The Primado Surgical System is an AC-electrically powered total surgical system that allows for the use of multiple attachments to perform delicate bone dissection in neurological, maxillofacial, craniotomy, and spinal procedures, and in the ear, nose, and throat area. A wide range of attachments includes a perforator drive, craniotome, straight/angle/malleable drills, and microsaws.
Device Story
Primado Surgical System is an AC-electrically powered surgical motor system; used for delicate bone dissection in neurological, maxillofacial, craniotomy, spinal, and ENT procedures. System components include control unit, foot controller, micromotor, motor cord, and various attachments (perforator drive, craniotome, drills, microsaws). Operated by surgeons in clinical settings; foot controller regulates motor speed. Device provides mechanical power to attachments for cutting bone and soft tissue. Benefits include versatility through interchangeable attachments for diverse surgical applications. Device is non-sterile; requires sterilization prior to use.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
AC-electrically powered surgical motor system. Components: control unit, foot controller, micromotor, attachments. Materials: surgical stainless steel, diamond. High speed up to 100,000 rpm; low speed up to 234 rpm; oscillating/sagittal saw up to 20,000 cpm; reciprocating saw up to 19,300 cpm. Standards: IEC 60601-1, IEC 60601-1-2. Non-sterile; requires sterilization.
Indications for Use
Indicated for patients requiring delicate bone dissection during neurological, maxillofacial, craniotomy, spinal, and ear, nose, and throat surgical procedures.
Regulatory Classification
Identification
An electric cranial drill motor is an electrically operated power source used with removable rotating surgical cutting tools or drill bits on a patient's skull.
{0}------------------------------------------------
K083112
pg 1 of 3
#### 510(k) SUMMARY
### Nakanishi Inc. Primado Surgical System
## Name of Device and Name/Address of Sponsor
| Trade or Proprietary Name: | Primado Surgical System |
|----------------------------|-------------------------------------------------------------------------------------------------------------------|
| Common Name: | electric surgical motorsystem; ENT surgical drill system |
| Classification Name: | motor, drill, electric; ear, nose, and throat electric or pneumatic surgical drill (21 CFR 882.4360 and 874.4250) |
| Product Code: | HBC, ERL |
Nakanishi Inc. 700 Shimohinata Kanuma-shi Tockigi-ken 322-8666 Japan
| Contact: | Ms. Tomoko Hirabayashi |
|----------------|------------------------|
| Telephone: | 011-81-289-64-3380 |
| Facsimile: | 011-81-289-62-6665 |
| Date Prepared: | April 23, 2009 |
#### Intended Use
The Primado Surgical System is an AC-electrically powered total surgical system that allows for the use of multiple attachments to perform delicate bone dissection in neurological, maxillofacial, craniotomy, and spinal procedures, and in the ear, nose, and throat area. A wide range of attachments includes a perforator drive, craniotome, straight/angle/malleable drills, and microsaws.
#### Technological Characteristics and Substantial Equivalence
The Primado Surgical System is an AC-electrically powered total surgical system that allows for the use of multiple attachments to perform delicate bone dissection in neurological procedures. The device consists of a control unit, foot controller, micromotor, motor cord, AC power cord, perforator drive, craniotome, straight/angle/malleable drills, and microsaws. The device does not come sterile and all parts that contact the patient as well as the AC power cord should be sterilized prior to the first use and after each use. The Primado has already been cleared for marketing for "delicate bone dissection in the ear, nose, and throat areas" under K#080722.
The Primado Surgical System is substantially equivalent in terms of safety and effectiveness to NSK's own previously authorized Primado device (K#080722) and the Advantage Drive System (K#002523). A chart comparing the Primado to the predicate devices is attached.
APR 2 9 2009
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows handwritten text that appears to be part of a document or page numbering system. The text includes the identifier "K083112" at the top, followed by "pg. 2 of 3" below it. The handwriting is in black ink and has a slightly tilted orientation.
