PRONTO V3 EXTRACTION CENTER, MODEL 5003, PRONTO LP EXTRACTION CATHETER, MODEL 5010, SKYWAY RX SUPPORT CATHETER,
K083784 · Vascular Solutions, Inc. · QEZ · Mar 30, 2009 · Cardiovascular
Device Facts
Record ID
K083784
Device Name
PRONTO V3 EXTRACTION CENTER, MODEL 5003, PRONTO LP EXTRACTION CATHETER, MODEL 5010, SKYWAY RX SUPPORT CATHETER,
Applicant
Vascular Solutions, Inc.
Product Code
QEZ · Cardiovascular
Decision Date
Mar 30, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PRONTO extraction catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature. The SKYWAY support catheter is intended to be used in conjunction with steerable guidewires in order to access discreet regions of the arterial and or coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The SKYWAY OTW also may be used to subselectively infuse/deliver therapeutic agents. The TWIN-PASS dual access catheter is intended to be used in conjunction with steerable guidewires in order to access discrete regions of the coronary and peripheral arterial vasculature, to facilitate placement and exchange of guidewires and other interventional devices, for use during two guidewire procedures and to subselectively infuse/deliver diagnostic or therapeutic agents.
Device Story
Catheter family includes extraction (Pronto V3/LP) and support/access (Skyway RX/OTW, Twin-Pass) devices. Extraction catheters use dual-lumen design; larger lumen allows thrombus aspiration via syringe/stopcock; smaller lumen accommodates guidewires. Support/access catheters provide guidewire support; facilitate guidewire exchange; Twin-Pass allows dual-guidewire procedures; Skyway OTW/Twin-Pass allow subselective infusion. All devices are hydrophilically coated. Used in interventional procedures by physicians to navigate coronary/peripheral vasculature. Pronto catheters include 70um filter basket for thrombus analysis. Stiffening mandrels/stylets included for delivery support. Clinical benefit: improved vessel access, thrombus removal, and therapeutic delivery.
Clinical Evidence
Bench testing only. Verification testing and biomaterial assessments performed to qualify hydrophilic coating modifications.
Technological Characteristics
Dual-lumen (Pronto/Twin-Pass) or single-lumen (Skyway) catheters; hydrophilically coated; compatible with 0.014" guidewires; 6F guide catheter compatible; includes accessories (syringes, stopcocks, 70um filter basket, stiffening mandrels/stylets).
Indications for Use
Indicated for removal of fresh, soft emboli and thrombi from coronary and peripheral vessels (Pronto V3/LP); access to coronary/arterial vasculature, guidewire placement/exchange, and subselective infusion of therapeutic/diagnostic agents (Skyway/Twin-Pass).
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
May 20, 2021
Vascular Solutions, Inc. Charmaine Sutton Acting VP Of Regulatory 6464 Sycamore Court Minneapolis, Minnesota 55369
Re: K083784
Trade/Device Name: Pronto V3 Extraction Center, Pronto LP Extraction Catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: OEZ
Dear Charmaine Sutton:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated March 30, 2009. Specifically, FDA is updating this SE Letter as an administrative correction because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Gregory O'Connell, OHT2: Office of Cardiovascular Devices, (301) 796-6075. Gregory. Oconnell(@FDA.HHS.gov.
Sincerely,
Image /page/0/Picture/9 description: The image shows a digital signature. On the left side, the name "Gregory W. O'connell -S" is written in a large font. On the right side, the text "Digitally signed by Gregory W. O'connell -S Date: 2021.05.20 09:35:42 -04'00'" is written in a smaller font. The signature indicates that the document was signed by Gregory W. O'connell on May 20, 2021, at 09:35:42 with a time zone offset of -04'00'.
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern around the symbol. The caduceus is positioned to the right of the text, creating a balanced and recognizable emblem.
Public Health Service
MAR 3 0 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Vascular Solutions, Inc. c/o Ms. Charmaine Sutton, RAC 6464 Sycamore Court Minneapolis, MN 55369
Re: K083784
> Trade/Device Name: Pronto V3 extraction catheter, Pronto LP extraction catheter, Skyway RX support catheter, Skyway OTW support catheter, and Twin-Pass dual access catheter
Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: DXE, DQY Dated: March 16, 2009 Received: March 17, 2009
Dear Ms. Sutton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{2}------------------------------------------------
Page 2 - Ms. Charmaine Sutton
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Duna R. Turner
Image /page/2/Picture/5 description: The image shows a handwritten symbol or character. It appears to be a stylized or cursive form, possibly resembling a letter or a unique mark. The symbol is drawn with thick, dark lines, suggesting it was created with a pen or marker. The overall impression is that of a signature, initial, or abstract design.
