← Product Code [LLZ](/productcode/LLZ) · K083691

# BONEVUE (K083691)

_Micromri, Inc. · LLZ · Feb 25, 2009 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K083691

## Device Facts

- **Applicant:** Micromri, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Feb 25, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

BoneVue is a software tool for 3D visualization of bone micro-architecture using MRI datasets with the possibility of manual or automated measurements of descriptive parameters regarding cortical and trabecular morphology. BoneVue is a workflow tool intended to assist a trained physician in the assessment of high-resolution MRI datasets of bone.

## Device Story

BoneVue is a software tool for 3D visualization and analysis of high-resolution MRI bone datasets. Input: MRI datasets of bone tissue. Processing: Software generates 3D models of trabecular structures; performs manual or automated measurements of cortical morphology (inner/outer diameter, thickness) and trabecular morphology (bone volume/total volume, thickness, number, separation). Output: 3D visualizations (surface rendering, maximum intensity projection, 'bone plug' representative sample) and quantitative morphological parameters. Used in clinical settings by trained physicians to assist in qualitative and quantitative assessment of bone micro-architecture. Output aids physician decision-making by providing structural data to support clinical evaluation of bone health.

## Clinical Evidence

Bench testing only. The device underwent performance testing to verify that it met established acceptance criteria for its visualization and measurement functions. No clinical data was required or provided.

## Technological Characteristics

Software-only radiological image processing tool. Compatible with commercially available MRI pulse sequences and imaging coils. Does not control image acquisition or contact the patient. Provides 3D visualization (surface rendering, maximum intensity projection) and quantitative morphological analysis. Operates as a workflow tool for physician interpretation.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Voxar 3D ([K070831](/device/K070831.md))
- Aquarius Workstation ([K011142](/device/K011142.md))
- EVMS Enterprise Visual Medical System ([K053281](/device/K053281.md))
- Vitrea Version 4.0 Medical Image Processing Software ([K071331](/device/K071331.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KC083691/

510 (k) BoneVue Software v1

December 11, 2008

#### XII. 510(k) SUMMARY .

美女女女

# FEB 2 5 2009

This 510(K) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).

## Submitter's Information:

| Manufacturer: | MicroMRI, Inc.           |
|---------------|--------------------------|
|               | 580 Middletown Boulevard |
|               | Suite D-150.             |
|               | Langhorne, PA 19047      |

| Contact:   | Richard Elrath                                   |
|------------|--------------------------------------------------|
|            | Manager Quality Assurance and Regulatory Affairs |
|            | MicroMRI, Inc.                                   |
| Phone:     | 267 212-1119                                     |
| Facsimile: | 267 212-1101                                     |
| Email:     | relrath@micromri.com                             |

## Trade Name, Common Name and Classification:

| Trade Name:          | BoneVue                                |
|----------------------|----------------------------------------|
| Common Names:        | Image processing software              |
| Classification Name: | System, Image Processing, Radiological |
| Product Code:        | LLZ                                    |

## Predicate Devices:

| 510k Reference No. & Date    | Device Name                                                 | Manufacturer                                                                  |
|------------------------------|-------------------------------------------------------------|-------------------------------------------------------------------------------|
| K070831<br>May 22, 2007      | Voxar 3D™                                                   | Barco View MIS<br>2 Anderson Place<br>Edingburgh, EH6 5NP,<br>UK              |
| K011142<br>May 8, 2001       | Aquarius Workstation™                                       | TeraRecon, Inc.<br>2955 Campus Drive,<br>Suite 325<br>San Mateo, CA 94403     |
| K053281<br>September 3, 2004 | EVMS™ Enterprise<br>Visual Medical System                   | Emageon UV, Inc.<br>131 Wilson Street<br>Suite 700<br>Madison, WI 53703       |
| K071331<br>May 25, 2007      | Vitrea® Version 4.0<br>Medical Image<br>Processing Software | Vital Images, Inc.<br>5850 Opus Parkway,<br>Suite 300<br>Minnetonka, MN 55343 |

Confidential and Proprietary

Page 39 of 218

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image contains the word "microMRI" in a stylized font. Below the word "microMRI" is the phrase "bone micro-architecture". The text is in a simple sans-serif font. The image appears to be a logo or title for a product or service related to bone micro-architecture imaging.

