U-RIGHT TD-4249/U-RIGHT TD-4250 BLOOD GLUCOSE MONITORING SYSTEM

K083570 · Taidoc Technology Corporation · CGA · Mar 11, 2009 · Clinical Chemistry

Device Facts

Record IDK083570
Device NameU-RIGHT TD-4249/U-RIGHT TD-4250 BLOOD GLUCOSE MONITORING SYSTEM
ApplicantTaidoc Technology Corporation
Product CodeCGA · Clinical Chemistry
Decision DateMar 11, 2009
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

The U-RIGHT TD-4249/ U-RIGHT TD-4250 Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and is not intended for use on neonates. The alternative site testing in the U-RIGHT TD-4249/ U-RIGHT TD-4250 Blood Glucose Monitoring System can be used only during steady-state blood glucose conditions.

Device Story

System measures glucose in fresh capillary whole blood via finger or alternative sites (palm, forearm, upper-arm, calf, thigh). Used by healthcare professionals or patients at home to monitor diabetes control effectiveness. Not for diagnosis or screening. Operates via electrochemical biosensor technology; provides quantitative glucose concentration output on device display. Assists in diabetes management; alternative site testing requires steady-state blood glucose conditions.

Clinical Evidence

No clinical data provided. Substantial equivalence established via design control activities, risk analysis, and verification/validation of the modified PCB and physical housing.

Technological Characteristics

Handheld blood glucose meter; electrochemical sensing principle. Modified PCB includes two integrated circuits for voice function language selection. Device dimensions and button functionality remain consistent with predicate. Sterilization and material specifications not detailed.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary whole blood (finger, palm, forearm, upper arm, calf, thigh) for healthcare professionals and patients with diabetes mellitus to monitor glycemic control. Not for diabetes diagnosis, screening, or neonatal use.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Submission Summary (Full Text)

{0} SPECIAL 510(k): Device Modification ODE Review Memorandum (Decision Making Document is Attached) To: THE FILE RE: DOCUMENT NUMBER k083570 This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary): 1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) k073119 TaiDoc Technology corporation's Clever Check TD-4232 Blood Glucose Test System. 2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use). 3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed. This change was for a change in the appearance of the device. The size is similar and the number of buttons and functions has not changed, but the appearance of the meter has been slightly modified. Additionally, the printed circuit board (PCB) was changed. The only difference between the old PCB and the new one is that the old one only has one integrated circuit for selecting the language for voice functions while the new one has two integrated circuits. 4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics. 5. A Design Control Activities Summary which includes: a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied c) A declaration of conformity with design controls. The declaration of conformity should include: i) A statement signed by the individual responsible, that, as required by the risk analysis, all verification and validation activities were performed by the designated individual(s) and the results demonstrated that the predetermined acceptance criteria were met, and ii) A statement signed by the individual responsible, that the manufacturing facility is in conformance with design control procedure requirements as specified in 21 CFR 820.30 and the records are available for review. 6. A Truthful and Accurate Statement, a 510(k) Summary or Statement and the Indications for Use Enclosure (and Class III Summary for Class III devices). The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...