ATTRIUS, MODEL PET150A6001, ATTRIUS L, MODEL PET150A8001, TRUESIGHT, MODEL NSP-P6C, TRUESIGHT, MODEL NSP-P8
K090178 · Neusoft Positron Medical Systems Co., Ltd. · KPS · Apr 24, 2009 · Radiology
Device Facts
Record ID
K090178
Device Name
ATTRIUS, MODEL PET150A6001, ATTRIUS L, MODEL PET150A8001, TRUESIGHT, MODEL NSP-P6C, TRUESIGHT, MODEL NSP-P8
Applicant
Neusoft Positron Medical Systems Co., Ltd.
Product Code
KPS · Radiology
Decision Date
Apr 24, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Indications for Use
The PET scanner is intended to be used for whole body, multislice, positron emission tomography diagnostic imaging. The PET scanner is intended to be used to image individual patients in a controlled medical facility, such as a hospital, medical center, or mobile medical unit environment.
Device Story
PET scanner images distribution of positron-emitting radionuclides in human body to visualize function/metabolism. System comprises gantry, detector system, front-end electronics, patient table, transmission source, and workstations. Inputs: radioactive isotope emissions; outputs: cross-sectional tomographic images. Operation: attenuation correction (corrects for tissue absorption/scatter) and image reconstruction (filtered back projection or iterative reconstruction) generate quantitative images. Used in hospitals/clinics by trained personnel. Healthcare providers use images for diagnostic assessment. Benefits: non-invasive functional/metabolic imaging. Features automated rod source loading, increased patient weight capacity (200 kg), and vertical table movement. Data transferred via Gigabit Ethernet to workstation for processing.
Clinical Evidence
Bench testing only. Performance verified using NEMA NU2 phantoms and test procedures. Safety verified via IEC 60601-1 and IEC 60601-1-2 testing. Software validated per FDA guidance.
Technological Characteristics
PET scanner; Ge-68 transmission source; Gigabit Ethernet connectivity; OSEM image reconstruction; Linux/RTAI operating system; 200kg patient table capacity; automated rod source loading; 400ns discriminator dead time; 60ns coincidence dead time; 30ns encoder dead time. Complies with IEC 60601-1 and NEMA NU2 standards.
Indications for Use
Indicated for whole body, multislice, positron emission tomography diagnostic imaging in patients within controlled medical facilities (hospitals, clinics, mobile units).
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
K090178
Page 1 of 4
510(k) (Traditional) Submission
Section 5 510(k) Summary
NEUSOFT POSITRON MEDICAL SYSTEMS CO., LTD.
# 510(k) Summary
Date: _ 01/21/2009
# APR 2 4 2009
# Company:
1
| 510(k) owner | Neusoft Positron Medical Systems Co., Ltd. |
|--------------|----------------------------------------------------------------|
| Address | No.16 Shiji Road, Hunnan New District, Shenyang, P.R. China |
| Telephone | 86-24-83662197 |
| Fax | 86-24-83661796 |
| Contact | Sun Yan, Chief Quality Engineer, Quality Management Department |
| U.S. Agent | Delphi Consulting Group |
| Contact | J. Harvey Knauss |
| Telephone | 713-723-4080 |
| Fax | 713-723-0786 |
#### Device:
| Trade Name | Attrius™, Attrius L™, Truesight |
|----------------------|--------------------------------------------|
| Common Name | Positron Emission Tomography (PET) Scanner |
| Classification Name | System, tomography, computed, emission |
| Device Class | 21 CFR 892.1200 Class II |
| Product Code | KPS |
| Classification panel | Radiology |
| Predicate Device | K022001, mPower, Model 712, Positron Corp. |
# Device description summary:
The Positron Emission Tomography (PET) scanners from Neusoft Positron Medical Systems Corporation image the distribution of positron-emitting radionuclides in the human body. These software-controlled medical devices use positron emitting radioactive isotopes to create cross-sectional tomographic images that show function or metabolism, rather than anatomy, as in conventional scanning techniques.
The PET scanner's software generates quantitative images through Attenuation Correction, which corrects for loss of information due to absorption and scatter by overlying soft tissue, and through theoretical, measured, or hybrid (segmented attenuation correction) Image Reconstruction, which generates transaxial images from acquired data using a filtered back projection or iterative reconstruction process.
{1}------------------------------------------------
The PET scanner consists of Detector System, Gantry, Front-end Electronics, Patient Table, Transmission Source, Operator's Control Workstation, Image Processing Workstation and optional Remote Display Workstation.
## Intended Use:
The PET scanner is intended to be used for whole body, multi-slice, Positron Emission Tomography diagnostic imaging.
### Environment of Use / Patient Population:
The PET scanner is intended to be used for imaging individual patients in a controlled medical facility, such as a hospital, medical clinic, or mobile medical unit environment.
