BELTBEAT WF-400

K083715 · Bonus Communicaciones Srl · DPS · Sep 15, 2009 · Cardiovascular

Device Facts

Record IDK083715
Device NameBELTBEAT WF-400
ApplicantBonus Communicaciones Srl
Product CodeDPS · Cardiovascular
Decision DateSep 15, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2340
Device ClassClass 2
AttributesPediatric

Indications for Use

This monitoring system is intended to be used by clinicians for ECG signals monitoring of ambulatory and nonambulatory adult and pediatric patients in health care facilities. This product is intended to be used by trained health care professionals. It is not intended for home use.

Device Story

BeltBeat is an ambulatory 5-lead ECG monitoring system for healthcare facilities. System comprises wearable transmitter (BeltBeat WF400) and central monitoring software (BeltBeat STC). Transmitter acquires ECG signals; transmits via WLAN to central station. Software displays real-time ECG waveforms and heart rate numerics for up to 12 patients simultaneously; generates alarms/alerts for clinical changes; provides full data recording for offline analysis. Operated by trained healthcare professionals in settings like cardiac rehab, postoperative recovery, emergency departments, and telemetry. Wireless connectivity allows patient mobility within WLAN coverage. System benefits include continuous cardiac surveillance, centralized monitoring efficiency, and configurable alarm parameters to support clinical decision-making.

Clinical Evidence

Bench testing only. No clinical data provided. System validation included verification of performance against standards EC13:2002/(R)2007, 60601-1-4 +A1:2000, and 60601-2-27:2005.

Technological Characteristics

5-lead ECG telemetry system. Components: BeltBeat WF400 wearable transmitter, BeltBeat STC monitoring software. Connectivity: WLAN. Energy: Battery-powered transmitter. Standards: EC13:2002/(R)2007, 60601-1-4 +A1:2000, 60601-2-27:2005, ISO 13485:2003, ISO 14971:2007. Software: Centralized monitoring for up to 12 patients, configurable alarms, real-time waveform display, and data recording.

Indications for Use

Indicated for ECG signal monitoring of ambulatory and nonambulatory adult and pediatric patients (≥10 kg) in healthcare facilities. Contraindicated for home use.

