whereby the reference electrode is incorporated within a negative feedback loop and driven with the inverse of the common-mode signal derived from the two signal electrodes. This acts to reduce common-mode signal and improves signal to noise ratio. The ECG channel also has an "active screen (shield) drive", which ensures that the cable screen (shield) is maintained close to the common-mode potential, further negating interference effects. This drive signal is not a patient contact signal.
The ECG channel signal is digitised and processed for printout and display.
Maternal heart rate information is naturally useful for monitoring of the mother. However, the FM800E also cross-checks the maternal HR derived from the ultrasound channel(s) and provides a warning if there is similarity. This helps prevent false monitoring of maternally derived HR signals, which is a major risk in fetal monitoring.
Fetal ECG provides a more reliable signal during the intrapartum stage.
The ECG channel is electrically isolated from the main circuitry of the FM800E by transformer and optical isolators. The type of transducer in use (Maternal or Fetal) is identified by electrical coding of the connections so that from the user perspective, selection is "automatic".
#### 5.4.2.4 Fetal Event Marker
A simple push button switch is provided to capture fetal movements sensed by the mother. Pressing the switch causes a message to be briefly displayed on the screen, and a marker to be printed on the tocograph trace.
The Fetal Event Marker input relies on the switch insulation from the circuitry of the FM800E, which is referenced to mains earth.
#### 5.4.2.5 Maternal SpO2
This facility is incorporated in the FM830E version only (i.e. not the FM820E). It uses a third party module, which provides a data signal to the FM830E's processing system.
In use, a sensor, in the form of a spring-loaded clip, is placed on the mother's finger. The sensor emits two wavelengths of light, one red and one infrared, through body tissue to a photodetector. The principle of operation is on the basis that oxygen-rich blood absorbs less red light than oxygen depleted blood. The maternal oxygen saturation and heart rate are derived and passed to the FM830E's processing system in the form of a serial data stream. This information is displayed on the screen and printed on the chart.
Apart from being used to monitor the well-being of the mother, as with MECG, the FM830E also crosschecks the SpO2 derived MHR with the fetal HR derived from the ultrasound and/or FECG channel(s) and provides a warning in the event of coincidence.
{4}------------------------------------------------
Interface with the patient is in the form of low energy light transmission only, electrical isolation. In addition, this section is isolated from the main electronics by transform optical isolators.
#### 5.4.2.6 Maternal Non-Invasive Blood Pressure
This facility is incorporated in the FM830E version only (i.e. not the FM820E). It uses a third party module, which provides a serial stream to the FM830E's processing system.
In use, an inflatable cuff is placed around the mother's arm. A measurement cycle begins with the cuff being inflated; the system senses both the cuff pressure, and the oscillometric signal generated by arterial pulses. From these, the systolic blood pressures are determined, along with the MHR. The cycle ends with the cuff being deflated. Measurements can be made manually, on an asrequired basis, or the FM830E can be set to take measurements periodically.
The NIBP module incorporates fail-safe mechanisms to prevent over-inflation of the cuff, including a separate safety microprocessor, which separately monitors cuff pressure and can shutdown the pump and open the valves in the event of a fault.
The NIBP information is displayed on the screen and printed on the chart. The MHR information is only valid at the time of the measurement and is displayed for a limited time - but only if no continuous method of obtaining MHR (ECG or SpO2) is available.
Interface with the patient is via an air tube only, which provides total electrical isolation.
#### 5.4.2.7 Display
The FM800E uses a colour TFT Liquid Crystal Display with a resolution of 320x240 dots (1/4 VGA). This is used to display the following information:
- Fetal heart rate (FHR) from ultrasound channels 1 and/or 2. .
- Uterine contractions from the Toco/IUP channel. .
- Fetal or Maternal heart rate (FHR or MHR) from fetal or maternal ECG. ●
- Maternally sensed Fetal Events in the form of an on-screen message. .
- MECG or FECG waveform trace or "baby" graphic. .
- CTG Trace. The unit stores in excess of 24 hours of trace data and this can be recalled and � displayed on screen.
- Maternal oxygen saturation and heart rate (FM830E only). .
- Maternal blood pressure and heart rate (FM830E only). .
In general, information is colour matched to the connectors used for the different channels - e.g. green for ultrasound, pink for toco/IUP, white for ECG.
