K090181 · Biolase Technology, Inc. · GEX · Feb 11, 2009 · General, Plastic Surgery
Device Facts
Record ID
K090181
Device Name
WATERLASE MD
Applicant
Biolase Technology, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 11, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
General Indications * - I Class I, II, III, IV and V cavity preparation - I Caries removal - Hard tissue surface roughening or etching - 트 Enameloplasty, excavation of pits and fissures for placement of sealants * For use on adult and pediatric patients Root Canal Hard Tissue Indications - B Tooth preparation to obtain access to root canal - Root canal preparation including enlargement - Root canal debridement and cleaning Root Canal Disinfection - Laser root canal disinfection after endodontic treatment Endodontic Surgery (Root Amputation) Indications - 트 Flap preparation = incision of soft tissue to prepare a flap and expose the bone. - 지 Cutting bone to prepare a window access to the apex (apices) of the root(s). - Apicoectomy amputation of the root end. - Root end preparation for retrofill amalgam or composite. - Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation. Bone Surgical Indications - Cutting, shaving, contouring and resection of oral osseous tissues (bor - Osteotomy Soft Tissue Indications including Pulpal Tissues* Incision, excision, vaporization, ablation and coagulation of oral soft assussion - Excisional and incisional biopsies - Exposure of unerupted teeth - Fibroma removal - Flap preparation = incision of soft tissue to prepare a flap and expose the bone. - Flap preparation = incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue 피 impactions) - Frenectomy and frenotomy - Gingival troughing for crown impressions 트 - 트 Gingivectomy - 이 Gingivoplasty - 해 Gingival incision and excision - 트 Hemostasis - 지 implant recovery - 트 Incision and drainage of abscesses - Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during ■ apical surgery - 을 Leukoplakia - l Operculectomy - Oral papillectomies - Operculectomy - Oral papillectomies - Pulpotomy - Pulp extirpation - Pulpotomy as an adjunct to root canal therapy - Root canal debridement and cleaning - Reduction of gingival hypertrophy - NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation. - I Soft tissue crown lengthening - Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa - I Vestibuloplastv * For use on adult and pediatric patient Laser Periodontal Procedures - Full thickness flap - Partial thickness flap I - Split thickness flap - Laser soft tissue curettage - Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket - Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining I junctional epithelium - Removal of granulation tissue from bony defects - Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility) - Osteoplasty and esseous recontouring (removal of bone to correct osseous defects and create ■ physiologic osseous contours) - Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.) - Osseous crown lengthening
Device Story
Waterlase MD is an Er, Cr: YSGG dental laser system. Device uses laser energy for hard/soft tissue dental procedures; input is operator-controlled laser settings. New Turbo handpiece added to existing system; minor software update adds 'Select' icon and unassigned preset fields for dentist-stored settings. Operated by dentists in clinical settings. Output is laser energy for tissue interaction; assists in surgical precision, hemostasis, and tissue removal. Benefits include minimally invasive tissue management and reduced need for traditional mechanical tools. No changes to operational software or hardware core.
Clinical Evidence
No clinical data presented. Substantial equivalence based on design controls, biocompatibility testing of new handpiece materials, and comparison to previously cleared predicate devices.
Technological Characteristics
Er, Cr: YSGG laser surgical instrument. Turbo handpiece materials identical to MD Gold handpiece; biocompatibility verified per ISO 13485-2003 and 21 CFR 820.30. System includes GUI updates for user-defined presets. Class II device.
Indications for Use
Indicated for adult and pediatric patients requiring dental hard and soft tissue procedures, including cavity preparation, caries removal, endodontic surgery, bone surgery, periodontal procedures, and soft tissue oral surgery (incision, excision, ablation, coagulation). Contraindicated for patients with conditions where local anesthesia or surgical procedures are contraindicated, such as specific heart diseases, bleeding disorders, or immune deficiencies.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
KOGOLFI
Special 510k Summary of Safety and Effectiveness Waterlase® MD New Accessor Biolase Technology, Inc January 23, 2009
FER = 1 2009
# Special 510(k) Summary of Safety and Effectiveness Information
Regulatory Authority:
CONFIDENTIAL :
Company:
Contact:
Safe Medical Devices Act of 1990, 21 CFR 807.92
Biolase Technology, Inc. 4 Cromwell Irvine, CA 92618
Ms. Ioana M. Rizoiu Biolase Technology, Inc. 4 Cromwell Irvine, CA 92618 Tel: (949) 226-8144 Fax: (949) 273-6680
Trade Name:
Waterlase® MD
Common Name:
Er, Cr: YSGG laser
Classification Name:
Classification Code:
Surgical laser instrument 79 GEX, a Class II device
Equivalent Devices:
Waterlase® MD Biolase Technology, Inc. K031140, July 7, 2004
#### Device Description:
The Waterlase MD is a dental laser device previously cleared by the FDA for hard and soft tissue dental indications (K031140, K071363). The only changes from the previously cleared device are the addition of the Turbo handpiece, which combines minor software and graphics changes that will be added to the Waterlase MD laser system. This software does not change the operational software, but adds a "Select" icon on the control panel, and unassigned pre-sets fields for use by the dentist to store preferred setting selections. There are no other hardware or software changes to the Waterlase MD device pending herein when compared to the device cleared under K031140 and K071363.
