K083095 · Ultrasonix Medical Corporation · IYN · Oct 31, 2008 · Radiology
Device Facts
Record ID
K083095
Device Name
SONIXTOUCH ULTRASOUND SYSTEM
Applicant
Ultrasonix Medical Corporation
Product Code
IYN · Radiology
Decision Date
Oct 31, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The SonixTouch Ultrasound Imaging System is intended for the following applications: Ophthalmic, Abdominal, Cardiac, Intraoperative (specific), Intraoperative Neurological, Fetal, Pediatric, Small Parts, Neonatal / Adult Cephalic, OB/GYN, Transesophageal, Transrectal, Transvaginal, Peripheral Vascular, Musculoskeletal conventional, Musculoskeletal superficial, Pelvic, Nerve block, Vascular Access, Transcranial, Intravascular, Laparoscopic. The system also provides the ability to measure anatomical structures {fetal, abdominal, pediatric, small organ, cardiac, transvaginal, peripheral vessel, musculo-skeletal} and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Device Story
SonixTouch is a mobile, software-controlled diagnostic ultrasound system; acquires harmonic ultrasound echo data; displays in B-Mode, M-Mode, Pulsed/Continuous Doppler, Color Doppler, Amplitude Doppler, and Elastography. Used in clinics/hospitals by physicians/technicians. Features include ECG display, measurement/calculation packages for fetal biometry and anatomical structures, cine review, image zoom, biopsy guidance, and DICOM storage. System uses piezoelectric transducers (linear, convex, phased, microconvex, TEE) to transmit/receive sound waves. Backend processing generates real-time images. Elastography mode extracts tissue mechanical properties by applying manual pressure; color-codes stiff vs. soft structures. Output displayed on flat panel; assists clinicians in diagnosis and interventional guidance (biopsy, nerve block, vascular access).
Clinical Evidence
Bench testing only. Verification protocol (00.046.088_Ultrasonix_OP-SP_2.7.0) confirmed the elasticity imaging algorithm can differentiate structures with different rigidity. Acoustic output limits and safety compliance verified against IEC 60601-1 and IEC 60601-2-37 standards.
Indicated for diagnostic ultrasound imaging or fluid flow analysis in ophthalmic, abdominal, cardiac, intraoperative, neurological, fetal, pediatric, small parts, neonatal/adult cephalic, OB/GYN, transesophageal, transrectal, transvaginal, peripheral vascular, musculoskeletal, pelvic, nerve block, vascular access, transcranial, intravascular, and laparoscopic applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Siemens S2000 Diagnostic Ultrasound System (K072786)
Volcano S5i Family of Ultrasound Systems (K061215)
Submission Summary (Full Text)
{0}------------------------------------------------
K0r3095
### 510 (k) Summary for the SonixTouch Ultrasound Scanner
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with the Safe Medical Devices Act of 1990 revisions to 21 CFR, Part 807.92, Content and format of a 510(k) summary.
### 1.0 Submitter Information
### 1.1 Submitter
Ultrasonix Medical Corporation 130-4311 Viking Way Richmond, British Columbia Canada V6V 2K9 (t) 604-279-8550 (f) 604-279-8559
OCT 3 1 2008
## 1.2 Contact
Chas Yu, Quality Assurance Manager (t) 604-279-8550 x 152 (f) 604-279-8559 (e) chas.yu@ultrasonix.com
### 1.3 Date Prepared
September 1, 2008
### 2.0 Device Name
### 2.1 Common Name
Ultrasound Imaging System
### 2.2 Proprietary Name
SonixTouch Ultrasound Scanner
### 2.3 Classification Name
| | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
{1}------------------------------------------------
### 2.4 Classification
Class II
### 2.5 Predicate Device:
SONIX MDP Ultrasound Scanner (K080935) SONIX Ultrasound Scanner (K061827) SonoSite MicroMaxx High Resolution Ultrasound System (K053069) Philips HD 11 Diagnostic Ultrasound System (K062247) GE Logiq P5 and A5 (K060993) Imacor Zura TEE System (K080223) Siemens S2000 Diagnostic Ultrasound System (K072786) Volcano S5i Family of Ultrasound Systems (K061215)
### 2.6 Reason for submission:
### Name change request
N/A
### New product clearance for:
SonixTouch Ultrasound Scanner and Transducers:
L14-5/38 C5-2/60 C5-2/40 EC9-5/10 SA4-2/24 SA3-1/24 L14-5W/60 PA7-4 BPSL9-5/55 BPC8-4/10 C7-3/50 4DC7-3/40 4DL14-5/38 4DEC9-5/10 L9-4/38 MC9-4/12 HST15-8/20 L15-8/26 TEM7-3/9 TEM10-7/5 TEEIMA LAP
{2}------------------------------------------------
## 2.7 Device description
The SonixTouch Ultrasound Scanner is a new multi-purpose mobile, software controlled diagnostic ultrasound system with on-screen thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-Mode, M-Mode, Pulsed(PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, or Harmonic imaging on a Flat Panel Display. The user interface includes specialized controls, a minimized computer keyboard, and touch panel on an ergonomic console.
