K083877 · Neusoft Medical Systems Co., Ltd. · ITX · Jan 13, 2009 · Radiology
Device Facts
Record ID
K083877
Device Name
FLYING DIAGNOSTIC ULTRASOUND SYSTEM
Applicant
Neusoft Medical Systems Co., Ltd.
Product Code
ITX · Radiology
Decision Date
Jan 13, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: [Applications listed in tables include: Ophthalmic, Fetal/Obstetric, Abdominal, Intra-operative (vascular, epicardial, neuro), Pediatric, Small Organ (thyroid, scrotum, prostate, breast), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Gynecological, Cardiac (Adult, Pediatric), Peripheral vessel, Musculo-skel (conventional, superficial). Modes include: B, M, Pulse Wave Doppler, Continuous Wave Doppler, Color Doppler, Combined (Duplex, Triplex), and others including Color Power Angio, 3D, Panoramic, Harmonics, Directional Angio Imaging, Tissue Doppler Imaging.]
Device Story
FLYING Diagnostic Ultrasound System; console-based system with display and multiple transducers (L12-3, L12-5 50, 15-6L, S4-2, S8, C5-2, C8-4V, E6509, D5009V). System generates electronic voltage pulses sent to transducers; piezo-electric arrays convert pulses into ultrasonic pressure waves. Waves reflect off body tissues; transducers convert echoes back to electrical signals. System processes signals to generate B-mode, M-mode, Doppler (PWD, CWD, Color, Tissue), 3D, and harmonic images. Used in clinical settings by healthcare professionals for diagnostic imaging and fluid flow analysis. Output displayed on screen for clinical interpretation to aid in diagnosis and patient management.
Clinical Evidence
Bench testing only. No clinical data provided. Substantial equivalence is based on technological characteristics, performance testing, and compliance with safety standards for diagnostic ultrasound systems.
Technological Characteristics
System console with power supply and electronic circuitry; piezo-electric transducer arrays. Modes: B, M, PWD, CWD, Color Doppler, 3D, Panoramic, Harmonics, Tissue Doppler. Connectivity: Transducer ports. Electrical safety: Complies with FDA Track 3 limits. Materials: Biocompatible. Software: Signal processing for image reconstruction and Doppler shift analysis.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in ophthalmic, fetal, abdominal, pediatric, intraoperative (vascular, epicardial, neuro), musculoskeletal, peripheral vascular, small organ, cardiac (adult, pediatric), endocavity (trans-rectal, trans-vaginal), adult/neonatal cephalic, and gynecological applications. Prescription use only.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}------------------------------------------------
K083872
# JAN 1 3 2009
510(k) Premarket Notification
FLYING Diagnostic Ultrasound system
#### 3.2 Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 as implemented in 21 C.F.R. 8807.92.
The submitter of this premarket notification is:
Tian Yanfang
Manager of Quality Management Department No,16 Shiji Road,Hunnan District,Shenyang,China,110179 T: +86 24 83660649 F: +86 24 83780480
This summary was prepared on May 19, 2008.
The proprietary name of the device is the FL YING Diagnostic Ultrasound System. In combination with transducers - L12-3, L12-5 50, 15-6L, S4-2, S8, C5-2, D5009V, E6509, C8-4V- are commonly known as a diagnostic ultrasound system and transducers.
These devices are classified as follows:
Ultrasonic Pulsed Doppler Imaging System 90IYN
Ultrasonic Pulsed Echo Imaging System 90IYO
90ITX Diagnostic Ultrasound Transducer
As stated in 21 CFR, parts 892.1550, 892.1560 and 892.1570, each of these generic types of devices have been classified as Class II.
The FLYING is a diagnostic ultrasound device. It consists of a system console containing o the power supply and electronic circuitry required to generate the image, a display screen. and a connection to the separate transducers. It is substantially equivalent to the currently marketed M2540 ultrasound systems and transducers cleared in K014191 and Boris platform diagnostic ultrasound system cleared in K030455.
{1}------------------------------------------------
The FLYING system and transducers function in a manner identical to all Philips ultrasound systems and transducers. The system circuitry generates an electronic voltage pulse, which is transmitted to the transducer. In the transducer, a piezo-electric array converts the electronic pulse into an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The differing acoustic properties of the tissues in the body reflect some of the transmitted energy back to the transducer, where it is converted back to electrical signals and sent back to the system. In the system, advanced signal processing technologies convert the returned signals into images of the Doppler functions of this system process the Doppler shift frequencies from the echoes of moving targets (such as blood), to detect and graphically display the Doppler shifts of these tissues as flow.
