K083358 · Farco Pharma GmbH · MUI · Jan 9, 2009 · Radiology
Device Facts
Record ID
K083358
Device Name
LUBRICANO-ULTRASOUND
Applicant
Farco Pharma GmbH
Product Code
MUI · Radiology
Decision Date
Jan 9, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
Lubricano is an in vivo biocompatible and bioexcretable sterile couplant intended for use as a scanning gel in surgical procedures, biopsies and similar sterile applications. Lubricano is used to couple sound waves between the patient and medical imaging electronic transducers during intraoperative and intracavitary medical diagnostic ultrasound imaging procedures, such as transcutaneous ultrasound image guided biopsy and aspiration, intraoperative ultrasound imaging, intracavity ultrasound imaging, and gel infusion sonography. Lubricano is unit dose packaged, sterilized and intended for use in all diagnostic ultrasound procedures that currently use an ultrasound coupling gel or other fluid, alone or in combination with a transducer cover, where sterility, in vivo biocompatibility and bioelimination are required.
Device Story
Lubricano Sterile Gel is a sterile, water-soluble acoustic coupling agent; composed of hydroxyethylcellulose, glycerol, and purified water; free from fats, latex, disinfectants, and preservatives. Packaged in 10 mL unit-dose syringes. Used by clinicians during intraoperative and intracavitary diagnostic ultrasound imaging (e.g., biopsies, aspirations, gel infusion sonography). Device applied to patient or transducer to facilitate sound wave transmission between patient and imaging transducer; ensures smooth movement of instruments; adheres to mucosa. Provides sterile, biocompatible, and bioexcretable interface for procedures where sterility is required. Benefits patient by enabling diagnostic imaging in sterile environments while minimizing risk of irritation or toxicity.
Clinical Evidence
No clinical data. Safety profile established via in vivo and in vitro testing confirming the gel is non-toxic, non-irritating, and non-sensitizing, comparable to predicate devices.
Technological Characteristics
Water-soluble gel; composition: hydroxyethylcellulose, glycerol, purified water. Form factor: 10 mL unit-dose syringe. Sterilized. No electronic components or software.
Indications for Use
Indicated for patients undergoing diagnostic ultrasound procedures requiring sterile, biocompatible, and bioexcretable coupling, including surgical procedures, biopsies, aspiration, and gel infusion sonography. Used for intraoperative and intracavitary imaging.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
UltraBio – In Vivo Biocompatible Bioeliminated Sterile Ultrasound Imaging Couplant (K042619)
UltraBio – In Vivo Biocompatible Bioeliminated Sterile Ultrasound Imaging Couplant (K033178)
{0}------------------------------------------------
l0r3358
#### 510(k) Summary
JAN - 9 2009
Submitted by Farco-Pharma GmbH Pharmazeutische Präparate Address: Gereonsmühlengasse 1 - 11 50670 Köln, Germany Telephone: +49 221 594061 Kornelia Ely-Koort, Regulatory Affairs Dept. Contact Name:
Date Submitted: December 11, 2008
Lubricano® Sterile Gel - Ultrasound Trade Name: Common Name: Ultrasound gel Product Code / Regulation: MUI (21 C.F.R. 892.1570)
Description: Lubricano® Sterile Gel is a sterile, water-soluble gel composed of hydroxyethylcellulose, glycerol, and purified water, which is contained in a 10 mL syringe. Lubricano Sterile Gel is free from fats, latex, disinfectants and preservatives. Lubricano Sterile Gel ensures that catheters and instruments move easily, and it adheres well to the mucosa.
Intended Use:Lubricano is an in vivo biocompatible and bioexcretable sterile couplant intended for use as a scanning gel in surgical procedures, biopsies and similar sterile applications.
> Lubricano is used to couple sound waves between the patient and medical imaging electronic transducers during intraoperative and intracavitary medical diagnostic ultrasound imaging procedures, such as transcutaneous ultrasound image guided biopsy and aspiration, intraoperative ultrasound imaging, intracavity ultrasound imaging, and gel infusion sonography.
