K083391 · Diagnostic Hybrids, Inc. · JZO · May 21, 2009 · Immunology
Device Facts
Record ID
K083391
Device Name
THYROID STIMULATING IMMUNOGLOBULIN REPORTER ASSAY
Applicant
Diagnostic Hybrids, Inc.
Product Code
JZO · Immunology
Decision Date
May 21, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5870
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K083391 · May 21, 2009
THYROID STIMULATING IMMUNOGLOBULIN REPORTER ASSAY
Diagnostic Hybrids, Inc.
Retrospective clinical serum samples from patients with Graves' disease and other thyroid conditions
Retrospective clinical samples were used to establish clinical sensitivity and specificity, and to verify the assay cutoff against diagnostic information provided by physicians.
Clinical Performance Study (Sites 1-COH, 2-MN, and 3-NC); Retrospective comparative study
Patients undergoing thyroid screening or evaluation for thyroid disease; Sample Size: 530 total specimens (299 at sites 1/2; 231 at site 3); Number of Sites: 3
Clinical Sensitivity and Specificity Study; Retrospective diagnostic accuracy study
249 characterized specimens (Graves' disease patients, other autoimmune diseases, and healthy controls); Sample Size: 249
Not applicable for this study
Clinical sensitivity and specificity
Indications for Use
The Thyretain™ TSI Reporter BioAssay is intended for the qualitative detection in serum of thyroid stimulating autoantibodies to the thyroid stimulating hormone receptors (TSHRs) on the thyroid. The detection of these stimulating autoantibodies, in conjunction with other clinical and laboratory findings, may be useful as an aid in the differential diagnosis of patients with Graves' disease.
Device Story
Thyretain™ TSI Reporter BioAssay is a cell-based chemiluminescent assay for detecting thyroid-stimulating autoantibodies (TSI) in human serum. The device utilizes genetically engineered Chinese hamster ovary (CHO) cells expressing chimeric human TSH receptors and a cAMP-induced luciferase reporter gene. Patient serum is incubated with a confluent monolayer of these cells; TSI binding to the TSH receptor triggers a signaling cascade, increasing intracellular cAMP and inducing luciferase production. After cell lysis, luciferase levels are measured via luminometer. The output is a percentage of the Reference Control (%SRR); values ≥140% indicate the presence of TSI. The assay is performed in clinical laboratories by trained technicians using standard laboratory equipment (incubator, luminometer, biosafety cabinet). Results assist clinicians in the differential diagnosis of Graves' disease by specifically identifying stimulating antibodies, distinguishing them from blocking antibodies that other assays might detect.
Clinical Evidence
Clinical performance was evaluated using 50 patients with Graves' disease and 199 controls (healthy and other autoimmune diseases). The assay demonstrated a clinical sensitivity of 92.0% (95% CI: 81.2–96.9%) and a clinical specificity of 99.5% (95% CI: 97.2–99.9%). Method comparison studies against the predicate device across three sites showed overall percent agreement ranging from 90.5% to 91.3%.
Technological Characteristics
Cell-based chemiluminescent bioassay. Components: CHO Mc4 cell line (chimeric human TSHR), cell attachment solution, growth medium, reaction buffer, luciferase substrate/buffer. Detection via luminometer measuring Relative Light Units (RLU). Qualitative output based on SRR% (Sample/Reference Ratio) cutoff of ≥140%.
Indications for Use
Indicated for qualitative detection of thyroid stimulating autoantibodies (TSI) in human serum to aid in the differential diagnosis of Graves' disease in patients suspected of having thyroid dysfunction.
Regulatory Classification
Identification
A thyroid autoantibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the thyroid autoantibodies (antibodies produced against the body's own tissues). Measurement of thyroid autoantibodies may aid in the diagnosis of certain thyroid disorders, such as Hashimoto's disease (chronic lymphocytic thyroiditis), nontoxic goiter (enlargement of thyroid gland), Grave's disease (enlargement of the thyroid gland with protrusion of the eyeballs), and cancer of the thyroid.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k083391
B. Purpose for Submission:
New Device
C. Measurand:
Thyroid Stimulating Hormone Receptor (TSHR) Autoantibodies
D. Type of Test:
Cell-based qualitative chemiluminescent assay
E. Applicant:
Diagnostic Hybrids, Inc.
