STEREOS WORKSTATION

K090050 · Biospace Med · LLZ · Mar 5, 2010 · Radiology

Device Facts

Record IDK090050
Device NameSTEREOS WORKSTATION
ApplicantBiospace Med
Product CodeLLZ · Radiology
Decision DateMar 5, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device, Real-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K090050 · Mar 5, 2010STEREOS WORKSTATIONBiospace MedRetrospective patient image data set (175 patients)The device uses a model of bone structures derived from a retrospective clinical data set of 175 patients (normal, moderate scoliosis, and severe scoliosis) to provide 3D measurement tools for spine analysis.Retrospective clinical data; Scoliosis; 3D measurement model

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
A priori image data set; Retrospective analysis of clinical patient images175 patients (91 normal, 47 moderate idiopathic scoliosis, 37 severe idiopathic scoliosis); Sample Size: 175Not applicable for this studyDevelopment of bone structure model for 3D measurement tools

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
3D measurement of spine———A priori image data set from 175 patients (91 normal patients, 47 patients with moderate idiopathic scoliosis and 37 patients with severe idiopathic scoliosis), and dry isolated vertebrae data.—Accuracy studies performed by using either anthropomorphic phantoms or patient images acquired with EOS system in hospitals.—

Indications for Use

The sterEOS Workstation is intended for use in the fields of musculoskeletal radiology and orthopedics in both pediatric and adult populations as a general PACS device for acceptance, transfer, display, storage, and digital processing of 2D X-ray images of the musculoskeletal system including interactive 2D measurement tools. When using 2D X-ray images obtained with the Biospace EOS System (K071546), the sterEOS Workstation provides interactive 3D measurement tools to aid in the analysis of scoliosis and related disorders and deformities of the spine in adult patients as well as pediatric patients 7 years and older. The 3D measurement tools include interactive analysis based on a model of bone structures derived from an a priori image data set from 175 patients (91 normal patients, 47 patients with moderate idiopathic scoliosis and 37 patients with severe idiopathic scoliosis), and dry isolated vertebrae data. The model of bone structures is not intended for use in patients with a Cobb's angle > 50 degrees and is not intended for use to assess individual vertebral abnormalities.

Device Story

Workstation functions as PACS for 2D musculoskeletal X-ray images; accepts, transfers, displays, stores, and processes digital images. When paired with Biospace EOS System images, provides interactive 3D measurement tools for scoliosis and spinal deformity analysis. Operates via clinician-driven interface in clinical settings. Uses a priori bone structure model derived from 175-patient dataset and dry vertebrae to assist analysis. Output aids orthopedic/radiology decision-making regarding spinal deformities. Benefits include enhanced visualization and measurement of spinal alignment in pediatric (7+) and adult patients.

Clinical Evidence

Bench testing only. Accuracy studies of 3D measurement tools performed using anthropomorphic phantoms and patient images acquired with the EOS system. Results validated 3D measurement tools and demonstrated equivalent performance in both adult and pediatric populations.

Technological Characteristics

Supports DICOM 3.0 images. Runs on off-the-shelf hardware using Windows XP operating system. Features interactive 2D/3D measurement tools and bone structure modeling. User interface follows standard clinical workflow patterns.