## Device Description:

BoneVye evaluates high-resolution MRI datasets containing bone tissue and provides 3D visualization of trabecular structures as well as measurements of descriptive parameters regarding cortical and trabecular morphology.

The following cortical measurements are reported: average cortical inner diameter, average cortical outer diameter, and average cortical thickness.

The following trabecular measurements are reported: average measures of bone volume/total volume, trabecular thickness, trabecular number, and trabecular separation.

The 3D visualization module is used to display a high resolution 3D model of the trabecular bone and its micro-architecture. The 3D visualization helps a trained physician make a qualitative assessment of bone micro-architecture, which may be viewed from different angles. BoneVue allows standard surface rendering views as well as standard maximum intensity projection views.

Additionally, BoneVue offers the display of a standard "bone plug", a surface rendering of a central cylinder of trabecular tissue, as a representative sample of the bone microarchitecture.

### Indications for Use:

BoneVue is a software tool for 3D visualization of bone micro-architecture using MRI datasets with the possibility of manual or automated measurements of descriptive parameters regarding cortical and trabecular morphology. BoneVue is a workflow tool intended to assist a trained physician in the assessment of high-resolution MRI datasets of bone.

### Technological Characteristics:

BoneVue software does not control image acquisition and can be used with a variety of commercially available pulse sequences and imaging coils. BoneVue does not contact the patient, nor does it control any life sustaining devices. A trained physician interprets the data and information being displaved.

### Performance Testing:

BoneVue has been successfully tested and has met acceptance criteria previously established in accordance with documented procedures.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "MicroMRI" in a stylized font. Below the word "MicroMRI" is the phrase "bone micro-architecture". The text is in a light color, and the background is white.

510 (k) BoneVue Software v1 December 11, 2008

## Conclusion:

The 510(k) Pre-Market Notification contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate devices. BoneVue has been developed, and will be manufactured in accordance with the MicroMRI's established Quality Policy Manual, which meets all of the requirements of 21 CFR Part 820, Quality System Regulation and ISO 13485, Medical Devices - Quality Management Systems - Requirements for regulatory purposes. The submission contains the results of a hazard analysis and the "Level of Concern" for potential hazards has been classified as "Minor".

{3}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## FB 2 5 2009

Mr. Richard Elrath Manager of Quality Assurance and Regulatory Affairs MicroMRI, Inc. 580 Middletown Boulevard, Suite D-150 LANGHORNE PA 19047

Re: K083691

Trade/Device Name: BoneVue

Regulation Number: 21 CFR 892.2050

Regulation Name: Picture archiving and communications system

Regulatory Class: II

Product Code: LLZ

Dated: December 8, 2008

Received: December 15, 2008

## Dear Mr. Elrath:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set , forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for vour device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology)                      | (240) 276-0120 |
| Other          |                                  | (240) 276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.

Sincerely yours,

Laurie B. Morris

anine M. Morris Acting Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

510 (k) BoneVue Software v1

December 11, 2008

## XIII. STATEMENT of INDICATIONS FOR USE

Indications for Use

510(k) Number (if known): _ NA

Device Name: _BoneVue

Indications for Use:

BoneVue is a software tool for 3D visualization of bone micro-architecture using MRI datasets with the possibility of manual or automated measurements of descriptive parameters regarding cortical and trabecular morphology. BoneVue is a workflow tool intended to assist a trained physician in the assessment of high-resolution MRI datasets of bone.

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

| Concurrence/of CDRH, Office of Device Evaluation (ODE) |  |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices

08369 510(k) Number _

Confidential and Proprietary

---

**Source:** [https://fda-staging.innolitics.com/device/K083691](https://fda-staging.innolitics.com/device/K083691)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
