K083920 · Vasonova, Inc. · LJS · May 1, 2009 · General Hospital
Device Facts
Record ID
K083920
Device Name
FLOWPICC CATHETER
Applicant
Vasonova, Inc.
Product Code
LJS · General Hospital
Decision Date
May 1, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Indications for Use
VPS catheters are indicated for short or long-term peripheral access to the central venous system for intravenous therapy and injection of contrast media. For blood sampling, infusions, or therapies, use a 4F or larger catheter. The maximum recommended injection rate is 2cc/sec for the 3F catheter and 5cc/sec for the 4F and 5F catheters. The VPS Catheter is indicated for use by itself or with the VPS System (Stylet and Console) sold separately.
Device Story
VPS Catheter is a radiopaque polyurethane central venous access device (3F-5F, single/dual lumen). Used with VPS Stylet, which contains a Doppler sensor and ivECG sensing wire. Stylet transmits physiological signals to a separate VPS Console to assist in catheter placement. Used in clinical settings by healthcare professionals for intravenous therapy, contrast injection, and blood sampling. System provides real-time physiological feedback to guide catheter tip positioning, potentially reducing malposition risks and improving procedural efficiency.
Clinical Evidence
Bench testing only. Testing included static/burst/dynamic pressure, flexural fatigue, flexibility, flow rate, collapsibility, elongation, tensile/torque strength, air leakage, priming volume, and mechanical hemolysis. No clinical data presented.
Technological Characteristics
Materials: medical grade radiopaque polyurethane. Dimensions: 3F/4F/5F, 50-55cm length. Sensing: Doppler sensor and ivECG wire in stylet. Connectivity: Wired connection to external VPS Console. Sterilization: Sterile, single-use.
Indications for Use
Indicated for patients requiring short or long-term peripheral access to the central venous system for intravenous therapy, contrast media injection, blood sampling, and infusions. 4F or larger catheters required for blood sampling/infusions. Contraindicated for injection rates exceeding 2cc/sec (3F) or 5cc/sec (4F/5F).
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
VasoNova VPS Catheter
Bard PerQCath
Submission Summary (Full Text)
{0}------------------------------------------------
#### ร์. 510(k) Summary K083920, page 1 of 2
| | MAY - 1 200 |
|----------------------|-------------------------------------------------------------|
| Sponsor: | VasoNova Inc.<br>1368 Bordeaux Drive<br>Sunnyvale, CA 94089 |
| Contact Person: | Kim Tompkins |
| Phone Number: | 408.738.7006 |
| Fax Number: | 650.644.2456 |
| Prepared: | March 27, 2009 |
| Trade Name: | VPSTM Catheter, sizes 3F and 4F name subject to change |
| Common Name: | Central venous catheter |
| Classification Name: | Percutaneous, implanted, long-term intravascular catheter |
| Classification: | II |
| Product Code: | LJS 21 CFR 880.5970 |
| Advisory Panel: | General Hospital |
| Predicate Device: | VasoNova VPS Catheter<br>Bard PerQCath |
## Device Description
The VPS Catheter and VPS Stylet are designed for use together and with the VPS Console. The VPS Catheter is preloaded with the Stylet and supplied in a tray. The VPS Catheter/Stylet is sterile, single use, non-pyrogenic and non-toxic.
The VPS Catheters are single lumen (3 and 4 Fr) or dual lumen (5 Fr) central venous access catheters made from soft, radiopaque medical grade polyurethane and packaged with the VPS Stylet in a tray with the accessories necessary for a percutaneous microintroducer placement (Modified Seldinger or Seldinger technique). The VPS Catheters feature a reverse-taper open ended design with a working length of 50 or 55 cm. VPS Catheters are marked at 1 cm intervals and labeled every 5 cm for the entire length.
The VPS Stylet is a 6 foot long polymeric tube which contains a Doppler sensor at the distal tip and an intravascular electrocardiogram (ivECG) signal sensing wire. The Doppler sensor and the ivECG signal sensing wire, when connected to the VPS Stylet are used to detect and transmit physiological information to the VPS Console (available separately). The VPS Stylet has an outer diameter of 0.019 inches and is designed to be used with the compatible VPS Catheter with an inner lumen of 0.021 inches.
## Intended Use
VPS catheters are indicated for short or long-term central or peripheral access to the central venous system for intravenous therapy, power injection of contrast media, central venous monitoring and blood sampling.
{1}------------------------------------------------
# Performance Data
In vitro testing demonstrates that the subject device meets all acceptance criteria. Competed testing included:
- Static, burst and dynamic pressure .
- � Physical characteristics
- Flexural fatigue and flexibility .
- . Flow rate
·
- Flexural fatigue, flexibility, collapsibility and elongation .
- Tensile and torque strength .
- Freedom from air leakage .
- Priming volume .
- . Mechanical hemolysis
## Substantial Equivalence
VasoNova VPS Catheters have the same intended use, technological characteristics and principles of operation as its predicate devices. The subject and predicate catheters are similar in that they are fabricated from polyurethane, are 50-55cm in length, have markings every 1-5 cm, have one or two lumens, may have labeled power injectability, and similar flow rates and maximum pressures. Performance data demonstrate that the subject device is as safe and effective as the predicate devices.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol consists of three stylized human figures intertwined, representing the department's focus on health and well-being. The logo is presented in black and white.
#### Public Health Service
MAY - 1 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Kim Tompkins, RN, MBA VP, Regulatory/Quality/Clinical VasoNova Incorporated 1368 Bordeaux Drive, Suite 100 Sunnyvale, California 94089
Re: K083920
Trade/Device Name: VasoNova VSP Catheter Regulation Number: 21 CFR 880.5970 Regulation Name: Percutancous, Implanted, Long-term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: April 27, 2009 Received: April 29, 2009
Dear Ms. Tompkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration
{3}------------------------------------------------
Page 2- Ms. Tompkins
and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biomctrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Turner
Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### 4. 510(k) Indications for Use
K083920 510(k) Number (if known):
Device Name:
VasoNova VPS Catheter
Indications for use:
VPS Catheters are indicated for short or long-term peripheral access to the central venous system for intravenous therapy and injection of contrast media. For blood sampling, infusions, or therapies, use a 4F or larger catheter. The maximum recommended injection rate is 2cc/sec for the 3F catheter and 5cc/sec for the 4F and 5F catheters. The VPS Catheter is indicated for use by itself or with the VPS System (Stylet and Console) sold separately.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-the-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
s al for
on Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K083920
Page _ 1_of _ 1
Page 5 of 149
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.