K083752 · Glustitch, Inc. · KMF · Feb 23, 2009 · General Hospital
Device Facts
Record ID
K083752
Device Name
GLUSEAL 90, MODEL GLU9010
Applicant
Glustitch, Inc.
Product Code
KMF · General Hospital
Decision Date
Feb 23, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5090
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
GluSeal® 90 liquid adhesive bandage is intended to cover minor cuts, scrapes, burns, and minor irritations of the skin and help protect them from infection.
Device Story
GluSeal® 90 is a cyanoacrylate-based liquid adhesive bandage. Applied topically to minor skin wounds; polymerizes rapidly upon contact with moisture (hydroxyl ions) to form a protective film. Used in clinical or home settings to cover minor cuts, scrapes, burns, and irritations. Provides a physical barrier to protect wounds from infection. Film remains adhered until natural re-epithelialization occurs or mechanical displacement. Benefits include rapid application and wound protection.
Clinical Evidence
No clinical data. Bench testing only. Studies included hydrolytic byproduct analysis, thermal effects (exotherm) study, and adhesive strength properties. Results showed GluSeal® 90 performed similarly to the control article (Indermil) and met specifications.
Technological Characteristics
Formulated cyanoacrylate monomer. Polymerizes to a thin film in <1 minute. Biocompatibility tested per ANSI/AAMI/ISO 10993. No specific material standards or software components described.
Indications for Use
Indicated for covering minor skin cuts, scrapes, burns, and irritations to protect against infection. Contraindicated for eye application, infected areas, draining wounds, mucosal surfaces, and patients with cyanoacrylate hypersensitivity.
Regulatory Classification
Identification
A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.
{0}------------------------------------------------
#### 12 2017 11:51 人 > Section 5: 510k) Su imary
. "",
. . . .
. . ..
The Summary of Safety and Effectiveness on the GluSeal® 90 reflects data available and represented at the time the submission was prepared, but caution should be exercised in interpreting the data. The results of future studies and or tests may require alterations of the conclusions or recommendations set forth.
and the state and the comments of
Q
ાં છે.
22 45 23 33 1 1 1 1 1
83752
| Applicant | Don Blacklock<br>GluStitch, Inc.<br>7188 Progress Way, #307<br>Delta, British Columbia<br>Canada V4G 1M6 |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | FEB 23 2009 |
| Telephone | 800-667-2130 |
| Facsimile | 877-450-4000 |
| Date | December 18, 2008 |
| Name | GluSeal® 90 |
| Classification | Liquid Bandage, 21 CFR 880.5090 |
| Predicate: | • K063202 Procurement Technology Systems, LLC, ProDerma, market<br>clearance date May 2, 2007; and<br>• K002338 Closure Medical Corporation's Liquidermtm Liquid<br>Adhesive Bandage, market clearance date January 29, 2001. |
| Description: | The GluSeal® 90 is a formulated compound of the cyanoacrylate series.<br>The compound, which exists in monomeric form in the plastic containers,<br>polymerizes extremely rapidly in the presence of anions, especially of<br>hydroxyl ions [in the presence of water]. It has the ability to adhere to<br>moist living tissues. It demonstrates a favorable tissue response and<br>reveals no toxic or foreign body reaction in humans. |
| Intended Use | GluSeal® 90 liquid adhesive bandage is intended to cover minor cuts,<br>scrapes, burns, and minor irritations of the skin and help protect them from<br>infection. |
| Contraindication: | Do not apply GluSeal® 90 adhesive to the eye(s). If contact with the eye(s)<br>occurs, keep the eye(s) closed and covered, and immediately contact an<br>ophthalmologist. No attempt should be made to open the eye(s). The<br>adhesive will lose its adhesion over time, between one and four days and<br>the eye(s) will open spontaneously with no damage.<br>Do not use on infected areas, or wounds that are draining. |
| Warning | Do not use on mucosal surfaces (e.g., oral cavity, lips).<br>Do not use if hypersensitive to cyanoacrylate. |
| Technological<br>Characteristics | When applied on a wound, the GluSeal® 90 product polymerizes to form a<br>thin, protective film, within less than a minute. The applied GluSeal® 90<br>has a high degree of adhesion strength. The GluSeal® 90 film remains<br>adhered to the tissue surface until the underlying tissue is sloughed through<br>natural re-epithelialization or until mechanically displaced. |
| Performance<br>Testing | The Hydrolytic Byproducts Analyses indicated that no other significant<br>degradation products were produced in the GluSeal® 90 formulated<br>cyanoacrylate. Additionally, the GluSeal® 90 samples produced similar<br>levels of formaldehyde as found in the Indermil samples.<br><br>The Thermal Effects Study concluded that the GluSeal® 90 had very little<br>exotherm of the adhesive. Furthermore, the GluSeal® 90 performed<br>similarly to the control article Indermil during these tests.<br><br>The Adhesive Strength Properties Study concluded that the GluSeal® 90<br>appears to have similar properties to the Indermil commercial tissue<br>adhesive |
| Substantial<br>Equivalency<br>Information | The GluSeal® 90 formulation has been subjected to the appropriate<br>biocompatibility testing in accordance with ANSI/AAMI/ISO 10993 and<br>the results have confirmed that the product is safe for its intended use.<br><br>GluSeal® 90 has also been subjected to mechanical and performance tests<br>demonstrating equivalence to the predicated devices. |
| Conclusion | Testing as shown that the GluSeal® 90 performs to its specifications,<br>operates as intended, is safe and effective, and is substantially equivalent to<br>legally marketed devices. |
.
{1}------------------------------------------------
# 510k) Summary continue _ KO 83 752
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
ir
The Summary of Safety and Effectiveness on the GluSeal® 90 rchects data available and represented at the time the submission was prepared, but caution should be exercised in interpreting the data. The results of future studies and or tests may require alterations of the conclusions or recommendations set forth.
--- .
# 5 / 002
"
প
of
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars extending from its back, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged around the upper portion of the circle.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## FEB 2 3 2009
GluStitch, Inc. % Mr. Don Blacklock President 307-7188 Progress Way Delta, British Columbia V4G 1M6 Canada
Re: K083752 Trade/Device Name: GluSeal® 90 Regulation Number: 21 CFR 880.5090 Regulation Name: Liquid bandage Regulatory Class: I Product Code: KMF
Dated: December 17, 2008
Received: December 17, 2008
Dear Mr. Blacklock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Don Blacklock
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biomctric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely a
fp
Mark N. Melkerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
ection 4: Indication for Use Summary
| 510(k) Number (if known): | | AQ2 7 C<br>ิโครที่ 1988 - 1998 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 | |
|---------------------------|--|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | | | |
GluSeal® 90 Device Name:
Indications For Use:
GluSeal® 90 liquid adhesive bandage is intended to cover minor cuts, scrapes, burns, and minor irritations of the skin and help protect them from infection.
Prescription Use _ (Per 21 CFR 801.109) OR
Over-The-Counter-Use
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Rest . and Neurological General, Rest
510(k) Number 14053152
4 / 001
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.