← Product Code [ODG](/productcode/ODG) · K090196

# EG-3670URK (K090196)

_Pentax Medical Company · ODG · May 19, 2009 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K090196

## Device Facts

- **Applicant:** Pentax Medical Company
- **Product Code:** [ODG](/productcode/ODG.md)
- **Decision Date:** May 19, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The EG-3670URK, Ultrasound Video Gastroscope, is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Track including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.

## Device Story

EG-3670URK is an ultrasound video gastroscope; features flexible insertion tube, control body, and umbilical connectors for video processor and ultrasound scanner. Inputs: optical image data via CCD; ultrasonic pulses/reflections via radial array ultrasound transducer. Operation: endoscope illuminates body cavity; CCD captures optical images; transducer delivers/receives ultrasonic pulses. Outputs: endoscopic video and ultrasound images displayed via external processor/scanner. Used in clinical settings by physicians for visualization and therapeutic access (e.g., biopsy via working channel). Benefits: enables combined optical/ultrasonic diagnostic and therapeutic procedures in upper GI tract.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including system performance testing and acoustic power data.

## Technological Characteristics

Flexible gastroscope with radial array ultrasound transducer and CCD imaging sensor. Includes working channel for accessory introduction. Connectivity: dedicated umbilical connectors for video processor and ultrasound scanner. Immersible design (excluding ultrasound connector).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- EG-3670URK, Ultrasound Video Gastroscope ([K061068](/device/K061068.md))
- Hi Vision 900 Diagnostic Ultrasound Scanner ([K063518](/device/K063518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Summary

K 090196

Submitter Information:

Pentax Medical Company, A Division of Pentax America, Inc. 102 Chestnut Ridge Road Montvale, New Jersey 07645-1856

MAY 19 2009

## Name of Device

| Trade Name:          | EG-3670URK, Ultrasound Video Gastroscope                                       |
|----------------------|--------------------------------------------------------------------------------|
| Classification Name: | Endoscopic Ultrasound System, Gastro-Urology (780DG) (876.1500); track3, tier2 |

## Predicated Device(s) Information

| Model, Description                          | Manufacturer       | PMN #   |
|---------------------------------------------|--------------------|---------|
| EG-3670URK, Ultrasound Video Gastroscope    | Pentax Corporation | K061068 |
| Hi Vision 900 Diagnostic Ultrasound Scanner | Hitachi            | K063518 |

## Device Description

The EG-3670URK, Ultrasound Video Gastroscope, must be used with a Pentax Video processor (a software controlled device) and must be used with an Ultrasound Scanner (a software controlled device). The endoscope has a flexible insertion tube, a control body, PVE umbilical connector, and ultrasound scanner umbilical connector. The PVE connector will be attached to the Video Processor and has connections for illumination, video signals, air/ water/ and suction. The ultrasound umbilical connector will be attached to the ultrasound scanner unit. The control body includes controls for up/ down/ left/ right angulation, air/ water delivery, suction selection control, balloon insufflation/ evacuation, and an accessory inlet port. The endoscope contains light carrying bundles to illuminate the body cavity, a charge couple device (CCD) to collect endoscopic image data, and a radial array ultrasound transducer to collect ultrasonic image data. The instrument contains a working channel through which biopsy devices, or other devices, may be introduced. The video processor contains a lamp that light and is focused at the PVE connector light guide prong. The endoscope light carrying bundles present the light to the body cavity and the CCD collects endoscopic image data. Image data and other screen display information are formatted and presented to the video outputs of the video processor for display. The ultrasound transducer delivers ultrasonic pulses, reflections of the pulses are received and the signals are passed to the ultrasound scanner for processing and display. The instrument is immersable (with the use of supplied cleaning accessories) except for the ultrasound scanner connector (as described in the endoscope operator manual cleaning instructions).

### Intended Use:

The EG-3670URK, Ultrasound Video Gastroscope, is intended to provide optical visualization of, ultrasonic visualization of, and therapeutic access to, the Upper Gastrointestinal Track including but not restricted to the organs, tissues, and subsystems: Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The instrument is introduced per orally when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.

### Comparison to Predicated Device(s)

The submission for substantial equivalence was made to allow the use of the 510k cleared endoscope, EG-3670URK, with the 510k cleared ultrasound scanner, Hi Vision 900. The submission included mechanical drawings, material lists, labeling, the identification of standard set components and optional accessories, tables to summarize the comparisons to the predicated device(s), and system performance testing, and acoustic power data. The submission for substantial equivalence is not based on an assessment of clinical perfermance data.