F. Proprietary and Established Names:
Thyretain™ TSI Reporter BioAssay
G. Regulatory Information:
1. Regulation section:
21 CFR §866.5870, Thyroid autoantibody immunological test system
2. Classification:
Class II
3. Product code:
JZO, System, test, thyroid autoantibody
4. Panel:
Immunology (82)
H. Intended Use:
1. Intended use(s):
The Thyretain™ TSI Reporter BioAssay is intended for the qualitative detection in serum of thyroid stimulating autoantibodies to the thyroid stimulating hormone receptors (TSHRs) on the thyroid. The detection of these stimulating autoantibodies, in conjunction with other clinical and laboratory findings, may be useful as an aid in the differential diagnosis of patients with Graves' disease.
2. Indication(s) for use:
Same as Intended use
3. Special conditions for use statement(s):
For prescription only
4. Special instrument requirements:
Luminometer; Humidified, 5% CO₂, 35°C to 37°C Incubator; Bio-safety Cabinet Class II; -70°C or lower freezer or liquid nitrogen Dewar; Light Microscope
I. Device Description:
The Thyretain™ TSI Reporter BioAssay consists of cryovials containing CHO Mc4 cells cryogenically preserved in cryoprotective medium containing DMSO (stored at ≤-70°C); 100 mL bottle of Cell Attachment Solution; 100 mL growth medium (Hamm's F12 culture medium with 10% FBS); 500 mL reaction buffer; positive, reference and normal control set (0.5 mL vial each); and a luciferase reagent set (1
{1}
vial luciferase substrate and 10 mL vial luciferase assay buffer solution).
# J. Substantial Equivalence Information:
1. Predicate device name(s): KRONUS TSH Receptor Antibody (TRAb) Coated Tube (CT) Assay Kit
2. Predicate 510(k) number(s): k032134
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | The Thyretain™ TSI Reporter Bioassay is intended for the qualitative detection of thyroid stimulating autoantibodies to the thyroid stimulating hormones receptors (TSHRs) on the thyroid. The detection of these stimulating autoantibodies, in conjunction with other clinical and laboratory findings, may be useful as an aid in the differential diagnosis of patients with Graves' disease. | KRONUS TSDH Receptor Antibody (TRAb) Coated Tube (CT) Assay Kit is designed to measure human serum autoantibodies to the thyroid stimulating hormone (TSH or thyrotropin) receptor. The TRAb CT kit is useful as an aid in the differential diagnosis of Graves' disease. |
| Sample matrix | Serum | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Assay Format | Qualitative | Qualitative and quantitative |
| Assay principle | Cell-based Chemiluminescent Assay | Radioreceptor Assay |
| Solid Phase | CHO Mc4 cell monolayer (96-well microplate) | TSHR-coated tube |
| Analyte | Autoantibodies to Thyroid stimulating hormone (TSI-hyperthyroidism) | Autoantibodies to Thyroid stimulating hormone (TSI-hyperthyroidism) and thyroid blocking immunoglobulins (TBI-hypothyroidism) |
| Calibration | NISBC Standard 03/192 or similar standard | NIBCC Standard 90/672 or similar calibrator |
| Signal | Optical density | Radioactive |
| Detection instrument | Luminometer | Gamma counter set for \( ^{125}\mathrm{I} \) |
| Unit of measure | %SRR | U/L |
{2}
3
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Cut-off | Positive: ≥ 140% of Reference Control
Negative: ≤140% of Reference Control | Positive: >15% inhibition
Indeterminate: 11-15% inhibition
Negative: <11% inhibition |
K. Standard/Guidance Document referenced (if applicable):
1. CLIA EP 17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline.
2. CLSI: EP 12-A: User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline.
3. Statistical Guidance on Reporting Results from Studies Evaluating Diagnostic Tests.
L. Test Principle:
In this assay, patient sera, reference control, positive and normal controls are added to the genetically engineered Chinese hamster ovary (CHO) cells which express chimeric form of the human TSHR and a cyclic adenosine monophosphate (cAMP) induced luciferase reporter gene. The cells are seeded and grown for 15-18 hours to a confluent monolayer in a 96-well plate prior to the addition of the samples. CHO cell monolayer is incubated with patient serum for 3 hours to allow the binding of TSI immunoglobulin to the chimeric human TSHR on the cell surface. This binding induces a signaling cascade resulting in increased production of intra-cellular cAMP which induces the production of luciferase. After incubation, the cells are lysed and luciferase levels are measured with a luminometer. A significant increase in luminescence over the Reference Control indicates the presence of TSI antibodies in the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Intra-assay Precision:
Intra-assay precision was conducted by a single user over a 20 day period, with each sample tested in triplicate. The resulting RLU values were averaged per sample. Each plate of cells analyzed per day contained 16 samples (3 high responding samples, 3 medium responding samples, 3 low responding samples, 3 normal samples, 2 reference controls, 1 positive control and 1 blank replicate). The average variation for each sample was calculated (CV%) across each plate. The result was summarized in the following table:
| Sample | High | Mid | Low | Normal | Positive Control |
| --- | --- | --- | --- | --- | --- |
| N | 3 | 3 | 3 | 3 | 1 |
| Mean %SRR | 525 | 310 | 184 | 62 | 291 |
| SD | 9 | 6 | 9 | 4 | * |
| CV% | 1.8 | 1.9 | 4.7 | 5.9 | * |
{3}
Average intra-plate variation was calculated at 4.7% (%CV).