The display is also used for configuration with the front panel "soft" buttons. Parameters that can be changed include:
- . Alert (alarm) thresholds.
- Time and date. .
- Patient details.
- Printer settings (speed, headings)
- International settings (language, date format etc.). .
- "Baby" graphic and FECG/MECG waveform display.
- "Easinote" defaults (pre-entered chart annotation text).
- Interface settings.
{5}------------------------------------------------
#### 5:4.2.8 Printer
The FM800E incorporates a chart recorder that provides a hard copy printout on continuous "fan fo thermal paper. Long life (25 year) archival paper is available.
Print speed can be set to 1, 2 or 3 cm/s.
The monitor incorporates a number of dedicated front panel buttons and indicators for printer control as follows:
- . Printer on/off (with indicator)
- Paper Advance t
- "Easinotes" annotation
The chart content varies according to the model (FM820E or FM830E), and which transducers and sensors are connected, but can include:
- Patient and hospital information (header)
- Date and time
- Fetal heart rate (FHR) from ultrasound channels 1 and/or 2. t
- Uterine contractions from the Toco/IUP channel. 0
- Fetal or Maternal heart rate (FHR or MHR) from fetal or maternal ECG. .
- Maternally sensed Fetal Events in the form of printed markers.
- "Easinote" annotation text.
- Cross-channel heart rate coincidence warnings.
- Maternal oxygen saturation and heart rate (FM830E only).
- Maternal blood pressure and heart rate (FM830E only).
Impending paper exhaustion is signalled by a coloured band pre-printed on the paper. A sensor detects when the paper is exhausted, or the printer tray is removed; up to ten minutes of trace data are stored in memory, giving adequate time to change the paper pack. The buffered data are then printed, ensuring that there are no gaps in the printout.
#### 5.4.2.9 Interfaces
The FM800E incorporates the following interfaces on the rear panel:
- RS232 connection to Central Monitoring System.
- RS232 connection to external device (future provision).
- Ethernet connection to Central Monitoring System.
The RS232 and Ethernet links are electrically isolated from the main FM800E circuitry.
#### 5.4.2.10 User Alerts
The monitor has the facility to provide audio/visual alerts to draw the operator's attention to any of the following occurrences:
- Ultrasound: Signal loss. Heart rate goes above or below preset limits.
- ECG: Disconnection. Signal loss. Heart rate goes above or below preset limits.
- MSpO2: Signal loss. Maternal oxygen saturation falls below a preset limit. .
- NIBP: Systolic or diastolic blood pressure go outside preset limits.
- Maternal & Fetal heart rate coincidence.
All limits are determined by the user and are not themselves indicative of potential harm or the need for clinical intervention.
{6}------------------------------------------------
#### 5.4.2.11 Audio Output
The FM800E incorporates a built-in loudspeaker. This is used for audible feedback of:
- Ultrasound.
- MECG/FECG. .
- g Sp02.
Each has a characteristic sound, recognisable by clinicians, and helps in optimising the position of sensors and transducers in order to give the best signal.
The loudspeaker is also used to provide user feedback in the form of "beeps" to warn of alarm conditions etc.
There are three dedicated buttons on the front panel associated with the loudspeaker:
- Channel select (US, ECG or Sp02). The selected channel is apparent from the display.
- Volume Increase .
- Volume Decrease
#### 5.5 STATEMENT OF INTENDED USE
Huntleigh Healthcare Ltd Sonicaid FM800 Encore series fetal monitors are indicated for use in noninvasive and invasive monitoring of fetal and maternal vital signs during the intrapartum and antepartum periods.
Sonicaid FM820E provides comprehensive fetal monitoring facilities, offering twin ultrasound fetal heart rate, fetal or maternal ECG channel, external and internal uterine activity monitoring and maternally sensed fetal movements.
Sonicald FM830E additionally provides maternal monitoring with the facility for simultaneous monitoring of maternal pulse oximetry and non-invasive blood pressure.
The device is intended for use by healthcare professionals for monitoring fetal and adult patients in clinical and hospital-type facilities. It is not intensive care units, operating rooms or in transport monitoring applications.
The FM820/830 Encore is not intended for use with patients fitted with cardiac pacemakers, during defibrillation, while undergoing surgery, or while MRI scanning is taking place.
