FORTEX PEDICLE SCREW SYSTEM

K090224 · X-Spine Systems, Inc. · MNH · Jul 30, 2009 · Orthopedic

Device Facts

Record IDK090224
Device NameFORTEX PEDICLE SCREW SYSTEM
ApplicantX-Spine Systems, Inc.
Product CodeMNH · Orthopedic
Decision DateJul 30, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Fortex Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The Fortex Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

Device Story

The Fortex Pedicle Screw System is a spinal fixation device used by surgeons to provide immobilization and stabilization of spinal segments as an adjunct to fusion. The system consists of titanium alloy rods, pedicle screws (cannulated or non-cannulated), and optional cross bars. The device is implanted during surgery to stabilize the thoracic, lumbar, and sacral spine. It functions as a mechanical construct to support the spine while autogenous bone graft facilitates fusion. Once solid fusion is achieved, the implants are intended for removal. The system provides structural support to treat various spinal pathologies, including deformities and instabilities, thereby aiding in patient recovery and spinal alignment.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Constructed from titanium alloy conforming to ASTM F136. System includes 5.5 mm diameter rods, self-tapping pedicle screws (4.75 mm to 8.25 mm diameters), and cross bar assemblies. Configurations include cannulated and non-cannulated screws. Mechanical fixation device; no energy source or software.

Indications for Use

Indicated for skeletally mature patients requiring spinal fusion for severe spondylolisthesis (Grade 3-4, L5-S1) or acute/chronic instabilities/deformities (degenerative spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, spinal tumor, pseudarthrosis) of the thoracic, lumbar, and sacral spine.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K090224 ## Special 510(k): Device Modification Fortex Pedicle Screw System #### 510(k) Summary ### ADMINISTRATIVE INFORMATION JUL 80 2009 X-spine Systems, Inc. Manufacturer Name: 452 Alexandersville Rd. Miamisburg, OH 45342 > Telephone (937) 847-8400 FAX (937) 847-8410 Official Contact: David Kirschman, MD Chief Medical Officer DEVICE NAME Classification Names: Orthosis. Spondylolisthesis Spinal Fixation Orthosis, Pedicle Spinal Fixation Fortex Pedicle Screw System Trade/Proprietary Name: Pedicle Screw Spinal System Common Name: ESTABLISHMENT REGISTRATION NUMBER X-spine Systems, Inc. has submitted an Establishment Registration to FDA. The Establishment Registration number is 3005031160. The owner/operator number for X-spine Systems, Inc. is 9063903. DEVICE CLASSIFICATION FDA has classified pedicle screw spinal systems as Class II devices (21 CFR 888.3070). The product code for Orthosis, Spondylolisthesis Spinal Fixation is (MNH). The product code for Orthosis, Spinal Pedicle Fixation is (MNI). These device classifications are reviewed by the Orthopedic Devices Branch. #### INTENDED USE The Fortex Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. The Fortex Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the Page 10 of 41 Prg 1 of 2. {1}------------------------------------------------ ### Special 510(k): Device Modification following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). #### DEVICE DESCRIPTION #### Fortex Rod Fortex rods are 5.5 mm diameter solid cylinders with spherically rounded ends, provided in 40 mm, 60 mm, 70 mm, 80 mm, 120 mm, 120 mm, 160 mm, 180 mm, 200 mm, and 300 mm lengths. #### Fortex Pedicle Screw Assembly Each Fortex pedicle screw assembly consists of a pedicle screw, yoke, and screw cap. Selftapping pedicle screw assemblies are provided in diameters of 4.75mm, 5.5 mm, 6.5 mm, 7.5 mm and 8.25mm. All screw assemblies are provided in lengths of 30 mm, 35 mm, 45 mm, 50 mm, and 55 mm. Screws are provided in canulated and non-canulated configurations. #### Fortex Cross Bar Assembly The Fortex cross bar assembly is an optional component and can be used for additional stabilization. Cross bar assemblies are available in lengths from 25 mm to 81 mm. #### Material composition The rods, pedicle screws and cross bars of the Fortex Pedicle Screw System are made of titanium alloy conforming to ASTM F136. ## STATEMENT OF TECHNOLOGICAL COMPARISON X-spine Systems, Inc. has submitted information to demonstrate that, for the purposes of FDA's regulation of medical devices, the Fortex Pedicle Screw System is substantially equivalent in indications and design principles to the predicate Capless Pedicle Screw System (K052847) and Capless LI Pedicle Screw System (K072282). Page 11 of 41 Page 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle-like symbol with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the symbol. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 X-Spine Systems, Inc. c/o Dr. David Kirschman, M.D. Chief Medical Officer 452 Alexandersville Rd Miamisburg, OH 45342 JUL 3 0 2009 Re: K090224 Trade/Device Name: Fortex™ Pedicle Screw System Regulation Number: 21 CFR 888. 3070 Regulation Name: Pedicle screw spinal system . Regulatory Class: Class II Product Code: MNH, MNI Dated: June 26, 2009 Received: June 30, 2009 Dear Dr. Kirschman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Dr. David Kirschman forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Sincerely yours, Mark A. Mellersen Mark M. Melkerson Director Division of Surgical, Orthopaedic. and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Special 510(k): Device Modification Fortex Pedicle Screw System #### Indications for Use ## 510(k) Number (if known): Ki9 D224 Device Name: Fortex™ Pedicle Screw System Indications for Use: The Fortex Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion. In addition, the Fortex Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ex-2 (exT for MLM) (Division Sign-Off) Division of S. . greal, Orthopedic, and Restorative Devices 510(k) Number K090224 Page 8 of 41
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...