The BD MAX™ GBS Assay as implemented on the BD MAX™ System is a qualitative in vitro diagnostic test designed to detect Group B Streptococcus (GBS) DNA in Lim Broth cultures after incubation for greater than or equal to (>)18 hours, obtained from vaginal and rectal swab specimens from antepartum pregnant women. The test incorporates automated DNA extraction to isolate the target nucleic acid from the specimen and real-time polymerase chain reaction (PCR) to detect a 124 bp region of the cfb gene sequence of the Streptococcus agalactiae chromosome. Results from the BD MAX GBS Assay can be used as an aid in determining colonization status in antepartum women. The BD MAX™ GBS Assay does not provide susceptibility results. Cultured isolates are necessary for performing susceptibility testing as recommended for penicillin-allergic women. Subculture to solid media for additional testing when indicated.
Device Story
BD MAX GBS Assay is an in vitro diagnostic test for detecting Group B Streptococcus (GBS) DNA in enriched Lim Broth cultures. Input: 15 µL aliquot of incubated vaginal/rectal swab specimen. Process: Automated DNA extraction and real-time PCR amplification of a 124 bp region of the cfb gene on the BD MAX System. Output: Qualitative detection of GBS DNA. Used in clinical laboratories by trained personnel. System automates extraction, amplification, and interpretation, eliminating manual intervention post-loading. Internal process control monitors for inhibition and system/reagent failures. Results aid clinicians in determining GBS colonization status in antepartum women to guide prophylactic decisions during birth, reducing risk of neonatal GBS infection.
Clinical Evidence
Prospective 3-site clinical study (n=601 compliant specimens) compared BD MAX GBS Assay to culture-based reference method. Primary endpoints: sensitivity and specificity. Results: 95.0% sensitivity (95% CI: 90.0-98.0%) and 96.7% specificity (95% CI: 94.7-98.2%).
Technological Characteristics
Nucleic acid amplification assay system; real-time PCR; automated DNA extraction; targets 124 bp cfb gene sequence; uses Scorpion probes; internal process control included; random access platform; qualitative output.
Indications for Use
Indicated for detection of GBS DNA in Lim Broth cultures (incubated >18 hours) from vaginal and rectal swabs of antepartum pregnant women to aid in determining colonization status. Not for susceptibility testing.
Regulatory Classification
Identification
Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identify Streptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Streptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.
{0}------------------------------------------------
# MAY 2 7 2010
## K090191 510k Summary BD MAX GBS Assay
Per 21 CFR Sec. 807.92
Supplement Date: May 24, 2010
Submitted by: HandyLab, Inc.( a wholly owned subsidiary of Becton, Dickinson and Company) 5230 S. State Road Ann Arbor, MI, 48108, USA
Contact Information: Mrs. Martha J. Rumford Director QA/RA Martha_Rumford@BD.com (734) 663-4719 Ext: 225 FAX (734) 663-7837
The name of the device:
BD MAX™ GBS Assay Trade name:
Common or usual name: Group B Strep Assay
| Product Codes: | NJR |
|----------------------|-----------------------------------------------------------------------------------------|
| Classification name: | NUCLEIC ACID AMPLIFICATION ASSAY SYSTEM,<br>GROUP B STREPTOCOCCUS, DIRECT SPECIMEN TEST |
| Regulation number: | 866.3740 |
CLASS I
Classification Panel MICROBIOLOGY
Predicate Device(s): K062948 Cepheid Smart GBS Assay
### DEVICE DESCRIPTION
#### INTENDED USE
-i-i
The BD MAX™ GBS Assay as implemented on the BD MAX™ System is a qualitative in vitro diagnostic test designed to detect Group B Streptococcus (GBS) DNA in Lim Broth cultures after incubation for greater than or equal to (>)18 hours, obtained from vaginal and rectal swab specimens from antepartum pregnant women. The test incorporates automated DNA extraction to isolate the target nucleic acid from the specimen and real-time polymerase chain reaction (PCR) to detect a 124 bp region of
{1}------------------------------------------------
the cfb gene sequence of the Streptococcus agalactiae chromosome. Results from the BD MAX GBS Assay can be used as an aid in determining colonization status in antepartum women.
The BD MAX™ GBS Assay does not provide susceptibility results. Cultured isolates are necessary for performing susceptibility testing as recommended for penicillin-allergic women. Subculture to solid media for additional testing when indicated.
