← Product Code [LLZ](/productcode/LLZ) · K090045

# COLOR LCD MONITOR, RADIFORCE RX320 (K090045)

_Eizo Nanao Corporation · LLZ · Jan 28, 2009 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K090045

## Device Facts

- **Applicant:** Eizo Nanao Corporation
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jan 28, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

RadiForce RX320 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. RadiForce RX320 does not support the display of mammography images for diagnosis.

## Device Story

RadiForce RX320 is a 21.2-inch color LCD medical display; used by trained medical practitioners for viewing digital diagnostic images (X-ray, MRI). Device receives digital video input via DVI-D interface; transforms input signals into high-definition visual output. Features Digital Uniformity Equalizer (DUE) to compensate for luminance non-uniformity; improved brightness (900 cd/m2) and contrast (1000:1) compared to predicate. Operates in clinical environments; supports optional photo-sensor and calibration software for luminance management. Output allows clinicians to interpret medical images for diagnostic decision-making; provides high-resolution visualization to aid patient diagnosis.

## Clinical Evidence

No clinical data. Substantial equivalence is based on bench testing and technical comparison of display performance characteristics (brightness, contrast, response time, uniformity).

## Technological Characteristics

54cm (21.2") TFT Color LCD panel; 2048 x 1536 native resolution; 900 cd/m2 brightness; 1000:1 contrast ratio; 20ms response time. Connectivity via DVI-D. Includes Digital Uniformity Equalizer (DUE) for luminance compensation. Optional photo-sensor and calibration software. Complies with EN60601-1, UL60601-1, CSA C22.2 No. 601-1, FCC-B, VCCI-B.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- RadiForce R31 and R31-C ([K052344](/device/K052344.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

16090045

# 510(k) Summarv as required by 807.92

# 1. Company Identification

EIZO NANAO CORPORATION 153 Shimokashiwano-cho. Hakusan-shi. Ishikawa-ken. 924-8566. Japan Tel: +81-76-274-2468 Fax: +81-76-274-2484

JAN 2 8 2009

- 2. Official Correspondent Hiroaki Hashimoto (Mr.) Manager of Product Safety
3. Date of Submission January 6, 2009

4. Device Trade name Color LCD Monitor, RadiForce RX320

### 5. Common/Usual Name

Image display system, medical image workstation, image monitor/display, and others

### 6. Classification Number

Medical displays classified in Class II per 21 CFR 892.2050.

### 7. Predicate Device

Manufacturer : EIZO NANAO CORPORATION Device Name : Color LCD Monitor Model Name : RadiForce R31 and R31-C 510(k) No. : K052344

### 8. Description of Device

RadiForce RX320 is a 54cm (21.2") Color LCD display for medical image viewing. RX320 displays high-definition medical imaging.

### 9. Intended Use

RadiForce RX320 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. RadiForce RX320 does not support the display of mammography images for diagnosis.

### 10. Technological Characteristics

RadiForce RX320 is substantially equivalent to R31 and R31-C (K052344). Additional product innovations include Digital Uniformity Equalizer (DUE), which enables compensates for luminance non-uniformity. RX320 improved the brightness and contrast of the LCD module, and modified the calibration software. The brightness improved in 900 cd/m2 from 400 cd/m2. The contrast improved by it.

Comparison table of the principal characteristics of 2 devices is shown in the Attachment 1.