The FLYING is intended for diagnostic ultrasound imaging and fluid flow analysis. It is intended for diagnostic ultrasound imaging in B. M. Pulse Wave Doppler, Continuous Wave Doppler, Panoramic, Color Power Angio (CPA, formerly Angio), 3D. Harmonics, Directional Angio Imaging, Tissue Doppler Imaging and combined mode (see table 4.3.2). It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Ophthalmic.Fetal.Abdominal.Pediatric.Intraoperative (vascular, evicardial. neuro),Musculoskeletal, Peripheral Vascular, Small Organ, Cardiac (Adult, Pediatric),Endocavity (Trans-rectal, Trans-vaginal),Adult and Neonatal Cephalic, Gynecological
The FLYING is substantially equivalent in safety and effectiveness to the predicates identified above:
- . Both the predicate device and the FLYING are indicated for the diagnostic ultrasonic imaging and fluid flow analysis.
- Both the predicate device and the FLYING have the same gray-scale and Doppler ● capabilities.
- Both the predicate device and the FLYING use essentially the same technologies for . imaging, Doppler functions and signal processing.
- Both the predicate device and the FLYING have acoustic output levels below the Track 3 . FDA limits.
- . Both the predicate device and the FLYING are manufactured under equivalent quality systems.
- Both the predicate device and the FL YING are manufactured of materials with equivalent . biosafety. The materials have been evaluated and found to be safe for this application.
- a Both the predicate device and FLYING are designed and manufactured to the same electrical and physical safety standards.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Neusoft Medical Systems Co., Ltd. % Mr. Tamas Borsai Division Manager, Medical Division TÜV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470
Re: K083877
Trade/Device Name: FLYING Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYO, ITX. and IYN Dated: December 23, 2008 Received: December 29, 2008
Dear Mr. Borsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
JAN 1 3 2009
This determination of substantial equivalence applies to the following transducers intended for use with the FLYING Diagnostic Ultrasound System, as described in your premarket notification:
Transducer Model Number
S4-2 Sector S8 Sector L12-3 Linear Array L12-5 50 Linear 15-6L Linear Array
{3}------------------------------------------------
## C5-2 Curved Linear C8-4v Curved Linear E6509 Endocavity D5009V Non-imaging pencil
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Paul Hardy at (240) 276-3666.
Sincerely vours.
Vorut Whay
Co Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
ﮩ
{4}------------------------------------------------
FLYING Diagnostic Ultrasound system
#### Indications for Use Tables 4.3.2
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
| 510(k) Number: | K083877 | | | | | | | | |
|-----------------------------|--------------------------------------------------------------------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|---------------------|--|
| Device name: | FLYING Diagnostic Ultrasound System | | | | | | | | |
| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | | | Mode of Operation | | | | | | |
| General | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | |
| (Track I<br>Only) | | | | | | | | | |
| Ophthalmic | Ophthalmic | N | N | N | N | N | N | N | |
| | Fetal/Obstetric | N | N | N | N | N | N | N | |
| | Abdominal | N | N | N | N | N | N | N | |
| | Intra-operative<br>(vascular/epicardial) | N | N | N | N | N | N | N | |
| | Intra-operative (Neuro) | N | N | N | | N | N | N | |
| | Laparoscopic | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | N | N | N | N | N | N | N | |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | N | N | N | | N | N | N | |
| | Neonatal Cephalic | N | N | N | | N | N | N | |
| | Adult Cephalic | N | N | N | N | N | N | N | |
| | Trans-rectal | N | N | N | | N | N | N | |
| | Trans-vaginal | N | N | N | | N | N | N | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Intra-luminal | | | | | | | | |
| | Other (Gynecological) | N | N | N | N | N | N | N | |
| | Cardiac Adult | N | N | N | N | N | N | N | |
| Cardiac | Cardiac Pediatric | N | N | N | N | N | N | N | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Other (Fetal) | N | N | N | | N | N | N | |
| Peripheral<br>Vessel | Peripheral vessel | N | N | N | N | N | N | N | |
| | Other (Specify) | | | | | | | | |
| | Musculo-skel (conventional) | N | N | N | | N | N | N | |
| | Musculo-skel (superficial) | N | N | N | | N | N | N | |
| | N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | |
* Other modes: Color Power Angio, 3-D Imaging, Panoramic, Harmonics (Tissue & Contrast), , Directional Angio Imaging, Tissue Doppler Imaging
Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color
Previous submission: No previous 510(k)s are associated with this product
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
htqui mothy
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
ﺮ ﮐﮯ
{5}------------------------------------------------
### FLYING Diagnostic Ultrasound system
| 510(k) Number: | System: FLYING Diagnostic Ultrasound System<br>Transducer: S4-2 Sector transducer | | | | | | | |
|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | Ophthalmic | P | P | P | P | P | P | P |
| Fetal<br>Imaging<br>& Other | Fetal/Obstetric | P | P | P | P | P | P | P |
| | Abdominal | P | P | P | P | P | P | P |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | P | P | P | P | P | P | P |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Gynecological) | P | P | P | P | P | P | P |
| Cardiac | Cardiac Adult | P | P | P | P | P | P | P |
| | Cardiac Pediatric | P | P | P | P | P | P | P |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Fetal) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | E | E | E | E | E | E | E |
| | Other (Specify) | | | | | | | |
### DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Musculo-skel (conventional) Musculo-skel (superficial)
* Other modes include: Color Power Angio, 3D, Panoramic, Harmonics, Directional Angio Imaging, Tissue Doppler Imaging
