K083264 · Raysearch Laboratories AB · MUJ · Jan 22, 2009 · Radiology
Device Facts
Record ID
K083264
Device Name
RAYAUTOPLAN 1.0
Applicant
Raysearch Laboratories AB
Product Code
MUJ · Radiology
Decision Date
Jan 22, 2009
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
Ray Autoplan is a software that, based on the planned dose distribution from a TomoPlan treatment planning system, generates a family of individually optimized IMRT plans and presents their characteristics to the user in a GUI Among the generated plans, the user selects and climically approves a treatment plan and exports it electronically to DICOM, for subsequent treatment of the pattent The intended users of RayAutoplan shall be clinically qualified staff, such as medical physicists, medical doctors or dosimetrists
Device Story
RayAutoplan is a standalone software program for radiation treatment planning. It takes patient descriptions (CT images, regions of interest) and a reference dose distribution from a TomoTherapy treatment planning system as input. The system generates a family of individually optimized IMRT plans using a three-dimensional collapsed cone dose engine. The software displays 2D/3D dose distributions, dose difference maps, dose-volume histograms (DVHs), and dose statistics in a GUI. Clinically qualified staff (physicians, physicists, dosimetrists) review these plans, select one for approval, and export the final plan via DICOM to an R&V system for patient treatment. The device assists clinicians in efficient plan generation and selection, potentially improving treatment planning workflows.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on technological characteristics and performance comparisons to the predicate device.
Technological Characteristics
Standalone software; runs on Windows XP or Vista. Uses a three-dimensional collapsed cone dose engine for IMRT plan calculation. Features GUI for plan visualization (2D/3D dose, DVHs, statistics). Connectivity via DICOM for data import/export. No specific hardware materials or energy sources as it is a software-only device.
Indications for Use
Indicated for use by clinically qualified staff (medical physicists, doctors, dosimetrists) to generate and optimize IMRT treatment plans based on reference dose distributions from a TomoPlan system for subsequent patient treatment.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
| Title | Document ID | Version |
|----------------------------------|-------------|---------|
| 510(k) Application - RayAutoplan | RSL-D-58-10 | 1.0 |
# 5. 510(k) Summary RayAutoplan
- 5 1 510(k) owner
RaySearch Laboratories AB Sveavägen 25, plan 9 111 34 Stockholm Sweden
+46 (8) 54506130 Tel Fax +46 (8) 54506139
- 5 2 Contact person
Anders Murman
- 5 3 Preparation date
09/01/2008
# 5 4 Trade name
RayAutoplan, alias t-RayAutoplan, alias SharePlan
RayAutoPlan is the product name RaySearch will use moving this device into the world market
t-RayAutoplan was the name of the product during the development project, and when seen in the documentation related to this application, it is the same as RayAutoplan, by definition
SharePlan is the trade name TomoTherapy Inc will use on the world market, as being the exclusive distributor of this device This name is not further used in the documentation of this application
## 5 5 Common name
Radiation treatment planning system
## 5 6 Classification name
Medical charged-particle radiation therapy system (21 CFR 892 5050, Product Code MUJ)
#### 5 7 Predicate devices
Oncentra MasterPlan v3 1 510(k) number K081281
# 5 8 Device description
RayAutoplan is a stand-alone software program that generates a number of treatment plans, based on the dose distribution exported from a TomoTherapy treatment planning system
Detailed information of RayAutoplan can be found in reference 1, Customer requrement specifications
However, the most important workflow is described below
| Flow of Events | | |
|----------------|------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| | User | System |
| 1 | The user launches RayAutoplan | 3 The system imports the data and checks<br>consistency of in-data |
| 2 | The user selects the transferred<br>TomoTherapy case from a list of<br>TomoTherapy cases | |
| 4 | The user selects machine and treatment<br>energy | |
| 5 | The user selects MLC delivery mode<br>(SMLC/DMLC) | |
| 6 | The user specifies beam configuration | |
| 7 | The user specifies which plan family or<br>families that should be generated | |
KOB3uey
JAN 2 2 2009
{1}------------------------------------------------
| Title | Document ID | |
|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|
| 510(k) Application - RayAutoplan | RSL-D-58-10 | 10 |
| 8 The user starts the automatic plan generation | 9 The system generates plans<br>10 The system displays plans as<br>- 2D and 3D dose and patient displays<br>- Dose difference maps<br>- DVHs (several overlaid in same graph)<br>- Dose statistics<br>- Plan data (machine, beams, segments and so on) | |
| 11 The user reviews the Linac plans<br>12 The user selects a plan for approval and generates plan documentation | 13 The system creates a report containing<br>- Patient info<br>- Plan info<br>- Dose info | |
