COOK VASCULAR VITAL-JECT POWER INJECTABLE SAFETY INFUSION SET
K083482 · Cook Vascular, Inc. · FPA · Apr 2, 2009 · General Hospital
Device Facts
Record ID
K083482
Device Name
COOK VASCULAR VITAL-JECT POWER INJECTABLE SAFETY INFUSION SET
Applicant
Cook Vascular, Inc.
Product Code
FPA · General Hospital
Decision Date
Apr 2, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cook Vascular VITAL-JECT Power Injectable Safety Infusion Set is intended to be used to administer solutions and medications into vascular implant ports. When used with a power Injectable vascular implant port, the Cook Vascular Safety Infusion Set is indicated for power injection of contract media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s using contrast with a maximum viscosity of 11.8cP. In addition, to minimize the risk of accidental needle stick after use, the attached safety guard fully encapsulates the needle when manually activated during withdrawal.
Device Story
The VITAL-JECT is a safety infusion set for vascular implant ports; consists of protector cap, female luer lock, PVC tubing, pinch clamp, needle-free y-site, butterfly wing, safety needle guard, and 19 ga AISI 304 stainless steel needle. Used by clinicians to administer medications or power-inject contrast media. Safety guard manually activated during withdrawal to encapsulate needle, reducing accidental stick risk. Device performance validated for power injection up to 5 mL/s with 11.8 cP viscosity media. Benefits include safe vascular access and reduced sharps injury risk.
Clinical Evidence
Bench testing only. Tests included dynamic failure flow, instantaneous burst, static pressure, and life cycle power injection. Sterilization validation performed via ETO cycle with bioburden, endotoxin, and EO residual testing.
Indicated for administration of solutions and medications into vascular implant ports and power injection of contrast media (max 5 mL/s, max viscosity 11.8 cP) when used with power-injectable ports. Includes safety guard to minimize accidental needle stick risk.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
K083.482
## 510(k) Summary
Submitted by:
Jim Ferguson Quality Systems Manager Cook Vascular, Incorporated 1186 Montgomery Lane Vandergrift. Pa 15690 724-845-8621, XT 2227 April 10, 2006
Device:
| Trade name: | Cook Vascular VITAL-JECT <sup>TM</sup> Power Injectable Safety Infusion Set |
|-------------|-----------------------------------------------------------------------------|
|-------------|-----------------------------------------------------------------------------|
Intravascular Administration Set, 880.5440 Proposed Classification:
### Predicate Devices:
The Cook Vascular VITAL-JECT™ Power Injectable Safety Infusion Set is similar in terms of intended use, and exactly the same as materials of construction and technology characteristics to the predicate devices that have been found substantially equivalent.
#### Device Description:
The Cook Vascular VITAL-JECT Power Injectable Safety Infusion Set is intended to administer solutions and medications into vascular implant ports. In addition, to minimize the risk of accidental needle stick after use, the attached safety guard fully encapsulates the needle when manually activated during withdrawal.
When used with a power Injectable vascular implant port, the Cook Vascular Safety Infusion Set is indicated for power injection of contract media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s using contrast with a maximum viscosity of 11.8cP.
The device consists of a protector cap, female luer lock, PVC tubing, two each 4 inch tubing length, a pinch clamp, with a needle free y-site, a butterfly wing, a safety needle guard, an AISI 304 (19 ga) stainless steel needle and a needle sheath.
### Substantial Equivalence:
The components and the processes used to manufacture these solution administration sets are the exact same as the currently legally marketed by Command Medical, HuberPRO™ Safety Huber Infusion Set. (K033515). The packaged Infusion set will be supplied from Command Medical and Cook Vascular will label, sterilize and market the product.
● Cook Vascular, Inc. ● 510(k) Premarket Submission ● VITAL-JECT™ Power Injectable Safety Infusion Set ● Confidential
APR - 2 2009
{1}------------------------------------------------
### Test Data:
The Cook Vascular, Inc. VITAL-JECT™ Power Injectable Safety Infusion Set was subjected to the flowing tests to assure reliable design and performance under the specified testing parameters. These tests were comprised of:
- Dynamic Failure Flow Test 1.