## Inter-Assay Precision (Intra-Day and Inter-Day):
Inter-Assay Precision was assessed using the same set of samples. Six replicates of each, high positive, medium positive, low positive, normal samples, 2 TSI positive control, and 4 reference control samples were tested per day, respectively. The results were summarized in the following table:
| Sample (plate to plate) | High | Mid | Low | Normal | Positive Control |
| --- | --- | --- | --- | --- | --- |
| N | 6 | 6 | 6 | 6 | 2 |
| Mean (%SRR) | 509 | 313 | 183 | 61 | 302 |
| SD | 19 | 8 | 8 | 3 | 15 |
| CV% | 3.6 | 2.6 | 4.2 | 5.0 | 5.0 |
| Sample (day to day) | | | | | |
| N | 120 | 120 | 120 | 120 | 40 |
| Mean (%SRR) | 446 | 263 | 158 | 52 | 267 |
| SD | 51 | 34 | 23 | 8 | 33 |
| CV% | 11.5 | 12.8 | 14.5 | 15.7 | 12.3 |
## Site-to-Site Reproducibility:
Site-to-site reproducibility was assessed at 3 trained sites (with 2 technicians at site 3) that performed 2 runs per day with each of four samples in triplicate over an eight day period. Two days were excluded from site 2 due to a technical failure. The total number of replicates for each sample is 180 (3x2x8x4 - 3x2x2x1). The data was summarized in the following table:
| Reproducibility
Average SRR% (%CV) | | | | |
| --- | --- | --- | --- | --- |
| Specimen | Site 1 | Site 2 | Site 3 | |
| | | | T1 | T2 |
| A | 270.09%
(11.47%) | 439.8%
(10.51%) | 280.10%
(13.56%) | 289.49%
(10.55%) |
| B | 293.11%
(14.72%) | 495.14%
(14.59%) | 374.26%
(14.70%) | 345.13%
(11.8%) |
| C | 48.23%
(14.94%) | 67.89%
(14.66%) | 44.33%
(20.21%) | 48.60%
(22.46%) |
| D | 158.6%
10.62%) | 198%
(10.96%) | 143.6%
(14.43%) | 142.7%
(12.12%) |
The overall percents of coefficient of variation (CV%) for the four specimens were 23.7%, 23.7%, 24.6%, and 17.9% respectively.
{4}
An additional smaller study was performed using three samples near the cut off at two sites twice a day for five-days.
| Reproducibility
Average SRR% (%CV) | |
| --- | --- |
| Site 2 | Diagnostic
Hybrids Inc. |
| Specimen E | Specimen E |
| 194%
(9.3%) | 170%
(17.7%) |
| Specimen F | Specimen F |
| 106%
(20.2%) | 106%
(20.7%) |
| Specimen G | Specimen G |
| 102%
(21.6%) | 107%
(19.4%) |
Each site’s data were analyzed to determine the Reproducibility and Repeatability of the panel samples. Sample E had a positive ratio (Number Positive/Total Number Tested) of 60/60, Samples F and G had negative ratios (Number Negative/Total Number Tested) of 60/60. The overall coefficient of variation (%CV) for the three samples was 15.0%, 20.3%, and 20.5%, respectively.
b. Assay temperature:
The effect of temperature on cell lysis was determined to be critical during the reproducibility study. The assay was tested at different temperature to confirm the effect of the temperature on assay result. No statistical difference was observed when the lysis occurred between 20 - 25°C; however, a decrease of approximately 20% in %SRR was observed if the lysis temperature was dropped from 20°C to 19°C. The following notation was added to the assay procedure in the package insert: “Note: It is critical to maintain the cell lysis temperature above 20°C. The test result will be affected if the lysis temperature is dropped to 19°C or less.”
c. Cell density:
Cell density was determined to be critical to the assay performance. The impact of cell density was studied on duplicate plates that were seeded with three concentrations of cells, which will reach to 90% confluent, 100% confluent and over-100% confluent (more densely packed) respectively, after 18 hours of incubation. Each plate was treated with a positive, reference and normal controls as well as with serum containing high, med and low levels of TSI. No difference was found on %SRR among three plates containing different cell confluent levels when high and med levels of TSI were tested; however, a 19% decrease in %SRR was observed in wells planted with densely packed cells when serum containing a low level of TSI was tested.
