TC ENDO PORT

K083218 · Transcardiac Therapeutics, Inc. · GCJ · Mar 6, 2009 · Gastroenterology, Urology

Device Facts

Record IDK083218
Device NameTC ENDO PORT
ApplicantTranscardiac Therapeutics, Inc.
Product CodeGCJ · Gastroenterology, Urology
Decision DateMar 6, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The TC Endo Port has applications in abdominal, thoracic, laparoscopic and gynecologic minimally invasive procedures to establish a path of entry for endoscopic instruments.

Device Story

TC Endo Port is a sterile, single-use trocar system (5-12 mm diameter) for minimally invasive surgery; consists of sleeve and plastic obturator with dilator tip. Sleeve features dual seals (annular and internal) to minimize gas/fluid leakage and a stopcock/luer valve for insufflation/desufflation or fluid management. Unlike predicate devices using barbs, TC Endo Port utilizes inflatable balloons along the shaft for securement and hemostatic sealing. Used by surgeons in OR settings to provide access for endoscopic instruments. Benefits include customizable securement for varying patient needs and maintenance of pneumoperitoneum.

Clinical Evidence

No clinical tests were conducted. Substantial equivalence is based on non-clinical bench testing, including sterilization validation, valve leak/pressure testing, balloon inflation/deflation/burst strength, and mechanical tensile strength of interfaces.

Technological Characteristics

Sterile, single-use trocar (5-12 mm). Materials: plastic obturator. Securement: inflatable balloons. Seals: dual-seal system. Connectivity: standard luer lock fittings. Sterilization: EtO gas. Mechanical testing: tensile strength (15 N), pressure withstand (450 mmHg).

