ACTHERM DIGITAL CLINICAL THERMOMETER PROBE COVERS, MODEL(S) MEDACCU1010, MEDACCU1010-1, MEDACCU1020
K083419 · Actherm, Inc. · FLL · Jul 16, 2009 · General Hospital
Device Facts
Record ID
K083419
Device Name
ACTHERM DIGITAL CLINICAL THERMOMETER PROBE COVERS, MODEL(S) MEDACCU1010, MEDACCU1010-1, MEDACCU1020
Applicant
Actherm, Inc.
Product Code
FLL · General Hospital
Decision Date
Jul 16, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
The Actherm Digital Clinical Thermometer Probe Covers medACCU1010, medACCU1010-1 and medACCU1020 are accessories for digital thermometers, including the models of Actherm Digital Clinical Thermometers. They are indicated for use as barners between digital thermometers and users' rectum or oral cavities to avoid the possible contamination and infection during temperature measuring. These probe covers are intended for single use when temperature taking.
Device Story
Disposable probe covers for digital clinical thermometers; act as physical barrier between thermometer probe and patient oral or rectal cavity; prevent cross-contamination and infection during temperature measurement. Models medACCU1010 and medACCU1020 used for oral or rectal measurements; model medACCU1010-1 includes lubrication for rectal use. Used in clinical or home settings; operated by healthcare providers or patients. Single-use design; provides hygienic protection during temperature taking.
Clinical Evidence
Bench testing only. Compliance with ASTM E 1104-98 (Standard Specification for Clinical Thermometer Probe Covers and Sheaths) and ISO 10993-1 (biocompatibility) demonstrated.
Technological Characteristics
Disposable probe covers; various shapes/sizes; model medACCU1010-1 includes lubrication. Compliant with ASTM E 1104-98. Biocompatible per ISO 10993-1.
Indications for Use
Indicated for use as a barrier between digital thermometers and a patient's oral or rectal cavities to prevent cross-contamination during temperature measurement. Intended for single-use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
ACON LABORATORIES, INC. ACON Digital Thermometer Probe Covers (K063418)
Submission Summary (Full Text)
{0}------------------------------------------------
Aetherium
K08 3419
1 of 2
Chapter 15 - 1/2
# Chapter 15 510 (K) Summary
# JUL 1 6 2009
#### 510 (k) Summary
This summary of 510 (k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### Submitter's Identification:
Actherm Inc. 6F, No.18, Jhanye 2nd Rd, Hsinchu Science Park, Hsinchu 30078, TAIWAN, R.. O.C.
#### Contact Person:
Mr. Richard HSIEH
TEL: +886-3-6669596 #888
FAX: +886-3-6669697 E-mail: richard@actherm.com.tw actherm@actherm.com.tw
Date Summary Prepared: July 8, 2009
Manufacturer: Actherm Inc. 6F, No.18, Jhanye 2nd Rd, Hsinchu Science Park, Hsinchu 30078, TAIWAN, R.O.C.
#### Device Name:
Digital Clinical Thermometer Probe Covers
#### Trade Name:
Actherm Digital Clinical Thermometer Probe Covers. The Trade models name medACCU1010, medACCU1010-1 and medACCU1020.
#### Classification Name:
Clinical Electronic Thermometer (per 21 CFR 880.2910)
#### Predicate Device Information:
BANTA HEALTHCARE GROUP, LTD (Rite Aid Brand) Sanitherm Oral Disposable Thermometer Sheaths 510(k) Number is K983406
ACON LABORATORIES, INC. ACON Digital Thermometer Probe Covers 510(k) Number is K063418
| Code: FDA-007 | .<br>Version: A | Revision Status: | And and with the count of the first and the first and the first and the first and the first and<br>Issuing Date: July 8. 2009 |
|---------------|-----------------|------------------|-------------------------------------------------------------------------------------------------------------------------------|
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a stylized signature or logo, possibly of a person or company. The text appears to read "Aetherium" in a cursive or calligraphic style. There is a thick, dark line underlining the name, adding emphasis and a sense of grounding to the design.
K083419
2 of 2
Chapter 15 - 2/2
#### Device Description:
The Digital Clinical Thermometer Probe Covers medACCU1010 and medACCU1010-1 are manufactured in the same shapes and sizes. The only difference between them is medACCU1010-1 coated with lubrication; therefore it is only used for rectal measurements for digital thermometers. The Digital Clinical Thermometer Probe Covers medACCU1020 are manufactured in different shapes and sizes from medACCU1010 and medACCU1010-1. The disposable digital thermometer probe covers medACCU1010 and medACCU1020 are used for either oral or rectal measurements for digital thermometers.
#### Indications for Use:
The Actherm Digital Clinical Thermometer Probe Covers medACCU1010, medACCU1010-1 and medACCU1020 are accessories for digital thermometers, including the models of Actherm Digital Clinical Thermometers. They are indicated for use as barners between digital thermometers and users' rectum or oral cavities to avoid the possible contamination and infection during temperature measuring. These probe covers are intended for single use when temperature taking.
#### Comparison to Predicate Devices:
The Actherm Digital Clinical Thermometer Probe Covers medACCU1010, medACCU1010-1 and medACCU1020 are substantially equivalent to the following predicate devices: Sanitherm Oral Disposable Thermometer Sheaths (K983406) and ACON Digital Thermometer Probe Covers (K063418).
#### Performance Data:
Actherm Digital Clinical Thermometer Probe Covers medACCU1010, medACCU1010-1 and medACCU1020 meet the ASTM Standard Specification for Clinical Thermometer Probe Covers and Sheaths (ASTM E 1104-98), as well as ISO 10993-1 biocompatibility testing,
#### Conclusion:
It is concluded that the Actherm Digital Clinical Thermometer Probe Covers medACCU1010, medACCU1010-1 and medACCU1020 are substantially equivalent to marked products, functions as intended, and with exception to the risks, which have been evaluated within the Hazard Analysis, no additional potential hazards have been found. The assessment of user benefit to health from use as intended is much higher than the probable risks of injury or illness. The study results also demonstrated that the Actherm Digital Clinical Thermometer Probe Covers medACCU1010, medACCU1010-1 and medACCU1020 are safe, effective and easy-to-use.
Code: FDA-007
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Richard Hsieh Actherm Incorporated 6F, No. 18 Jhanye 2nd Road Hsinchu Science Park Hsinchu CHINA (TAIWAN) 30078
# JUL 1 6 2009
Re: K083419
Trade/Device Name: Actherm Digital Thermometer Probe Covers, Models medACCU1010, medACCU1010-1, and medACCU1020
Regulation Number: 21 CFR 880.2910
Regulation Name: Clinical Electronic Thermometer
Regulatory Class: II
Product Code: FLL
Dated: July 8, 2009 Received: July 14, 2009
Dear Mr. Hsieh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 – Mr. Hsieh
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Payne
Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): K083419
# Device Name: Actherm Digital Thermometer Probe Covers, Models medACCU1010, medACCU1010-1, and medACCU1020
#### Indications For Use:
Clinical Thermometer Probe Covers medACCU1010, Digital Actherm The medACCU1010-1 and medACCU1020 are accessories for digital thermometers, including the models of Actherm Digital Clinical Thermometers. They are indicated for use as barriers between digital thermometers and users' rectum or oral cavities to avoid the possible contamination and infection during temperature measuring. These probe covers are intended for single use when temperature taking.
Prescription Use (Part 21 CFR 801 Subpart D) · AND/OR
. -
6:5 + 1
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH-Office of Deviee Evaluation (ODE)
*(Division Sign-Off)*
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: KD83419
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.