The system has an electrocardiography (ECG) display feature and support for a 3-lead ECG cable assembly. The systems provide measurement capabilities for anatomical structures and fetal biometry that provide information used for clinical diagnostic purposes. The system has a PW and CW audio output feature and cine review, image zoom, labeling, biopsy, measurements and calculations, image storage and review, printing, and recording capabilities. The systems include a Digital Imaging and Communications (DICOM) module which enables storage.
The system is designed for use in linear, convex and phased array scanning modes, and supports linear, convex, microconvex and phased array probes.
The biopsy kits are accessories to the SonixTouch Ultrasound Scanner. These accessories are made up of a polymeric bracket. There are features on the bracket that prevent the bracket from being oriented incorrectly when attached to the transducer. The brackets are not sterile and will be covered with a sterile sheath prior to use. These brackets are designed to accept and retain the needle quides in a mechanically secure way through the medium of the sterile sheath. The brackets are reusable. The needle guide is a separate sterile polymeric part that attaches to the bracket through a sterile sheath. The needle quides will support various sized needle guides are sold in sterile kits that contain multiple needle guides, sterile sheaths, ultrasound transmission gel, and bands.
| Frequency Range | 2-15MHz |
|------------------|-------------------|
| Transducer types | Linear array |
| | Curved array |
| | TEE array |
| | Intracavity array |
| | Phased array |
{3}------------------------------------------------
The SonixTouch Ultrasound Scanner is designed to comply with the following standards and the system follows Track 3.
| EN 60601-1 | European Norm, Medical Electrical Equipment |
|----------------|--------------------------------------------------------------------------------------|
| UL 2601-1 | Underwriters Laboratories Standards, Medical<br>Electrical Equipment |
| C22-2 No 601.1 | Canadian Standards Association, Medical<br>Electrical Equipment |
| EM 60601-1-2 | European Norm, Collateral Standard,<br>Electromagnetic Compatibility |
| IEC 60601-2-37 | Particular requirements for the safety of<br>ultrasonic medical diagnostic equipment |
| AIUM | Acoustic Output Labeling Standard for Diagnostic<br>Ultrasound Equipment - Jan 1998 |
| AIUM | Standard for Real-Time Display of Thermal and<br>Mechanical Acoustic Output Indices |
### 3.0 Summary of Intended Uses
The SonixTouch Ultrasound Imaging System is intended for the following applications: Ophthalmic, Abdominal, Cardiac, Intraoperative (specific), Intraoperative Neurological, Fetal, Pediatric, Small Parts, Neonatal / Adult Cephalic, OB/GYN, Transesophageal, Transrectal, Transvaginal, Peripheral Vascular, Musculoskeletal conventional, Musculoskeletal superficial, Pelvic, Nerve block, Vascular Access, Transcranial, Intravascular, Laparoscopic.
The system also provides the ability to measure anatomical structures {fetal, abdominal, pediatric, small organ, cardiac, transvaginal, peripheral vessel, musculo-skeletal} and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
### 4.0 Comparison to Predicate Device
The SonixTouch Ultrasound Scanner is substantially equivalent to the predicate devices with respect to intended use/indications for use, principles of operation and technological characteristics.
The SONIX Touch Ultrasound Scanner includes a digital beamformer that is similar in function to the predicate devices beamformer. It allows transmitting and receiving signals through the ultrasound transducers. The ultrasound transducers are similar to the ones used on predicate devices.