Indications for Use

Indicated for patients undergoing abdominal, thoracic, laparoscopic, or gynecologic minimally invasive procedures requiring an entry path for endoscopic instruments.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K083218 Pg 1 of 5 ## 510(K) SUMMARY # MAR - 6 2009 ### 1) Submitter's Name: TransCardiac Therapeutics Inc 3355 Lenox Road NE Suite 415 Atlanta, GA 30326 Tel: 404-816-4520 Ext. 105 Fax: 404-816-4530 #### Contact Person (s): #### Primary Contact Person David Smith · Chief Technology Officer Tel: 404-816-4520 ext 105 Fax: 404-816-4530 Cell: 404-353-3717 Establishment registration number: TBD #### Date the summary was prepared: 10-27-2008 #### 2) Device Identification: | Proprietary Name: | TC Endo Port | |------------------------|------------------------------| | Common Name: | Endoscope and Accessories | | Classification Name: | Endoscope and Accessories | | Classification Number: | 21 CFR 876.1500 | | Classification Panel: | Gastroenterology and Urology | | Product Code: | GCJ | ## Secondary Contact Person Lina Lattouf Vice President Scientific & Regulatory Tel: 404-816-4520 ext 108 Fax: 404-816-4530 {1}------------------------------------------------ #### 3) Legally marketed device to which equivalence is claimed: | Manufacturer: | Ethicon Endo-Surgery | |------------------------|------------------------| | Device Name: | Endopath Trocar System | | 510(K) Identification: | K032676. | #### 4) DEVICE DESCRIPTION TC Endoport trocars are sterile single patient use instruments consisting of a sleeve and obturator in sizes ranging from 5-12 mm in diameter. The TC Endoport obturator is made from plastic, and has a dilator tip which gently moves aside any internal viscera that may be adjacent to the abdominal or thoracic wall. The trocar sleeve contains two seals, an outer annular seal that accommodates instruments from 5-12 mm in diameter, and an internal seal. Together these seals minimize gas or fluid leakage when instruments are inserted or withdrawn through the trocar. A stopcock valve, or valved luer are provided with standard luer lock fittings, to provide for gas insufflation, desufflation, or infusion and aspiration of fluids. The device depiction section provides a more detailed description of the device, and the predicate. #### 5) Statement of Intended Use The TC Endo Port has applications in abdominal, thoracic, laparoscopic and gynecologic minimally invasive procedures to establish a path of entry for endoscopic instruments. #### 6) Technological Characteristics #### a) Summary of technological characteristics of the TC Endo Port compared to the predicate device The technological characteristics of the new device are different from the predicate in that the trocar securement mechanism is created through the use of inflatable balloons placed along the trocar shaft instead of barbs as in the predicate device. This provides for a secure hemostatic seal, and allows for the device to be customized for different patient needs. {2}------------------------------------------------ ## b) Assessment of Performance Data ## 1) Summary of the non-clinical performance data The following non-clinical tests were performed as the basis for establishing substantial equivalence to the predicate device, as well as safety and effectiveness of the indication for use: | Test Name | Sample<br>Size | Test Description | Required<br>Results | |-----------------------------------------|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Sterilization<br>Exposure | All | The samples are exposed<br>to a 2X sterilization cycle. | All samples must<br>be capable of<br>withstanding a<br>2Xsterilization<br>cycle. | | Simulated<br>implant | 30<br>1 Pred | Simulate placing the device<br>using an implant simulator.<br>Measure insertion forces<br>and compare to predicate<br>device | All samples must<br>be capable of being<br>implanted in<br>simulated<br>conditions using<br>the instructions for<br>use. Insertion<br>forces must be<br>equal to or less<br>than predicate<br>device | | Valve tool pass<br>verification | 30 | Pass tools of various size<br>through the distal valve and<br>verify that tools can be<br>accommodated without<br>damage or excessive<br>leakage | Must pass the<br>range of<br>recommended tools<br>through the valve<br>without damaging<br>the valve or<br>causing the loss of<br>more than 10ml of<br>fluid | | Valve cyclic<br>test | 30 | Pass tools through the<br>valves repeatedly until<br>failure occurs or maximum<br>number of tool passes<br>requirement is reached | Must pass the<br>maximum sized<br>tool through the<br>valve ten times<br>without damaging<br>the valve or<br>causing the loss of | | | | | more than 10ml of<br>fluid | | Valve pressure<br>withstand | 30 | Determine that valves meet<br>the minimum pressure<br>requirement without failure.<br>Compare to predicate<br>device | Must withstand a<br>fluid pressure of<br>450mmHg without<br>leaking | | Balloon<br>diameter<br>verification | 30 | Measure the inflated<br>diameter and length of the<br>distal and proximal balloons<br>when fully inflated | Profile of balloons<br>must be within<br>15%-25% of<br>expected values<br>from design | | Balloon volume<br>loss | 30 | Determine that volume loss<br>of balloons over time is<br>within specification | Balloons cannot<br>lose more than<br>0.5ml over one<br>hour | | Balloon<br>inflation/<br>deflation time | 30 | Measure time required to<br>inflate and deflate balloons | Balloons must<br>inflate and deflate<br>in under 15<br>seconds | | Balloon cyclic<br>test | 30 | Repeatedly inflate/deflate<br>balloons until failure occurs<br>or minimum number of<br>cycles required is reached | Balloons and<br>inflation lumens<br>must withstand 40<br>inflation/deflation<br>cycles to 1 atm<br>without failure. | | Balloon<br>minimum burst<br>strength | 30 | Measure burst pressure and<br>volume. Verify that<br>minimum is achieved | Balloon should<br>withstand 2X SF<br>over 1 atm | | Cannula to hub<br>strength | 30 | Measure tensile forces.<br>Verify that minimum is<br>achieved | Material interface<br>should resist a<br>tensile force of up to<br>3.4 lb (15 N). | | Inflation lumen<br>to hub strength | 30 | Measure tensile forces.<br>Verify that minimum is<br>achieved | Material interface<br>should resist a<br>tensile force of up to<br>3.4 lb (15 N). | | Valve body to<br>cannula<br>strength | 30 | Measure tensile forces.<br>Verify that minimum is<br>achieved | Material interface<br>should resist a<br>tensile force of up to<br>3.4 lb (15 N). | | Dilator<br>insertion/remov<br>al force | 30 | Measure insertion removal<br>forces, verify that forces are<br>less than tensile forces | Material interface<br>should resist a<br>tensile force of up to<br>3.4 lb (15 N). | {3}------------------------------------------------ K083218 pg4 ofs ・ {4}------------------------------------------------ The TC Endo port and components are packaged in industry-standard, EtO gaspermeable sterile barrier materials using industry-standard package sealing processes. The packaging is an inner tray containing the port, and introducer components sealed with a Tyvek lid within an outer tray that is also sealed with a Tyvek lid. 2) No clinical tests were conducted as part of this products development. ## 3) SUMMARY OF CONCLUSIONS The TC Endo Port and the predicate devices have identical intended uses and fundamental scientific technology. The subject and predicate devices are substantially similar in configuration, dimensions, and materials. No new questions of safety or effectiveness were raised during the performance testing of the device. The TC Endo Port has been determined to be substantially equivalent to the predicate device. 895 075 {5}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES · USA Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 TransCardiac Therapeutics, Inc. % Mr. David Smith VP and Chief Technology Officer 3355 Lenox Road North East, Suite 415 Atlanta, Georgia 30326 - - - MAR - 6 2009 Re: K083218 Trade/Device Name: TC Endo Port Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: February 6, 2009 Received: February 9, 2009 Dear Mr. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {6}------------------------------------------------ #### Page 2 - Mr. David Smith This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, fuchti, m.D. for Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ K08/3218 pg 1041 #### INDICATIONS FOR USE 510(k) Number (if known): Device Name: TC Endo Port Indications for Use: The TC Endo Port has applications in abdominal, thoracic, laparoscopic and gynecologic minimally invasive procedures to establish a path of entry for endoscopic instruments. AND/OR Over-The-Counter Use Prescription Use × (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Niel RP Sgden for nixon (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number K083218 14
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...