ATL CO2 MEDICAL LASER SYSTEMS, MODELS ATL-150 AND ATL-250
K083918 · Advanced Technology Laser Co., Ltd. · GEX · Feb 2, 2009 · General, Plastic Surgery
Device Facts
Record ID
K083918
Device Name
ATL CO2 MEDICAL LASER SYSTEMS, MODELS ATL-150 AND ATL-250
Applicant
Advanced Technology Laser Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 2, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The ATL Family of CO2 Medical Laser Systems (and the delivery accessories that are used to deliver laser energy) are indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology.
Device Story
Microprocessor-controlled CO2 laser system; delivers 10.6um light energy via sealed laser tube. Models ATL-150 (15W) and ATL-250 (25W). Components include control system, user interface, power supply, laser emission/delivery system, cooling system, and safety features. Operated by clinicians in surgical settings for soft tissue management (ablation, vaporization, excision, incision, coagulation). Laser energy delivered via accessories to target tissue; clinical effect achieved through thermal interaction. Benefits include precise tissue removal and hemostasis across multiple medical specialties.
Clinical Evidence
No clinical data required. Substantial equivalence established through bench testing and adherence to recognized international safety and performance standards for medical laser equipment.
Technological Characteristics
CO2 laser system (10.6um wavelength); 15W (ATL-150) or 25W (ATL-250) max power. Sealed laser tube. Microprocessor-controlled. Standards: IEC 60825-1 (laser safety), IEC 60601-1 (general safety), IEC 60601-2-22 (therapeutic laser safety), IEC 60601-1-2 (EMC). Complies with 21 CFR 1040.10/11 for Class IV lasers. ISO 9001/13485 certified manufacturing.
Indications for Use
Indicated for soft tissue ablation, vaporization, excision, incision, and coagulation in dermatology, plastic surgery, general surgery, gynecology, podiatry, dental, and ENT procedures. Specific applications include treatment of wrinkles, skin tags, acne scars, keloids, tattoos, carcinomas, warts, cysts, hemorrhoids, blepharoplasty, hair transplant site prep, periodontal/oral surgery, podiatric lesions (ingrown/fungal nails, verrucae), and ENT/gynecological lesions (leukoplakia, condyloma, dysplasia).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
SLIM Evolution Family of CO2 Lasers and Delivery Device Accessories (K063001)
Submission Summary (Full Text)
{0}------------------------------------------------
K083918
Image /page/0/Picture/1 description: The image shows the logo for Advanced Technology Laser Co., LTD. The logo includes the company's initials, ATL, in a stylized font inside of a circle. Below the company name is the text "510(k) Summary".
FEB = 2 2009
I 2
# Exhibit #B 510(k) Summary
ATL Family of CO2 Medical Laser Systems Advanced Technology Laser Co., Ltd
(As required by 21 CFR 807.92)
1. Date Prepared: November 4, 2008
## 2. Sponsor Information
Advanced Technology Laser Co .; Ltd 920 Jian-chuan Road, Bldg. A2, Level 5 Shanghai, 200240, China
Contact Person Mingxia Xi, Director for Regulatory Affairs Phone: (+86) 21 5471-2151 Fax: (+86) 21 5471-2152
# 3. Submission Correspondent
Ms. Diana Hong
Mr. Lee Fu
Shanghai Mid-Link Business Consulting Co., Ltd Suite 8D, No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, 200030, China Tel: +86-21-64264467 Fax: +86-21-64264467-809
Email: Diana,hong@mid-link.net
## 4. Proposed Device Information
Device Common or Usual Name: Carbon dioxide laser;
Device Trade or Proprietary Name: ATL Family of CO2 Medical Laser Systems; Model: ATL-150, ATL-250;
Classification Name: laser instrument, surgical, powered;
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letters "ATL" in a stylized, bold font. The letters are arranged in a slightly overlapping manner, with the "A" on the left, the "T" in the middle, and the "L" on the right. The entire logo is enclosed within a circle, with a small portion of the circle missing at the bottom. The logo appears to be a simple, yet distinctive design.
ADVANCED TECHNOLOGY LASER CO., LTD. 510(k) Summary
Regulation Number: 21 CFR 878.4810; Product Code: GEX
Panel: 878 General and Plastic Surgery
## 5. Predicate Device
SLIM Evolution Family of CO2 Lasers and Delivery Device Accessories (K063001)
## 6. Device Description
The ATL Family of CO2 Medical Laser Systems is a microprocessor-controlled CO2 laser system using a sealed laser tube providing either 15 watts maximum power (ATL-150) or 25 watts maximum power (ATL-250).
The ATL Family of CO2 Medical Laser Systems is intended to be used to deliver 10.6um CO2 light energy in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties.
The ATL Farnily of CO2 Medical Laser Systems covers control system, user interface, power supply, laser emission and delivery system, cooling system and safety features.