The backend processing is also similar to the predicate devices and yields an ultrasound image in realtime for diagnosis purposes.
### Elastography
The Elastography imaging mode is similar to the Siemens S2000 Elasticity imaging mode implementation (also called eSie Touch Elasticity Imaging).
{4}------------------------------------------------
The following slides explain the concept used for Elastography imaging:
Elastography is a method to extract the mechanical properties of tissue
# Comparison of Ultrasound Imaging Techniques
Image /page/4/Picture/3 description: The image shows a medical ultrasound. The ultrasound displays a series of waveforms and grayscale images, likely representing blood flow or tissue density. The image also contains some text, but it is not legible. The ultrasound is likely being used for diagnostic purposes.
B-mode Observe internal structure types and boundaries
> Doppler modes Observe internal flow and tissue motion
Elastography Observe mechanical properties of tissue
The imaging method involves applying a mechanical pressure with the hand on the transducer. The actual imaging sequence is similar to the B-mode sequence except that the system will acquire the RF signal instead of acquiring B mode data. The acoustic output transmission is identical to B-mode. The algorithm extracts a strain value information for every point on the image. The Elastography image then color-codes the stiff versus softer structures. The clinical benefits of elastography imaging are still under evaluation and no clinical diagnosis claim can be made other than being able to tell whether or not a structure inside the patient is stiffer than another one. The submission includes a document "00.046.088_Ultrasonix_OP-SP_2.7.0_Release_Validation_Protocol_B_061308_ElastoValidation_Corina" included in the additional information dated October 14, 2008 which provides the verification protocol as well as expected results for elastography imaging. The report verifies the elasticity imaging algorithm (images actually in the validation protocol as well) can differentiate different structures with different rigidity
{5}------------------------------------------------
# Elastography Imaging Technique
Image /page/5/Picture/1 description: The image shows a dark screen with some text and graphical elements. The text "Soft" and "Hard" are visible on the left side of the screen, possibly indicating settings or options. There are some scattered bright spots and lines, which could be part of a user interface or some kind of data visualization. Due to the low resolution and contrast, it is difficult to discern the exact content or purpose of the screen.
stiff structure displaced but RF data unchanged color coded in Blue
softer structure displaced and RF data unchanged color coded in Red
## Elastography verification data
The attached document "00.046.088_Ultrasonix_OP-
SP 2.7.0 Release Validation Protocol B 061308_ElastoValidation_Corina" includes verification protocol as well as expected results for elastography imaging. The report verifies that our elasticity imaging algorithm (images actually in the validation protocol as well) passes and can differentiate different structures with different rigidity.
### 5.0 Technological characteristics
The technological characteristics are substantially similar to that of the predicates. The device operates identically to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as 2D or M-mode images. Doppler shift caused by blood flow is displayed as Color Flow, or as spectrum analysis. The modes of this device (2D, PW Doppler, Color Flow Mapping Doppler, Power Doppler, Continuous Wave Doppler) are the same as the predicate devices identified in item 2.5. Transducer patient contact materials are biocompatible.
The beam forming architecture is very similar to that of the predicate devices. The receiving and processing hardware is similar but innovative in that it is a programmable system made of 2 building blocks, which can be reconfigured to operate the system in any imaging mode.
The parameters used to adjust image quality are the same as that seen in the predicates. This includes the use of TGC gain, depth control, base control and angling, among others.
### 6.0 Safety considerations
As track 3 ultrasound device, the SonixTouch Ultrasound Scanner is designed to comply with the "Standard For Real Time Display Of Thermal And Mechanical Acoustic Output Indices On Diagnostic Ultrasound Equipment (1992)" published by the National Electrical Manufacturers Association as UD-3.
{6}------------------------------------------------
With respect to limits on acoustic outputs, the SonixTouch Ultrasound Scanner complies with the guideline limits set in the September 30, 1997 revision of 510(k) Diagnostic Ultrasound Guidance.
With regard to general safety, the SonixTouch Ultrasound Scanner is designed to comply with IEC 601-1 (1988) Medical Electrical Equipment, Part 1: General Requirements for Safety, and IEC 60601-2-37: Particular Requirements For The Safety Of Ultrasonic Medical Diagnostic And Monitoring Equipment.
