MICRODOT XTRA

K083273 · Cambridge Sensors Limited · NBW · Nov 19, 2009 · Clinical Chemistry

Device Facts

Record IDK083273
Device NameMICRODOT XTRA
ApplicantCambridge Sensors Limited
Product CodeNBW · Clinical Chemistry
Decision DateNov 19, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2
AttributesPediatric

Indications for Use

microdot® Xtra Blood Glucose Monitoring System microdot® Xtra Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood from a fingerstick. It is intended for use by persons with diabetes or by health care professionals in home settings or healthcare facilities. microdot ® Xtra Blood Glucose Monitoring System is intended for self testing of qlucose in capillary whole blood from a fingerstick by persons with diabetes or by health care professionals in home settings or healthcare facilities. It is not intended for use on neonates or in the diagnosis of or screening for diabetes mellitus. microdot® Xtra Blood Glucose Meter The microdot® Xtra Blood Glucose Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood from a fingerstick. It is intended for use by persons with diabetes or by health care professionals in home settings or healthcare facilities. microdot® Xtra Test Strips The microdot® Xtra Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood from a fingerstick. It is intended for use by persons with diabetes or by health care professionals in home settings or healthcare facilities. microdot® Control solutions The microdot® Control solutions are intended for use with microdot® Xtra Blood Glucose Meter and microdot® Xtra Test Strips as a quality control check to verify the accuracy of the blood glucose test results.

Device Story

System measures glucose in capillary whole blood; utilizes enzymatic conversion via glucose dehydrogenase; NAD+ cofactor reduced to NADH; mediator compound re-oxidizes NADH; meter detects resulting micro-current proportional to glucose concentration. Device is auto-coding; factory-calibrated; requires no user coding. Used by patients or healthcare professionals in home/clinical settings. Output displayed on meter screen for glucose monitoring. Supports quality control checks via control solutions. Modification from predicate involves transition to auto-coding and modified test strip port.

Clinical Evidence

No clinical data provided; substantial equivalence established through bench testing and comparison of technological characteristics to the predicate device.

Technological Characteristics

Glucose oxidase/dehydrogenase enzymatic sensing; electrochemical detection (micro-current). Auto-coding system with factory-set calibration. Components: meter, test strips, control solutions, lancing device. Standard: CLSI EP6-A for linearity evaluation.