Prepared by: Paul Silva

Signature: Paul Schulz

Date: 01.24.2009

Control Number: EG-3670URK.HIVision900

Page I of 1

Revision: c

{1}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

#### Public Health Service

#### MAY 19 2009

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Paul Silva Director of Regulatory Affairs PENTAX Medical Company 102 Chestnut Ridge Road MONTVALE NJ 07645-1856

Re: K090196

Trade/Device Name: EG-3670URK Ultrasound Video Gastroscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODG and ITX Dated: January 24, 2009 Received: February 4, 2009

Dear Mr. Silva:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration ..

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology)                      | (240) 276-0120 |
| Other          |                                  | (240) 276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

| 510(k) Number (if known): | K090196    |
|---------------------------|------------|
| Device Name:              | EG-3670URK |

EG-3670URK, Ultrasound Video Gastroscope

## Endoscope Intended Use Statement

Ellussely Intended Use Battoscope, is intended to provide optical visualization of, ultrasonic I is 10 10 10 10 10 10 10 12 12 12 12 12 12 12 12 12 12 11 12 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 Visanzation of, and subsystems: Esophagus, Stomach, Duodenum, Small Bowel, and underlying areas. The organs, usused, and unory counter indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.

# Diagnostic Ultrasound Indications For Use Statement

| System:                    | Hi Vision 900                                                                                   |   |   |     |     |               |                   |
|----------------------------|-------------------------------------------------------------------------------------------------|---|---|-----|-----|---------------|-------------------|
| Probe:                     | EG-3670URK                                                                                      |   |   |     |     |               |                   |
|                            | Intended Use: Diagnostic Ultrasound imaging or fluid flow analysis of the human body as follows |   |   |     |     |               |                   |
| Clinical Application       | Mode of Operation                                                                               |   |   |     |     |               |                   |
| General<br>(Track I only)  | Specific<br>(Track I & III)                                                                     | B | M | PWD | CWD | Color Doppler | Amplitude Doppler |
| Ophthalmic                 |                                                                                                 |   |   |     |     |               |                   |
| Fetal Imaging<br>and other | Fetal                                                                                           |   |   |     |     |               |                   |
|                            | Abdominal                                                                                       |   |   |     |     |               |                   |
|                            | Intra-operative (Spec.)                                                                         |   |   |     |     |               |                   |
|                            | Intra-operative (Neuro.)                                                                        |   |   |     |     |               |                   |
|                            | Laproscopic                                                                                     |   |   |     |     |               |                   |
|                            | Pediatric                                                                                       |   |   |     |     |               |                   |
|                            | Small Organ                                                                                     |   |   |     |     |               |                   |
|                            | Neonatal Cephalic                                                                               |   |   |     |     |               |                   |
|                            | Adult Cephalic                                                                                  |   |   |     |     |               |                   |
|                            | Trans-rectal                                                                                    |   |   |     |     |               |                   |
|                            | Trans-vagina                                                                                    |   |   |     |     |               |                   |
|                            | Trans-urethral                                                                                  |   |   |     |     |               |                   |
|                            | Trans-esoph. (non-Card.)                                                                        |   |   |     |     |               |                   |
|                            | Musculo-skel. (Convert.)                                                                        |   |   |     |     |               |                   |
|                            | Musculo-skel. (Superfic.)                                                                       |   |   |     |     |               |                   |
|                            | Intra-luminal                                                                                   |   |   |     |     |               |                   |
|                            | Endoscopy                                                                                       | N | N | N   |     | N             | N                 |
| Cardiac                    | Cardiac Adult                                                                                   |   |   |     |     |               |                   |
|                            | Cardiac Pediatric                                                                               |   |   |     |     |               |                   |
|                            | Trans-esophageal (card.)                                                                        |   |   |     |     |               |                   |
|                            | Other (spec.)                                                                                   |   |   |     |     |               |                   |
| Peripheral<br>Vessel       | Peripheral vessel                                                                               |   |   |     |     |               |                   |
|                            | Other (Spec.)                                                                                   |   |   |     |     |               |                   |

E = added under Appendix E P = previously cleared by FDA; N = new application:

> (Please do not write below this line - continue on another page if needed) Concurrence of CDRH, Office of Device Evaluation (ODE)

Halid Temman

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devic

510(k) Number

Prescription Use (Per 21 CFR 801.109)

---

**Source:** [https://fda-staging.innolitics.com/device/K090196](https://fda-staging.innolitics.com/device/K090196)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).