K083309 · Zargis Medical Corp. · DQD · Feb 13, 2009 · Cardiovascular
Device Facts
Record ID
K083309
Device Name
ZARGIS ACOUSTIC CARDIOSCAN, VERSION 9.33
Applicant
Zargis Medical Corp.
Product Code
DQD · Cardiovascular
Decision Date
Feb 13, 2009
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1875
Device Class
Class 2
Indications for Use
The Zargis Acoustic Cardioscan; (ZAC), is an electronic auscultatory device, intended to provide support to the physician in the evaluation of patients' heart sounds. The product acquires and records the acoustic signals of the heart and analyzes these signals. The analysis procedure will identify specific heart sounds that may be present. Identified sounds include S1, S2, and suspected murmurs. The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis. The interpretations of heart sounds offered by the Zargis Acoustic Cardioscan are only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data.
Device Story
Zargis Acoustic Cardioscan (ZAC) is an electronic auscultatory system; acquires/records heart acoustic signals via Bluetooth-enabled electronic stethoscope; transmits data to PC via USB transceiver. Software analyzes signals to identify S1, S2, and suspected murmurs. Used in clinical settings by physicians or trained personnel. Output displayed on PC for physician review; supports clinical decision-making by providing automated sound identification. Not a sole diagnostic tool; requires physician over-read and integration with other patient data. Benefits include enhanced auscultation support.
Clinical Evidence
Bench testing only. Software verification performed at module and system levels. Integrated ZAC system evaluated via graphic user interface testing. Clinical software validation performed to ensure performance of enhanced software algorithm.
Technological Characteristics
Electronic auscultatory device; Bluetooth wireless connectivity; USB transceiver for PC interface. Software-based signal analysis. Complies with FCC requirements for Bluetooth transceiver.
Indications for Use
Indicated for use in clinical settings by physicians or trained personnel under physician orders to support evaluation of heart sounds in patients. Not intended as a sole diagnostic tool; requires physician over-read.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Nonin Medical, Inc. Model 4100 Patient Oximeter Module (K043359)
Submission Summary (Full Text)
{0}------------------------------------------------
K083309
R1/2
#### 510(k) Summary of Safety and Effectiveness FEB 1 3 2009
## Zargis Acoustic Cardioscan
Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Zargis Medical Corporation is required to submit with this Premarket Notification either an " ... adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Zargis Medical chooses to submit a summary of information respecting safety and effectiveness.
| Trade Name: | Zargis Acoustic Cardioscan (ZAC) |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Heart Sound Analyzer |
| Regulation Number: | 21 CFR 870.1875, 870.2390 |
| Classification Name: | Stethoscope, Electronic; Phonocardiograph |
| Product Code: | DQD, DQC |
| Regulatory Class: | Class II |
| Submitter Information: | Zargis Medical Corporation<br>2 Research Way, 1st Floor<br>Princeton, NJ 08540<br>Tel: 609.734.4747<br>Fax: 609.228.5000 |
| Summary Prepared By: | John Kallassy<br>CEO |
| Date Prepared: | November 3, 2008 |
| Predicate Devices: | Zargis Acoustic Cardioscan, K031517, K042128, K060197<br>AViTA Bluetooth Blood Pressure Monitor, K072137<br>CardioBelffM Electrode Belt, K063044<br>Nonin Medical, Inc. Model 4100 Patient Oximeter Module<br>K043359 |
Device Description: The ZAC is an electronic auscultatory device intended to acquire, record, and analyze heart sounds.
The complete system is a CD comprising software and instructions for use with an included USB transceiver for connection with a PC and capable of wireless
{1}------------------------------------------------
communication with a Bluetooth-enabled electronic stethoscope.
Intended Use: The Zargis Acoustic Cardioscan; (ZAC), is an electronic auscultatory device, intended to provide support to the physician in the evaluation of patients' heart sounds.
The product acquires and records the acoustic signals of the heart and analyzes these signals. The analysis procedure will identify specific heart sounds that may be present. Identified sounds include S1, S2, and suspected murmurs. The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.
The interpretations of heart sounds offered by the Zargis Acoustic Cardioscan are only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data.
Substantial Equivalence: The modified ZAC is similar in design/technological characteristics, indications for use, and performance characteristics to the currently cleared ZAC product and other medical devices capable of wireless transmission of physiological data through use of Bluetooth technology. In addition, non-clinical performance testing has been conducted to demonstrate the performance of the modified ZAC device and that it meets its intended use.
Safety and Performance: A comprehensive list of verification and validation testing was performed in accordance with Zargis' Design Control procedures.
Software verification of the modified components was performed at the module and system level. The integrated ZAC system was further evaluated to verify the graphic user interface software and the ZAC system as an integrated product.
Validation of the ZAC was performed to ensure that the ZAC device consistently fulfills its intended use and the needs of the user. A clinical software validation was performed to insure the performance of the enhanced software algorithm.
The Bluetooth transceiver complies with all relevant FCC requirements.
Conclusion: Based upon the indications for use, technological characteristics and safety and performance testing, the Zargis Acoustic Cardioscan (ZAC) has been shown to be substantially equivalent to the currently cleared predicate devices under the Federal Food, Drug, and Cosmetic Act.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# FEB 1 3 2009
Zargis Medical Corp. c/o Mr. John Kallassy, CEO 2 Research Way, 1st floor Princeton, NJ 08540
Re: K083309
> Trade/Device Name: Zargis Acoustic Cardioscan Regulation Number: 21 CFR 8.70.1875 Regulation Name: Electronic Stethoscope Regulatory Class: Class II Product Code: DQD, DQC Dated: January 13, 2009 Received: January 14, 2009
Dear Mr. Kallassy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a preinarket approval application (PMA): You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Page 2 - Mr. John Kallassy
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram D. Zuckerman, M.D. Director, Division of Cardiovascular Devices Office of Device Evaluation >> Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATION FOR USE
510(k) Number (if known): Not Assigned
Device Name:
Zargis Acoustic Cardioscan
Indications for Use:
The Zargis Acoustic Cardioscan, (ZAC) is an electronic auscultatory device, intended to provide support to the physician in the evaluation of heart sounds in patients.
The product will acquire and record the acoustic signals of the heart and analyze these signals. The analysis procedure will identify specific heart sounds. that may be present. Identified sounds include S1, S2, and suspected murmurs.
The device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician. It is not intended as a sole means of diagnosis.
The interpretations of heart sounds offered by the Zargis Acoustic Cardioscan are only significant when used in conjunction with physician over-read as well as consideration of all other relevant patient data.
| Prescription Use: | X | AND/OR | Over-The Counter Use: |
|----------------------------|-----------------|--------|-----------------------|
| (Per 21 CFR 801 Subpart D) | (Per 21 CFR 801 | | |
| SubpartC) | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off) | for B Zuckerman |
|------------------------------------|-----------------|
| Division of Cardiovascular Devices | 2/13/09 |
| 510(k) Number | K083309 |
|---------------|---------|
|---------------|---------|
016
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.