{7}------------------------------------------------
## 5.6 SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPA PREDICATE DEVICES
Diagnostic Products Division
いらい
Note that further predicate device comparison information can be found in section 12 of this submission.
#### 5.6.1 Comparison with Sonicaid FM830 (k002150)
This summary compares the Sonicaid FM830E to the legally marketed predicate device, the Sonicaid FM830 (k002150). The FM820E contains a subset of the FM830E (and is similar to the FM820).
| CHARACTERISTIC | Predicate Device<br>Sonicaid FM830 (k002150) | Submitted Device<br>Sonicaid FM830 Encore | RATIONALE | |
|----------------------------------------------------|------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Human Interface | | | | |
| Display | Monochrome Electro-<br>luminescent 320x240 | Colour LCD 320x240 | Colour display => clearer user<br>interface. Reduced power<br>consumption. | |
| Controls | Membrane type push<br>buttons with dedicated and<br>"soft" keys. | Membrane type push<br>buttons with dedicated and<br>"soft" keys. | Same. | |
| Printer | 128mm Thick Film Thermal<br>Array, 8 dots/mm. | 128mm Thick Film Thermal<br>Array, 8 dots/mm. | Same. | |
| Connectors | | | | |
| Front | Ultrasound<br>1&2 | U'sound 1& 2: Nikolay<br>12-pole, Yellow & Blue | U'sound 1& 2: Nikolay<br>12-pole, Green x2 | New ultrasound transducers not<br>compatible with FM800 |
| | Toco/IUP | Nikolay 12-pole, Pink | Nikolay 12-pole, Pink | Same. Transducers are compatible |
| | FECG | Nikolay 12-pole, Blue (shared<br>with US2). | Nikolay 12-pole, White<br>(shared with MECG) | FECG no longer shared with US2.<br>Electrode system is compatible. |
| | MECG | Nikolay 12-pole, Blue | Nikolay 12-pole, White<br>(shared with FECG) | Connectors keyed differently.<br>MECG electrode system is<br>compatible. |
| | MSpO2 | Odu 12-Pole, Metal | Odu 12-Pole, Plastic | Plastic connector improves patient<br>isolation. |
| | NIBP | Pneumatic | Pneumatic | Same. |
| Rear | RS232 (CRS) | 9-Way "D-type" (F) x1 | 9-Way "D-type" (F) x1 | Same. |
| | RS232<br>(Unallocated) | 9-Way "D-type" (F) x2 | 9-Way "D-type" (F) x1 | Same type, quantity reduced due<br>to redundancy. |
| | RS485 (Axis) | 6-Pole PCC x1 | | Axis system now obsolete. |
| | VGA<br>(Ext. Monitor) | 15-Way HD "D-type" (F) x1 | | External monitor not supported -<br>clinically redundant. |
| Side | Ethernet | | RJ45 x1 | Improves CRS connection options. |
| | Event Marker | 6.35mm Jack x2 | 6.35mm Jack x2 | Same |
| Power Requirements | | | | |
| Voltage | Auto-switches in two ranges:<br>100-120 & 200-240. | 90-240Vac. | Increased supply flexibility. | |
| Consumption | 100VA max. | 100VA max. | Same. | |
| Ultrasound | | | | |
| | Transducers - Physical | 74x27mm (Dia x Height)<br>7-element piezo | 79x27mm (Dia x Height)<br>8-element piezo | Similar. |
| | Transducers -<br>Frequency | 1.5MHz (US1) and 2MHz<br>(US2) | 1MHz (both channels) | Improved pickup & discrimination<br>in conjunction with new software.<br>Product rationalisation. |
| | | | | |
| CHARACTERISTIC | Predicate Device<br>Sonicaid FM830 (k002150) | Submitted Device<br>Sonicaid FM830 Encore | RATIONALE | |
| Signal Processing | Analogue/Digital | Analogue/Digital | Similar (new DSP algorithms<br>introduced for improved detection<br>over a wider range of conditions). | |
| Measured Acoustic<br>Output (Ispta)<br>FDA Track 1 | 1.5MHz: 11mW cm-2<br>2.0MHz: 7.3mW cm-2 | 2.1mW cm-2 | Superior signal processing has<br>enabled a reduction in acoustic<br>power output. | |
| Audio Feedback | Analogue | Digital | At 1MHz, raw analogue Doppler<br>frequency is too low for good<br>audibility. Improved artefact<br>rejection. | |
| Safety | Insulated transducer | Insulated transducer | Same | |
| Tocograph | | | | |