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
{3}------------------------------------------------
510(k) Number (if known): _ _ _ 003789
Skyway® OTW support catheter Device Name: Skyway® RX support catheter
Indications for Use:
The SKYWAY support catheter is intended to be used in conjunction with steerable guidewires in order to access discreet regions of the arterial and or coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The SKYWAY OTW also may be used to subselectively infuse/deliver therapeutic agents.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Posted November 13, 2003)
3
{4}------------------------------------------------
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Twin-Pass® dual access catheter
Indications for Use:
The TWIN-PASS dual access catheter is intended to be used in conjunction with steerable guidewires in order to access discrete regions of the coronary and peripheral arterial vasculature, to facilitate placement and exchange of guidewires and other interventional devices, for use during two guidewire procedures and to subselectively infusc/deliver diagnostic or therapeutic agents.
Prescription Use X (Part 21 CFR 801 Subpart D)
ﻴﺔ
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Posted November 13, 2003)
{5}------------------------------------------------
510(k) Number (if known): K083784
.
Pronto® V3 extraction catheter Device Name: Pronto® LP extraction catheter
Indications for Use:
The PRONTO extraction catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
. При 1910 года продательно продобродности при полности и моремника по примество полнения по странция по подательности по полнения по полнения по подательные поведника с
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
5
(Posted November 13, 2003)
R.L. Jones
Division Sign-Off) avision of Cardiovascular Devices
510(k) Number_KO83784
{6}------------------------------------------------
Vascular Solutions, Inc.
Traditional 510(k) Premarket Notification Hydrophilic Coating Modification for Pronto V3, Pronto LP, Skyway and Twin-Pass Catheters
1683784
#### 3 510(k) Summary
Date Prepared: December 18, 2008
## Submitter's Name / Contact Person
#### Manufacturer
Vascular Solutions, Inc. 6464 Sycamore Court Minneapolis, MN 55369 USA Establishment Registration # 2134812
## Contact Person
Charmaine Sutton, RAC Acting VP of Regulatory Tel: 763.656.4349 (direct); Fax: 763.656.4253 Email: csutton@vascularsolutions.com
| Common Name<br>(Classification) | Trade Name | Predicate Device |
|-----------------------------------------------|---------------------------------|----------------------------------------|
| Embolectomy<br>catheter<br>(21 CFR 870.5150) | Pronto® V3 extraction catheter | K063371 Pronto V3 extraction catheter |
| | Pronto® LP extraction catheter | K072810 Pronto LP extraction catheter |
| Percutaneous<br>catheter<br>(21 CFR 870.1250) | Skyway® RX support catheter | K060327 Skyway RX support catheter |
| | Skyway® OTW support catheter | K060327 Skyway OTW support catheter |
| | Twin-Pass® dual access catheter | K060327 Twin-Pass dual access catheter |
#### General Information
## Device Description
The Pronto V3 extraction catheter is a dual lumen, hydrophilically coated, rapid exchange catheter with related accessories. The smaller wire lumen of the catheter is able to accommodate guidewires that are ≤ 0.014"/0.36mm in diameter. The larger extraction lumen allows for the removal of thrombus by use of the included syringe through the extension line and stopcock. The catheter has a rounded distal tip with a protected, sloped opening of the extraction lumen to facilitate advancement of the catheter into the blood vessel and maximize extraction of thrombus through the extraction lumen. The proximal end of the catheter has a standard luer adapter that attaches to the included extension line, stopcock and syringes. The catheter has a working length of 140cm and is compatible with standard 6F guide catheters. A 70um filter basket is included for assistance in filtering the blood removed during the procedure for laboratory analysis of any thrombosis.