The devices' acoustic output limits are:
| ISPTA (d) | 720mW/cm² |
|-----------------------|--------------------|
| TIS/TIB/TIC | 0.1 - 4.0 (Range) |
| Mechanical Index (MI) | 1.9 (Maximum) |
| ISPPA (d) | 0-700W/cm² (Range) |
The limits are the same as predicate Track 3 devices.
{7}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, and the words "HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle. The logo is black and white and has a simple, clean design. The eagle is a symbol of strength and freedom, and the logo represents the department's mission to protect the health of all Americans and provide essential human services.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 5 - 2009
Ultrasonix Medical Corporation % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K083095
Trade/Device Name: SonixTouch Ultrasound Scanner Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: October 16, 2008 Received: October 17, 2008
Dear Mr. Job:
This letter corrects our substantially equivalent letter of October 31, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the SonixTouch Ultrasound System, as described in your premarket notification:
#### Transducer Model Number
SA4-2/24 Phased Array 2.8 MHz 24 mm SA3-1/24 Phased Array 2.0 MHz 24mm PA7-4 Phased Array 5.0 MHz C5-2/60 Convex 3.2 MHz 60mm Radius C5-2/40 Convex 3.2 MHz 40mm Radius C7-3/50 Convex 5.0 MHz 50mm Radius EC9-5/10 Microconvex Endocavity 6.6 MHz 10mm Radius L14-5/38 Linear 8 MHz 38mm L9-4/38 Linear 6.5 MHz 38mm L14-5W/60 Linear 8 MHz 38mm
{8}------------------------------------------------
BPSL9-5/55/10L Linear ECBP 7.5 MHz BPSL9-5/55-10C Microconvex ECBP 6.5 MHz 4DC7-3/40mm 3D/4D Curved Transabdominal 3.5 MHz 4DEC9-5/10mm 3D/4D Endocavity 6.5 MHz MC9-4/12mm Mircoconvex 6.0 MHz 4DL14-5/38mm 3D/4D Linear 7.5 MHz HST15-8/20mm Hockey Stick Linear 10.0 MHz L15-8/26mm Linear 10.0 MHz TEM7-3/9mm Transesophageat 5.0 MHz TEM10-7/5mm Transesophageat 7.0 MHz TEEIMA Transesophageal 7.0 MHz LAP Laproscopic
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on vour responsibilities under the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Sunder Rajan at (240) 276-3666.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{9}------------------------------------------------
## Indications for Use
510(k) Number (if known): Device Name: SonixTouch Ultrasound Scanner Indications For Use:
The SonixTouch Ultrasound Scanner is intended for the following applications: Cardiac, Intraoperative (specific), Intraoperative Ophthalmic, Abdominal, Neurological, Fetal, Pediatric, Small Parts, Neonatal/ Adult Cephalic, OB/GYN, Peripheral Vascular, Transvaginal, Transrectal, Transesophageal, Musculoskeletal conventional, Musculoskeletal superficial, Pelvic, Nerve Block, Vascular Access, Transcranial.
The system also provides the ability to measure anatomical structures {fetal, abdominal, pediatric, small organ, cardiac, transvaginal, peripheral vessel, musculoskeletal} and provides calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helsleuen
Page 1 of
(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices
510(k) Number K083095
{10}------------------------------------------------
### ULTRASOUND INDICATIONS FOR USE TABLES
### SonixTouch Ultrasound Scanner – Diagnostic Ultrasound Indications for Use Form
#### Intended use:
### Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Fetal | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Abdominal | N | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative Neurological | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Pediatric | N | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Cardiac | N | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Transesophageal | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Transrectal | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Transvaginal | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Transurethral | | | | | | | | | | |
| Transcranial | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Peripheral Vascular | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Laparoscopic | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| MSK Conventional | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| MSK Superficial | N | N | N | N | | N | N | N | N (*1) | N (*2) |
| Vascular Access | N | N | N | N | | N | N | N | N (*1) | N (*2, *4) |
| Nerve Block | N | N | N | N | | N | N | N | N (*1) | N (*2, *3) |
| Other (specify) | | | | | | | | | | |
#### N = New indication; P = Previously cleared
#### Additional Comments:
Small Organ: Breast, Thyroid, Testicle
Intraoperative: Abdominal organs and vascular
- *1. BIM, BIPWD, BICWD, BICFMPWD, BIAD/PWD, BIOPD/PWD, BICFWCWD, B/AD/CWD, B/DPD/CWD
- 1. DiM, Dr VVD, Drover Dopler (DPD), Imaging for guidance of biopsy, Panoramic Imaging, Harmonic Imaging, Harmonic Imaging, Tissue Doppler Imaging, Compound Imaging, Freehand 3D Imaging, Live 3D/4D Imaging.