## 7. Intended Use
The ATL Family of CO2 Medical Laser Systems (and the delivery accessories that are used to deliver laser energy) are indicated for use in surgical applications requiring the ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology,
#### 6. Substantial Equivalence
The ATL Family of CO2 Medical Laser Systems share the same indications for use and safety compliance, similar design features, functional features, and therefore are substantially equivalent to the predicate device, the SLIM Evolution Family of CO2 Lascrs and Delivery Device Accessories (K063001). In addition, a review of the predicate device demonstrate that the ATL Family of CO2 Medical Laser Systems is safe and effective as the predicate device as they share
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a logo with the letters "ATL" inside of a circle. The letters are stylized and appear to be connected. The logo has a slightly distressed or textured appearance, as if it has been printed or stamped. The circle is not perfectly smooth, adding to the overall vintage or worn look of the design.
#### ADVANCED TECHNOLOGY LASER CO., LTD. 510(k) Summary
equivalent wavelengths, and are used to perform the same indicated surgical procedures. Therefore the proposed device is substantially equivalent (SE) to the predicate device.
## 7. Testing
The ATL Family of CO2 Medical Laser Systems is designed, tested and will be manufactured in accordance with both mandatory and voluntary standards, including:
- IEC 60825-1: 2007, Safety of laser products - Part 1: Equipment classification, requirements and user's guide.
- . IEC 60601-2-22:2007, Medical Electrical Equipment - Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment.
- IEC 60601-1:1988+A1:1991+A2:1995; Medical Electrical Equipment Part 1: General requirements for safety.
- . IEC60601-1-2:2001+A1:2004, Medical Electrical Equipment - Part 1: General requirements for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
The devices also comply with European Medical Directive 93/42/EEC and the US Federal Performance Standards 21 CFR 1002.10 Requirements (21CFR 1040.10 and 21CFR 1040.11 for Class IV Laser Products), Part 820 - Quality System Regulation, and have passed ISO9001 and ISO13485 System Certification.
#### Non-Clinical Conclusion:
Laboratory testing was conducted to validate and verify that the proposed device. ATL Family of CO2 Medical System met all design specifications and was substantially equivalent to the predicate device.
Clinical Conclusion: No Clinical Information is required.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
#### Public Health Service
FEB = 2 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Advanced Technology Laser Co., Ltd % Underwriters Laboratories, Inc. Mr. Morten Simon Christensen 455 E. Trimble Road San Jose, California 95131-1230
Re: K083918
Trade/Device Name: ATL Family of CO2 Medical Laser Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: January 16, 2009
Received: January 23, 2009
Dear Mr. Christensen:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Morten Simon Christensen
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours.
Mark M. Millerson
- Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
ADVANCED TECHNOLOGY LASER CO., LTD. 510(k) Submission Report - Indications for Use Statement
# Section II Indications for Use Statement
്കാം വേണ്ടിക്കുന്നത്. അവലംബം കോട്ടിക വിക്കുന്നത്. കാലക്കുന്നത് ഇവ കാലത്തിലെ സമ്മാനമായിരുന്നു. അവലംബം സമ്മാനമായ സമ
## 510(k) Number: Device Name:
#### ATL Family of CO2 Medical Laser Systems
#### Indications for Use:
The ATL Family of CO2 Medical Laser Systems (and the delivery accessories that are used to deliver laser energy) are indicated for use in surgical applications requiring the ablation, vaporization, excision, incision, and coagulation of soft tissue in dermatology and plastic surgery, general surgery, gynecology, podiatry, dental and otorhinolaryngology.
Dermatology, Plastic Surgery and General Surgery procedures including but not limited to -
- A Treatment of furrows and wrinkles
- A Removal of skin tags, actinic keratosis, acne scars, keloids, tattoos, telangiectasia, squamous and basel cell carcinoma, warts and uneven pigmentation,
- A Treatment of cysts, abscesses, hemorrhoids and other soft tissue applications.
- A Blepharoplasty
- V Site preparation for hair transplants
Dental procedure including but not limited to
- ア Periodontal procedures such as - gingivectomy, removal of hyperplasias, gingivonlasty (incision and excision)
- > Oral Surgery procedures such as - aphous ulcer excision, frenectorny, benignimalignant lesion ablation, operculectomy and homeostasis
Podiatry procedures including but not limited to -
- > Ablation, vaporization and excision of soft tissue lesions such as ingrown nail, fungal nail, porokeratoma, matrixectomy and verrucae vulgares.
Otorhinolaryngology (ENT) procedures including but not limited to -
- ア Treatment of leukoplakia of larynx, nasal obstruction, rhinophyma, verrucae vulgares, choanal atresia, LAUP and papillomatosis polyps.
#### Gynecology
- V Treatment of condyloma acuminata, cervical intracpithelial neoplasia, leukoplakia and vulvar/vaginal intraepithelial neoplasia, cervical dysplasja.
1
| Prescription Use | X | AND/OR | Over-The-Counter Use | |
|-----------------------------|---|--------|------------------------|--|
| (Part 21 CFR 801 Subpart D) | | | (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of ODRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Page 1 of 1
Division of General, Restorative, and Neurological Devices
510(k) Number 1108351Y
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.