#### SUMMARY AND EXPLANATION OF THE PROCEDURE
A vaginal and rectal swab is collected and transported to the laboratory using standard bacterial swab transport systems containing a non-nutritive transport medium (e.g. Amies or Stuart). In the lab, the swab is removed from the transport medium and placed into selective Lim Broth [Todd-Hewitt Broth supplemented with colistin (10μg/mL) and nalidixic acid (15µg/mL)]. After incubation of inoculated Lim Broth culture for >18 hours at 37 ℃ in ambient air or 5 % CO2, a 15 µL aliquot of the broth is mixed with BD MAX GBS Sample Preparation Reagent and processed on the BD MAX System using the BD MAX GBS Assay. The BD MAX System automatically extracts the target nucleic acid and amplifies a section of the cfb gene sequence of the GBS chromosome, if present. The BD MAX GBS Assay includes an Internal Process Control to monitor for the presence of potential inhibitory substances as well as system or reagent failures that may be encountered during the entire process.
Group B Streptococcus (GBS) is a Gram positive bacterium found in the lower intestinal tract of 10-30 % of all healthy adults. A person who carries GBS but does not show signs of GBS disease is said to be "colonized" with GBS. GBS colonization is not contagious and GBS are commonly found bacteria associated with the human body. Under certain circumstances, GBS can invade the mody and cause serious infection; this is referred to as Group B streptococcal disease. GBS can cause severe disease in a newborn and is known to be the leading cause of life threatening bacterial infections in newborns. A number of strains of the pathogen circulate in the community, and approximately 80 % of newborn infections are acquired during birth by vertical (mother-to-baby) transmission. Research has shown that GBS colonizes the anogenital mucosa of 25-40 % of healthy women, and despite the introduction of antibiotic prophylaxis in the US, GBS still causes approximately 2,500 cases of newborn infections per year. Of those infected, 25 % incur serious long-term consequences such as deafness, blindness and neurological damage. Approximately 100 babies die annually in the U.S. in the first three months of life as a result of GBS infection and half of these deaths occur during the first week of life.1
The current standard of care for preventing neonatal GBS disease is screening pregnant women at 35-37 weeks of gestation to determine their GBS colonization status. 1 Most GBS testing is performed by culture and can take up to 48 hours for definitive identification of GBS following the initial >18 hour incubation of vaginal and rectal swabs in a selective broth medium. The BD MAX GBS Assay, as implemented on the BD MAX System, can provide results from up to 24 specimens in approximately 120 minutes after the initial >18 hour incubation/enrichment step. The BD MAX GBS Assay streamlines and simplifies the testing process by eliminating the need for operator intervention from the time the sample is placed onto the BD MAX System until results are available.
Image /page/1/Picture/6 description: The image shows a close-up of a heavily pixelated and distorted image. The image appears to contain some text, but the pixelation makes it difficult to read. The text seems to be arranged vertically, and the overall impression is one of low image quality and significant data loss.
{2}------------------------------------------------
# Substantial Equivalence Table
・
| Item Being | BD MAX | Smart | Same | Different | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-----------|---|
| Compared | GBS Assay | GBS Assay | | | |
| Intended Use | The BD MAX™ GBS Assay as<br>implemented on the BD MAX TM<br>System is a qualitative<br>in vitro diagnostic test designed<br>to detect Group B<br>Streptococcus (GBS) DNA in<br>Lim Broth cultures after<br>incubation for greater than or<br>equal to (>)18 hours, obtained<br>from vaginal and rectal swab<br>specimens from antepartum<br>pregnant women. The test<br>incorporates automated DNA<br>extraction to isolate the target<br>nucleic acid from the specimen<br>and real-time polymerase chain<br>reaction (PCR) to detect a 124<br>bp region of the cfb gene<br>sequence of the Streptococcus<br>agalactiae chromosome.<br>Results from the BD MAX GBS<br>Assay can be used as an aid in<br>determining colonization status<br>in antepartum women.<br>The BD MAX™ GBS Assay<br>does not provide susceptibility<br>results. Cultured isolates are<br>needed for performing<br>susceptibility testing as<br>recommended for penicillin-<br>allergic women. Subculture to<br>solid media for additional<br>testing when indicated. | Same - Assay testing is<br>indicated for<br>identification of GBS<br>colonization in<br>antepartum women<br>Smart GBS Assay<br>testing is also indicated<br>for in intrapartum<br>women<br>The Smart GBS Assay<br>does not provide<br>susceptibility results.<br>Culture isolates are<br>needed for performing<br>susceptibility testing as<br>recommended for<br>penicillin-allergic | x | X | |
| Analyte | Group B Streptococci<br>DNA | women.<br>Same | X | | |
| Specimen Type | Vaginal and Rectal | Same | X | | |