#### Comparison of Major Technological Characteristics:
The main design differences between the predicate device, the mPower™ PET scanner, and the candidate device, the NPMS PET scanner, are listed below:
- The system hardware dead time of the candidate device is reduced relative to the 1. predicate device, e.g., the predicate device has a Discriminator dead time period of 520 nanoseconds, whereas the candidate device Discriminator dead time period is 400 nanoseconds; the predicate device has a coincidence dead time period of 160 nanoseconds and encoder dead time period of 42 nanoseconds, whereas the candidate device has a coincidence dead time period of 60 nanoseconds and encoder dead time period of 30 nanoseconds.
- The predicate device Front End Electronics include 64 Discriminator boards, 8 2. Encoder boards, 4 LAFD boards, 1 Coincidence board, and 1 Arbitration and Transfer board, whereas the candidate device integrates the same functionality into 8 Bank Encoder Boards (BEB) and 1 Data Assembly Board (DAB) .
- 3. The predicate device uses 34 fans and blowers to pull cooling air through the Gantry and Front End Electronics enclosures. TThe candidate device uses 4 fans located in the top cover of the Gantry to pull cooling air through the Gantry.
- The predicate device rod source loading mechanism requires manual intervention to 4. switch between the high activity transmission rod source and low activity emission calibration rod source whereas the candidate device provides an automated rod source loading mechanism that loads either the high activity transmission rod source or the low activity emission calibration source without manual intervention.
{2}------------------------------------------------
- The predicate device has a single LED display on the front cover of the Gantry along 5. with two pushbutton control panels. The candidate device has two LCD displays combined with pushbutton control panels on the front cover of the Gantry, along with two pushbutton control panels on the sides of the Gantry (accessible from behind the Gantry), and an LED display on the rear cover of the Gantry.
- The predicate device provides a stop switch which makes the device slow down and 6. stop gradually, whereas the candidate device provides emergency stop switches which shut off the power of table, septa and wobble, and so the moving components of the device stop quickly.
- 7. The predicate device patient table weight limit is 159 kg. The candidate device patient table weight limit is 200 kg.
- 8. The predicate device patient table is a fixed height design. The candidate device patient table can move over a 24cm vertical range from a loading level to a working level.
- റ്റ. The predicate device utilizes one Ge-68 4mCi transmission rod source. The candidate device utilizes one Ge-68 5mCi transmission source.
- 10. The candidate device includes an updated control workstation and image processing workstation.
- 11. The predicate device data acquisition system is a rack mounted VMEbus-based real time computer running the VxWorks real-time operating system on a MC68030-based single board computer. The data acquisition and processing functions are executed on a set of proprietary VME/VSBbus compatible boards. The candidate device provides a Gigabit Ethernet interface on the Data Assembly Board that transfers the event data and system control data to the control workstation where the data are processed by software.
- 12. The candidate device provides the same 30 party OSEM image reconstruction software used on the predicate device. In addition, the candidate device uses the following 30 party software: Linux Operating System, RTAI (Real Time Application Interface). These 30 party software packages were not used on the predicate device.
- 13. The performance test data of the candidate device have been obtained using NEMA NU2 phantoms, under the test procedure of NEMA NU2 standards and detailed in section 18 of the submission.
{3}------------------------------------------------
#### NEUSOFT POSITRON MEDICAL SYSTEMS CO., LTD.
# Testing
The following tests can demonstrate the candidate device is as safe, as effective, and performs as well as or better than the predicate device:
- Radiation measurement ●
- NEMA NU2 Performance test
- IEC 60601-1 Safety tests .
- IEC 60601-1-1 Safety tests .
- IEC 60601-1-2 Electromagnetic Compatibility test
- Software validation, US FDA Software Used in Medical Devices Guidance
## Conclusions:
All the above differences have been verified and determined to not affect safety and effectiveness of the device. We submit that the candidate device and the predicate device are substantially equivalent.
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. The logo is black and white.
#### Public Health Service
APR 2 4 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Sun Yan Chief Quality Engineer, Quality Management Department Neusoft Positron Medical Systems Co., Ltd. No., 16, Shiji Road, Hunnan New District Shenyang, Liaoning. 110179 P. R. China
Re: K090178
Trade/Device Name: Attrius™, Attrius LTM, Truesight PET scanner · Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system
Regulatory Class: II Product Code: KPS
Dated: January 26, 2009 Received: January 26, 2009
# Dear Mr. Yan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate. commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
{5}------------------------------------------------
Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
21 CFR 876.xxx (Gastroenterology/Renal/Urology) (240) 276-0115 21 CFR 884.xxx (Obstetrics/Gynecology) (240) 276-0115 21 CFR 892.xxx (Radiology) (240) 276-0120 Other (240) 276-0100
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely wours,
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for use statement
510(k) Number:
K090178
Attrius™, Attrius L™, Truesight PET scanner Device Name:
Intended Use / Indications for Use:
The PET scanner is intended to be used for whole body, multislice, positron emission tomography diagnostic imaging.
Environment of Use / Patient Population:
The PET scanner is intended to be used to image individual patients in a controlled medical facility, such as a hospital, medical center, or mobile medical unit environment.
Prescription Use YES
Over-The-Counter Use NO
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|--------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal and Radiological Devices | |
| 510(k) Number | K090178 |
11
.
73
4-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.