| Transducers - Physical | Floating piston type with<br>adhesive membrane. | New moulding, similar to u/s<br>with overmoulded elastomer<br>faceplate. | Lower maintenance. Improved<br>patient comfort. Easier to clean,<br>reduced chance of infection. | |
| Transducers -<br>Functional | Double strain gauge/spring<br>assembly. | Double strain gauge/spring<br>assembly. | Same. | |
| Signal Processing | Analogue/digital. | Analogue/digital. improved<br>processing algorithms. | Product improvement | |
| Safety | Insulated transducer +<br>isolation Barrier. | Insulated transducer +<br>isolation Barrier. | Same. | |
| IUP | | | | |
| Compatible Probes | Intran Plus IUP 400<br>(k955443) | Intran Plus IUP 400<br>(k955443) | Same. | |
| Safety | Insulated transducer +<br>isolation Barrier. | Insulated transducer +<br>isolation Barrier. | Same. | |
| MECG/FECG | | | | |
| Compatible Electrodes<br>- MECG | Not known | Skintact F-TB (k040249) | Comparison not possible. | |
| Compatible Electrodes<br>- FECG | Graphic Controls (k905830) | Safelinc Fetal Spiral Electrode<br>(k904745)<br>GE Quik-Connect Spiral<br>Electrode (k911657) | Options are legally marketed and<br>are validated for use with the<br>submitted device. | |
| Signal Processing | Analogue/digital. | Analogue/digital (enhanced) | Performance improvement under<br>sub-optimal conditions. | |
| Safety | CF level protection. | CF level protection. | Same (BF is minimum<br>requirement). | |
| NIBP | | | | |
| OEM Technology | Suntech Alta | Suntech Advantage | Alta modules no longer produced.<br>Superseded by Advantage. | |
| Compatible Cuffs | Not known | Suntech APC series<br>(k051904) | Comparison not possible. | |
| Maternal Temp. | | | | |
| Function | Included on FM830 | Not included | Not clinically beneficial. | |
| Accessories | | | | |
| Wall Bracket | Not known | Huntleigh Healthcare<br>ACC223 | Comparison not possible. | |
| Trolley | Not known | Huntleigh Healthcare | Comparison not possible. | |
| CHARACTERISTIC | Predicate Device<br>Sonicaid FM830 (k002150) | Submitted Device<br>Sonicaid FM830 Encore | RATIONALE | |
| Software | | | | |
| CTG Trace | Not implemented | Trace data stored in memory<br>for review on screen - up to<br>72 hours. | Product improvement. | |
Huntleigh Healthcare Ltd.
Page 5.8 FM800E FDA 510(k) Premarket Notification
{8}------------------------------------------------
# Diagnostic Products Division Iन UNILLIG
·
ット..
:
、「
:
{9}------------------------------------------------
### 5.6.1 Comparison with RDT Tempus IC™ Patient Monitor (k082718)
The OEM technology for the Maternal SpO2 function differs in the Sonicaid FM830E from the Sonicaid FM830 (k002150). In respect of this function, the submitter claims Substantial Equivalence to another predicate device – the Tempus IC™ Patient Monitor (k082718) manufactured by Remote Diagnostic Technologies Limited of The Old Coach House, The Avenue, Farleigh Wallop, Hampshire, RG25 2HT, United Kingdom.
| CHARACTERISTIC | Predicate Device<br>Tempus ICTM (k082718) | Submitted Device<br>Sonicaid FM830 Encore | RATIONALE |
|--------------------|-------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| MSPO2 | | | |
| OEM Technology | BCI (Smiths Medical)<br>WW3711 "B2" | BCI (Smiths Medical)<br>WW3711 "B2" | Same. |
| Compatible Sensors | Proprietary. | BCI Adult Comfort Clip 3444<br>(k974697)<br>Nellcor Durasensor DS-100A<br>(k863784) | Use of BCI 3444 and Nellcor<br>DS-100A is supported by BCI<br>clinical trials (see section 12 for<br>further details). |
{10}------------------------------------------------
## 5.7 TESTS CARRIED OUT IN SUPPORT OF A DETERMINATION OF SU EQUIVALENCE
Diagnostic Products Division
#### 5.7.1 Non-Clinical Tests
Electrical safety is to be confirmed by testing to the requirements of IEC 60601-1: 1988 and EN 60601-1-1: 2000.
