NOVATION 12/14 CEMENTED FEMORAL STEMS

K083392 · Exactech, Inc. · LZO · Dec 17, 2008 · Orthopedic

Device Facts

Record IDK083392
Device NameNOVATION 12/14 CEMENTED FEMORAL STEMS
ApplicantExactech, Inc.
Product CodeLZO · Orthopedic
Decision DateDec 17, 2008
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis. congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion. Cemented femoral stems and cemented acetabular cups are intended for cemented fixation only. Press-fit femoral stems and acetabular cups are intended for press-fit fixation. Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon. Femoral heads and endoprostheses are intended for use in cemented and press-fit applications.

Device Story

The Novation Cemented Plus femoral stem is a modified orthopedic implant for hip arthroplasty. It functions as a prosthetic component to replace the femoral head/neck in cemented hip replacement procedures. The device features a trapezoidal distal geometry, collar, cement groove, and 'cobra' flange. It is designed to mate with previously cleared 12/14 femoral heads (CoCr, Zirconia, Alumina). The device is implanted by an orthopedic surgeon in a clinical/OR setting. It provides structural support and restores joint function, benefiting patients by addressing degenerative hip disease or fractures. The modification adds five additional sizes to the existing product line.

Clinical Evidence

No clinical data. Substantial equivalence is supported by engineering studies.

Technological Characteristics

Cemented femoral hip prosthesis; trapezoidal distal geometry; collar, cement groove, and cobra flange design; compatible with 12/14 taper femoral heads (CoCr, Zirconia, Alumina).

Indications for Use

Indicated for skeletally mature individuals requiring primary hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative hip conditions, or proximal femoral fractures. Also indicated for ankylosing spondylitis, congenital hip dysplasia, revision of failed reconstructions with sufficient bone stock, and restoration of mobility after fusion.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Exactech® Novation® Cemented Plus Femoral Stems Special 510(k) - 510(k) Summary of Safety and Effectiveness Sponsor: Exactech® Inc. 2320 N.W. 66th Court Gainesville, FL 32653 DEC 1 7 2008 Phone: (352) 377-1140 Fax: (352) 378-2617 FDA Establishment Number 1038671 Contact: Graham L. Cuthbert Regulatory Affairs Specialist Date: November 14, 2008 Trade or Proprietary or Model Name(s): Exactech® Novation® Cemented Plus Femoral Stems Common Name: Cemented Femoral Hip Prosthesis Classification Name: Prosthesis, hip, semi-constrained, metal/polymer, cemented Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented Information on devices to which Substantial equivalence is claimed: 510(k) Number Trade or Proprietary or Model Name Manufacturer K052787 Novation® 12/14 Cemented Femoral Stems Exactech, Inc. {1}------------------------------------------------ ### Indications for Use: All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis. congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion. - . . Cemented femoral stems and cemented acetabular cups are intended for cemented fixation only. - Press-fit femoral stems and acetabular cups are intended for press-fit fixation. ● Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon. - Femoral heads and endoprostheses are intended for use in cemented and . press-fit applications. #### Device Description: The proposed Novation Cemented Plus femoral stem is a modification to the existing Novation Cemented femoral stem devices previously cleared in K052787. The proposed stem has the same general design features (collar, cement groove, and "cobra" flange) as the predicate stem. but incorporates a more trapezoidal distal geometry. Additionally, the proposed device includes 5 additional sizes (standard offset Size 10, standard and extended offset Size 12 and 14) to the scope. The proposed stems mate with previously cleared 12/14 CoCr (#K041906), Zirconia ceramic (#K060107), and Alumina ceramic (#K032964 and #K051682) femoral heads. The predicate and proposed devices have intended use and basic fundamental scientific technology and share the following similarities: - · the same indications for use - · the same design features - · incorporate the same materials - · the same shelf life - · are packaged and sterilized using the same materials and processes. ### Substantial Equivalence Conclusion: Results of engineering studies referenced in this 510(k) submission demonstrate that the proposed Novation Cemented Plus femoral stem is substantially equivalent to the cleared predicate devices. {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is black and white. #### Public Health Service Food and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850 Exactech, Inc. % Mr. Graham L. Cuthbert Regulatory Affairs Specialist 2320 NW 66th Court Gainesville, Florida 32653 DEC 1 7 2008 Re: K083392 > Trade/Device Name: Exactech® Novation® Cemented Plus Femoral Stems Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, JDI Dated: November 14, 2008 Received: November 17, 2008 Dear Mr. Cuthbert: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Graham L. Cuthbert forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Mark N. Milliman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Exactech® Novation® Cemented Plus Femoral Stems Special 510(k) - Indications for Use 510(k) Number: K083392 Device Name: Exactech® Novation® Cemented Plus Femoral Stems ## INDICATIONS All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for hip replacement due to osteoarthritis, theumatoid arthritis, osteonecrosis, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for ankylosing spondylitis. congenital hip dysplasia, revision of failed previous reconstructions where sufficient bone stock is present, and to restore mobility resulting from previous fusion. - Cemented femoral stems and cemented acetabular cups are intended for ● cemented fixation only. - Press-fit femoral stems and acetabular cups are intended for press-fit ● fixation. Press-fit components without hydroxyapatite (HA) coating may also be used with bone cement at the discretion of the surgeon. - Femoral heads and endoprostheses are intended for use in cemented and press-fit applications. Prescription Use X (Part 21 CFR 801 Subpart D) and/or Over-The-Counter Use (21 CFR 807 Subpart C) Please do not write below this line - use another page if needed. Concurrence of CDRH, Office of Device Evaluation (ODE) Mark N. Milken (Division Sign-Off) (Division Sign-Old) Division of General, Restorative, Division of General, Devices Division of General Devices **510(k) Number** **__**
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...