- *3. Imaging for guldance of nerve block injections
- Imaging for guidance of central or peripheral lines. *4.
SONIX Touch User Manual
_
ﻨﺴ 1
Appendix C: Transdcuer Specifications
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
{11}------------------------------------------------
#### Diagnostic ULTRASOUND INDICATIONS FOR USE Form SA4-2/24 phased array 2.8 MHz 24mm transducer
#### Intended use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|---------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Cardiac | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Transcranial | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| MSK Conventional | | | | | | | | | | |
| MSK Superficial | | | | | | | | | | |
| Vascular Access | | | | | | | | | | |
| Nerve Block | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = New indication; P = Previously cleared
Additional Comments:
Small Organ: Breast, Thyroid, Testicle
Intraoperative: Abdominal organs and vascular
*1 B/M B/PWD. B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
eraries : 1000: Minister : 12:00:40
11. Bik, BPWD, B/CFM/PWD, B/DPWD, B/DPDP, Pancamic Imaging, Hamonic Imaging, Hamonic Imaging, Hamonic Imaging, Harmonic Imaging, Harmo
Appendix C: Transdcuer Specifications
2
| (Division Sign-Off) | |
|--------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal and Radiological Devices | |
| 510(k) Number | K083095 |
SONIX Touch User Manual
{12}------------------------------------------------
#### Diagnostic ULTRASOUND INDICATIONS FOR USE Form SA3-1/24 phased array 2.0 MHz 24mm transducer
Intended use:
### Diagnostic ultrasound imaging or fluid flow anaiysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Cardiac | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Transcranial | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| MSK Conventional | | | | | | | | | | |
| MSK Superficial | | | | | | | | | | |
| Vascular Access | | | | | | | | | | |
| Nerve Block | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = New indication; P = Previously cleared
#### Additional Comments:
Small Organ: Breast, Thyroid, Testicle
Intraoperative: Abdominal organs and vascular
*1. BIM, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/CFM/CWD, B/AD/CWD, B/AD/CWD, B/DPD/CWD
- *2. Elastography, Directional Power (DPD), Imaging for guidance of biopsy, Panoramle Imaging, Harmonic Imaging, Tissue Doppler Imaging, Compound Imaging, Freehand 3D Imaging.
| SONIX Touch User Manual | Appendix C: Transdcuer Specifications |
|--------------------------------------------------------------|---------------------------------------|
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal and Radiological Devices | |
510(k) Number K023045
{13}------------------------------------------------
### Diagnostic ULTRASOUND INDICATIONS FOR USE Form PA7-4 phased array 5.0 MHz transducer
#### Intended use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Cardiac | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Transcranial | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| MSK Conventional | | | | | | | | | | |
| MSK Superficial | | | | | | | | | | |
| Vascular Access | | | | | | | | | | |
| Nerve Block | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = New indication; P = Previously cleared
#### Additional Comments:
Small Organ: Breast, Thyroid, Testicle
Intraoperative: Abdominal organs and vascular
*1. B/M, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/OPD/CWD, B/OPD/CWD
"1. BM, BPWU, BIGHWFWD, BIGHMFWD, DIAL/ WD, Drom You Crem Porter Corporation Imaging, Hamonic Imaging, Hamonic Imaging, Hamonic Imaging, Namonic Imaging, Tissue Doppler Imaging, Compound Imaging, Freehand 3D Imaging.