Materials used for the Turbo handpiece are the same as those for the MD Gold handpiece. Biocompatibility testing of these materials was conducted in compliance with 21 CFR Part 820, Quality System Regulation, Subpart C, Sec. 820.30 - Design Controls, and ISO 13485-2003, Section 7.3 - Design and Development; documented results (on file) confirmed there is no biocompatibility risk presented by these materials.
{1}------------------------------------------------
### Indications for Use:
### General Indications *
- I Class I, II, III, IV and V cavity preparation
- I Caries removal
- Hard tissue surface roughening or etching
- 트 Enameloplasty, excavation of pits and fissures for placement of sealants
- * For use on adult and pediatric patients
### Root Canal Hard Tissue Indications
- Tooth preparation to obtain access to root canal ■
- 트 Root canal preparation including enlargement
- 지 Root canal debridement and cleaning
### Root Canal Disinfection
- 트 Laser root canal disinfection after endodontic treatment
### Endodontic Surgery (Root Amputation) Indications
- 트 Flap preparation = incision of soft tissue to prepare a flap and expose the bone.
- 트 Cutting bone to prepare a window access to the apex (apices) of the root(s).
- 1 Apicoectomy - amputation of the root end.
- 요 Root end preparation for retrofill amalgam or composite.
- . Removal of pathelogical tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
### Bone Surgical Indications
- 트 Cutting, shaving, contouring and resection of oral osseous tissues (bone)
- 비 Osteotomy
#### Soft Tissue Indications including Pulpal Tissues*
Incision, excision, vaporization, ablation and coagulation of oral soft tissues, including:
- 보 Excisional and incisional biopsies
- . Exposure of unerupted teeth
- l Fibroma removal
- 트 Flap preparation = incision of soft tissue to prepare a flap and expose the bone.
- . Flap preparation = incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue impactions)
- Frenectomy and frenotomy 트
- Gingival troughing for crown impressions 피
- B Gingivectomy
- Gingivoplasty 트
{2}------------------------------------------------
CONFIDENTIAL
- Gingival incision and excision
- Hemostasis
- 트 implant recovery
- I Incision and drainage of abscesses
- Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during 트 apical surgery
- 파 Leukoplakia
- . Operculectomy
- I Oral papillectomies
- Operculectomy
- 1 Oral papillectomies
- 에 Pulpotomy
- Pulp extirpation
- 8 Pulpotomy as an adjunct to root canal therapy
- D Root canal debridement and cleaning
- Reduction of gingival hypertrophy ■
- Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues
- (i.e., granulation tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) nust be submitted to a qualified laboratory for histopathological evaluation.
- Soft tissue crown lengthening
- Treatment of cariker sores, herpetic and aphthous ulcers of the oral mucosa ■
- Vestibuloplasty
- * For use on adult and pediatric patient
### Laser Periodontal Procedures
- Full thickness flap
- 제 Partial thickness flap
- Split thickness flap
- 트 Laser soft tissue curettage
- . Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket
- 제 Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining
- junctional epithelium
- Removal of granulation tissue from bony defects
- 트 Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleedingindex, probe depth, attachment loss and tooth mobility)
- 제 Osteoplasty and osseous recontouring (removal of bone to correct osseous defects and create physiologic osseous contours)
- 요 Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.)
- 트 Osseous crown lengthening
{3}------------------------------------------------
All Contraindications remain the same as for the previously cleared device, Waterlase®MD (K031140)
### Contraindications:
All clinical procedures performed with the Waterlase®MD must be subjected to the same clinical judgment and care as with traditional techniques. Patient risk must always be considered and fully understood before clinical treatment.
The clinician must completely understand the patient's medical history prior to treatment. Exercise caution for general medical conditions, which might contraindicate a local procedure. Such conditions may include, but are not limited to,
allergy to local or topical anesthetics, heart disease, bleeding disorders, sleep apnea or an immune system deficiency. Medical clearance from the patient's physician is advisable when doubt exists regarding treatment.