| Specimen<br>collection media<br>type | Amies or Stuart | Copan or Stuart | X | | |
| Sample<br>Preparation<br>Method | Sample Preparation for<br>Nucleic Acid extraction is<br>automated on BD MAX<br>System | Incorporates a manual<br>sample preparation<br>process | | x | |
| Sample<br>Processing | Enriched in overnight LIM | Direct swab method or<br>Enriched in overnight<br>LIM | X | X | |
| | Platform | BD MAX System<br>(random access) | Cepheid SmartCylcer Dx<br>(batch) | | x |
| | Assay Format | Real Time Fluorogenic detection of PCR amplification | Same | x | |
| | Probes | Scorpion | Taqman | | x |
| | Single Use | Yes | yes | x | |
| | User / skill required | Moderately complex<br>- no special skills required<br>- built in protocol<br>- no data interpretation required. | Highly complex<br>(Smart Cycle)<br>Moderately complex<br>(Gene Expert) | x | x |
| | Automatic Assay | Yes - built-in result interpretation | Yes | x | |
| | Internal Process Control | Extraction and PCR internal process control is a process monitor | Probe Check | | x |
| | External Control | Materials available commercially but not required to run the test | Materials available commercially but not required to run the test | x | |
ﺗ
:
BD MAX GBS Assay 510(k) summary
:
ﻤﺎﺗ
{3}------------------------------------------------
ist
Press
## PERFORMANCE DATA
#### Precision
Qualitative testing was performed over a 12 day period in order to determine within laboratory precision using the BD MAX GBS Assay. Precision was determined within instrument as well as across instruments. For consistency, testing was performed using the same lot of BD MAX GBS Assay. Panel members were prepared at five levels, which included four concentrations of GBS along with True Negative (TN) samples. The levels of the panel members were determined by relation to the Limit of Detection (LoD) of the assay. The Moderate Positive (MP) sample was at a concentration of ~3X LoD, the Low Positive (LP) sample was at a level of ~1.5X LoD, the High Negative 2 (HN-2) sample was at a ~10 fold dilution of the LoD and the High Negative 1 (HN-1) sample was at a ~100 fold dilution of the LoD. Four replicates of each panel member were tested over a 12 day period with two runs per day on three different instruments by multiple operators.
#### Reproducibility
Qualitative testing was performed in order to determine reproducibility using the BD MAX GBS Assay. Reproducibility was determined within site as well as across sites. Panel members were prepared at four leve); , which included three concentrations of GBS along with True Negative (TN) sampless. The levels of the panel members were determined by relation to the Limit of Detection (LoD) of the assay. The Moderate Positive (MP) sample was at a concentration of ~2X LoD, the Low Positive (LP) sample was at a level of ~1X LoD, the High Negative (HN) sample was at a concentration of
{4}------------------------------------------------
~100 fold dilution of the LoD. Six replicates of each panel member were tested at 3 sites. across 5 runs over a minimum of a 3 day period.
## Carry over/Cross Contamination
A study was conducted to investigate within-run carry-over and between-run carry-over. All High Positive samples that gave a and result were accurately identified as positive while all of the True Negative sampless were accurately identified as negative. IND results were due to PCR failure as netther Target or Process control were amplified. This study demonstrated the absence of carry-over and cross-contamination either within a run or between successive runs using the GBS assay on the BD MAX System.
#### Effectiveness of Control
The BD MAX GBS Assay was performed with variations to critical assay and/or process conditions on the automated BD MAX System to demonstrate the effectiveness of the internal process controls incorporated as part of the assay. The BD MAX GBS Assay was also performed in the presence of known inhibitory factors to demonstrate the effectiveness of the internal process control to monitor for inhibition. Of the 4 critical assay conditions that were tested on the automated Jaquar system, the BD MAX GBS Assay will fail if any one of the following 3 critical assay conditions fail - (i) absence of Reagent 1, (ii) absence of Reagent 2, and (iii) absence of Reagent 3 from the strip. As a result the incorporation of the internal process control controls for the occurrence of errors due to the use of 'bad' reagents or errors due to some fluidic system malfunction resulting in the non-aspiration of these reagents from the reservoirs. In addition, the presence of an inhibitory agent (spermicidal jelly) leads to inhibition of the internal process control resulting in a higher Ct value than a sample without the agent. The data demonstrates the effectiveness of the internal process control to monitor for inhibition that may prevent detection of low levels of target and potentially lead to False negative errors. For the samples that had the specimen omitted, the internal process control was still detected at normal levels becausa the internal process control pellet present in the test strip (and exogenous to the spectinen) amplified to give a valid result.
