← Product Code [KWA](/productcode/KWA) · K083642

# DEPUY PINNACLE METAL-ON-METAL ACETABULAR CUP LINERS (K083642)

_DePuy Orthopaedics, Inc. · KWA · Mar 6, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K083642

## Device Facts

- **Applicant:** DePuy Orthopaedics, Inc.
- **Product Code:** [KWA](/productcode/KWA.md)
- **Decision Date:** Mar 6, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3330
- **Device Class:** Class 3
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Pinnacle Metal-on-Metal Acetabular Cup Liners are indicated for use as the acetabular component in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for patients with congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis and disability due to previous fusion, where bone stock is inadequate for other reconstruction techniques. The Pinnacle Metal-on-Metal Acetabular Cup Liners are intended for use with DePuy Pinnacle Acetabular Shells and M-Spec Co-Cr-Mo femoral heads only.

## Device Story

Acetabular cup liner component for total hip replacement; used with DePuy Pinnacle Acetabular Shells and M-Spec Co-Cr-Mo femoral heads. Device provides articulating surface for hip joint. Modification to labeling only; no changes to device design. Contraindications added for chronic renal failure and females of childbearing age due to potential metal ion effects. Intended for surgical use by orthopedic surgeons.

## Clinical Evidence

No clinical data provided; submission is for labeling revisions only.

## Technological Characteristics

Metal-on-metal acetabular cup liner; Co-Cr-Mo alloy construction; semi-constrained hip joint prosthesis; uncemented acetabular component.

## Regulatory Identification

A hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. The femoral component is intended to be fixed with bone cement. The acetabular component is intended for use without bone cement (§ 888.3027).

## Predicate Devices

- DePuy Pinnacle Metal-on-Metal Acetabular Cup Liners ([K062426](/device/K062426.md))
- DePuy Pinnacle Metal-on-Metal Acetabular Cup Liners ([K023786](/device/K023786.md))
- DePuy Pinnacle Metal-on-Metal Acetabular Cup Liners ([K003523](/device/K003523.md))
- DePuy Pinnacle Metal-on-Metal Acetabular Cup Liners ([K002883](/device/K002883.md))
- Biomet Metal-on-Metal Hip Systems ([K082446](/device/K082446.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K083642 (pg 1/2)

# MAR 6 2009

# 510 (k) Summary (As required by 21 CFR 807.92 and 21 CFR 807.93)

#### NAME OF SPONSOR:

510(K) CONTACT:

DePuy Orthopaedics, Inc. 700 Orthopaedic Drive Warsaw, Indiana 46582 Establishment Registration Number: 1818910

Dawn Sinclair Regulatory Affairs Associate Telephone: (574) 372-5023 Facsimile: (574) 371-4987 Electronic Mail: Dsincla3@its.jnj.com

DATE PREPARED:

PROPRIETARY NAME:

COMMON NAME:

CLASSIFICATION:

December 5, 2008

87 KWA

DePuy Pinnacle® Metal-on-Metal Acetabular Cup Liners

Acetabular Cup Prosthesis

Class III per 21 CFR 888.3330, Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis

#### DEVICE PRODUCT CODE:

SUBSTANTIALLY EQUIVALENT DEVICE(S):

DePuy Pinnacle® Metal-on-Metal Acetabular Cup Liners, K062426 DePuy Pinnacle® Metal-on-Metal Acetabular Cup Liners, K023786 DePuy Pinnacle® Metal-on-Metal Acetabular Cup Liners, K003523 DePuy Pinnacle® Metal-on-Metal Acetabular Cup Liners, K002883 Biomet® Metal-on-Metal Hip Systems, K082446

{1}------------------------------------------------

K083642.(pg 2/2)

## DESCRIPTION:

The subject of this 510(k) consists of modifications to the Instructions for Use (IFU). The following two contraindications and potential adverse effect have been added to the IFU. DePuy Orthopaedics is making these IFU revisions to comply with a request from Health Canada, as well as to update the contents to reflect current industry practice. These are the only changes to the labeling and IFU; there are no changes to the device design.

Contraindications:

- Use is contraindicated in cases with chronic renal failure. .
- Females of childbearing age are contraindicated due to the unknown effects of . elevated levels of metal ions on the fetus.

Potential adverse effect:

- The potential long-term biological effects of metal wear debris and metal ion . production are not known.
No changes have been made to the Indications for Use or the Intended Use of this device.

# INDICATIONS AND INTENDED USE:

## Indications:

The Pinnacle Metal-on-Metal Acetabular Cup Liners are indicated for use as the acetabular component in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for patients with congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis and disability due to previous fusion, where bone stock is inadequate for other reconstruction techniques.

#### Intended Use:

The Pinnacle Metal-on-Metal Acetabular Cup Liners are intended for use with DePuy Pinnacle Acetabular Shells and M-Spec Co-Cr-Mo femoral heads only.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DePuy Orthopaedics, Inc. % Ms. Dawn Sinclair Regulatory Associate, Regulatory Affairs 700 Orthopaedic Dr. Warsaw, Indiana 46581

MAR 6 2009

Re: K083642

Trade/Device Name: DePuy Pinnacle Metal-on-Metal Acetabular Cup Liners Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint, metal/metal semi-constrained, with an uncemented acctabular component, prosthesis Regulatory Class: III Product Code: KWA Dated: December 8, 2008

Dear Ms. Sinclair:

Received: December 9, 2008

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Dawn Sinclair

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Millican

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use Statement

510 (k) Number (if known): 16083642 (po 1/1)

Device Name: DePuy Pinnacle® Metal-on-Metal Acetabular Cup Liners

#### Indications for Use:

The Pinnacle Metal-on-Metal Acetabular Cup Liners are indicated for use as the acetabular component in total hip replacement procedures for patients suffering severe pain and disability due to structural damage in the hip joint from rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, and nonunion of femoral fractures. Use of the prosthesis is also indicated for patients with congenital hip dysplasia, protrusio acetabuli, silipped capital femoral epiphysis and disability due to previous fusion, where bone stock is inadequate for other reconstruction techniques.

The Pinnacle Metal-on-Metal Acetabular Cup Liners are intended for use with DePuy Pinnacle Acetabular Shells and M-Spec Co-Cr-Mo femoral heads only.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

() x Sh Yo

(Please do not write below this line. Continue on another page if needed.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Posted November 13, 2003)

CDRH, Office of Device Evaluation (ODE)

Mark H. Milliken

Page 1 of 1

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

0000

510(k) Number

---

**Source:** [https://fda-staging.innolitics.com/device/K083642](https://fda-staging.innolitics.com/device/K083642)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).