Appendix C: Transdcuer Specifications
<sup>4</sup>
| (Division Sign-Off) |
|-----------------------------------------|
| Division of Reproductive, Abdominal and |
Brasion of Roproduct
510/k) Number 083 510(k) Number
SONIX Touch User Manual
{14}------------------------------------------------
#### Diagnostic ULTRASOUND INDICATIONS FOR USE Form C5-2/60 convex 3.2 MHz 60mm radius transducer
Intended use:
#### Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Abdominal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Trasncranial | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Laparoscopic | | | | | | | | | | |
| MSK Conventional | | N | N | N | | N | N | N | N (*1) | N (*2) |
| MSK Superficial | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Vascular Access | | | | | | | | | | |
| Nerve Block | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
#### N = New indication; P = Previously cleared
#### Additional Comments:
Small Organ: Breast, Thyroid, Testicle
Intraoperative: Abdominal organs and vascular
1. B/M, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
*1. BM, B/PWD, B/CWD, B/DPWVD, B/DPD/PWD, B/CFMCWD, B/AD/CWD, B/OPD/CWD
*2. Elastography, Directional Power Dopels (DPD), Maging, Penoramic Imaging, Hamoric Imaging, Hamor
| SONIX Touch User Manual | | Appendix C: Transdcuer Specifications |
|-------------------------|--|---------------------------------------|
|-------------------------|--|---------------------------------------|
5
| (Division Sign-Off) | |
|-----------------------------------------|---------|
| Division of Reproductive, Abdominal and | |
| Radiological Devices, | |
| 510(k) Number | K083095 |
{15}------------------------------------------------
#### Diagnostic ULTRASOUND INDICATIONS FOR USE Form C5-2/40 convex 3.2 MHz 40mm radius transducer
#### Intended use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Abdominal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Trasncranial | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Laparoscopic | | | | | | | | | | |
| MSK Conventional | | N | N | N | | N | N | N | N (*1) | N (*2) |
| MSK Superficial | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Vascular Access | | | | | | | | | | |
| Nerve Block | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = New indication; P = Previously cleared
#### Additional Comments:
Small Organ: Breast, Thyroid, Testicle
Intraoperative: Abdominal organs and vascular
- *1. BIM, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
11. Elastography, Directional Power (DPD), Imaging for guidance of biopsy, Panoramic Imaging, Harmonic Imaging, Tissue Doppler Imaging, Compound Imaging, Freehand 3D Imaging,
| Appendix C: Transdcuer Specifications | |
|---------------------------------------|-----------------------------------------|
| | 6 |
| | (Division Sign-Off) |
| | Division of Reproductive, Abdominal and |
| | SO |
4083095
Radiological Devices
510(k) Number
SONIX Touch User Manual
{16}------------------------------------------------
#### Diagnostic ULTRASOUND INDICATIONS FOR USE Form Č7-3/50 convex 5.0 MHz 50mm radius transducer
Intended use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Abdominal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Trasncranial | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Laparoscopic | | | | | | | | | | |
| MSK Conventional | | N | N | N | | N | N | N | N (*1) | N (*2) |
| MSK Superficial | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Vascular Access | | | | | | | | | | |
| Nerve Block | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = New indication; P = Previously cleared
Additional Comments:
Small Organ: Breast, Thyroid, Testicle
Intraoperative: Abdominal organs and vascular
*1. BIM, BIPWD, BICWD, BICFM/PWD, BIAD/PWD, BIDPD/PWD, BICFM/CWD, BI/DPD/CWD
*2. Elastography, Directional Power Doppler (DPD), Imaging for guldance of biopsy, Panoramic Imaging, Tissue Doppler Imaging, Compound Imaging, Freehand 3D Imaging.
SONIX Touch User Manual
Appendix C: Transdcuer Specifications
从 (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
{17}------------------------------------------------
#### Diagnostic ULTRASOUND INDICATIONS FOR USE Form EC9-5/10 microconvex endocavity 6.6 MHz 10mm radius transducer
Intended use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Transvaginal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Transurethral | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| MSK Conventional | | | | | | | | | | |
| MSK Superficial | | | | | | | | | | |
| Vascular Access | | | | | | | | | | |
| Nerve Block | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = New indication; P = Previously cleared
Additional Comments:
Small Organ: Breast, Thyroid, Testicle
Intraoperative: Abdominal organs and vascular
- *1. BM, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/DPD/PWD, B/CFM/CWD, B/DPD/CWD
- *2. Elastography, Directional Power (DPD), Imaging for guidance of biopsy, Panoramic Imaging, Harmonic Imaging, Tissue Doppler Imaging, Compound Imaging, Freehand 3D Imaging.