Indications for Use

Indicated for self-testing or professional use in home/healthcare settings for quantitative glucose measurement in fresh capillary whole blood from fingersticks. Not for neonates, diabetes diagnosis, or screening.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k083273 B. Purpose for Submission: Modification of predicate meter that required coding to new auto code meter in which there is one pre-programmed code. D. Type of Test: Quantitative (glucose oxidase) E. Applicant: Cambridge Sensors Limited F. Proprietary and Established Names: microdot Xtra Blood Glucose Meter G. Regulatory Information: 1. Regulation section: 21 CFR 862.1345 Glucose Test System 21 CFR 862.1660 Quality control material (assayed and unassayed) 2. Classification Class II – Glucose Class I - Controls 3. Product code: NBW – System, Test, Blood Glucose, Over the Counter CGA – Glucose Oxidase, Glucose JJX --Single analyte control (assayed and unassayed) {1} 4. Panel: 75, Clinical Chemistry H. Intended Use: 1. Intended use(s): See Indications for use below 2. Indication(s) for use: The microdot Xtra Blood Glucose Monitoring System is intended for self testing of glucose in capillary whole blood by persons with diabetes or by health care professionals in home settings or healthcare facilities. It is intended for monitoring of blood glucose levels only. It is intended not intended for use on neonates or in the diagnosis of or screening for diabetes mellitus Microdot Xtra Blood Glucose Meter The microdot Xtra Blood Glucose Meter is intended for the quantitative measurement of glucose in fresh capillary whole blood from a fingerstick. It is intended for use by persons with diabetes or by health care professionals in home settings or healthcare facilities. It is intended for monitoring of blood glucose levels only. microdot Xtra Test Strips The microdot® Xtra Test Strips are intended for the quantitative measurement of glucose in fresh capillary whole blood from a fingerstick. It is intended for use by persons with diabetes or by health care professionals in home settings or healthcare facilities. It is intended for monitoring of blood glucose levels only. microdot Control Solutions microdot Control Solutions are intended for use with the microdot Xtra Blood Glucose meter and microdot Xtra Test strips as a quality control check to verify the accuracy of the blood glucose test results 3. Special conditions for use statement(s): For over the counter use. Not intended for the diagnosis of or screening for diabetes and not intended for use on neonates. The device should not be used for patients who are dehydrated, in shock, critically ill or in a hyperosmolar state. {2} 4. Special instrument requirements: microdot Xtra Glucose Meter I. Device Description: The microdot Xtra Blood Glucose Monitoring System consists of the microdot Blood Glucose Meter, microdot Xtra Test Strips, microdot Xtra Control Solutions and a commercially available lancing device and lancets. J. Substantial Equivalence Information: 1. Predicate device name(s): microdot Blood Glucose Monitoring System 2. Predicate 510(k) number(s): k070524 3. Comparison with predicate: The modified device has the same performance characteristics as the predicate. The difference is the new device is an auto code meter and the test strip port has been modified to accommodate the microdot Xtra test strip. K. Standard/Guidance Document Referenced (if applicable): CLSI Guideline EP6-A – Evaluation of the linearity of Quantitative Analytical Methods L. Test Principle: The test is based on the enzymatic conversion of glucose. Glucose dehydrogenase converts glucose in the sample to gluconolactone, with concomitant reduction of the enzyme cofactor NAD+ to NADH. The NADH is re-oxidised to NAD+ by the mediator compound which in turn becomes reduced; re-oxidation of the mediator by the meter induces a micro current to flow, and the size of this micro current is directly proportional to the amount of glucose in the sample. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Precision and reproducibility were evaluated in k070524 b. Linearity/assay reportable range: Linearity was evaluated in k070524 {3} c. Traceability, Stability, Expected values (controls, calibrators, or methods): The controls supplied with this device were cleared under k070524. d. Detection limit: The detection limit of the microdot Xtra is 20 mg/dL, and was evaluated in k075024. e. Analytical specificity: Specificity was evaluated in k070524. f. Assay cut-off: Not applicable 2. Comparison studies: a. Method comparison with predicate device: Method comparison studies were evaluated in k070524 b. Matrix comparison: Not applicable. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable {4} 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: The expected blood glucose values for people without diabetes¹: Time Range, mg/dL Before meals 70-105 1 hour after meals Less than 160 2 hours after meals Less than 120 Between 2 and 4 a.m. Greater than 70 ¹Beaser, R.S. and Hill, Joan: The Joslin Guide to Diabetes, New York: Simon and Schuster (2005). N. Instrument Name: Microdot Xtra Blood Glucose Meter O. System Descriptions: 1. Modes of Operation: Each test strip is single use and must be replaced with a new strip for each additional reading. 2. Software: FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ☐ X ☐ or No ☐ 3. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected. 4. Specimen Sampling and Handling: This device is intended to be used with capillary whole blood from the finger only. Since the whole blood sample is applied directly to the test strip, there are no special handling 5 {5} or storage issues. 5. Calibration: Calibration is factory set by the manufacturer. No coding by the user is required. 6. Quality Control: One control level is provided with the device. Two additional control levels are available from authorized distributors. The user is instructed to run controls when the meter is first used in order to verify that they can use the meter correctly. In addition they are instructed to run a control when a new vial of test strips is opened, at least once a week, when they suspect the meter or strips are not working correctly, when test results are not consistent with the patient's symptoms or the patient does not think the results are accurate, or if the meter is dropped. The acceptable results ranges are shown on the test strip vial label. If the results are outside the expected range, the user is instructed to repeat the test. If the results continue to fall outside the expected range, the user is instructed to call customer service. P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: The new device is a modification of predicate meter that required coding, to a new auto code meter in which there is one pre-programmed code. The sponsor indicated the microdot extra test strip is identical in composition and manufacture to the microdot test strip and only lots of microdot test strips that have a specific calibration curve are designated as the microdot extra test strip. The unique calibration curve is determined through the QC release process which utilizes a library of multiple calibration codes. The qualification plan for the new auto code meter which included documentation of design inputs, risk analysis, design output, test procedures, verification and validation procedures, and documentation of formal design reviews was reviewed and found to be acceptable. Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...