ISOLYSER HOME SHARPS MANAGEMENT (HSM), MODEL 800M; ISOLY SHARPS MANAGEMENT SYSTEM (SMS), MODELS 2400M, 4000M, 10000M, 5.
K090157 · Wcm Waste & Compliance Management, Inc. · MMK · May 4, 2009 · General Hospital
Device Facts
Record ID
K090157
Device Name
ISOLYSER HOME SHARPS MANAGEMENT (HSM), MODEL 800M; ISOLY SHARPS MANAGEMENT SYSTEM (SMS), MODELS 2400M, 4000M, 10000M, 5.
Applicant
Wcm Waste & Compliance Management, Inc.
Product Code
MMK · General Hospital
Decision Date
May 4, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The Isolyser Sharps Management Systems SMSm is a single-use, disposable, over-the-counter sharps container intended for use in clinical and non-clinical settings for the disposal of contaminated sharps and/or medical/red-bag wastes. The SMSm is marketed in five (5) different models: (1) Home Sharps Management (HSM) 800m, .98 liters, 3.45 x 2.72 x 7.29 (in.); (2) Sharps Management System (SMS) 2400m, 3.3 liters, 5 x 5x 10 (in.); (3) Sharps Mariagement System (SMS) 4000m, 5.97 liters, 8.47 x 5.99 x 8.87 (in.); (4) Sharps Management System (SMS) 10000m, 10.47 liters, 8.51 x 5.99 x 14.32 (in.); and (5) Sharps Management System (SMS) 5.3m, 21.1 liters, 11.93 x 9.87 x 14.68 (in.). Each SMSm model is designed to safely and securely contain used medical sharps and/or medical/red-bag wastes prior to removal and subsequent disposal. The SMS 5.3m is intended only for the storage and disposal of filled approved sharps containers and/or medical/red-bag wastes. Once an empty SMSm container has been filled with used medical sharps and/or medical/red-bag wastes, the entire SMSm container is intended to be packaged and mailed-back to WCM Waste and Compliance Management, Inc. for proper disposal. The HSM 800m, SMS 2400m, SMS 4000m, SMS 10000m, and SMS 5.3m are intended for home use by home health care providers and for use by small quantity health care providers such as dentists, medical doctors, veterinarians, and laboratories. The SMSm sharps containers are not intended for reuse.
Device Story
Disposable, non-reusable sharps containers; used for safe containment of medical sharps and red-bag waste; operated by home healthcare providers, physicians, dentists, veterinarians, and laboratory staff. Inputs: used medical sharps/waste placed vertically into top opening. Operation: passive containment; no hand-retrieval required; puncture-resistant and leak-proof design. Output: filled container mailed back to WCM Waste and Compliance Management, Inc. for disposal. Benefits: provides secure storage and mail-back disposal path for small quantity generators; prevents accidental needlesticks.
Clinical Evidence
Bench testing only. Compliance with OSHA Bloodborne Pathogens Standard and ASTM F2132-01(2008) (excluding SMS 5.3m) demonstrated.
Technological Characteristics
Plastic, puncture-resistant, leak-proof containers. Dimensions range from 0.98L to 21.1L. Compliant with ASTM F2132-01(2008). Features BIOHAZARD warning labels. Non-reusable.
Indications for Use
Indicated for disposal of contaminated sharps and/or medical/red-bag wastes in home and clinical settings (doctors, dentists, veterinarians, labs). SMS 5.3m model indicated only for storage/disposal of already filled sharps containers/red-bag waste, not direct sharps disposal.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Sharps Compliance, Inc.'s Sharps Disposal by Mail System 15000
Submission Summary (Full Text)
{0}------------------------------------------------
K090157
Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
MAY - 4 2009
## Re: Isolyser SMSm Traditional 510(k) WCM Waste and Compliance Management, Inc. Carlsbad, CA 92011
## 510(k) Summary
#### Owner of Device
WCM Waste & Compliance Management, Inc. 6054 Corte Del Cedro Carlsbad, CA 92009-1514
## Contact
Sara Beth Watson Steptoe & Johnson LLP 1330 Connecticut Ave., NW Washington, DC 20036 Phone (202) 429-6460 Fax (202) 429-3902 swatson@steptoe.com
## Date Prepared
January 15, 2009
#### Name of Device
Common Name: Proprietary Name:
#### Sharps Container
Isolyser SMSm: Home Sharps Management (HSM) 800m; Sharps Management System (SMS) 2400m; Sharps Management System (SMS) 4000m; Sharps Management System (SMS) 10000m; and Sharps Management System (SMS) 5.3m.
Establishment Registration Number: 2032810 (2008) Classification Name: Regulation: Classification: Product Code: Panel:
Accessory to hypodermic single lumen needles 880.5570 Class II MMK General Hospital
{1}------------------------------------------------
K090157
## Device Uses and Description
The Isolyser Home Sharps Management (HSM) 800m; Sharps Management System (SMS) 2400m; Sharps Management System (SMS) 4000m; Sharps Management System (SMS) 10000m are disposable non-reusable sharps containers that are intended to provide for safe and effective disposal of medical sharps and/or medical/red-bag wastes. The Isolyser Sharps Management System (SMS) 5.3m is intended only for the storage and disposal of filled approved sharps containers and/or medical/red-bag wastes. The SMS 5.3m is not intended for the direct disposal of used medical sharps. The HSM 800m, SMS 2400m, SMS 4000m, SMS 10000m, and SMS 5.3m are marketed for home use by home healthcare providers and for use in the offices, exam. and patient rooms of small quantity generators such as medical doctors, dentists, veterinarians, and laboratories.
