Intended for use in reconstruction of natural or surgical periodontal defects of the oral and maxillofacial region, including sinus floor elevation and augmentation of the alveolar crest. Intended for filling into the site of a bony defect in combination with patient blood, autologous bone, membranes or sterile saline after removal of cysts or surgical removal of retained teeth.
Device Story
ArrowBone-B is a synthetic bone grafting material composed of porous, spherical granules of high-purity beta-tricalcium phosphate (beta-TCP). It is used by dental surgeons to fill and reconstruct periodontal and maxillofacial bony defects. The material is applied to the defect site, often mixed with patient blood, autologous bone, membranes, or sterile saline. As a biodegradable ceramic, it acts as a scaffold for new bone growth, gradually being absorbed and replaced by natural bone tissue. The device is provided as a sterile, biocompatible, and inert material, ensuring safety and structural support during the healing process.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing (x-ray diffraction, surface area analysis) and an animal implantation study conducted at Tokyo Medical and Dental University, which demonstrated the material's biocompatibility, biodegradation, and utility as a bone growth scaffold.
Technological Characteristics
Material: High-purity beta-tricalcium phosphate (beta-TCP) ceramic; Ca/P ratio 1.50. Form: Porous spherical granules. Standards: Complies with ASTM F 1088 (Standard Specification for Beta-Tricalcium Phosphate for Surgical Implants), ISO 10993 (biocompatibility), and ISO 11137 (radiation sterilization). Manufacturing: Wet synthesis, spray drying, and sintering.
Indications for Use
Indicated for patients requiring reconstruction of periodontal defects in the oral and maxillofacial region, including sinus floor elevation, alveolar crest augmentation, and filling of bony defects following cyst or tooth removal.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K083372
DEC - 8 2009
#### XIV. 510(k) Summary
Submitter: Satoshi Noake, International Division, BrainBase Corporation, Tokyo, Japan, Phone: +81-3-3778-0745, Fax: + 81-3-3778-4910.
- I. Classification Name and Number: Bone Grafting Material, Synthetic. Part 872.3930. Product code is LYC; Class II, special controls.
II. Proprietary Name(s): ArrowBone-ß
III. Registration No: 3005488486
- IV. Compliance with Performance Standards: No performance standards are applicable. However we followed "Class 11 Special Controls Guidance: Document: Dental Bone Grafting Material Devices", issued April 28, 2005 ArrowBone-B meets ASTM F 1088 "Standard Specification for Beta-Tricalcium Phosphate for Surgical Implants. ArrowBone-B meets ISO 10993 for biocompatibility and ISO 11137:1995, Sterilization of Health Care Products, requirements for validation and routine control, radiation sterilization. A risk analysis was prepared using ISO 14971:2000, "Application of risk management to medical devices."
- V. Description of the Device: ArrowBone-B consists of high purity Tricalcium Phosphate Ceramics of which the Ca/P ratio is 1.50. It is manufactured by a validated manufacturing process which guarantees pure-phase materials depending on the sintering temperature. This process is illustrated in Appendix I. Basically, it involves the well-established methods of wet synthesis from phosphoric acid and calcium hydroxide in dilute solution, spray drying to make spherical aggregates, calcination of the aggregates, x-ray diffraction analysis/identification of impurities, removal of fines by sieving, and molding of the particles into porous granules. These are then sintered at the selected temperature for the product which incinerates and removes the binder, hydroxypropyl cellulose, for which the specifications are shown in Appendix 1.6.
The chemical composition of this material is shown. It can be identified by x-ray powder diffraction analysis as pure-phase beta-Tricalcium Phosphate Ceramic in ArrowBone-B. The diffraction analysis is provided.. The identifying peaks as shown on these patterns for ßtricalcium phosphate are indicated by open circles. These data show that the crystalline structure of ArrowBone-B is pure-phase beta-Tricalcium Phosphate
- VI. Labels and Labeling: Draft labels for ArrowBone-B are provided, as are detailed instructions for use, warnings and contraindications.