Electromagnetic compatibility (EMC) is to be confirmed by testing to the requirements of IEC 60601-1-2: 2007.
Substantial equivalence to the FM800 predicate device (k002150) is confirmed by comparative bench testing. See document 7515101 - FM800E Comparative Bench Test Summary, which can be found in section 18 of this submission.
No specific tests have been carried out by Huntleigh Healthcare to support substantial equivalence to the Tempus IC 100 Patient Monitor (k082718). However, this uses the same OEM module (BCI WW3711) as used in the FM830E, and the aim of the bench tests carried out in this regard has been to verify that the raw data output from the OEM module is correctly displayed and printed on the FM830. See document 7514502 – FM800E MsP02 Verification Protocol & Results, which can be found in section 18 of this submission.
#### 5.7.2 Clinical Tests
No specific clinical tests were carried out to determine substantial equivalence. However, this submission references clinical trials carried out by the OEM providers of the SpO2 technology to validate the use of the specified finger sensors.
User evaluation trials are to be carried out to verify the effectiveness of the FM830E in clinical situations. This submission includes a copy of the proposed User Evaluation Trial protocol.
See section 20 of this submission for further details.
#### 5.8 CONCLUSIONS
#### 5.8.1 Safety
The Electrical Safety and EMC tests to be carried out on the Sonicaid FM830E will confirm that it is at least as safe as the Sonicaid FM830 (k002150) predicate device.
#### 5.8.2 Effectiveness
The FM830E has been designed to provide improved performance, clearer presentation of information and easier use compared with the FM830 predicate device (k002150). This is to be confirmed by User Evaluation trials in clinical situations in the home market.
#### 5.8.3 Performance
Comparative bench tests confirm that the performance of the Sonicaid FM830E is at least as good as the Sonicaid FM830 (k002150) predicate device.
The overall conclusion is that the FM820/830E is as safe, as effective, and performs as well as, or better than, the FM820/830 (k002150) predicate device.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three abstract shapes that resemble a human figure.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## JUL 15 2009
Mr. Graham Booth Technical Manager - Obstetric Products Huntleigh Healthcare Limited Diagnostic Products Division 35 Portmanmoor Road Cardiff, CF24 5HN UNITED KINGDOM
K090285 Re:
> Trade/Device Name: Sonicaid FM820 & FM830 Encore Series Fetal Monitors Regulation Number: 21 CFR §884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: June 9, 2009 Received: June 11, 2009
Dear Mr. Booth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical
{12}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Mannfacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Jadine M. Morris
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Eyaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
# Huntleigh Healthcare Limited | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
## Indications For Use
510(k) Number:
K090285
Device Name:
Sonicaid FM820/830 Encore
Indications for Use:
The Huntleigh Healthcare Ltd Sonicaid FM820 and FM830 Encore fetal monitors are indicated for use by trained healthcare professionals in non-invasive monitoring of physiological parameters in pregnant adult human females, and fetuses, during the intrapartum and antepartum periods of pregnancy. The devices are intended for use in clinical and hospital-type facilities. They are not intended for use in intensive care units, operating rooms or in transport monitoring applications.
Sonicaid FM820E is suitable for use when there is a need to monitor the following physiological parameters:
- . Single or twin fetal heart rates by means of ultrasound
- ◆ Fetal or maternal heart rate via ECG
- Uterine activity externally or internally sensed .
- . Fetal movement - maternally sensed.
Sonicaid FM830E is suitable for use when there is a need to monitor the following physiological parameters:
- . Single or twin fetal heart rates by means of ultrasound
- . Fetal or maternal heart rate via ECG
- Uterine activity externally or internally sensed .
- Fetal movement - maternally sensed.
- . Maternal heart rate and oxygen saturation via pulse oximetry
- . Maternal non-invasive blood pressure
Image /page/13/Figure/20 description: The image shows two options for drug use: prescription use and over-the-counter use. The prescription use option is marked with "YES" and includes the text "(Part 21 CFR 801 Subpart D)". The over-the-counter use option is marked with "NO" and includes the text "(Part 21 CFR 801 Subpart C)". The text "AND/OR" is placed between the two options.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Hulu Rumm
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices K090285
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.