The SMSm container is stable, puncture resistant (SMS 5,3m not tested) and leak-proof on the sides and bottom. Used medical sharps and/or medical/red-bag wastes are placed vertically into the opening on the top of the container. Once placed into the container, the SMSm does not require users to reach by hand into the container to retrieve contaminated sharps and/or medical/red-bag wastes. Once the container is full, the SMSm is intended to be mailed back to WCM Waste and Compliance Management, Inc. for proper disposal. The overall design and specifications for the Isolyser SMSm meet OSHA Bloodborne Pathogens Standard as well as the American Society for Testing and Materials Standard F2132-01(2008) (SMS 5.3m not tested). All SMSm containers have BIOHAZARD warning labels clearly visible with lettering in contrasting color and are affixed by adhesives to the sides of the SMSm container.
#### Predicate Devices
The Isolyser SMSm sharps containers are substantially equivalent in intended use, function and basic composition to the Isolyser SMS sharps containers approved in 510(k) K943626 and Sharps Compliance, Inc.'s Sharps Disposal by Mail System 15000 [510(k) unknown], Specifically, the HSM 800m, SMS 2400m, SMS 4000m, and SMS 10000m have identical technological characteristics to the Isolyser SMS containers approved in 510(k) K943626; the containers are identical in size, composition, and design. In contrast to the Isolyser SMS approved in 510(k) K943626, the Isolyser SMSm is marketed as an empty container and does not contain the ingredients necessary to create a polymer to encapsulate used medical sharps. The SMS 5.3m has technological characteristics similar to Sharp Compliance. Inc.'s Sharps Disposal by Mail 15000. Both containers are closable plastic 5-gallon receptacles intended for the disposal of filled approved sharps containers and/or medical/red-bag wastes. The SMS 5.3m and the Sharps Disposal by Mail 15000 are not intended for the direct disposal of used medical sharps. Once filled, the Isolyser SMSm containers are mailed to WCM Waste and Compliance Management, Inc. for proper disposal. This disposal is substantially equivalent to Sharps Compliance, Inc.'s Sharps Disposal By Mail System.
## Conclusion
The Isolyser SMSm is substantially equivalent to the predicate devices based on the descriptive data, compliance with standards, and indications for use.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/16 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with outstretched wings, symbolizing protection and service. The eagle is positioned to the right of the circular text, which reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
WCM Waste & Compliance Management, Incorporated C/o Ms. Sara Beth Watson Attorney ` Steptoe & Johnson LLP 1330 Connecticut Avenue, NW Washington, District of Columbia 20036
Re: K090157
Trade/Device Name: Isolyser SMSm Sharps Containers, Including Models: Home Sharps Management (HSM) 800m; Sharps Management System (SMS) 2400m; Sharps Management System (SMS) 4000m; Sharps Management System (SMS) 10000m, and Sharps Sharps Management System (SMS) 5.3m.
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II
Product Code: MMK
Dated: April 9, 2009
Received: April 10, 2009
Dear Ms. Watson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2- Ms. Watson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration
and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Canton V. Martin Jr.
S. B. B. S. M. A.
Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): K090157
Device Name: Isolyser SMSm Sharps Containers, including models: Home Sharps Management (HSM) 800m; Sharps Management System (SMS) 2400m; Sharps Management System (SMS) 4000m; Sharps Management System (SMS) 10000m, and Sharps Management System (SMS) 5.3m.
Indications For Use: The Isolyser Sharps Management Systems SMSm is a single-use, disposable, over-the-counter sharps container intended for use in clinical and non-clinical settings for the disposal of contaminated sharps and/or medical/red-bag wastes. The SMSm is marketed in five (5) different models: (1) Home Sharps Management (HSM) 800m, .98 liters, 3.45 x 2.72 x 7.29 (in.); (2) Sharps Management System (SMS) 2400m, 3.3 liters, 5 x 5x 10 (in.); (3) Sharps Mariagement System (SMS) 4000m, 5.97 liters, 8.47 x 5.99 x 8.87 (in.); (4) Sharps Management System (SMS) 10000m, 10.47 liters, 8.51 x 5.99 x 14.32 (in.); and (5) Sharps Management System (SMS) 5.3m, 21.1 liters, 11.93 x 9.87 x 14.68 (in.). Each SMSm model is designed to safely and securely contain used medical sharps and/or medical/red-bag wastes prior to removal and subsequent disposal. The SMS 5.3m is intended only for the storage and disposal of filled approved sharps containers and/or medical/red-bag wastes. Once an empty SMSm container has been filled with used medical sharps and/or medical/red-bag wastes, the entire SMSm container is intended to be packaged and mailed-back to WCM Waste and Compliance Management, Inc. for proper disposal. The HSM 800m, SMS 2400m, SMS 4000m, SMS 10000m, and SMS 5.3m are intended for home use by home health care providers and for use by small quantity health care providers such as dentists, medical doctors, veterinarians, and laboratories. The SMSm sharps containers are not intended for reuse.
Shula M. Murphy MD
ivision Sign-Off) ··vision of Anesthesiology, General Hospital *ection Control. Dental Devices
:0(k) Number: K090157
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
×
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.