{1}------------------------------------------------
VII. Substantial Equivalence: ArrowBone-B is substantially equivalent to several synthetic calcium phosphate materials. ArrowBone-B is nearly identical to Cerasorb Dental, Cerasorb M Dental, and CeraSorb P, which were cleared by Curasan AG in K051443, since these (as well as Arrowbone-{} consist of pure-phase Beta-Tricalcium Phosphate with a phase purity of more than 95% and comply with ASTM F 1088-04. ArrowBone-B differs only from Cerasorb products in its granular form in that the ArrowBone-B particles are of a porous spherical nature where as the Cerasorb forms are presented in various shaped porous particles including spherical. These synthetic products were cleared by the FDA in PMA800035 so they are well-established for dental use. Because they are synthetic they do not pose a risk of potential allergic reactions and are neither locally nor systemically toxic.
An extensive animal implantation study of ArrowBone-B was recently conducted by independent scientists at the Tokyo Medical and Dental University (See Section VIII. Animal Studies below for a review of the study). These scientists point out that many studies have shown that tricalcium phosphate is a biodegradable ceramic and useful as bone growth material. These observations demonstrate the biocompatibility and biodegradation of TCP, which was absorbed by new bone. These are the prime requirements of effective bone growth materials and show this device is substantially equivalent to currently marketed synthetic dental bone growth materials, and nearly identical to Ceresorb cleared in K051443.
ArrowBone-B is also substantially equivalent to Bioresorb Macro Pore, cleared by Oraltronics in K050260. We note that Curasan Cersorb, cleared in K014156 was cited in K050260 as substantially equivalent to Bioresorb. Ceresorb has a surface area of 0.17, Bioresorb has a surface area of 0.78 (m2.g-') almost 4.6 times as large. The surface area is 0.352 (m².gi') for ArrowBone-B. This is well within the range between Cerasorb and Bioresorb. The properties of ArrowBone-B are compared to Bioresorb and Cerasorb in Table 1. All are of interconnecting porous material with varying degrees of porosity and other physical characteristics. We believe this is more evidence of the substantial equivalence of ArrowBone-B to Bioresorb and Cerasorb.
The "510(k) "Substantial Equivalence" Decision-Making Process (Detailed) from ODE Guidance Memorandum #86-3 was followed as described below:
1. This product has the same Indications for Use as predicate devices:
Intended for use in reconstruction of natural or surgical periodontal defects of the oral and maxillofacial region, including sinus floor elevation and augmentation of the alveolar crest. .
Intended for filling into the site of a bony defect in combination with patient blood, autologous bone, membranes or sterile saline after removal of cysts or surgical removal of retained teeth.
{2}------------------------------------------------
2. The technological characteristics for this product are similar and in many cases nearly identical with those of the predicate devices and those currently on the market They are fine powders, relatively inert, biocompatible and previously used for periodontal purposes.
3. Descriptive information provided shows that the materials from which this device is made are well-established and well understood in the industry and among professional users.
4. The FDA "Decision-Making Process" chart was used and appears in Appendix VIII.
# (End of Summary)
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure with outstretched wings.
### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
DEC - 8 2009
BrainBase Corporation C/O Dr. H. Neal Dunning President Neal Dunning Associates 26420 Summer Greens Drive Bonita Springs, Florida 34135
Re: K083372
Trade/Device Name: ArrowBone-B Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: November 23, 2009 Received: November 24, 2009
Dear Dr. Dunning:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2- Dr. Dunning
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Nh for
Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{5}------------------------------------------------
## XII.1 Indications for Use
510(k) Number: K083372
Device Names: ArrowBone-B
### Indications for Use:
Intended for use in reconstruction of natural or surgical periodontal defects of the oral and maxillofacial region, including sinus floor elevation and augmentation of the alveolar crest.
Intended for filling into the site of a bony defect in combination with patient blood, autologous bone, membranes or sterile saline after removal of cysts or surgical removal of retained teeth.
Prescription Use X (Per 21 CFR 801 Subpart D)
or
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
RSBetz DDS for Dr. Kevin Muluy
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K083372
10
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.