### Conclusion:
The indications included herein are the same as the indications that have been previously cleared by the FDA for the Waterlase®MD (K031140, K071363). Substantial equivalency for the Waterlase MD with the Turbo Handpiece has been determined through comparison to the previously cleared Waterlase MD.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the three branches of government. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# FEB 1 1 2009
Biolase Technology, Inc. % Ms. Ioana Rizoiu VP, Clinical Research & Development 4 Cromwell Irvine, California 92618
Re: K090181
Trade/Device Name: Waterlase MD (Turbo Handpiece) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: II Product Code: GEX Dated: January 23, 2009 Received: January 26, 2009
Dear Mr. Christensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{5}------------------------------------------------
### Page 2 - Ms. Ioana Rizoiu
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Mulhausen
- Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
# 510(k) Number (if known):
# Device Name: Waterlase MD (Turbo Handpiece)
## Indications for Use:
### General Indications *
- Class I, II, III, IV and V cavity preparation .
- Caries removal 에
- Hard tissue surface roughening or etching
- Enameloplasty, excavation of pits and fissures for placement of sealants
* For use on adult and pediatric patients
### Root Canal Hard Tissue Indications
- B Tooth preparation to obtain access to root canal
- Root canal preparation including enlargement
- Root canal debridement and cleaning M
### Root Canal Disinfection
- Laser root canal disinfection after endodontic treatment 에
### Endodontic Surgery (Root Amputation) Indications
- 트 Flap preparation = incision of soft tissue to prepare a flap and expose the bone.
- 지 Cutting bone to prepare a window access to the apex (apices) of the root(s).
- Apicoectomy amputation of the root end.
- Root end preparation for retrofill amalgam or composite.
- Removal of pathological tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues (i.e., granulation tissue) from around the apex
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
### Bone Surgical Indications
- Cutting, shaving, contouring and resection of oral osseous tissues (bor
- Osteotomy
### Soft Tissue Indications including Pulpal Tissues*
Incision, excision, vaporization, ablation and coagulation of oral soft assussion
- Excisional and incisional biopsies
- Exposure of unerupted teeth
Division of General, Restorative, and Neurological Devices
510(k) Number_
{7}------------------------------------------------
- Fibroma removal
- Flap preparation = incision of soft tissue to prepare a flap and expose the bone. ■
- Flap preparation = incision of soft tissue to prepare a flap and expose unerupted teeth (hard and soft tissue 피 impactions)
- Frenectomy and frenotomy
- Gingival troughing for crown impressions 트
- 트 Gingivectomy
- 이 Gingivoplasty
- 해 Gingival incision and excision
- 트 Hemostasis
- 지 implant recovery
- 트 Incision and drainage of abscesses
- Laser soft tissue curettage of the post-extraction tooth sockets and the periapical area during ■ apical surgery
- 을 Leukoplakia
- l Operculectomy
- Oral papillectomies
- Operculectomy
- Oral papillectomies
- Pulpotomy
- Pulp extirpation
- Pulpotomy as an adjunct to root canal therapy
- Root canal debridement and cleaning
- Reduction of gingival hypertrophy
- -
NOTE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted to a qualified laboratory for histopathological evaluation.
- I Soft tissue crown lengthening
- Treatment of canker sores, herpetic and aphthous ulcers of the oral mucosa ■
- I Vestibuloplastv
* For use on adult and pediatric patient .
### Laser Periodontal Procedures
- Full thickness flap
- Partial thickness flap I
- Split thickness flap
- Laser soft tissue curettage
l mucosa
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number K090181
- Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket
- Removal of highly inflamed edematous tissue affected by bacteria penetration of the pocket lining I junctional epithelium
junct to root canal
ement and cleaning
-
- Removal of pathelegical tissues (i.e., cysts, neoplasm or abscess) and hyperplastic tissues
(i.e., granulation tissue) from around the apex
(i.e., granulation tissue) from around the apex
TE: Any tissue growth (i.e., cyst, neoplasm or other lesions) must be submitted
{8}------------------------------------------------
- Removal of granulation tissue from bony defects
- Sulcular debridement (removal of diseased, inflamed or necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss and tooth mobility)
- Osteoplasty and esseous recontouring (removal of bone to correct osseous defects and create ■ physiologic osseous contours)
- Ostectomy (resection of bone to restore bony architecture, resection of bone for grafting, etc.)
- Osseous crown lengthening
Prescription Use X Use (Part 21 CFR 801 Subpart D) Over-the-Counter
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
| 1 | |
|---|---|
| | œ |
| | |
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
K090181 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.