K083691 · Micromri, Inc. · LLZ · Feb 25, 2009 · Radiology
Device Facts
Record ID
K083691
Device Name
BONEVUE
Applicant
Micromri, Inc.
Product Code
LLZ · Radiology
Decision Date
Feb 25, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
BoneVue is a software tool for 3D visualization of bone micro-architecture using MRI datasets with the possibility of manual or automated measurements of descriptive parameters regarding cortical and trabecular morphology. BoneVue is a workflow tool intended to assist a trained physician in the assessment of high-resolution MRI datasets of bone.
Device Story
BoneVue is a software tool for 3D visualization and analysis of high-resolution MRI bone datasets. Input: MRI datasets of bone tissue. Processing: Software generates 3D models of trabecular structures; performs manual or automated measurements of cortical morphology (inner/outer diameter, thickness) and trabecular morphology (bone volume/total volume, thickness, number, separation). Output: 3D visualizations (surface rendering, maximum intensity projection, 'bone plug' representative sample) and quantitative morphological parameters. Used in clinical settings by trained physicians to assist in qualitative and quantitative assessment of bone micro-architecture. Output aids physician decision-making by providing structural data to support clinical evaluation of bone health.
Clinical Evidence
Bench testing only. The device underwent performance testing to verify that it met established acceptance criteria for its visualization and measurement functions. No clinical data was required or provided.
Technological Characteristics
Software-only radiological image processing tool. Compatible with commercially available MRI pulse sequences and imaging coils. Does not control image acquisition or contact the patient. Provides 3D visualization (surface rendering, maximum intensity projection) and quantitative morphological analysis. Operates as a workflow tool for physician interpretation.
Indications for Use
Indicated for trained physicians to assist in the assessment of high-resolution MRI datasets of bone, specifically for 3D visualization and manual or automated measurement of cortical and trabecular morphology.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Vitrea Version 4.0 Medical Image Processing Software (K071331)
Submission Summary (Full Text)
{0}------------------------------------------------
KC083691/
510 (k) BoneVue Software v1
December 11, 2008
#### XII. 510(k) SUMMARY .
美女女女
# FEB 2 5 2009
This 510(K) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
## Submitter's Information:
| Manufacturer: | MicroMRI, Inc. |
|---------------|--------------------------|
| | 580 Middletown Boulevard |
| | Suite D-150. |
| | Langhorne, PA 19047 |
| Contact: | Richard Elrath |
|------------|--------------------------------------------------|
| | Manager Quality Assurance and Regulatory Affairs |
| | MicroMRI, Inc. |
| Phone: | 267 212-1119 |
| Facsimile: | 267 212-1101 |
| Email: | relrath@micromri.com |
## Trade Name, Common Name and Classification:
| Trade Name: | BoneVue |
|----------------------|----------------------------------------|
| Common Names: | Image processing software |
| Classification Name: | System, Image Processing, Radiological |
| Product Code: | LLZ |
## Predicate Devices:
| 510k Reference No. & Date | Device Name | Manufacturer |
|------------------------------|-------------------------------------------------------------|-------------------------------------------------------------------------------|
| K070831<br>May 22, 2007 | Voxar 3D™ | Barco View MIS<br>2 Anderson Place<br>Edingburgh, EH6 5NP,<br>UK |
| K011142<br>May 8, 2001 | Aquarius Workstation™ | TeraRecon, Inc.<br>2955 Campus Drive,<br>Suite 325<br>San Mateo, CA 94403 |
| K053281<br>September 3, 2004 | EVMS™ Enterprise<br>Visual Medical System | Emageon UV, Inc.<br>131 Wilson Street<br>Suite 700<br>Madison, WI 53703 |
| K071331<br>May 25, 2007 | Vitrea® Version 4.0<br>Medical Image<br>Processing Software | Vital Images, Inc.<br>5850 Opus Parkway,<br>Suite 300<br>Minnetonka, MN 55343 |
Confidential and Proprietary
Page 39 of 218
{1}------------------------------------------------
Image /page/1/Picture/2 description: The image contains the word "microMRI" in a stylized font. Below the word "microMRI" is the phrase "bone micro-architecture". The text is in a simple sans-serif font. The image appears to be a logo or title for a product or service related to bone micro-architecture imaging.
## Device Description:
BoneVye evaluates high-resolution MRI datasets containing bone tissue and provides 3D visualization of trabecular structures as well as measurements of descriptive parameters regarding cortical and trabecular morphology.
The following cortical measurements are reported: average cortical inner diameter, average cortical outer diameter, and average cortical thickness.
The following trabecular measurements are reported: average measures of bone volume/total volume, trabecular thickness, trabecular number, and trabecular separation.
The 3D visualization module is used to display a high resolution 3D model of the trabecular bone and its micro-architecture. The 3D visualization helps a trained physician make a qualitative assessment of bone micro-architecture, which may be viewed from different angles. BoneVue allows standard surface rendering views as well as standard maximum intensity projection views.
Additionally, BoneVue offers the display of a standard "bone plug", a surface rendering of a central cylinder of trabecular tissue, as a representative sample of the bone microarchitecture.
### Indications for Use:
BoneVue is a software tool for 3D visualization of bone micro-architecture using MRI datasets with the possibility of manual or automated measurements of descriptive parameters regarding cortical and trabecular morphology. BoneVue is a workflow tool intended to assist a trained physician in the assessment of high-resolution MRI datasets of bone.
### Technological Characteristics:
BoneVue software does not control image acquisition and can be used with a variety of commercially available pulse sequences and imaging coils. BoneVue does not contact the patient, nor does it control any life sustaining devices. A trained physician interprets the data and information being displaved.
### Performance Testing:
BoneVue has been successfully tested and has met acceptance criteria previously established in accordance with documented procedures.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "MicroMRI" in a stylized font. Below the word "MicroMRI" is the phrase "bone micro-architecture". The text is in a light color, and the background is white.
510 (k) BoneVue Software v1 December 11, 2008
## Conclusion:
The 510(k) Pre-Market Notification contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate devices. BoneVue has been developed, and will be manufactured in accordance with the MicroMRI's established Quality Policy Manual, which meets all of the requirements of 21 CFR Part 820, Quality System Regulation and ISO 13485, Medical Devices - Quality Management Systems - Requirements for regulatory purposes. The submission contains the results of a hazard analysis and the "Level of Concern" for potential hazards has been classified as "Minor".
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## FB 2 5 2009
Mr. Richard Elrath Manager of Quality Assurance and Regulatory Affairs MicroMRI, Inc. 580 Middletown Boulevard, Suite D-150 LANGHORNE PA 19047
Re: K083691
Trade/Device Name: BoneVue
Regulation Number: 21 CFR 892.2050
Regulation Name: Picture archiving and communications system
Regulatory Class: II
Product Code: LLZ
Dated: December 8, 2008
Received: December 15, 2008
## Dear Mr. Elrath:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set , forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for vour device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.
Sincerely yours,
Laurie B. Morris
anine M. Morris Acting Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
510 (k) BoneVue Software v1
December 11, 2008
## XIII. STATEMENT of INDICATIONS FOR USE
Indications for Use
510(k) Number (if known): _ NA
Device Name: _BoneVue
Indications for Use:
BoneVue is a software tool for 3D visualization of bone micro-architecture using MRI datasets with the possibility of manual or automated measurements of descriptive parameters regarding cortical and trabecular morphology. BoneVue is a workflow tool intended to assist a trained physician in the assessment of high-resolution MRI datasets of bone.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
| Concurrence/of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
08369 510(k) Number _
Confidential and Proprietary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.