Appendix C: Transdcuer Specifications
| (Division Sign-Off) |
|---------------------|
|---------------------|
Division of Reproductive, Abdominal and
Radiological Devices
| 510(k) Number | 4083095 |
|---------------|---------|
|---------------|---------|
SONIXIX Touch User Manual
{18}------------------------------------------------
#### Diagnostic ULTRASOUND INDICATIONS FOR USE Form L14-5/38 linear 8 MHz 38mm transducer
Intended use:
#### Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | | N | N | N (*1) | N (*2) |
| Abdominal | | N | N | N | | | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | N | N | N | | | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | | N | N | N | | | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | | | N | N | N (*1) | N (*2) |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | | N | N | N (*1) | N (*2) |
| Laparoscopic | | | | | | | | | | |
| MSK Conventional | | N | N | N | | | N | N | N (*1) | N (*2) |
| MSK Superficial | | N | N | N | | | N | N | N (*1) | N (*2) |
| Vascular Access | | N | N | N | | | N | N | N (*1) | N (*2, *4) |
| Nerve Block | | N | N | N | | | N | N | N (*1) | N (*2, *3) |
| Other (specify) | | | | | | | | | | |
N = New indication; P = Previously cleared
#### Additional Comments:
SONIX
#### Small Organ: Breast, Thyroid, Testicle
#### Intraoperative: Abdominal organs and vascular
-
- 11. BM, B/PWD, B/C/WD, B/O/PWD, B/DPPWD, B/DP/PWD, B/O/CWO, B/DPD/CWD
*2. Elastography, Directional Power Dopels ( DPD), Maging for guitance of biopsy, Panoramic Imaging,
- *3. Imaging for guidance of nerve block injections
- *4. Imaging for guidance of central or peripheral lines.
Touch User Manual Appendix C: Transdcuer Specifications
| | 9 |
|--|---------------------|
| | (Division Sign-Off) |
Division of Reproductive. Abdominal andDivision of Reproductive, Abdollian
510(k) Number ____________________________________________________________________________________________________________________________
{19}------------------------------------------------
#### Diagnostic ULTRASOUND INDICATIONS FOR USE Form L9-4/38 linear 6.5 MHz 38mm transducer
Intended use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Abdominal | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Peripheral Vascular | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Laparoscopic | | | | | | | | | | |
| MSK Conventional | | N | N | N | | N | N | N | N (*1) | N (*2) |
| MSK Superficial | | N | N | N | | N | N | N | N (*1) | N (*2) |
| Vascular Access | | N | N | N | | N | N | N | N (*1) | N (*2, *4) |
| Nerve Block | | N | N | N | | N | N | N | N (*1) | N (*2, *3) |
| Other (specify) | | | | | | | | | | |
N = New indication; P = Previously cleared
Additional Comments:
Small Organ: Breast, Thyroid, Testicle
Intraoperative: Abdominal organs and vascular
- *1. BIM, B/PWD, B/CWD, B/CFM/PWD, B/AD/PWD, B/CFM/CWD, B/CFM/CWD, B/AD/CWD, B/DPD/CWD
- 11. BM, BFWD, DVGYD, DVGT WT WO, Droin 105, Droi MSO, Dron Sor and Partic Partic Imaging, Harmonic Imaging, Harmonic Imaging, Tissue Doppler Imaging, Compound Imaging, Freehand 3D Imaging,
- *3. Imaging for guidance of nerve block injections
- *4. Imaging for guidance of central or peripheral lines.
| Appendix C: Transdcuer Specifications | SONIX Touch User Manual |
|--------------------------------------------------------------|-------------------------|
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal and Radiological Devices | |
| 510(k) Number | K083095 |
{20}------------------------------------------------
#### Diagnostic ULTRASOUND INDICATIONS FOR USE Form L14-5W/60 linear 8 MHz 38mm transducer
Intended use:
#### Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | |
|-----------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Doppler | Combined<br>(Specify) | Other<br>(Specify) |
| Ophthalmic | | | | | | | | | | |
| Fetal | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Abdominal | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Intraoperative (specify) | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | |
| Pediatric | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Small Organ (specify) | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Neonatal Cephalic | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Adult Cephalic | | N | N | N | N | N | N | N | N (*1) | N (*2) |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.