{1}------------------------------------------------

# Appendix 1: Comparison Table with Predicate Device

| Items                           | R31                                                                                                                                           | R31-C    | RX320                                                                                                                                                                           |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                   | K052344                                                                                                                                       |          | Not Provided                                                                                                                                                                    |
| Panel Protector                 | Not provided                                                                                                                                  | Provided | Optional                                                                                                                                                                        |
| Panel Size and<br>Type          | 53 cm (20.8") TFT Color LCD panel                                                                                                             |          | 54cm (21.2") TFT Color<br>LCD panel                                                                                                                                             |
| Pixel Pitch                     | 0.207 x 0.207mm                                                                                                                               |          | 0.21075 x 0.21075 mm                                                                                                                                                            |
| Cabinet Color                   | Black                                                                                                                                         |          | Same as R31 and R31-C                                                                                                                                                           |
| Display Colors                  | 10-bit: 1.06 billion (maximum)<br>8-bit: 16.77 million from a palette of 1.06<br>billion                                                      |          | Same as R31 and R31-C                                                                                                                                                           |
| Viewing Angles                  | H: 170°, V: 170°                                                                                                                              |          | Same as R31 and R31-C                                                                                                                                                           |
| Scanning<br>Frequency<br>(H, V) | 31-127kHz, 59-61Hz<br>(VGA Text: 69-71Hz,<br>QXGA: 36-61Hz)<br>Frame synchronous mode: 59-61Hz                                                |          | 31-127kHz, 29-61Hz<br>(VGA Text: 69-71Hz)<br>Frame synchronous mode:<br>59-61Hz                                                                                                 |
| Native Resolutions              | 2048 x 1536 (portrait)<br>/1536 x 2048 (landscape)                                                                                            |          | Same as R31 and R31-C                                                                                                                                                           |
| Brightness                      | 400 cd/m²                                                                                                                                     |          | 900 cd/m²                                                                                                                                                                       |
| Contrast Ratio                  | 400 : 1 (typical)                                                                                                                             |          | 1000 : 1 (typical)                                                                                                                                                              |
| DOT Clock                       | 215 MHz                                                                                                                                       |          | Same as R31 and R31-C                                                                                                                                                           |
| Response Time                   | 50 ms (typical)                                                                                                                               |          | 20 ms (typical)                                                                                                                                                                 |
| Active Display Size<br>(H x V)  | 318 x 424 mm                                                                                                                                  |          | 323.7 x 431.6 mm                                                                                                                                                                |
| Viewable Image<br>Size          | Diagonal: 529 mm                                                                                                                              |          | Diagonal: 540 mm                                                                                                                                                                |
| Luminance<br>Calibration        | Software (Optional)<br>Photo-sensor (Optional)                                                                                                |          | Software (Optional)<br>Photo-sensor (Optional)<br>Digital Uniformity<br>Equalizer                                                                                               |
| Input Signals                   | DVI Standard 1.0                                                                                                                              |          | Same as R31 and R31-C                                                                                                                                                           |
| Input Terminals                 | DVI-D 24 pin                                                                                                                                  |          | Same as R31 and R31-C                                                                                                                                                           |
| USB Ports /<br>Standard         | 1 upstream, 2 downstream                                                                                                                      |          | Same as R31 and R31-C                                                                                                                                                           |
| Power                           | AC100-120V, 200-240V, 50/60Hz                                                                                                                 |          | Same as R31 and R31-C                                                                                                                                                           |
| Power<br>Management             | DVI-DMPM                                                                                                                                      |          | Same as R31 and R31-C                                                                                                                                                           |
| Dimensions<br>(W x H x D)       | With Stand:<br>368 x 515.5 mm<br>- 597.5 x 209 mm<br>Without Stand:<br>368 x 486 x 88.5 mm                                                    |          | Same as R31 and R31-C                                                                                                                                                           |
| Certifications &<br>Standards   | TUV/GM, CE Medical Device Directive, CB<br>(EN60601-1), cTUVus (UL2601-1, CSA<br>C22.2 No. 601-1), VCCI-B, FCC-B,<br>Canadian ICES-003-A, CCC |          | CE (Medical Device<br>Directive), TUV/GM<br>(EN60601-1), cTUVus (UL<br>60601-1, CSA C22.2 No.<br>601-1), CB(IEC60601-1),<br>VCCI-B, FCC-B, Canadian<br>ICES-003-B, c-Tick, RoHs |

*The software used in RX320 is modified, refer to the "12. Information of Software used in RX320".

i : :

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Hiroaki Hashimoto Manager of Product Safety EIZO NANAO CORPORATION 153 Shimokashiwano-cho Hakusan, Ishikawa-ken 924-8566 JAPAN

# JAN 2 8 2009

Re: K090045

Trade/Device Name: Color LCD Monitor, (RadiForce RX320) Regulation Number: 21 CFR 892,2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 6, 2009 Received: January 7, 2009

## Dear Mr. Hashimoto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 80); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-011 |
|----------------|----------------------------------|---------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | (240) 276-011 |
| 21 CFR 892.xxx | (Radiology)                      | (240) 276-012 |
| Other          |                                  | (240) 276-010 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppor/index.html.

Sincerely yours.

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number:

K090045

Color LCD Monitor, RadiForce RX320 Device Name :

Indications for Use:

RadiForce RX320 is intended to be used in displaying and viewing digital images for diagnosis of X-ray or MRI, etc. by trained medical practitioners. RadiForce RX320 does not support the display of mammography images for diagnosis.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Tom M. Whay

(Division Sign-Off) Division of Reproductive, Abdominal a Radiological Devices 510(k) Number

5-1

---

**Source:** [https://fda-staging.innolitics.com/device/K090045](https://fda-staging.innolitics.com/device/K090045)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com