{5}
A following warning note was added into the package insert: “Individual wells containing piled or layered cells should be disqualified from use based on over-confluence.”
d. Linearity/assay reportable range:
Not applicable.
f. Traceability, Stability, Expected values (controls, calibrators, or methods):
Reference Control was the bovine Thyroid-Stimulating Hormone (bTSH) standardized against the rhTSH Standard (WHO Standard NIBSC 03/192, Thyroid-Stimulating Hormone, Recombinant, Human, for Bioassay).
Positive control:
Positive control values between lots provided in the kit are different due to differences in each manufactured antibody lot. An extra warning sentence was therefore added to the package insert as following:
“It is good practice to examine the results of the Positive and Normal Controls before examining the test results of the specimens. However different lots might have a different positive control range. Please check the positive control reference range label for the test range before the test. If one or both of the controls fail to perform as expected, review the steps and conditions under which the test was performed to determine the cause(s). Do not report results until controls perform as expected.”
Specimen stability:
Four specimens containing two positive (near the cut-off) and two high positive samples were stored at 2°C - 8°C (up to 72 hours), -20°C (up to 3 months) and compared to samples stored at -80°C. No apparent decrease of the signal was observed.
Freeze and thaw:
Four specimens, one negative and three positives, were tested repeatedly for up to 4 cycles. No statistical difference in the results was shown when a sample was repeatedly frozen and thawed for a maximum of three cycles.
g. Detection limit:
The Limit of Blank (LoB) was determined by 20 blank measurements and the average of the 19th and 20th ranked measurement was 62.75%. Twenty measurements of a low positive were carried out and the standard deviation was calculated at 16%.
The Limit of Detection (LoD) of 89.14% SRR was calculated based on the formula provided in CLSI EP17-A. [LoD = LoB + cβ SDs (62.75 + 1.649*16)]. The LoD was verified in 25 measurements of a positive specimen diluted to the claimed LoD (5 measurements of 5 positive specimens over 5-days). 24 out of 25 (96%) measurements showed %SRR value higher than
6
{6}
LoB.
h. Analytical specificity:
Cross-Reactivity:
Follicle Stimulating Hormone (FSH), Luteinizing Hormone (LH), and human Chorionic Gonadotropin (hCG) share the same α-chain with TSH. The cross-reactivity study was performed by testing different amounts of FSH, LH and hCG in normal human serum and TSI containing sera from confirmed Graves' Disease (GD) subjects with low positive and high positive %SRR. No cross-reactivity was observed with FSH up to 2000 mIU/mL, LH up to 625 mIU/mL, and hCG up to 40,625 mIU/mL in the serum. No cross-reactivity was observed in studies of spiked sera with TSH up to 0.35 mIU/mL (the normal physiological range in healthy adults is 0.0003 – 0.0030 mIU/mL).
Autoimmune Specimen Cross-Reactivity:
Serum from patients with auto-antibodies associated diseases [10 patients with Rheumatoid Arthritis (RA), 10 patients with Systemic Lupus Erythematosus (SLE), and 16 patients with Hashimoto's thyroiditis (Hm)] were tested and found to have no increased cAMP induced luciferase in the TSI Reporter Assay except one Hm sample. This sample which had a SRR% of 150%, was found to have a TSH level at or above -0.35 mIU/mL. A statement that any specimens with TSH levels above this range need to be disqualified for use is included in the Package Insert.
Interference:
Normal sera and TSI positive sera (one in high positive and one in low positive) were tested in the presence of different amounts of hemoglobin, bilirubin and lipemic samples. No interference was seen in samples spiked with hemoglobin up to 250 mg/dL, bilirubin levels up to 36.6 mg/dL, or lipemic levels up to 1168 mg/dL.
i. Assay cut-off:
The assay cut-off was established using 74 samples from 30 patients with diagnosed GD and 44 normal patients with no known or clinically diagnosed thyroid disease. The %SRR is calculated by the ratio of RLU (Relative Light Unit) from sample to RLU from reference control. The SRR% data obtained for each of these subjects were analyzed using receiver operating characteristic (ROC) curve analysis. The cutoff was verified in pre-clinical testing with an additional 50 GD positive sera from non-treated patients and 140 normal sera. The patient serum is considered positive for the presence of TSI if the SRR% measured ≥140% over the Reference Control.