# Substantial Equivalency Chart
| ITEMS | NEW DEVICE | PREDICATE<br>DEVICE | PREDICATE<br>DEVICE |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Classification<br>No. | HBC, ERL | ERL | HBC, GEY |
| 2. C.F.R. Section<br>No. | 882.4360 and<br>874.4250 | 874.4250 | 882.4360 and<br>878.4820 |
| 3. Classification<br>Panel | Neurology and Ear,<br>Nose & Throat | Ear, Nose & Throat | Neurology and<br>General and Plastic<br>Surgery |
| 4. Classification<br>Name | Motor, Drill,<br>Electric | Drill, Surgical, ENT<br>(Electric Or<br>Pneumatic) Including<br>Handpiece | Motor, Drill, Electric<br>and Motor, Surgical<br>Instrument, Ac-<br>Powered |
| 5. Class | II | II | II |
| 6. 510(k) Number | N/A | K#080722 | K#002523 |
| 7. Manufacturer | NAKANISHI INC. | NAKANISHI INC. | LINVATEC CORP |
| 8. Proprietary<br>Name | Primado Surgical<br>System | Primado<br>Neurological Drill | Advantage Drive<br>System |
| 9. Description | Electrically<br>powered drill and<br>surgical motor<br>system and<br>attachments<br>operated by foot<br>controller | Identical | Electrically powered<br>surgical instrument<br>system and<br>handpieces operated<br>by foot controller |
| 10. Intended use | Delicate bone<br>dissection in<br>neurological,<br>maxillofacial,<br>craniotomy, and<br>spinal procedures,<br>and in the ear nose,<br>and throat areas | Delicate bone<br>dissection in the ear,<br>nose, and throat areas | Cutting of soft tissue<br>and bone in<br>arthroscopic, foot,<br>hand, medial<br>sternotomy,<br>neurosurgical,<br>orthopedic,<br>otolaryngological,<br>oral/maxillofacial,<br>plastic/<br>reconstructive, and<br>spinal surgical<br>procedures |
| 11. Compliance<br>with standards | IEC 60601-1<br>IEC 60601-1-2 | Identical | IEC 601-1<br>IEC 601-1-2<br>UL2601-1 |
| 12. High speed | Up to 100,000 rpm | Identical | Identical |
| 13. Low speed | Up to 234 rpm | Identical | Up to 30 rpm |
| 14. Oscillating<br>saw | Up to 20,000 cpm | Identical | Identical |
| 15. Sagittal saw | Up to 20,000 cpm | Identical | Identical |
| 16. Reciprocating<br>saw | Up to 19,300 cpm | Identical | Up to 17,000 cpm |
| 17. Patient contact<br>materials | Surgical stainless<br>steel; diamond | Identical | Unknown |
{2}------------------------------------------------
40560854.doc
.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The eagle is black, and the text is also black.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Nakanishi, Inc. c/o Keith A. Barritt Fish & Richardson P.C. 1425 K Street, N.W. Suite 1100 Washington, DC 20005
APR 2 9 2009
Re: K083112
Trade/Device Name: Primado Surgical System Regulation Number: 21 CFR 882.4360 Regulation Name: Electric cranial drill motor Regulatory Class: Class II Product Code: HBC Dated: March 10, 2009 Received: March 11, 2009
Dear Mr. Barritt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B. Eglehus, MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a handwritten text that appears to be a combination of letters and numbers. The text is oriented diagonally, with the characters slightly distorted and overlapping. The text includes the letters 'K' and 'o' along with the numbers '8', '3', '1', and '2'. The overall impression is that the text is a code or identifier of some kind.
## Indications for Use 1. 1. 1. 1.
: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
510(k) Number (if known): 083112
.. .. ... .. . . . .
Device Name: Primado
Indications For Use:
The Primado is an AC-electrically powered total surgical system that allows for the use of multiple attachments to perform delicate bone dissection in neurological, maxillofacial, craniotomy, and spinal procedures, and in the ear nose, and throat area. A wide range of attachments includes a perforator drive, craniotome, straight/angle/malleable drills, and microsaws.
Prescription Use x (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices | |
|-----------------------------------------------------------------------------|---------|
| 510(k) Number | K083112 |
4-1
Prescription Use (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.