#### Interfering Substances
The BD MAX GBS Assay was tested in the presence of both endogenous and exogenous interfering agents to characterize the ability of the assay to detect GBS DNA under these conditions. The study was performed at GBS concentrations of 300 CFU/mL and 3000 CFU/mL of Sample Preparation Reagent. Interference was also studied in the presence of high concentrations of 127 relevant non-target organisms to determine if the detection of GBS at 300 CFU/mL was affected by the presence of these organisms. The list of organisms and concentration tested are the same as listed in Analytical Specificity section. The following exogenous interfering substances were tested: miconazole (fungicide), hemorrhoid cooling gel, spermicidal foam (nonoxynol 9),spermicidal gel (nonoxynol 9), contraceptive gel, deodorant spray, lubricating gel, moisturizing lotion, body oil and body powder. A complete swab of exogenous agent, similar to the collection of a GBS swab, was added to neqative LIM broth and released into the specimen. The specimen (15ul) with the interfering agent was added to the Sample Preparation Reagent tube. The following endogenous substances were tested: human DNA (1.55 x 103 ng /mL Sample Preparation Reagent), whole blood (10% in Lim), urine
Image /page/4/Picture/8 description: The image shows a blurry, black and white graphic. The graphic appears to be a logo or emblem, but the details are not clear due to the low resolution and blurriness. The shapes are abstract and difficult to discern, but there are some vertical lines and curved shapes.
{5}------------------------------------------------
(30% in Lim), mucous (one Swab in Lim), amniotic fluid (10% in Lim), and feces (one swab in Lim).
Interference (1/3 replicates) was observed in the presence of Corynebacterium xerosis, Serratia marcescans and EBV when tested at a GBS target concentration of 300 CFU/mL of Sample Preparation Reagent.
The BD MAX GBS Assay was able to detect GBS at a concentration of 300 CFU/mL of Sample Preparation Reagent in the presence of all interfering agents tested except body powder and feces where one of the three replicates was called negative. At 3000 CFU/mL of Sample Preparation Reagent no interference was observed with these agents.
## Analytical Sensitivity Limit of Detection Study
The Limit of Detection (LoD) of the Book MAX GBS Assay is 200 CFU/mL Sample Preparation Reagent (2x104 CFU/mL enriched Lim broth). Hit rate (95% positivity) method was used to determine the LoD. Pooled Clinical Neqative samples spiked with GBS culture was used and individual clinical negative specimen spiked with GBS culture were used in the determination of LoD.
#### Microbial Variants
The ability of the BD MAX GBS Assay to detect multiple GBS serotypes was demonstrated using 12 different strains of GBS bacteria listed in Table 4. The BD MAX GBS Assay was able to detect all major serotypes of GBS at 300 CFU/mL Sample Preparation Reagent (3x104 CFU/mL incubated Lim broth culture).
#### Analytical Specificity
The BD MAX GBS Assay was performed on samples containing high levels of nontarget organisms, using the BD MAX System, to demonstrate the specificity of the assay for detection of Group B Streptococcus. A total of 127 organisms were tested (119 viable organisms and 8 genomic DNA), including 11 organisms phylogenetically similar to Group B Streptococcus and a wide variety of other organisms including viruses, fungi and parasites that are known to infect the urogenital tract or are part of urogenital microflora.. The Internal Process Control was detected in all specimens. None of the 11 phylogenetically related streptococcal isolates tested positive with the BD MAX GBS Assay. Of the remaining strains tested only one (Moraxella osloensis) was positive in four of nine replicates.
