K083913 · Chesson Laboratory Associates, Inc. · KMF · Jul 31, 2009 · General Hospital
Device Facts
Record ID
K083913
Device Name
CHESSON LABS LIQUID BANDAGE
Applicant
Chesson Laboratory Associates, Inc.
Product Code
KMF · General Hospital
Decision Date
Jul 31, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5090
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
To cover intact skin or minor cuts, scrapes, burns, and irritations of the skin, to help keep them clean and dry and help protect them from infection
Device Story
Liquid polymer spray applied topically to affected skin areas; solvent evaporates to form a non-toxic, hydrophobic, elastomeric polymer coating. Acts as a wound sealant and moisture barrier; permeable to oxygen; flexible and transparent. Used by physicians or patients (OTC) to cover minor wounds, burns, and irritations. Provides a tough, pliable seal that prevents cracking and peeling during healing. Benefits patient by maintaining a clean, dry environment and protecting against infection.
Clinical Evidence
Bench testing only. Microbial barrier effectiveness validated via modified ASTM F1608 with five bacterial species. Biocompatibility evaluated per ISO 10993 (cytotoxicity, sensitization, irritation, acute systemic toxicity, genotoxicity, mutagenicity, implantation). Physical testing included molecular weight, purity, residual components, water content, drying time, viscosity, moisture vapor transmission rate, and tensile strength.
Indicated for covering intact skin, minor cuts, scrapes, burns, and skin irritations in patients requiring a protective barrier to keep wounds clean, dry, and protected from infection.
Regulatory Classification
Identification
A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.
{0}------------------------------------------------
K083913
Image /page/0/Picture/1 description: The image shows the logo for ChessonLabs. The logo consists of a chess knight piece inside of a circle. Below the circle is the text "ChessonLabs" in a sans-serif font.
1/4
JUL 3 I 2009
510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92.
510(k) owner's name: Chesson Laboratory Associates, Inc.
Address: 3100 Tower Blvd, Suite 117, Durham, NC 27707
Phone and fax numbers: (919) 419-4900; (919) 419-4923
Name of contact person: Lance L. Swick, PhD
Date of summary: 31 October 2008
{1}------------------------------------------------
K0839 1.3
Name of device:
Trade or proprietary name: to be determined
Common or usual name: liquid bandage
Classification name: liguid bandage; skin protectant
Legally marketed device to which equivalence is claimed: Pro-Derma Liquid Bandage (K063202) or Curad Spray Bandage (K022645)
Description of device:
This product is a liguid polymer in a spray format. The solution is topically applied to the affected area and the carrier solvent evaporates allowing polymerization of the remaining constituents to form a non-toxic, hydrophobic, elastomeric polymer coating that provides a barner against moisture but that is permeable to oxygen. The intended use is as a liquid bandage, as this coating is biocompatible and forms a tough yet pliable seal, preventing it from cracking and peeling as the skin heals.
Intended use of device: to cover intact skin or minor cuts, scrapes, burns, and irritations of the skin, to help keep them clean and dry and help protect them from infection
Performance characteristics:
The ability of the dried product film to act as a microbial barrier was validated by testing in a modification of ASTM F1608 (Standard Test Method for Microbial Ranking of Porous Packaging Materials) with five species of bacteria. The film formed an effective barrier to all organisms tested.
Biocompatibility of the product was evaluated per ISO-10993 for cytotoxicity, sensitization, and irritation, acute systemic toxicity, genotoxicity, mutagenicity, and implantation, Twelve different types of tests were run and the product was shown to be biocompatible.
Bench testing of the bulk formulation and packaged product included molecular weight, purity, and residual components, water content, setting/drying time, viscosity, percent solids, color and particulates, heat of polymerization, as well as moisture vapor transmission rate and tensile strength of the dried film, and distribution/ship tests of the packaged product. All product lots met specifications. Performance characteristics of the product are similar to those of the predicate devices.
| Parameter | Device | Predicate<br>Device | Predicate Device | Predicate Device |
|-----------------|------------------------------------------|-------------------------------------------|------------------------|----------------------------------------|
| Device Name | TBD | Pro-Derma<br>Liquid<br>Bandage | Curad Spray<br>Bandage | New-Skin |
| Company<br>Name | Chesson<br>Laboratory<br>Associates, Inc | Procurement<br>Technology<br>Systems, LLC | Beiersdorf AG | Medtech Products,<br>Inc (distributor) |
| 510(k) # | TBD | K063202 | K022645 | Pre-1976 |
Summary Comparison of New Device to Predicate Device
Chesson Laboratory Associates, Inc.
