K083342 · Evera Medical, Inc. · HPZ · Mar 12, 2009 · Ophthalmic
Device Facts
Record ID
K083342
Device Name
ORBITAL RECONSTRUCTIVE IMPLANT II
Applicant
Evera Medical, Inc.
Product Code
HPZ · Ophthalmic
Decision Date
Mar 12, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.3320
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ORI II is intended for augmentation or restoration in the craniofacial region.
Device Story
Non-absorbable, inert, sterile, porous implant for craniofacial/orbital augmentation; composed of porous silicone elastomer and expanded polytetrafluoroethylene (ePTFE). Available in various lengths/widths. Implanted by surgeons during procedures to restore orbital volume or correct enophthalmos following trauma or tumor excision. Provides structural support to orbital tissues. Bench testing confirms mechanical performance (tensile, shear, compression) and biocompatibility per ISO-10993.
Indicated for augmentation, reconstruction, or restoration in and around the orbit of the eye, including reconstruction following orbital trauma or tumor excision, treatment of orbital volume deficiencies, or correction of enophthalmos.
Regulatory Classification
Identification
An eye sphere implant is a device intended to be implanted in the eyeball to occupy space following the removal of the contents of the eyeball with the sclera left intact.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ocular peg which is supplied sterile only, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Solid Silicone and Silicone Foam Implants (K950806)
Submission Summary (Full Text)
{0}------------------------------------------------
KD83342
MAR 1 2 2009
#### 510(K) SUMMARY FOR THE 2.0 ORBITAL RECONSTRUCTIVE IMPLANT II
#### Submission Date: November 7, 2008
# Submitter Information:
| Company Name: | Evera Medical, Inc. |
|------------------|---------------------------------------------------------------|
| Company Address: | 353 Vintage Park Drive<br>Suite F<br>Foster City, CA 94404 |
| Contact Person: | Randy Kesten<br>650-287-2884<br>randy.kesten@everamedical.com |
# Device Information:
| Trade Name: | Orbital Reconstructive Implant II (ORI II) |
|--------------|--------------------------------------------|
| Common Name: | Orbital Implant |
Classification Name: 21 CFR § 886.3320
Classification Code: HPZ
Device Class: Class II
# Predicate Device(s):
| Trade Name: | Orbital Reconstructive Implant |
|---------------|------------------------------------------------|
| Manufacturer: | Evera Medical, Inc. |
| K Number: | K070130 |
| Product Code: | HPZ |
| Trade Name: | Medpor Plus Orbital Volume Replacement Implant |
| Manufacturer: | Porex Surgical, Inc. |
| K Number: | K021357 |
| Product Code: | HPZ |
| Trade Name: | Oculo-Plastik Universal-ePTFE |
| Manufacturer: | Oculo-Plastik, Inc. |
| K Number: | K934834 |
{1}------------------------------------------------
| Product Code: | HPZ |
|------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name:<br>Manufacturer:<br>K Number:<br>Product Code: | Solid Silicone and Silicone Foam Implants<br>MIRA, Inc.<br>K950806<br>HQX |
| Device Description: | The Orbital Reconstructive Implant II (ORI II) is a non-<br>absorbable, inert, sterile, porous, implant composed of<br>porous silicone elastomer and expanded<br>polytetrafluoroethylene (ePTFE). This polymeric implant is<br>available in a range of lengths and widths to accommodate<br>the surgical application and the needs of the individual<br>surgeon practicing medicine. |
| Intended Use: | The ORI II is intended for augmentation or restoration in<br>the craniofacial region. |
| Indications for Use: | The ORI II is indicated as an implant for augmentation,<br>reconstruction, or restoration in and around the orbit of the<br>eye, such as in reconstruction following orbital trauma or<br>tumor excision, to treat orbital volume deficiencies, or in<br>the correction of enophthalmos. |
| Comparison to<br>Predicate Device: | The ORI II has the same intended use and technological<br>characteristics as the predicate devices. Slight differences<br>in design and performance from the cited predicates do not<br>affect either the safety and/or effectiveness of the ORI II<br>for its intended use. As compared to the ORI predicate<br>(K070130), ORI II uses the same materials in its<br>construction: polydimethylsiloxane elastomer ("silicone<br>elastomer") and expanded polytetrafluoroethylene. The<br>slight design difference between ORI II and the ORI<br>predicate is that the silicone elastomer is processed to form<br>a porous core layer of the implant, while in ORI the<br>silicone was formed into a film core of the implant. The<br>slight performance difference between ORI II and the ORI<br>predicate is that ORI II exhibits approximately 50% greater<br>tensile strength than ORI, while the elongation at break of<br>ORI II is approximately 25% of that of ORI. |
{2}------------------------------------------------
Non-clinical performance data demonstrated in bench testing of ORI II are substantially equivalent to those of the predicates. The following testing was performed to measure the performance of ORI II as compared to the predicate devices: Tensile load at break; Tensile elongation at break; Elastic (Young's) modulus; Compression deflection; Uniaxial compressive modulus; Compressive set; Suture pullout; Shear modulus; Ultimate shear strength; and Shear load at break. Where applicable, test methods used the following testing standards: ASTM-D624; ASTM-D1777; ASTM-F88; ASTM-D412; and ASTM-D3787. The results of this testing showed that ORI II was demonstrated to be safe and effective for its intended use. ORI II will be manufactured in accordance with the product performance specifications derived from these test results.
The biocompatibility of ORI II was evaluated based on the long-term implant section of the ISO-10993 standard and shown to be biocompatible.
The safety and effectiveness evaluations based on biocompatibility and biomechanical performance data provided in this 510(k) demonstrate that the ORI II is substantially equivalent to the cited predicate devices.
### Conclusion:
The results of these evaluations of the ORI II support the conclusion that it is safe and effective for its intended use and that it is substantially equivalent to the cited predicate device(s) with regards to its safety and effectiveness.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circular pattern around the eagle. The eagle is depicted in black, and the text is also in black.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 2 2009
Evera Medical, Inc. c/o Randy Kesten Chief Technical Officer 1191 Chess Drive, Suite G Foster City, CA 94404
Re: K083342
Trade/Device Name: Orbital Reconstructive Implant II (ORI II) Regulation Number: 21 CFR 886.3320 Regulation Name: Implant, Eye Sphere Regulatory Class: II Product Code: HPZ Dated: February 10, 2009 Received: February 12, 2009
### Dear Mr. Kesten:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Trug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your devive con be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, TDA move publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements aft the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the eestonic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Malvina B. Egleston, M.D.
Malvina B. Eydelman, M.D Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### 1.0 STATEMENT OF INDICATIONS FOR USE
KO83342 510(k) Number (if known):
Device Name:
Orbital Reconstructive Implant II
Indications for Use:
The Orbital Reconstructive Implant II is indicated as an implant for augmentation, reconstruction, or restoration in and around the orbit of the eye, such as in reconstruction following orbital trauma or tumor excision, to treat orbital volume deficiencies, or in the correction of enophthalmos.
Caution: Federal law restricts this device to sale by or on the order of a medical practitioner licensed by the law of the State in which he / she practices to use or order the use of the device.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daryl L. Caubna
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices
CONFIDENTIAL
510(k) Number_k 08334 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.