| 14 The user approves the plan for delivery | 15 Prints plan documentation and exports | |
Alternative workflows exist, including automatic plan generation and beam commissioning
To execute, RayAutoplan needs the following unput
- . A patient description, normally CT images and regions of interests
- A dose distribution serving as a reference dose, which should be benchmarked ●
As output, RayAutoplan produces a inac IMRT treatment plan and the corresponding dose distribution from an accurate calculation of the dose engine
plan to DICOM archive or R&V system
The software runs on a Windows XP or Windows Vista platform
## 5 9 Intended use
Ray Autoplan is a software that, based on the planned dose distribution from a TomoPlan treatment planning system, generates a family of individually optimized IMRT plans and presents their characteristics to the user in a GUI Among the generated plans, the user selects and climically approves a treatment plan and exports it electronically to DICOM, for subsequent treatment of the pattent
The intended users of RayAutoplan shall be clinically qualified staff, such as medical physicists, medical doctors or dosimetrists
## 5 10 Technological characteristics summary
The technological characteristics are the same for Ray Autoplan as for Oncentra MasterPlan v3 1 Both devices produce IMRT treatment plans with corresponding dose distributions computed using a firee dimensional collapsed cone dose engine Both devices have a function of electronic clinical approval of treatment plans by trained and authorized staff, and finally export in DICOM format for commencing treatment or archiving
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo. The logo features a stylized bird-like symbol with three curved lines representing its wings. The bird is oriented towards the right. Encircling the bird is text that reads "DEPARTMENT OF HUMAN SERVICES (USA)" in a circular arrangement. The text is in all capital letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
RaySearch Laboratories AB % Mr Jay Y Kogoma Senior Staff Engineer Intertek Testing Services 2307 E Aurora Rd . Unit B7 TWINSBURG OH 44087
JAN 2 2 2009
Re K083264
Trade/Device Name RayAutoplan 1 0 Regulation Number 21 CFR 892 5050 Regulation Name Medical charged-particle radiation therapy system Regulatory Class II Product Code: MUJ Dated January 5, 2008 Received January 6, 2008
# Dear Mr Kogoma
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA)_ You may, therefore, market the device, subject to the general controls provisions of the Act The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteratıon
If your device is classified (see above) into etther class II (Special Controls) or class III (PMA), it may be subject to such additional controls Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In addition, FDA may publish further announcements concerning your device in the Federal Register
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies You must comply with all the Act's requirements, including but not limited to registration and listing (21 CFR Part 807), labeling (21 CFR Part 801, good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820), and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification The FDA finding os substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter
| 21 CFR 876 xxx | (Gastroenterology/Renal/Urology) | (240) 276-0115 |
|----------------|----------------------------------|----------------|
| 21 CFR 884 xxx | (Obstetrics/Gynecology) | (240) 276-0115 |
| 21 CFR 892 xxx | (Radiology) | (240) 276-0120 |
| Other | | (240) 276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807 97) For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474 For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464 You may obtain other general information on your responsibilities under the Act from the Division of Small Manufactures, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http //www fda gov/cdrh/industry suppor/index.html
Sincerely yours,
Lamme M. Morre
anıne M N Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
| Title | Document ID | Version |
|----------------------------------|-------------|---------|
| 510(k) Application - RayAutoplan | RSL-D-58-10 | 1.0 |
# 4. Indications for Use Statement
510(k) Number (if known) - Device Name RayAutoplan 1 0
2083264
Indications For Use
RayAutoplan is a software that, based on the planned dose distribution from a TomoPlan treatment planning system, generates a family of individually optimized IMRT plans and presents their characteristics to the user in a GUI Among the generated plans, the user selects and clinically approves a treatment plan and exports it electronically to DICOM, for subsequent treatment of the pattent
The intended users of RayAutoplan shall be clinically qualified staff, such as medical physicists, medical doctors or dosimetrists
Prescription Use YES (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use NO (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hales Reem
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.