K083541 · Advanced Imaging Research, Inc. · MOS · Feb 24, 2009 · Radiology
Device Facts
Record ID
K083541
Device Name
NEONATE HEAD AND BODY COILS
Applicant
Advanced Imaging Research, Inc.
Product Code
MOS · Radiology
Decision Date
Feb 24, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Neonate (Head, Body) Coil is a quadrature transmit/receive coil used for obtaining diagnostic images and spectra of the (pre- and term) infant head in 1,5 and 3,0 Tesla General Electric, Siemens Medical Solutions and Philips Medical Systems' Magnetic Resonance Imaging (MRI) scanners and magnetic resonance compatible incubator systems. Indications for use are same as that for standard MR imaging.
Device Story
Neonate (Head, Body) Coil is a 12-element quadrature transmit/receive MRI coil. Device captures RF signals from infant head/body during MRI scanning. Coil elements and circuitry are enclosed in rigid polycarbonate/polyurethane housing to prevent patient/environment exposure. Used in clinical settings (e.g., radiology departments) by trained MRI technicians/physicians. Device integrates with 1.5T and 3.0T MRI scanners from GE, Siemens, and Philips. Output consists of diagnostic images/spectra (T1, T2, proton density, diffusion, angiography, spectroscopy) displayed on MRI console for physician interpretation. Benefits include specialized sizing for infants, facilitating diagnostic imaging of neonates. Safety features include pin-diode isolation to prevent RF burns and cable routing to prevent resonance loops.
Clinical Evidence
Bench testing only. Testing demonstrated that 3.0T Neonate Head Coils met predetermined acceptance criteria for performance and safety.
Technological Characteristics
12-element quadrature transmit/receive coil. Materials: Polycarbonate, polyurethane plastic. Connectivity: Interfaces with 1.5T/3.0T MRI scanners. Safety: Fast active pin-diode circuitry for transmit/receive isolation and RF burn prevention; RF power deposition limited by scanner-embedded coil-specific files and SAR algorithms.
Indications for Use
Indicated for imaging of the infant head and body (torso) in pre-term and term infants. Used in conjunction with MRI systems or MRI-compatible incubators. Images are for interpretation by a trained physician to assist in diagnosis.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/3 description: The image shows a handwritten string of characters. The string appears to be a combination of letters and numbers. The characters are 'K083541'.
# FEB 2 4 2009
1. Manufacturer : Advanced Imaging Research, Inc. 4700 Lakeside Avenue, Suite 400 (Physical Location) Cleveland, OHIO 44114 PO Box 603220 (Use for Correspondence) Cleveland, OHIO 44103 1262 E 49th Street, Suite 400 (Courier Address) Cleveland, OHIO 44114 Phone: 216-426-1461 216-426-1180 Fax: 2. Registration Number: 3004036149 3. Contact: Ravi Srinivasan, President Phone: 216-426-1461 Fax: 216-426-1180 email: ravi(@advimg.com 4. Device Name: Magnetic Resonance Diagnostic Device Proprietary Name: Neonate Head Coil, Neonate Body Coil ડી Type of Submission: Traditional 6. Classification of Device: 21CFR 892.1000 Class of Device: Class II Product Code; 90 MOS (Magnetic Resonance Specialty Coil) 7. Intended Use: The Neonate (Head, Body) Coil is a quadrature transmit/receive coil used for obtaining diagnostic images and spectra of the (pre- and term) infant head in 1,5 and 3,0 Tesla General Electric, Siemens Medical Solutions and Philips Medical Systems' Magnetic Resonance Imaging (MRI) scanners and magnetic resonance compatible incubator systems. Indications for use are same as that for standard MR imaging.
- 8. Device Description:
The Neonate (Head, Body) Coil is a 12 element quadrature,
2
{1}------------------------------------------------
p.4
transmit/receive coil. Coil elements and associated circuitry are enclosed in a rigid former to prevent any exposure to patient or environment. The coil design facilitates scanning of pre-term and term newborn babies with different head and body sizes and maximizes comfort and ease of use.
Neonate Head Coil for GE, Siemens and Philips 3.0T Neonate Head and Body Coils for Philips 1.5T
3.0T Philips T/R Head Coil 3.0T Siemens T/R Head and Knee Coils (manufactured by [formerly USA Instruments], GE Medical, Aurora, OH) 1.5T Neonate Body Coil (K023929)
In general, Neonate Coils are similar to Advanced Imaging Research's FDA cleared 1.5T T/R Neonate (Head, Body) Coils in all aspects (K023929), as follows:
Intended use - Proton 1H 1D, 2D, 3D T1, T2 weighting, proton density, chemical shift, diffusion weighted imaging, angiography, functional MRI, chemical shift imaging, spectroscopy
Indications for Use - The Neonate Coils when used in conjunction with the MRI systems or Neonate coils when used with the MRI compatible incubator and in conjunction with the MRI systems provide imaging of the infant head and body (torso). When interpreted by a trained physician, these images provide information that can be useful in the determination of the diagnosis.
> Coil Enclosure Material - Polycarbonate, polyurethane plastic ..
Coil Design - 12 element, quadrature transmit/receive design
Transmit/Receive Switching - Fast active pin-diode circuitry
Prevention of RF Burns - Transmit and receive channels are isolated with pin-diodes; coil elements and associated circuitry are enclosed in plastic housing
Radio-Frequency Absorption - Radio-frequency power deposition is limited in the coil specific file embedded in the MRI and by SAR algorithm of the MRI scanner
#### ਹੈ. Marketed Device:
10.
Comparison to Predicate:
{2}------------------------------------------------
Formation of Resonance Loops - Active diodes and RF fuses isolate the coil elements from whole body RF coil transmit; length and routing of output cable does not permit looping
11. Summary of Studies:
Testing was performed to demonstrate that the 3.0 T Neonate Head Coils meet the predetermined acceptance criteria
### Conclusion:
It is the opinion of Advanced Imaging Research that the Neonate (Head, Body) Coils herein are substantially equivalent to Advanced Imaging Research's Neonate (Head, Body) Coils (K023929). Testing and usage of the 3.0T Neonate Coils does not result in any new potential hazards.
4
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 9 4 2009
Mr. Ravi Srinivasan, M.S. President Advanced Imaging Research, Inc. 4700 Lakeside Avenue, Suite 400 CLEVELAND OH 44114-3834
Re: K083541
Trade/Device Name: Neonate Coils Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: November 21, 2008 Received: January 15, 2009
Dear Mr. Srinivasan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and . adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean
that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.suppot/index.html.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
、
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ K083541
Device Name: _________________________________________________________________________________________________________________________________________________________________ Neonate Coils
Indications For Use;
The Neonate Colls when used in conjunction with the MRI systems or Neonate colls when used with the MRI compatible incubator and in conjunction with the MRI systems provide imaging of the infant head and body (torso). When interpreted by a trained physician, these images provide Information that can be useful in the determination of the diagnosis.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) :
Concurrence of ODRH, Office of Device Evaluation (ODE)
Signature
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number J
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.