Combined modes:. Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color Previous submission: K014191 for Ophthalmic, Fetal, Abdominal, Pediatric, Adult Cephalic, Adult &
Pediatric Cardiac. K043535 for Gynecological, Peripheral Vessel.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
forni M. Whan
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 510(k) Number
{6}------------------------------------------------
| 510(k) Number: | System: FLYING Diagnostic Ultrasound System | | | | | | | |
|-----------------------------|--------------------------------------------------------------------------------------------------|---|---|-------------------|-----|------------------|-----------------------|---------------------|
| Transducer: | S8 Sector transducer | | | | | | | |
| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | | | Mode of Operation | | | | |
| General | Specific | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| (Track I<br>Only) | (Tracks I & III) | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal/Obstetric | P | P | P | | P | P | P |
| | Abdominal | P | P | P | P | P | P | P |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | P | P | P | P | P | P | P |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | | | | | | | |
| | Neonatal Cephalic | P | P | P | | P | P | P |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Gynecological) | P | P | P | P | P | P | P |
| | Cardiac Adult | P | P | P | P | P | P | P |
| Cardiac | Cardiac Pediatric | P | P | P | P | P | P | P |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Fetal) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | P | P |
| | Other (Specify) | | | | | | | |
| | Musculo-skel (conventional). | | | | | | | |
| | Musculo-skel (superficial) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Other modes: Color Power Angio, 3D, Panoramic, Directional Angio Imaging, Tissue Doppler Imaging
Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color Previous submission: K014191
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
Lori M. Holm
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devic
510(k) Number
{7}------------------------------------------------
| 510(k) Number: | | | | | | | | |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------|---|---|-----|-----|------------------|-----------------------|---------------------|
| System: | FLYING Diagnostic Ultrasound System | | | | | | | |
| Transducer: | L12-3 Linear ArrayTransducer<br>Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| General<br>(Track I Only) | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Ophthalmic | | | | | | | |
| | Fetal/Obstetric | P | P | P | | P | P | P |
| | Abdominal | P | P | P | | | P | |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | P | P | | P | P | P |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | P | P | P | | P | P | P |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Fetal) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | P | P |
| | Other (Specify) | | | | | | | |
| | Musculo-skel (conventional) | P | P | P | | P | P | P |
| | Musculo-skel (superficial) | P | P | P | | P | P | P |
| | N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | |
Imaging Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color
Previous submission: K014191
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
Vornt Artz
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
Radiological Devices
510(k) Number
{8}------------------------------------------------
| 510(k) Number: | | | | | | | | |
|------------------------------|--------------------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| | System: FLYING Diagnostic Ultrasound System | | | | | | | |
| Transducer: | L12-5 50 Linear Transducer | | | | | | | |
| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal/Obstetric | E | E | E | | E | E | E |
| | Abdominal | E | E | E | | E | E | E |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | P | P | P | | P | P | P |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | P | P | P | | P | P | P |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Fetal) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | P | P | P | | P | P | P |
| | Other (Specify) | | | | | | | |
| | Musculo-skel (conventional) | P | P | P | | P | P | P |
| | Musculo-skel (superficial) | P | P | P | | P | P | P |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: Color Power Angio, Panoramic, Directional Angio Imaging Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color
Previous submission: K991671 for Intraoperative (Abdominal and Vascular Small Parts,) Musculo-skeletal (conventional and Superficial), Pediatric, Peripheral Vascular
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
Arni M. Nohr
(Division Sign Off) Division of Reproductive, Abdominal, and Radiological Devices
510(k) Number
{9}------------------------------------------------
510(k) Number: System: FLYING Diagnostic Ultrasound System Transducer: 15-6L Linear array transducer Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Clinical Application Mode of Operation General Specific B PWD CWD M Color Combined Other* (Track I (Tracks I & III) Doppler (Specify) (Specify) Only) Ophthalmic Ophthalmic Fetal/Obstetric Abdominal Intra-operative P P P P P P (vascular/epicardial) Intra-operative (Neuro) P P P P P -P Laparoscopic Pediatric Fetal P P P P . P Imaging & Other Small Organ (thyroid, scrotum, P P P b P P prostate, breast) Neonatal Cephalic Adult Cephalic Trans-rectal Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Intra-luminal Other (Gynecological) Cardiac Adult Cardiac Cardiac Pediatric Trans-esoph. (Cardiac) Other (Fetal) Peripheral Peripheral vessel P P P P P ﺔ Vessel Other (Specify) Musculo-skel (conventional) P P P P P p Musculo-skel (superficial) N= new indication; P= previously cleared by FDA; E= added under Appendix E *Other modes: Amplitude Doppler, Panoramic, Directional Angio Imaging Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color Previous submission: K014191
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
Lorgai In Witz
(Division Sign-Off) (Division of Reproductive, Abdominal, and Radiological Devi 510(k) Number
{10}------------------------------------------------
.