Liquid Bandage
Page 2 of 4
{2}------------------------------------------------
K083845
34
| Class | II | Same | Same | Same |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 21 CFR<br>number | 880.5090 | Same | Same | Same |
| Code | KMF | Same | Same | Same |
| Product Type | Device | Same | Same | Same |
| Description | Liquid bandage | Same | Same | Same |
| Format | Topical spray;<br>Single-use<br>applicator | Single-use<br>applicator | Topical spray | Same |
| Ancillary<br>components | None | Same | Same | Same |
| Intended Use | to cover intact skin<br>or minor cuts,<br>scrapes and<br>irritations of the<br>skin, to help keep<br>them clean and dry<br>and help protect<br>them from infection | to cover minor<br>cuts, scrapes<br>and minor<br>irritations of<br>the skin and<br>help protect<br>them from<br>infection | To provide a<br>covering over minor<br>cuts and scrapes<br>that are clean and<br>dry | To protect cuts<br>and scrapes,<br>prevent and<br>protect blisters,<br>helps prevent the<br>formation of<br>calluses, covers<br>painful hangnails |
| End User | Physician, OTC | OTC | OTC | OTC |
| Frequency of<br>Use | Once daily or as<br>needed to keep<br>wound covered | Same | Same | Same |
| Method of<br>Use | Spray on or topical<br>application | Topical<br>application | Spray on | Same |
| Mechanism of<br>Action | Wound sealant,<br>moisture barrier | Same | Same | Same |
| Materials /<br>components | Acetone and<br>methyl ethyl<br>ketone solvents,<br>polyetheramine.<br>N,N'-dialkylamino-<br>diphenylmethane,<br>polyether polyol<br>ethylene glycol,<br>propylene glycol,<br>diphenylmethane<br>diisocyanate | 2-Octyl and N-<br>Butyl<br>Cyanoacrylate<br>monomers,<br>plasticizer and<br>D&C violet #2<br>pigment | poly(methylacrylate-<br>isobutene-mono-<br>isopropylmaleate),<br>ethyl acetate, n-<br>pentane, carbon<br>dioxide, menthol | Acetone, Alcohol,<br>Isobutane<br>Propane, Oil of<br>Cloves, Pyroxylin<br>Solution, 8<br>Hydroxyquinoline |
| Testing | Biocompatibility<br>per ISO 10993 | Same | Same | Unknown |
| Performance<br>characteristics | Flexible,<br>transparent, water-<br>tight, but allows<br>passage of water<br>vapor; seals out<br>germs | Same | Same | Dries rapidly to<br>form a tough<br>protective cover<br>that is antiseptic,<br>flexible,<br>waterproof, and<br>lets skin breathe.<br>Completely covers<br>the entire wound<br>to keep out dirt<br>and germs |
Chesson Laboratory Associates, Inc.
Liquid Bandage
Page 3 of 4
{3}------------------------------------------------
K088913
uli
## Comparison statement
The Chesson Labs product is similar to the predicate devices in that they are polymers in organic solvents. They provide the same functions as a liquid bandage, are for over-thecounter (OTC) consumer use, have similar claims, and they have the same indications for use. All of these products are liquid solutions applied directly to the wound, which upon contact with the skin dry to form a clear barrier film. They have many of the same functional characteristics, such as water-tightness, water vapor permeability, oxygen permeability, flexibility, and barrier film duration.
Results of biocompatibility and design verification testing provided in the submission have demonstrated that the Chesson Labs Liquid Bandage has the same performance, safety and technical characteristics and therefore raises no new no new issues of safety or effectiveness.
5.004
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is black and white and appears to be a scanned image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Chesson Laboratory Associates, Inc. % Mr. Scott E. Neuville President and CEO 3100 Tower Boulevard, Suite 117 Durham, North Carolina 27707
'JUL 3 1, 2009
Re: K083913
Trade/Device Name: Regulation Number: 21 CFR 880.5090 Regulation Name: Liquid bandage Regulatory Class: Class II Product Code: KMF Dated: June 3, 2009 Received: June 4, 2009
Dear Mr. Neuville:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate. commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Scott E. Neuville
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrl/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark A. Milbers
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Indications for Use 1083913 510(k) Number (if known):
Device Name:
Indications for Use:
To cover intact skin or minor cuts, scrapes, burns, and irritations of the skin, to help keep them clean and dry and help protect them from infection
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use **X**
(21 CFR 801 Subpart C)
Ok
Daniel Krone for Mxu
(Division Sign-Off)
Page 1 of 1
Division of Surgical, Orthopedic, and Restorative Devices 3913
Liquid Bang Number.
Chesson Laboratory Associates, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.