CARPENTIER- EDWARDS PHYSIO II ANNULOPLASTY, MODEL 5200
Applicant
Edwards Lifesciences, LLC
Product Code
KRH · Cardiovascular
Decision Date
Jan 23, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Carpentier-Edwards Physio II Annuloplasty Ring is intended for the correction of mitral valve Insufficiency, or mixed mitral insufficiency and stenosis, where treatment does not necessitate a replacement of the natural mitral valve. The Carpentier-Edwards Physio if Annuloplasty Ring is intended to meet the challenges of modern valvuloplasty by maintaining the physiologic annular shape and motion. The annuloplasty ring is designed to follow the functional changes which occur during the cardiac cycle, thereby maintaining coaptation and valve integrity in systole while permitting good hemodynamics in diastole. The decision to undertake annuloplasty can be made only after visual analysis of the lesion present. The most favorable conditions for annuloplasty using a prosthetic ring are a combination of the distended natural valve ring associated with supple valve cusps and normal chordae tendineae. The remodeling annuloplasty technique with a Carpentier-Edwards Physio II Annuloplasty Ring, Model 5200, may be used in all acqured or congental mitral insufficiencies with dilatation and deformation of the fibrous mitral annulus, with the exception of severe congenital malformations (e g , AV canal or hypoplastic commissures) or severe degenerative valvular diseases where there is considerable excess tissue. For Type I mitral insufficiencies with no subvalvular lesions and normal vaivular movements, this ring technique used alone is sufficient. However, the ring technique must be associated with mitral valvuloplasty repar in Type II insufficiencies with a prolapsed valve due to elongation or rupture of the chordae tendineae or papillary muscle and in Type III imitation of valvular movements due to fusion of the commissures or chordae, or chordal hypertrophy.
Device Story
Carpentier-Edwards Physio II Annuloplasty Ring (Model 5200) is a prosthetic implant used during mitral valve repair surgery. Device functions as a remodeling tool to support the mitral annulus, maintaining physiologic shape and motion throughout the cardiac cycle. It promotes valve coaptation during systole and hemodynamics during diastole. Implanted by cardiac surgeons in an operating room setting. Device provides structural support to the fibrous mitral annulus in cases of dilatation or deformation. Benefits include preservation of the natural valve and restoration of valve integrity.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Constructed of cobalt-chromium alloy and polyester film bands; silicone sewing ring margin covered with polyester cloth. Ring form factor designed for mitral annular remodeling. Non-electronic, mechanical implant.
Indications for Use
Indicated for patients with acquired or congenital mitral insufficiency or mixed mitral insufficiency and stenosis requiring repair rather than valve replacement. Contraindicated in severe congenital malformations (e.g., AV canal, hypoplastic commissures) or severe degenerative valvular diseases with considerable excess tissue.
Regulatory Classification
Identification
An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”
Predicate Devices
Carpentier-Edwards Physio™ Annuloplasty Ring (K926138)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with the letter 'E' in a stylized font, superimposed on a background of circles and squares. The letter 'E' is white, while the background is black. Below the logo, the word 'Edwards' is written in a serif font.
K083470 JAN 2 3 2009
## 510(k) Summary
| Submitter | Edwards Lifesciences LLC |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Erica A Walters<br>Regulatory Affairs Associate<br>erica_walters@edwards.com<br>(ph) 949-250-2438<br>(fax) 949-250-3630 |
| Date Prepared | November 20, 2008 |
| Trade Name | Carpentier-Edwards Physio II™ Annuloplasty Ring,<br>Model 5200 |
| Classification Name | Class II, CFR 870 3800 Annuloplasty Ring, KRH |
| Predicate Device(s) | Carpentier-Edwards Physio™ Annuloplasty Ring<br>(K926138) |
| Device Description | The Carpentier-Edwards Physio II annuloplasty ring,<br>model 5200, is constructed of cobalt-chromium<br>alloy/polyester film bands with a silicone sewing ring<br>margin covered with a polyester cloth |
| Indications for Use | The Carpentier-Edwards Physio II annuloplasty ring,<br>model 5200, is intended for the correction of mitral<br>valvular insufficiency where the lesions are not so severe<br>as to require total valve replacement |
| Comparative Analysis | It has been demonstrated that the Carpentier-Edwards ,<br>Physio II annuloplasty ring, model 5200, is comparable to<br>its predicate device in design, intended use, materials,<br>and principal of operation |
| Functional/Safety Testing | The Carpentier-Edwards Physio II annuloplasty ring,<br>model 5200, has successfully completed design<br>verification testing |
| Conclusion. | The Carpentier-Edwards Physio II annuloplasty ring,<br>model 5200, is substantially equivalent to its predicate<br>device |
ﻟﺴﻴﺎﺭ ﺍ
{1}------------------------------------------------
Public Health Service
Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Edwards Lifesciences, LLC Erica A Walters Regulatory Affairs Associate One Edwards Way Irvine CA 92614
JAN 23 2009
Re K083470
Trade/Device Name Carpentier-Edwards Physio II annuloplasty ring, model 5200 Regulation Number 21 CFR 870 3800 Regulation Name Annuloplasty ring Regulatory Class Class II Product Code KRH Dated November 20, 2008 Received November 24, 2008
t,
Dear Ms Walters
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA) You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteratıon
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In addition, FDA may publish further announcements concerning your device in the Federal Register
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must
{2}------------------------------------------------
comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801), good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050 This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474 For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464 You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http //www fda gov/cdrh/industry/support/index html
Sincerely yours,
R buhner
Bram D Zuckerman, M D Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Carpentier-Edwards Physio II Annuloplasty Ring, Model 5200 Device Name
Indications For Use
The Carpentier-Edwards Physio II Annuloplasty Ring is intended for the correction of mitral valve Insufficiency, or mixed mitral insufficiency and stenosis, where treatment does not necessitate a replacement of the natural mitral valve
The Carpentier-Edwards Physio if Annuloplasty Ring is intended to meet the challenges of modern valvuloplasty by maintaining the physiologic annular shape and motion The annuloplasty ring is designed to follow the functional changes which occur during the cardiac cycle, thereby maintaining coaptation and valve integrity in systole while permitting good hemodynamics in diastole
The decision to undertake annuloplasty can be made only after visual analysis of the lesion present The most favorable conditions for annuloplasty using a prosthetic ring are a combination of the distended natural valve ring associated with supple valve cusps and normal chordae tendineae
The remodeling annuloplasty technique with a Carpentier-Edwards Physio II Annuloplasty Ring, Model 5200, may be used in all acqured or congental mitral insufficiencies with dilatation and deformation of the fibrous mitral annulus, with the exception of severe congenital malformations (e g , AV canal or hypoplastic commissures) or severe degenerative valvular diseases where there is considerable excess tissue
For Type I mitral insufficiencies with no subvalvular lesions and normal vaivular movements, this ring technique used alone is sufficient However, the ring technique must be associated with mitral valvuloplasty repar in Type II insufficiencies with a prolapsed valve due to elongation or rupture of the chordae tendineae or papillary muscle and in Type III imitation of valvular movements due to fusion of the commissures or chordae, or chordal hypertrophy
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
R. Holmes
(Division Sign-Off)
Division of Cardiovascular Devices
umber K083470
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.