- 2. Instantaneous Burst Test
- Static Pressure Test 3.
- 4. Life Cycle Power Injection Test
An analysis of the test results lead to the conclusion that the Vital-Ject infusion set is capable of being power injected at the establish 5mL/s with a given media viscosity of 11.8cP.
The results of these tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its use as a VITAL-JECT™ Power Injectable Safety Infusion Set
### Sterilization Information:
The infusion sets were validated to a standard 100% ETO Cook sterilization cycle with the Infusion Port Universal Tray set. Bioburden Testing, Endotoxin Testing, EO residual testing, and sterility testing was performed. EO Residual testing and sterility testing was performed on the Infusion Port body (model IP-S6110) which has been determined to provide the greatest challenge when testing for EO residuals and sterility. The test results for Endotoxin testing. EO Residual testing and sterility testing all indicated acceptable results. Bioburden testing on the infusion sets using combination membrane filtration and 3-layer pour plates was used. All individual tested infusion sets were within specification with the exception of one. The cfuldevice for this sample measured 1126. This result is above the alert level of 300 colony forming units per device. The other eight test samples cfu level ranged from 3 to 56 cfuldevice. The average Bioburden level for all nine samples is 149 cfu with is far below the 300 cfu alert limit. In addition, the sterilization validation and Endotoxin results for the Infusion sets indicate that the established sterilization process is capable of producing acceptable results.
Based on the fact the Cook Vascular, Inc. VITAL-JECT Power Injectable Safety Infusion Set utilizes the same design, components and manufacturing processes as currently legally marketed products, the Cook Vascular, Inc. VITAL-JECT Power Injectable Safety Infusion Set is safe and effective when used as intended.
• Cook Vascular, Inc. • 510(k) Premarket Submission • VITAL-JECT™ Power Injectable Safety Infusion Set • Confidential
12
{2}------------------------------------------------
# Device Description
The components and the processes use to manufacture these solution administration sets are the same as the currently legally marketed by Command Medical, K033515 HuberPRO™ Safety Huber Infusion Set.
Cook Vascular, Inc. will label and sterilize the final product. The proposed device is manufactured exactly the same as the currently marketed device. (K033515)
Drawings are included in Attachment B.
## Functional Features
A comparison of features and specifications of the Cook Vascular, Inc. VITAL-JECT™ Power Injectable Safety Infusion Set are included on page 7.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services in the USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. In the center of the seal is an abstract emblem that resembles an eagle or other bird in flight. The emblem is stylized and composed of curved lines, giving it a modern and symbolic appearance.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## APR - 2 2009
Mr. Jim Ferguson, Jr. Cook Vascular Incorporated 1186 Montgomery Lane Vandergrift, Pennsylvania 15690
K083482 Re:
> Trade/Device Name: Cook Vascular VITAL-JECT™ Power Injectable Safety Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: March 12, 2009 Received: March 17, 2009
Dear Mr. Ferguson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 -- Mr. Ferguson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Cinthan O.nne for
Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
1083482
510(k) Number: Device Name: Cook Vascular, Inc. Vital-Ject Power Injectable Safety Infusion Set Indications for Use:
The Cook Vascular VITAL-JECT Power Injectable Safety Infusion Set is intended to be used to administer solutions and medications into vascular implant ports. When used with a power Injectable vascular implant port, the Cook Vascular Safety Infusion Set is indicated for power injection of contract media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s using contrast with a maximum viscosity of 11.8cP.
In addition, to minimize the risk of accidental needle stick after use, the attached safety guard fully encapsulates the needle when manually activated during withdrawal.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Anesthesiology, General Hospital
infection Control, Dental Devices
| 510(k) Number: | K083482 |
|----------------|---------|
|----------------|---------|
| Prescription Use <span></span> X | OR Over-The-Counter Use <span></span> __ |
|----------------------------------|------------------------------------------|
|----------------------------------|------------------------------------------|
· Cook Vascular, Inc. · 510(k) Premarket Submission · Power Injectable Infusion Set · Confidential ·
б
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.