The Pronto LP extraction catheter is a dual lumen, hydrophilically coated rapid exchange catheter with related accessories. The smaller wire lumen of the catheter is able to accommodate guidewires that are <0.014"/0.36mm in diameter. The larger extraction lumen comes pre-loaded with a stylet that resists kinking during delivery but is removed to allow for the removal of
MAR 3 0 2009
{7}------------------------------------------------
thrombus by aspiration. The catheter has a rounded distal tip with a protected, sloped opening of the extraction lumen to facilitate advancement of the catheter into the blood vessel and maximize extraction of thrombus the extraction lumen. The proximal end of the catheter has a Yadapter hub containing a self-sealing stylet port, as well as a side-port for the attachment of the included extension line. The catheter has a working length of 137cm and is compatible with standard 6F guide catheters. A 70um filter basket is included for assistance in filtering the blood removed during the procedure for laboratory analysis of any thrombosis.
The Skyway OTW (over-the-wire) and Skyway RX (rapid exchange) support catheters are single lumen, hydrophilically coated catheters designed for use in the arterial vasculature. The catheters provide support for 0.014"/0.36mm guidewires during interventional procedures, and allow for the exchange of one distally located guidewire for another one while maintaining access to distal vasculature. The Skyway RX has a rapid exchange port that allows for the insertion of the catheter over a short guidewire; a stiffening mandrel provides support and pushability during catheter insertion.
The Twin-Pass dual access catheter is a hydrophilically coated, dual lumen catheter designed for use in the arterial vasculature. The catheter provides support for 0.014"/0.36mm guidewires during interventional procedures, and the dual lumen design allows for the delivery of a second guidewire into distal vasculature while leaving the initial guidewire in place. A stiffening mandrel provides support and pushability during catheter insertion.
#### Intended Use / Indications
The PRONTO extraction catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.
The SKYWAY support catheter is intended to be used in conjunction with steerable guidewires in order to access discreet regions of the arterial and or coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The SKY WAY OTW also may be used to subselectively infuse/deliver therapeutic agents.
The TWIN-PASS dual access catheter is intended to be used in conjunction with steerable guidewires in order to access discrete regions of the coronary and peripheral arterial vasculature, to facilitate placement and exchange of guidewires and other interventional devices, for use during two guidewire procedures and to subselectively infuse/deliver diagnostic or therapeutic agents.
### Substantial Equivalence and Summary of Studies
Pronto, Skyway and Twin-Pass catheters are substantially equivalent in intended use / indications and design to their predicate devices. Hydrophilic coating design differences have been qualified through biomaterial assessments and verification testing, the results of which did not raise any new safety or performance questions.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with outstretched wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.
MAR 3 0 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Vascular Solutions, Inc. c/o Ms. Charmaine Sutton, RAC 6464 Sycamore Court Minneapolis, MN 55369
Re: K083784
> Trade/Device Name: Pronto V3 extraction catheter, Pronto LP extraction catheter, Skyway RX support catheter, Skyway OTW support catheter, and Twin-Pass dual access catheter Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy catheter Regulatory Class: Class II Product Code: DXE, DQY Dated: March 16, 2009 Received: March 17, 2009
Dear Ms. Sutton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations; Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{9}------------------------------------------------
Page 2 - Ms. Charmaine Sutton
CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Drma. R. Wulner
Image /page/9/Picture/5 description: The image shows a stylized, handwritten letter 'W'. The letter is formed with curved lines, giving it a flowing appearance. The base of the 'W' has a slight slant, and the overall design is simple and elegant.
Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
510(k) Number (if known): KO83784
Skyway® OTW support catheter Device Name: Skyway® RX support catheter
Indications for Use:
The SKYWAY support catheter is intended to be used in conjunction with steerable guidewires in order to access discreet regions of the arterial and or coronary vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The SK YWA Y OTW also may be used to subselectively infuse/deliver therapeutic agents.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Posted November 13, 2003)
{11}------------------------------------------------
510(k) Number (if known):
Device Name: Twin-Pass® dual access catheter
Indications for Use:
The TWIN-PASS dual access catheter is intended to be used in conjunction with steerable guidewires in order to access discrete regions of the coronary and peripheral arterial vasculature, to facilitate placement and exchange of guidewires and other interventional devices, for use during two guidewire procedures and to subselectively infuse/deliver diagnostic or therapeutic agents.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Posted November 13, 2003).
{12}------------------------------------------------
510(k) Number (if known): K083784
Pronto® V3 extraction catheter Device Name: Pronto® LP extraction catheter
Indications for Use:
The PRONTO extraction catheter is indicated for the removal of fresh, soft emboli and thrombi from vessels in the coronary and peripheral vasculature.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Posted November 13, 2003)
ima R.L. Jones
Division Sign-Off) Ivision of Cardiovascular Devices
510(k) Number_KO83784
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.