K090130 · Atmos, Inc. · OMP · Apr 2, 2009 · General, Plastic Surgery
Device Facts
Record ID
K090130
Device Name
ATMOS S041 WOUND
Applicant
Atmos, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Apr 2, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The ATMOS S041 Wound™ is a suction device intended for aspiration and collection of secretions and body fluids from wounds and is indicated for patients who would benefit from a suction device particularly as the device may promote wound healing. ATMOS S041 is appropriate for use on the following wounds: - Pressure ulcers - Diabetic/neuropathic ulcers - Venous insufficiency ulcers - Traumatic wounds - Post-operative and dehisced surgical wounds - Explored fistulas - Skin flaps and grafts
Device Story
Portable, battery-operated suction pump for wound drainage; removes bodily fluids/secretions via hose system into collection jar. Electromotive membrane pump creates vacuum; microprocessor-controlled unit manages battery charging and prevents overcharging/overheating. Hydrophobic bacterial filter prevents fluid ingress to pump. Used in clinical or home settings; operated by patients or clinicians. Output is vacuum-assisted drainage; aids wound healing. Device is non-sterile; uses third-party disposable accessories (hoses, containers, catheters).
Clinical Evidence
Bench testing only. Software verification and product validation tests performed; met all acceptance criteria. Compliance with safety standards (ISO 10993-1, EN 60601-1-2, EN ISO 14971, IEC 60601-1) documented. No clinical data provided.
Technological Characteristics
Electromotive membrane pump; microprocessor-controlled battery charging and thermal protection. Portable form factor with carrying strap/bag. Hydrophobic bacterial filter. Connectivity: Standalone. Standards: ISO 10993-1, EN 60601-1-2, EN 61000-3-2, EN 61000-3-3, EN ISO 14971, IEC 60601-1.
Indications for Use
Indicated for patients requiring aspiration and collection of wound secretions/fluids to promote healing. Applicable to pressure, diabetic/neuropathic, venous insufficiency, traumatic, post-operative/dehisced wounds, fistulas, and skin flaps/grafts. Contraindicated for eschar, untreated osteomyelitis, malignancy, untreated malnutrition, exposed arteries/veins/organs/nerves/vasculature/bone/tendons/anastomotic sites, necrotic tissue, and non-enteric unexplored fistulas.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
K090130 Page 1/3
## 510(k) Summary of Safety and Effectiveness
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
Date Prepared:
December 10, 2008
Submitter's Information: 21 CFR 807.92(a)(1) Mr. Marco Schlegel President 3717 Huckleberry Road Allentown, PA 18104 Toll Free: 866-862-8667 Fax: 610-351-6827 Email: MSchlegel@Atmosmed.com
APR - 2 2009
Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2) ATMOS S041 Wound™ Product Name: Common Name: Powered suction pump pump, portable, aspiration (manual or powered) Classification Name: 878.4780, Class II Product Code: M
Predicate Device: 21 CFR 807. 92(a)(3)
| 510(k) Number | K061367 |
|-----------------------------------|----------------------------------------------------------------------------------|
| Device Name | BlueSky VISTA™ Wound Vacuum System |
| Applicant | BlueSky Medical Group, Inc.<br>5924 Balfour Ct., Suite 102<br>Carlsbad, CA 92008 |
| Regulation Number | 878.4780 |
| Classification Product Code | OMP |
| Decision Date | 08/10/2006 |
| Decision | Substantially equivalent (SE) |
| Classification Advisory Committee | General & Plastic Surgery |
Device Description: 21 CFR 807 92(a)(4)
The ATMOS S 041 Wound is a device to remove bodily fluids for wound drainage and can be used by patients over an extended period of time as recommended by a physician. The device has a rechargeable battery and therefore is portable. A microprocessor controlled electronic battery charging unit in the suction device guarantees the safe charging of the battery, and overcharging of the battery.
The ATMOS S 041 Wound itself is not a sterile device but can use disposable (sterile) items from other manufacturers but these items are not a part of the S 041, (e.g. various hose sets, containers, trocars and catheters) which have been designed for the wound drainage.
Indications for Use: 21 CFR 807 92(a)(5)
The ATMOS S041 Wound™ is a suction device intended for aspiration and collection of secretions and body fluids from wounds and is indicated for patients who would benefit from a suction device particularly as the device may promote wound healing. ATMOS S041 is appropriate for use on the following wounds:
- · Pressure ulcers
- · Diabetic/neuropathic ulcers
- · Venous insufficiency ulcers
- Traumatic wounds
- · Post-operative and dehisced surgical wounds
- Explored fistulas
- · Skin flaps and grafts
{1}------------------------------------------------
K090130
Image /page/1/Picture/1 description: The image shows the title "510(k) Summary of Safety and Effectiveness". The text is in bold font and is centered on the page. The image is a simple title with no other visual elements.
Contraindications
- The use of the ATMOS S 041 WOUND is contraindicated in the presence of:
- · Eschar
- · Untreated osteomyelitis
- · Malignancy in wound (with exception of palliative care to enhance quality of life)
- Untreated malnutrition
- Exposed arteries, veins or organs
- · Necrotic tissue
- · Non-enteric unexplored fistulas
- · Exposed nerves
- · Exposed anastomotic site
- · Exposed bone or tendons
- Exposed vasculature
The ATMOS S 041 WOUND is not designed for use with emergency medical services.