Indications for Use

Indicated for musculoskeletal radiology and orthopedics in pediatric (7+ years) and adult populations. Used for PACS functions (acceptance, transfer, display, storage, processing) of 2D X-ray images. Provides 3D measurement tools for scoliosis and spinal deformity analysis when used with Biospace EOS System images. Contraindicated for patients with Cobb's angle > 50 degrees and for assessment of individual vertebral abnormalities.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K090050 ## 510(k) SUMMARY ## Biospace Med's sterEOS Workstation MAR - 5 2010 # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Biospace Med 10 rue Mercoeur PARIS F-75011 FRANCE Phone: +33 1 55 25 60 60 Facsimile: + 33 1 55 25 60 61 Contact Person: Karine Chevrie Date Prepared: December 31, 2008 ## Trade Name: sterEOS Workstation ## Common or Usual Name: sterEOS Workstation #### Classification: 21 CFR 892.2050; radiological image processing system #### Product Code: LLZ ## Predicate Devices: - sterEOS Workstation (K080529) {1}------------------------------------------------ #### Device Description: The sterEOS Workstation is a general picture archiving and communications storage system for acceptance, transfer, display, storage, and digital processing of 2D X-ray images of the musculoskeletal system, including interactive 2D measurement tools. ﺮ: When used with 2D X-ray images obtained with the Biospace EOS System (K071546), the sterEOS Workstation provides interactive 3D measurement tools to aid in the analysis of scoliosis and related disorders and deformities of the spine. #### Indications for Use: The sterEOS Workstation is intended for use in the fields of musculoskeletal radiology and orthopedics in both pediatric and adult populations as a general PACS device for acceptance, transfer, display, storage, and digital processing of 2D X-ray images of the musculoskeletal system including interactive 2D and 3D reconstruction and measurement tools. When using 2D X-ray images obtained with the Biospace EOS System (K071546), the sterEOS Workstation provides interactive 3D measurement tools to aid in the analysis of scoliosis and related disorders and deformities of the spine in adult patients as well as pediatric patients 7 years and older. The 3D measurement tools include interactive analysis based on a model of bone structures derived from an a priori image data set from 175 patients (91 normal patients, 47 patients with moderate idiopathic scoliosis and 37 patients with severe idiopathic scoliosis), and dry isolated vertebrae data. The model of bone structures is not intended for use in patients with a Cobb's angle > 50 degrees and is not intended for use to assess individual vertebral abnormalities. #### Technological Characteristics: The sterEOS Workstation supports DICOM 3.0 formatted images. The sterEOS Workstation is based on the Windows XP operating system and runs on off-the-shelf hardware. The sterEOS Workstation user interface follows typical clinical workflow patterns to process, review, and analyze digital images. ### Performance Data: Accuracy studies of the 3D measurements were performed by using either anthropomorphic phantoms or patient images acquired with EOS system in hospitals. Results of these studies validate the 3D measurement tools of the software and demonstrate the equivalent performance of the device in both adult and pediatric populations. {2}------------------------------------------------ ## Substantial Equivalence: The sterEOS Workstation expands the indications for use of the company's cleared sterEOS device (K080529) to include patients 7 years and older. Performance data demonstrate that use of the device in the expanded patient population does not impact the sterEOS's safety. The modified sterEOS Workstation is otherwise identical to the cleared sterEOS Workstation. Accordingly, the sterEOS is substantially equivalent. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing another person, or possibly a stylized representation of an eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 # MAR - 5 2010 Biospace Med., Inc. % John J. Smith, M.D., J.D. Regulatory Counsel Hogan & Hartson LLP 555 Thirteen Street, N.W. WASHINGTON DC 20004 Re: K090050 Trade/Device Name: sterEOS Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 18, 2009 Received: November 18, 2009 Dear Dr. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {4}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Donald J. Bohn Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use Statement 510(k) Number (if known): Device Name: sterEOS Workstation The sterEOS Workstation is intended for use in the fields of musculoskeletal radiology and orthopedics in both pediatric and adult populations as a general PACS device for acceptance, transfer, display, storage, and digital processing of 2D X-ray images of the musculoskeletal system including interactive 2D measurement tools. When using 2D X-ray images obtained with the Biospace EOS System (K071546), the sterEOS Workstation provides interactive 3D measurement tools to aid in the analysis of scoliosis and related disorders and deformities of the spine in adult patients as well as pediatric patients 7 years and older. The 3D measurement tools include interactive analysis based on a model of bone structures derived from an a priori image data set from 175 patients (91 normal patients, 47 patients with moderate idiopathic scoliosis and 37 patients with severe idiopathic scoliosis), and dry isolated vertebrae data. The model of bone structures is not intended for use in patients with a Cobb's angle > 50 degrees and is not intended for use to assess individual vertebral abnormalities. Prescription Use (Part 21 C.F.R. 801 Subpart D) AND/OR Over-The-Counter Use (21 C.F.R. 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device signature (Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety 17 510(k) Number \\DC - 026967/000001 - 2838917 v2 of
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...