HYDROCOAT MESH

K090271 · Promethean Surgical Devices, Inc. · FTL · Jan 14, 2010 · General, Plastic Surgery

Device Facts

Record IDK090271
Device NameHYDROCOAT MESH
ApplicantPromethean Surgical Devices, Inc.
Product CodeFTL · General, Plastic Surgery
Decision DateJan 14, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The HydroCoat Mesh is indicated for use in general surgical procedures for reinforcement of soft tissue where weakness exists. The HydroCoat Mesh may be used for the repair of abdominal wall hernia, including inguinal, femoral, and incisional hernia, uterovaginal prolapse, and other fascial deficiencies that require support material.

Device Story

HydroCoat Mesh is a surgical implant for soft tissue reinforcement. Device consists of knitted monofilament polypropylene fiber coated with polyether urethane urea. Provided as flat, porous sheets in various shapes (rectangles, ovals, circles, etc.) and sizes; can be cut by surgeons to fit specific anatomical needs. Used in general surgical procedures to repair hernias and fascial deficiencies. Implanted by surgeons to provide mechanical support to weakened tissue. Benefits include elasticity in multiple directions and structural reinforcement of soft tissue defects.

Clinical Evidence

Bench testing only. Mechanical testing performed to compare strength characteristics against predicate devices.

Technological Characteristics

Polypropylene monofilament fiber coated with polyether urethane urea. Knitted mesh structure with 3mm² to 5mm² micropores. Thickness: 0.38mm to 0.76mm. Areal density: 37.3 g/m². Available in various pre-cut shapes and sizes. Non-absorbable polymeric surgical mesh.

Indications for Use

Indicated for patients requiring soft tissue reinforcement for abdominal wall hernias (inguinal, femoral, incisional), uterovaginal prolapse, and other fascial deficiencies.