Not to be used:
• In non-medical applications
- · In presence of combustible or explosive fluids or gases
#### Technological Characteristics: 21 CFR 807 92(a)(6)
The ATMOS S 041 Wound is a small suction unit. The device is operated by an electromotive membrane pump which is maintenance-free. When it operates, the pump creates a vacuum in the hose system and in the collection jar, with the help of which secretions can be sucked off through the set of hoses. The pump switches off after creation of the vacuum and then switches on again when the vacuum lies below a certain value.
The secretion is collected in a collection jar. A hydrophobic bacterial filter prevents the secretion being inadvertently drawn into the pump head. The device is fitted with a rechargeable battery. A microprocessor controlled electronic charging unit in the suction device guarantees the safe charging of the battery, and thus overcharging of the battery is avoided. The electronic unit prevents overheating of the suction device by a high temperature switch. For the mobile use a carrying strap is available. Useful accessories are a shoulder bag and a device support.
The ATMOS S 041™ device are substantially equivalent to other legally marketed devices in the United States and functions in a manner similar and is intended for the same use as the predicate device.
#### Brief summary of Non-clinical Tests and Results
Software verification and product validation tests were performed and meet the acceptance criterion. The ATMOS S 041™ device have been designed and tested to applicable safety standards (see below) and does not raise any new issues of safety, efficacy, or performance of the product.
| Number of standard | Std Date | Title of standard |
|----------------------------------------|--------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-1:1998 | 1998 | Biological Compatibility of medical devices - Part 1: Evaluation and<br>Testing1 |
| EN 60601-1-2 (2001) | 2001 | Medical electrical equipment -- Part 1-2: General requirements<br>for safety -Collateral standard: Electromagnetic compatibility -<br>Requirements and<br>tests |
| EN ISO 61000-3-2 (2000): | 2000 | Electromagnetic compatibility - Part 3-2: Limits - Limits for Harmonic<br>current emissions (equipment input current < 16A per phase). |
| EN ISO 61000-3-3 (1995) + A1<br>(2001) | 1995<br>2001 | Electromagnetic compatibility - Part 3: Limits - Section 3: Limitation<br>of voltage fluctuations and flicker in low-voltage supply systems for<br>equipment with rated current < 16A. |
<sup>1</sup> Result of evaluation: for ATMOS S 041 Wound: No biocompatibility tests are needed because not have parts that would normally touch patient.
{2}------------------------------------------------
K090130 Page 3/3
# 510(k) Summary of Safety and Effectiveness
| Number of standard | Std Date | Title of standard |
|------------------------------------|------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| CEN EN<br>980:1996+A1:1999+A2:2001 | 1999<br>2001 | Graphical symbols for use in the labeling of medical devices |
| EN ISO 14971 | 03/2001 | Medical devices - application of risk management to medical<br>devices |
| IEC 60601-1 | 1998,<br>1991,<br>1995 | Medical Electrical Equipment - Part 1: General Requirements for<br>Safety, 1988; Amendment 1, 1991-11, Amendment 2,<br>1995. (General) |
Conclusion: 21 CFR 807 92(b)(1)
The 510(k) for the ÀTMOS S 041 ™ device contains adequate information and data to enable FDA
to determine substantial equivalence to the predicate device. The device will be accordance with the voluntary standards listed in the enclosed voluntary standard survey.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black against a white background. The overall design is simple and conveys a sense of official authority.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR - 2 2009
ATMOS, Inc. % TUV Rheinland of North America, Inc. Tamas Borsai 12 Commerce Road Newton, Connecticut 06470
Re: K090130
Trade/Device Name: ATMOS S041 Wound™ Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: II Product Code: OMP Dated: March 19, 2009 Received: March 23, 2009
Dear Tamas Borsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
## Page 2 - Tamas Borsai
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a logally marketed predicate device results in a classification for your device and thus, permitts your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events. (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or a its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K090530
### Device Name: ATMOS S041 Wound™
#### Indications for Use:
The ATMOS S041 Wound™ is a suction device intended for aspiration and collection of secretions and body fluids from wounds and is indicated for patients who would benefit from a suction device particularly as the device may promote wound healing. ATMOS S041 is appropriate for use on the following wounds:
- · Pressure ulcers
- · Diabetic/neuropathic ulcers
- · Venous insufficiency ulcers
- · Traumatic wounds
- · Post-operative and dehisced surgical wounds
- · Explored fistulas
- · Skin flaps and grafts
## Contraindications
The use of the ATMOS \$ 041 WOUND is contraindicated in the presence of.
- Eschar
- · Untreated osteomyelitis
- · Malignancy in wound (with exception of palliative care to enhance quality of life)
- Untreated malnutrition
- · Exposed arteries, veins or organs
- · Necrotic tissue
- · Non-enteric unexplored fistulas
- · Exposed nerves
- · Exposed anastomotic site
- · Exposed bone or tendons
- · Exposed vasculature
The ATMOS S 041 WOUND is not designed for use with emergency medical services.
#### Not to be used:
· In non-medical applications
- · In presence of combustible or explosive fluids or gases
Prescription Use × (Part 21 CFR 801 Subpart D)
No. of the count
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
David Keane BellXU
**(Division Sign-Off)**
Division of General, Restorative,
and Neurological Devices
**510(k) Number** K090130
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.