..
| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
|------------------------------|--------------------------------------------------------------------------------------------------|---|---|-----|-----|-------------------|-----------------------|---------------------|
| Clinical Application | | | | | | Mode of Operation | | |
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal/Obstetric | P | P | P | | P | P | P |
| | Abdominal | P | P | P | | P | P | P |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | P | P | P | | P | P | P |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Gynecological) | P | P | P | | P | P | P |
| | Cardiac Adult | | | | | | | |
| Cardiac | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Fetal) | | | | | | | |
| Peripheral | Peripheral vessel | P | P | P | | P | P | P |
| Vessel | Other (Specify) | | | | | | | |
| | Musculo-skel (conventional) | | | | | | | |
| | Musculo-skel (superficial) | | | | | | | |
| | N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | |
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801 109)
Previous submission: K043535
longi In Whang
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devi
510(k) Number
ﻧﻴﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﺘﻲ ﺗﺮﻳﺐ ﺍﻟﺘﻲ ﺗﺮﻛﻴﺐ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﺘﻲ ﺗﺮﺍﺗﺐ ﺍﻟﺘﻲ ﺗﻌﺘﺒﺮ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﺘﻲ ﺗﻌﺘﺒﺮ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﺘﻲ ﺗﻌﺘﺒﺮ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﺘﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
{11}------------------------------------------------
510(k) Number: System: FLYING Diagnostic Ultrasound System
Transducer: C8-4v Curved Linear Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | Mode of Operation | | | | | |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---|-------------------|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal/Obstetric | P | P | P | | P | P | P |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | P | P | P | | P | P | P |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Gynecological) | P | P | P | | P | P | P |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Fetal) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Musculo-skel (conventional) | | | | | | | |
| | Musculo-skel (superficial) | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | |
| | *Other modes: SonoCT, X-Res, Color Power Angio, Panoramic, 3-D Imaging, Directional Angio Imaging | | | | | | | |
| | Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color, Dual | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
tom Whang
1381
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device
510(k) Number.
{12}------------------------------------------------
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
| 510(k) Number: | | | | | | | | |
|-----------------------------|--------------------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| | System: FLYING Diagnostic Ultrasound System | | | | | | | |
| Transducer: | E6509 Endocavity transducer | | | | | | | |
| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | |
| General | Specific | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| (Track I<br>Only) | (Tracks I & III) | | | | | | | |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal/Obstetric | P | P | P | | P | P | P |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Pediatric | | | | | | | |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | P | P | P | | P | P | P |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | P | P | P | | P | P | P |
| | Trans-vaginal | P | P | P | | P | P | P |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Fetal) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Musculo-skel (conventional) | | | | | | | |
| | Musculo-skel (superficial) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Other modes: Color Power Angio, Panoramic, Harmonics (Tissue & Contrast), 3-D Imaging, Directional Angio Imaging, Tissue doppler Imaging
Combined modes: Duplex = 2D + Doppler; Triplex = 2D + Doppler + Color
Previous submission: K014191
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of Center for Devices and Radiological Health, Office of Device Evaluation
Prescription Use (Per 21 CFR 801.109)
Agnus M. Witz
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
Page 22
510(k) Number
{13}------------------------------------------------
### FLYING Diagnostic Ultrasound system
| Clinical Application | | Mode of Operation | | | | | | |
|------------------------------|-----------------------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track I<br>Only) | Specific<br>(Tracks I & III) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal/Obstetric | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative<br>(vascular/epicardial) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (thyroid, scrotum,<br>prostate, breast) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Intra-luminal | | | | | | | |
| | Other (Gynecological) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Other (Fetal) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | P | P | | | |
| | Other (Specify) | | | | | | | |
# DIAGNOSTIC ULTRASOUND INDICATIONS FOR USE FORM
(PLEASE DO NOT WRITE BELOW THIS LINE…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.