Regulatory Classification

Identification

An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K093715 SEP 1 5 2009 Image /page/0/Picture/2 description: The image contains a logo with the words "CYGNUS Medical" in a bold, sans-serif font. Above the text is a graphic element that appears to be a stylized representation of a swan or similar bird, possibly in flight or with its wings spread. The overall design is simple and professional, suggesting a company in the healthcare or medical field. ## BeltBeat 510/k} Summary of Safety and Effectiveness | Submitter | Bonus Comunicaciones SRL | | | |-------------|-------------------------------------------------------------|-----------------------|--| | Address | Fonrouge 1561 - Ciudad Autónoma de Buenos Aires - Argentina | | | | Contact: | Héctor Foggia, RA Manager | | | | Phone: | +54 11 4630-1600 ext. 133 | Fax: +54 11 4630-1619 | | | E-mail: | hfoggia@bonuscom.com.ar | | | | Device | | | | | Trade Name: | BeltBeat | | | | Regulation No. | Class | Product<br>Code | Regulation Description | |----------------|-------|-----------------|------------------------------------------------------------------| | 870.2300 | II | DRT | Cardiac monitor (including cardiotachometer and rate<br>alarm) | | 870.2910 | II | DRG | Radiofrequency physiological signal transmitter and<br>receiver. | ## Device Definition The BeltBeat is a comprehensive ambulatory system solution for ECG monitoring in intermediate care units for adult and pediatic patients. The foundation of the system is a transmitter (BeltBeat WF400) that can capture and transmit ECG signals that are then processed and displayed on the Information Center (BealtBeat STC). The Information Center generates alarms and recordings, thus notifying clinicians of changes in patients' conditions. The monitoring software BeltBeat STC displays in a computer screen ECG waveforms and heart rate numerics, and may be configured to generate patient alarms and alert messages. Additionally, full recording of patient data is available. The software is capable of monitoring up to 12 patients (BeltBeat WF400 units) simultaneously. BeltBeat system is intended for 5-lead ECG monitoring. The most likely locations for patients monitored by this system are: cardiac rehabilitation facilities, postoperative recovery rooms, ambulatory surgery, intermediate care, emergency departments, telemetry departments and in-hospital transport. This system is not intended for home use, or any other use outside health care facilities. BeltBeat WF400 monitor is not intended for use with pediatric patients (or infants) weighing less than 22 lbs (10 kgs). {1}------------------------------------------------ # KOB3712 Image /page/1/Picture/1 description: The image shows a logo for "Bonus Medical". The logo features a stylized graphic above the text. The graphic appears to be an abstract representation of medical or healthcare-related imagery. The text "Bonus Medical" is written in a bold, sans-serif font, with "Bonus" stacked above "Medical". ## BeltBeat 510(k) Summary of Safety and Effectiveness #### Referenced Standards - EC13:2002/(R)2007 - 60601-1-4 +A1:2000 - 60601-2-27:2005 - ISO 13485:2003 - ISO 14971:2007 #### Indications for Use, Intended Use This monitoring system is intended to be used by clinicians for ECG signals monitoring of ambulatory and nonambulatory adult and pediatric patients in health care facilities. This product is intended to be used by trained health care professionals. It is not intended for home use. #### Principles of operation This monitoring system is intended to be used by clinicians for ECG signals monitoring of ambulatory and nonambulatory adult and pediatric patients, in health care facilities. A patient wearable device BettBeat WF400 acquires ECG signals and transmits them through a Wireless Local Area Network (WLAN), to be processed and displayed by the monitoring software BeltBeat STC. Patient should therefore remain within a defined WLAN coverage area, in order not to interrupt the ECG signals transmission and recording. Wireless technology provides flexibility and freedom of movement, increasing the patient's comfort. Time Optimization, since every central station allows up to 12 patients to be monitored . simultaneously. - Flexibility, the system can be configured to adapt to every users' needs. . - Safety and Control, by means of patient alarms and alert messages. - Stand-by mode available, for energy saving. ● - Versatility of printing, Several parameters can be configured before printing reports. BeltBeat is a telemetric cardiac monitoring system, designed with state-of-the-art technology for acquisition and display of ECG signals. A standard wireless connection allows real-time transmission to a monitoring central station. Multiple leads of up to 12 patients can be monitored simultaneously, and different configurations can be selected in a simple manner. Full recording of patient data is also available, allowing physicians to analyze cardiac off-line. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows a logo with the word "BONUS" in a stylized font. Below the word "BONUS" is the word "digital" in a smaller font. Above the word "BONUS" are two oval shapes, one above the other. The oval shapes appear to have some texture or pattern within them. #### Substantially Equivalent Devices - 1. "Welch Allyn Micropaq Monitor (Model 406 / 408) Software version 1.6X + Acuity Central Station" - 2. "Philips Transmitters M2601B + IntelliVue Information Center (Philips Telemetry System)" #### Conclusions All Verification, Validation and Testing (VV&T) documents - constitute a part of the device DMR and are available upon request. The Truthful and Accurate Statement complies with 21 CFR 807.87(j). We trust that the submitted information will enable the reviewer to process the material promptly. The time factor is crucial for Bonus' commercial interests. This dossier contains one paper copy of the 510(k) submission together with an electronic copy that is an exact duplicate of the paper copy. {3}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or other bird with its wings spread. The bird is composed of three curved lines that suggest movement or flight. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 SEP 1 5 2009 Bonus Comunicaciones SRL Mr. Héctor Foggia Regulatory Affairs Manager Fonrouge 1561 Ciudad Autonoma de Buenos Aires Argentina C1440CYO Re: K083715 Trade/Device Name: BeltBeat Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II (two) Product Code: DPS, DRG and DRT Dated: Undated Received: September 09, 2009 # Dear Mr. Foggia: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Héctor Foggia Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, W.M.T. - Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: BeltBeat #### Indications for Use: This monitoring system is intended to be used by clinicians for ECG signals monitoring of ambulatory and nonambulatory adult and pediatric patients in health care facilities. This product is intended to be used by trained health care professionals. It is not intended for home use. Prescription Use 1×1 (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) W.M.S.F. Page 1 of 1 (Posted November 13, 2003) (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...