Lubricano is unit dose packaged, sterilized and intended for use in all diagnostic ultrasound procedures that currently use an ultrasound coupling gel or other fluid, alone or in combination with a transducer cover, where sterility, in vivo biocompatibility and bioelimination are required.
| Substantial Equivalence: | Lubricano® Sterile Gel is substantially equivalent to the following devices:<br>SonoTech, Inc.'s UltraBio – In Vivo Biocompatible Bioeliminated Sterile Ultrasound Imaging Couplant (K042619 and K033178), and<br>Farco-Pharma GmbH's Lubricano® Sterile Gel (K081990). |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Results of in vivo and in vitro testing establishes the safety profile of the device as non-toxic, non-irritating and non-sensitizing, which is comparable to the predicate devices.
{1}------------------------------------------------
## 510(k) Summarv
Submitted by Farco-Pharma GmbH Pharmazeutische Präparate Address: Gereonsmühlengasse 1 - 11 50670 Köln, Germany Telephone: +49 221 594061 Contact Name: Kornelia Ely-Koort, Regulatory Affairs Dept.
Date Submitted: October 1, 2008
Lubricano® Sterile Gel - Ultrasound Trade Name: Common Name: Ultrasound gel Product Code / Regulation: MUI (21 C.F.R. 892.1570)
Description: Lubricano® Sterile Gel is a sterile, water-soluble gel composed of hydroxyethylcellulose, glycerol, and purified water, which is contained in a 10 mL syringe. Lubricano® Sterile Gel is free from fats, latex, disinfectants and preservatives. Lubricano® Sterile Gel ensures that catheters and instruments move easily, and it adheres well to the mucosa.
Intended Use:Sterile gel for acoustic coupling in ultrasound indications
Substantial Equivalence:
Lubricano® Sterile Gel is similar in intended use and technological characteristics to the predicate device reviewed as coupling gels used to couple ultrasound devices to the skin. The device is similar with respect to indications for use and physical characteristics to predicate device in terms of 510(k) substantial equivalency.
Results of in vivo and in vitro testing establishes the safety profile of the device as non-toxic, non-irritating and non-sensitizing, which is comparable to the predicate device.
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Lubricano® Sterile Gel – Ultrasound
Indications for Use:
Sterile gel for acoustic coupling in ultrasound indications
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal features the department's name in a circular arrangement around the perimeter. In the center is a stylized emblem of an abstract bird-like figure, with three curved lines representing its wings or body.
JAN - 9 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FARCO-PHARMA GmbH % Mr. Seth A. Mailhot Attorney Latham & Watkins LLP 555 Eleventh Street, NW WASHINGTON DC 20004-1304
Re: K083358
Trade/Device Name: Lubricano® Sterile Gel - Ultrasound Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: MUI Dated: December 11, 2008 Received: December 12, 2008
Dear Mr. Mailhot:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892.xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry.support/index.html.
Sincerely yours,
Laura B. Morris
anine M. M. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
### Indications for Use
#### 510(k) Number (if known):
Device Name: Lubricano® Sterile Gel -- Ultrasound
#### Indications for Use:
Lubricano is an in vivo biocompatible and bioexcretable sterile couplant intended for use as a scanning gel in surgical procedures, biopsies and similar sterile applications.
Lubricano is used to couple sound waves between the patient and medical imaging electronic transducers during intraoperative and intracavitary medical diagnostic ultrasound imaging procedures, such as transcutaneous ultrasound image guided biopsy and aspiration, intraoperative ultrasound imaging, intracavity ultrasound imaging, and gel infusion sonography.
Lubricano is unit dose packaged, sterilized and intended for use in all diagnostic ultrasound procedures that currently use an ultrasound coupling gel or other fluid, alone or in combination with a transducer cover, where sterility, in vivo biocompatibility and bioelimination are reauired.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sian-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Page 1 of 1
4-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.