#### Clinical Performance
Performance characteristics of the BD MAX GBS Assay were determined in a 3-site prospective investigational study. Specimens were collected by health care providers using the procedure recommended by Centers for Disease Control and Prevention that is described as follows: "Swab the lower vagina (vaginal introitus), followed by the rectum (i.e., insert swab through the anal sphincter) using the same swab or two different swabs." Swabs were sent for culture-based analysis to be performed by laboratories at three separate metropolitan locations in the U.S. Following incubation of the vaginal and rectal swab specimens for >18 hours in selective Lim Broth medium, a 15 µL aliquot of this enriched broth was tested using the BD MAX GBS Assay to
{6}------------------------------------------------
ascertain the clinical sensitivity and Wecificity of the BD MAX GBS Assay as compared to the reference culture method based on CDC recommendations.1
To perform the culture based analysis, the enriched Lim Broth specimens were then subcultured to a sheep blood agar plate and incubated up to 48 hours. Colonies suggestive of GBS were then Gram stained and tested for catalase production. Gram positive/catalase negative colonies were then specifically identified by the appropriate confirmatory method. Beta hemolytic GBS colonies were confirmed using a latex agglutination test method and gamma hemolytic GBS colonies were confirmed by performing a CAMP reaction. Of the 631 clinical specimens enrolled in the study, 601 were compliant and included in the statistical analyses (Table 2).
Table 2: Clinical performance statistics using the BD MAX GBS Assay
| | REFERENCE (CULTURE) | | | |
|---------------------|---------------------|----------|-------|-----|
| All Sites | Positive | Negative | Total | |
| BD MAX<br>GBS Assay | Positive | 133 | 15 | 148 |
| | Negative | 7 | 446 | 453 |
| | Total | 140 | 461 | 601 |
Table 3: Summary of Clinical perform statistics using the BD MAX GBS Assay
| | SENSITIVITY | SPECIFICITY | PREVALENCE 1 |
|--------------------|--------------------|--------------------|--------------------|
| Site 1 | 97.4 % (37/38) | 96.6 % (141/146) | 20.0 % (39/195) |
| Site 2 | 92.0 % (46/50) | 95.3 % (142/148) | 25.1 % (50/199) |
| Site 3 | 96.2 % (50/52) | 97.6 % (163/167) | 23.6 % (54/229) |
| Total<br>(95 % CI) | 95.0 % (133/140) | 96.7 % (446/461) | 23.0 % (143/623) |
| | CI (90.0 - 98.0 %) | CI (94.7 - 98.2 %) | CI (19.7 - 26.5 %) |
I Prevalence is based on all specimens with compliant culture reference method results.
The conclusions drawn from the nonclinical and clinical tests demonstrate that the device is safe, effective, and performs as well as or better than the legally marketed predicate device identified in paragraph (a)(3) of this section.
Image /page/6/Picture/8 description: The image shows a black and white drawing of a mushroom cloud. The cloud is tall and narrow, with a rounded top and a narrower base. There are two horizontal lines on either side of the cloud. The image is simple and stylized.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002
Ms. Martha J. Rumford Director QA/RA BD Diagnostics 5230 S. State Rd. Ann Arbor, MI 48108
Y 2 7 201
Re: K090191
Trade/Device Name: BD MAX™ GBS Assay Regulation Number: 21 CFR § 866.3740 Regulation Name: Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test Regulatory Class: Class II Product Codes: NJR, OOI Dated: May 24, 2010 Received: May 26, 2010 .
Dear Ms. Rumford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Ms. Rumford
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will-allow-you-to-begin-marketing-your-device.as-described.in-your.Section.510(k).premarket. notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Viro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Freddie M. Rodrigues.
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
{9}------------------------------------------------
# Indications for Use Form
510(k) Number (if known): K090191
Device Name: BD MAX™ GBS ASSAY
_ Indications for .Use :.... _________________________________________________________________________________________________________________________________________________
The BD MAX™ GBS Assay as implemented on the BD MAX™ System is a qualitative in vitro diagnostic test designed to detect Group B Streptococcus (GBS) DNA in Lim Broth cultures after incubation for greater than or equal to (>)18 hours, obtained from vaginal and rectal swar" specimens from antepartum pregnant women. The test incorporated automated DNA extraction to isolate the target nucleic acid from the specimen and real-time polymerase chain reaction (PCR) to detect a 124 bp region of the cfb gene sequence of the Streptococcus agalactiae chromosome. Results from the BD MAX™ GBS Assay can be used as an aid in determining colonization status in antepartum women.
The BD MAX™ GBS Assay does not provide susceptibility results. Cultured isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women. Subculture to solid media for additional testing when indicated.
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | |
|------------------------|--|
| (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
ਹੈ ਹੈ
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Offision Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k) K09019/
Page 1 of I
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.