7
{7}
| Cutoff Analysis on the TSI Reporter | | | | |
| --- | --- | --- | --- | --- |
| | Diagnosis | | | |
| | | Positive (Graves' Disease) | Negative (Healthy Controls) | Total |
| TSI Reporter | Positive | 46 | 0 | 46 |
| | Negative | 4 | 140 | 144 |
| | Total | 50 | 140 | 190 |
Sensitivity: 92% (46/50)
Specificity: 100% (140/140)
## 2. Comparison studies:
a. Method comparison with predicate device:
An initial study was performed at two testing sites with a total of 312 specimens to be evaluated by both the subject and predicate devices. One specimen was excluded due to insufficient quantity for testing. Twelve of these specimens were excluded from statistical analysis due to indeterminate results on the predicate device. The remaining 299 specimens were analyzed for positive and negative percent agreement.
| Sites 1 and 2 Results | | | | | |
| --- | --- | --- | --- | --- | --- |
| 310 specimen results | Predicate Device | | | | |
| | | | + | Indeterminate | - |
| TSI Reporter | + | | 120 | 7 | 18 |
| | - | | 8 | 5 | 153 |
| | total | | 128 | 12 | 171 |
| | | | 95% Confidence Interval | | |
| Positive Percent Agreement | | 93.8% (120/128) | | 88.25 – 96.8% | |
| Negative Percent Agreement | | 89.5% (153/171) | | 84.05 – 93.2% | |
| Overall Percent Agreement | | 91.3% (273/299) | | 87.5 – 94.2% | |
An additional study was performed at a third testing site using 247 specimens. Sixteen of these specimens were excluded from statistical analysis due to indeterminate results on the predicate device. The remaining 231 specimens were analyzed for positive and negative percent agreement.
| Site 3 Results | | | | |
| --- | --- | --- | --- | --- |
| 247 specimen results | Predicate Device | | | |
| | | | + | Indeterminate* |
| TSI Reporter | + | | 53 | 2 |
| | - | | 18 | 14 |
| | | | 71 | 16 |
| 95% Confidence Interval | | | | |
| Positive Percent Agreement | | 74.6% (53/71) | | 63.5 to 83.3% |
| Negative Percent Agreement | | 97.5% (156/160) | | 93.8 to 99.0% |
| Overall Percent Agreement | | 90.5% (209/231) | | |
{8}
The predicate device detects autoantibodies to the TSHR, of which there are two classes, stimulating (TSI-hyperthyroidism) and blocking (TBI-hypothyroidism). The predicate device is unable to distinguish between the two antibody types, whereas the TSI Reporter is specific for the TSI class. The patient populations tested in site 1 and 2 are different from site 3. Site 3 is a reference laboratory that receives specimens from all medical disciplines. The difference in patient populations tested at sites 1 and 2 vs. site 3 accounts for the decreased positive percent agreement value that is caused by the specificity difference between TSI Reporter and the predicate device at site 3.
b. Matrix comparison:
Not applicable. Serum is the only matrix.
3. Clinical studies:
a. Clinical Sensitivity and Specificity:
Sera from 50 Graves disease patients and 199 normal subjects were tested for assay sensitivity and specificity. The result is summarized in the following table:
| | Diagnosis | | | |
| --- | --- | --- | --- | --- |
| | | Positives (Graves Disease) | Negative (Other autoimmune diseases and healthy controls) | Totals |
| TSI Reporter | Positive | 46 | 1 | 47 |
| | Negative | 4 | 198 | 202 |
| | Total | 50 | 199 | 249 |
Clinical Sensitivity: 92.0% (46/50) 95%CI 81.2 – 96.9%
Clinical Specificity: 99.5% (198/199) 95%CI 97.2 – 99.9%
4. Clinical cut-off:
Same as Assay cut-off.
5. Expected values/Reference range:
A total of 140 normal samples with 72 female subjects and 68 male subjects were tested to determine the reference range of SRR%. The average SRR% for the female population is 34% (ranged from 9% to 116%) and the average SRR% for the male population is 37% (ranged from 5% to 116%). There are no apparent differences in the reference ranges for male and female populations. Expected value in normal population is negative.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.