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ - - 1. - Promethean HydroCoat Mesh 090271 JAN 1 4 2010 1 83 ## ADMINISTRATIVE INFORMATION Manufacturer Name: Promethean Surgical Devices 111 Roberts Street. Suite G2, East Hartford, CT 06108 Official Contact: Kenneth K. Kleinhenz Regulatory Affairs Telephone (858) 458-0900 Fax (858) 458-0994 DEVICE NAME Classification Name: Trade/Proprietary Name: HydroCoat Mesh Surgical Mesh, Polymeric ESTABLISHMENT REGISTRATION NUMBER This is our first device applications to FDA. We will register and pay the fee within 30 days of FDA's approval/clearance of this device. ## DEVICE CLASSIFICATION AND PRODUCT CODE As shown in 21CFR 878.3300, Surgical Mesh are polymeric screens intended to be implanted to reinforce soft tissues. These devices are classified as Class II. Surgical Mesh have been assigned Product Code FTL. #### INTENDED USE The HydroCoat Mesh is indicated for use in general surgical procedures for reinforcement of soft tissue where weakness exists. The HydroCoat Mesh may be used for the repair of abdominal wall hernia, including inguinal, femoral, and incisional hernia, uterovaginal prolapse, and other fascial deficiencies that require support material. {1}------------------------------------------------ ### DEVICE DESCRIPTION The HydroCoat Mesh is a polypropylene surgical mesh coated with polyurethane and intended for trauma and reconstructive surgical procedures involving soft tissues. The HydroCoat Mesh is fabricated from a knitted monofilament, medical-grade polypropylene fiber that links each porous window of the mesh, allowing for elasticity in both directions. The HydroCoat Mesh is available in various shapes and sizes for use in reinforcing soft tissues where weakness exists. The HydroCoat Mesh can be cut to most any shape or size for specific soft tissue reinforcement needs. The Promethean Surgical Devices HydroCoat Mesh is provided in various shapes such as rectangles, ovals, ellipticals and circles and will be provided in other shapes and sizes as needed for particular soft tissue reinforcement applications. The Promethean Surgical Devices HydroCoat Mesh is provided in sheets of various shapes and sizes as outlined in the table below: | Rectangular | Square | Elliptical | Oval | Round | Keyhole. | |-------------|-----------|------------|----------|-------------|-----------| | 12" x 12" | 12" x 12" | 12" x 10" | 12" x 9" | 8" diameter | 1.5" x 8" | | 12" x 10" | 10" x 10" | 10" x 8" | 10" x 7" | 6" diameter | 3" x 7" | | 10" x 8" | 8" x 8" | 8" x 6" | 8" x 5" | 5" diameter | | | 8" x 6" | 6" x 6" | 7" x 5" | 7" x 4" | 4" diameter | | | 7" x 5" | 4" x 4" | 6" x 4" | 6" x 3" | 3" diameter | | | 6" x 4" | | 5" x 3" | 5" x 2" | | | | 5" x 3" | | 4" x 2" | 4" x 1" | | | | 4" x 2" | | | | | | The thickness of the Promethean Surgical Devices HydroCoat Mesh ranges from 0.015" to 0.030" (0.38mm to 0.76mm) according to the region to be treated. The Promethean Surgical Devices HydroCoat Mesh is provided in flat sheets that contain micropores that range in size from 3mm2-to 5mm. The HydroCoat Mesh has an overall areal density of 1.1001 oz/yd2 (37.3 g/m2). #### Material Composition The Promethean Surgical Devices HydroCoat Mesh is fabricated from a polypropylene yarn that is coated with a polyether urethane urea. ### In Vitro Testing Mechanical testing was performed on the Promethean Surgical Devices HydroCoat Mesh which determined the HydroCoat Mesh to be substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions. {2}------------------------------------------------ Page 3 of 3 3 of 3 8090271 ## EQUIVALENCE TO MARKETED PRODUCT The Promethean Surgical Devices HydroCoat Mesh shares indications and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to pre-amendment devices: Mentor Suspend Sling (K980483). Artimplant AB SportMesh / Artelon Tissue Reinforcement (K071887), Artimplant AB SportMesh (K052830), the Mpathy Medical Minimesh (K041632), and the Ethicon Mersilene Mesh (pre-amendment); Class II medical devices that were cleared for marketing in the United States under K980483, K071887, K052830, K041632, and preamendment respectively. #### Indications for Use The HydroCoat Mesh and the predicate devices are substantially equivalent with respect to their indications for use as they are all indicated for the same general surgerv procedures requiring the reinforcement of soft tissues in the same peritoneal anatomy. #### Design and Materials The design of the Promethean Surgical Devices HydroCoat Mesh and the Mentor Suspend Sling (K980483), Artimplant AB SportMesh / Artelon Tissue Reinforcement (K071887), Artimplant AB SportMesh (K052830), the Mpathy Medical Minimesh (K041632), and the Ethicon Mersilene Mesh predicate devices are substantially equivalent as they are all thin, flat sheets of various shapes and sizes that can be cut to shape with surgical scissors and are all fabricated from a polymeric fiber that is knitted into a pattern that creates a meshed network of yarn with a uniform pattern of holes. The Promethean Surgical Devices HydroCoat Mesh, the Mpathy Medical Minimesh (K041632), and the Ethicon Mersilene Mesh devices are substantially equivalent as they share a common design feature of being fabricated from a polypropylene fiber. The Promethean Surgical Devices HydroCoat Mesh and the Mentor Suspend Sling (K980483), Artimplant AB SportMesh / Artelon Tissue Reinforcement (K071887), and Artimplant AB SportMesh (K052830) predicate devices are substantially equivalent as they all share the design feature of utilizing a polyurethane urea materials. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of a human figure, represented by three curved lines, symbolizing health and well-being. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 Promethean Surgical Devices, Inc. % Mr. Kenneth K. Kleinhenz Regulatory Affairs 111 Roberts Street, Suite G2 East Hartford, Connecticut 06108 JAN 1 4 2010 Re: K090271 Trade/Device Name: HydroCoat Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: Class II Product Code: FTL Dated: January 05, 2010 Received: January 06, 2010 Dear Mr. Kleinhenz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Mr. Kenneth K. Kleinhenz forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor.You/Industry/default.htm. Sincerely vours Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ K090271 pg 1 of 1 # Indications for Use Re: K090271 Device Name: HydroCoat Mesh Indications for Use: The HydroCoat Mesh is indicated for use in general surgical procedures for reinforcement of soft tissue where weakness exists. The HydroCoat Mesh may be used for the repair of abdominal wall hernia, including inguinal, femoral, and incisional hernia, uterovaginal prolapse and other fascial deficiencies that require support material. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel Keane (Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices Page 1 of 1 510(k) Number K090271
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...